Evaluation of light- and dark-adapted ERGs using a mydriasis-free, portable system: clinical classifications and normative data

verfasst von: Henry Liu, Xiang Ji, Sabrina Dhaliwal, Syeda Naima Rahman, Michelle McFarlane, Anupreet Tumber, Jeff Locke, Tom Wright, Ajoy Vincent, Carol Westall

Erschienen in: Documenta Ophthalmologica | Ausgabe 3/2018

Einloggen, um Zugang zu erhalten

Abstract

Purpose

The full-field electroretinogram (ff-ERG) is a widely used clinical tool to evaluate generalized retinal function by recording electrical potentials generated by the cells in the retina in response to flash stimuli and requires mydriasis. The purpose of this study was to determine the intra-visit reliability and diagnostic capability of a handheld, mydriasis-free ERG, RETeval (LKC Technologies, Gaithersburg, MD, USA), in comparison with the standard clinical ff-ERG by measuring responses recommended by the International Society for Clinical Electrophysiology of Vision (ISCEV).

Methods

This prospective, cross-sectional study included 35 patients recruited at the Hospital for Sick Children (median age = 17, range 11 months–69 years) who had undergone a clinical ff-ERG according to ISCEV standards. For RETeval (n = 35), pupils were undilated in most (n = 29) and sensor strip electrodes were placed under the inferior orbital rim. Stimulus settings on RETeval were equivalent to those used in the clinical ERG. Fifty-seven control participants (median age = 22, range 8–65 years) underwent undilated RETeval ERG to establish standard values for comparison. Patient waveform components with amplitudes < 5th percentile, or implicit times > 95th percentile of normal relative to control data were classified as abnormal for the RETeval system.

Results

The RETeval system demonstrated a high degree of within-visit reliability for amplitudes (ICC = 0.82) and moderate reliability for implicit times (ICC = 0.53). Cohen’s Kappa analysis revealed a substantial level of agreement between the diagnostic capability of RETeval in comparison with clinical ff-ERG (k = 0.82), with a sensitivity and specificity of 1.00 and 0.82, respectively. Pearson’s correlations for clinical ERG versus RETeval demonstrated a positive correlation for amplitudes across the rod (r = 0.65) and cone (r = 0.74) ERG waveforms. Bland–Altman plots showed no bias between the mean differences across all amplitude and implicit time parameters of the two systems.

Conclusions

The present study demonstrated that RETeval is a reliable tool with reasonable accuracy in comparison with the clinical ERG. The portable nature of RETeval system enables its incorporation at resource-limited centers where the ff-ERG is not readily available. The avoidance of sedation and pupillary dilation are added advantages of RETeval ERG.
Literatur
1.
Zurück zum Zitat McCulloch DL, Marmor MF, Brigell MG, Hamilton R, Holder GE, Tzekov R, Bach M (2015) ISCEV standard for full-field clinical electroretinography (2015 update). Doc Ophthalmol 130(1):1–12CrossRef McCulloch DL, Marmor MF, Brigell MG, Hamilton R, Holder GE, Tzekov R, Bach M (2015) ISCEV standard for full-field clinical electroretinography (2015 update). Doc Ophthalmol 130(1):1–12CrossRef
2.
Zurück zum Zitat Barber C (1994) Electrodes and the recording of the human electroretinogram (ERG). Int J Psychophysiol 16(2–3):131–136CrossRef Barber C (1994) Electrodes and the recording of the human electroretinogram (ERG). Int J Psychophysiol 16(2–3):131–136CrossRef
3.
Zurück zum Zitat Kato K, Kondo M, Sugimoto M, Ikesugi K, Matsubara H (2015) Effect of pupil size on flicker ERGs recorded with RETeval system: new mydriasis-free full-field ERG system. Invest Ophthalmol Vis Sci 56(6):3684–3690CrossRef Kato K, Kondo M, Sugimoto M, Ikesugi K, Matsubara H (2015) Effect of pupil size on flicker ERGs recorded with RETeval system: new mydriasis-free full-field ERG system. Invest Ophthalmol Vis Sci 56(6):3684–3690CrossRef
4.
Zurück zum Zitat Kato K, Kondo M, Nagashima R, Sugawara A, Sugimoto M, Matsubara H, McCulloch DL, Ikesugi K (2017) Factors affecting mydriasis-free flicker ERGs recorded with real-time correction for retinal illuminance: study of 150 young healthy subjects. Invest Ophthalmol Vis Sci 58(12):5280–5286CrossRef Kato K, Kondo M, Nagashima R, Sugawara A, Sugimoto M, Matsubara H, McCulloch DL, Ikesugi K (2017) Factors affecting mydriasis-free flicker ERGs recorded with real-time correction for retinal illuminance: study of 150 young healthy subjects. Invest Ophthalmol Vis Sci 58(12):5280–5286CrossRef
5.
Zurück zum Zitat Bradshaw K, Hansen R, Fulton A (2004) Comparison of ERGs recorded with skin and corneal-contact electrodes in normal children and adults. Doc Ophthalmol 109(1):43–55CrossRef Bradshaw K, Hansen R, Fulton A (2004) Comparison of ERGs recorded with skin and corneal-contact electrodes in normal children and adults. Doc Ophthalmol 109(1):43–55CrossRef
6.
Zurück zum Zitat Maa AY, Feuer WJ, Davis CQ, Pillow EK, Brown TD, Caywood RM, Chasan JE, Fransen SR (2016) A novel device for accurate and efficient testing for vision-threatening diabetic retinopathy. J Diabetes Complicat 30(3):524–532CrossRef Maa AY, Feuer WJ, Davis CQ, Pillow EK, Brown TD, Caywood RM, Chasan JE, Fransen SR (2016) A novel device for accurate and efficient testing for vision-threatening diabetic retinopathy. J Diabetes Complicat 30(3):524–532CrossRef
7.
Zurück zum Zitat Fukuo M, Kondo M, Hirose A, Fukushima H, Ikesugi K, Sugimoto M, Kato K, Uchigata Y, Kitano S (2016) Screening for diabetic retinopathy using new mydriasis-free, full-field flicker ERG recording device. Sci Rep 6:36591CrossRef Fukuo M, Kondo M, Hirose A, Fukushima H, Ikesugi K, Sugimoto M, Kato K, Uchigata Y, Kitano S (2016) Screening for diabetic retinopathy using new mydriasis-free, full-field flicker ERG recording device. Sci Rep 6:36591CrossRef
8.
Zurück zum Zitat Nakamura N, Fujinami K, Mizuno Y, Noda T, Tsunoda K (2016) Evaluation of cone function by a handheld non-mydriatic flicker electroretinogram device. Clin Ophthalmol (Auckland, NZ) 10:1175CrossRef Nakamura N, Fujinami K, Mizuno Y, Noda T, Tsunoda K (2016) Evaluation of cone function by a handheld non-mydriatic flicker electroretinogram device. Clin Ophthalmol (Auckland, NZ) 10:1175CrossRef
9.
Zurück zum Zitat Kurtenbach A, Kramer S, Strasser T, Zrenner E, Langrová H (2017) The importance of electrode position in visual electrophysiology. Doc Ophthalmol 134(2):129–134CrossRef Kurtenbach A, Kramer S, Strasser T, Zrenner E, Langrová H (2017) The importance of electrode position in visual electrophysiology. Doc Ophthalmol 134(2):129–134CrossRef
10.
Zurück zum Zitat Bland JM, Altman DG (1995) Comparing methods of measurement: why plotting difference against standard method is misleading. Lancet 346:1085–1087CrossRef Bland JM, Altman DG (1995) Comparing methods of measurement: why plotting difference against standard method is misleading. Lancet 346:1085–1087CrossRef
12.
Zurück zum Zitat Westall CA, Panton CM, Levin AV (1999) Time courses for maturation of electroretinogram responses from infancy to adulthood. Doc Ophthalmol 96(4):355–379CrossRef Westall CA, Panton CM, Levin AV (1999) Time courses for maturation of electroretinogram responses from infancy to adulthood. Doc Ophthalmol 96(4):355–379CrossRef
13.
Zurück zum Zitat Birch DG, Anderson JL (1992) Standardized full-field electroretinography: normal values and their variation with age. Arch Ophthalmol 110(11):1571–1576CrossRef Birch DG, Anderson JL (1992) Standardized full-field electroretinography: normal values and their variation with age. Arch Ophthalmol 110(11):1571–1576CrossRef
14.
Zurück zum Zitat Karpe G, Rickenbach K, Thomasson S (1950) The clinical electroretinogram. I. The normal electroretinogram above fifty years of age. Acta Ophthalmol 28(3):301CrossRef Karpe G, Rickenbach K, Thomasson S (1950) The clinical electroretinogram. I. The normal electroretinogram above fifty years of age. Acta Ophthalmol 28(3):301CrossRef
15.
Zurück zum Zitat Fulton AB, Hansen RM, Westall CA (2003) Development of ERG responses: the ISCEV rod, maximal and cone responses in normal subjects. Doc Ophthalmol 107(3):235–241CrossRef Fulton AB, Hansen RM, Westall CA (2003) Development of ERG responses: the ISCEV rod, maximal and cone responses in normal subjects. Doc Ophthalmol 107(3):235–241CrossRef
16.
Zurück zum Zitat Al-Otaibi H, Al-Otaibi MD, Khandekar R, Souru C, Al-Abdullah AA, Al-Dhibi H, Stone DU, Kozak I (2017) Validity, usefulness and cost of RETeval system for diabetic retinopathy screening. Transl Vis Sci Technol 6(3):3CrossRef Al-Otaibi H, Al-Otaibi MD, Khandekar R, Souru C, Al-Abdullah AA, Al-Dhibi H, Stone DU, Kozak I (2017) Validity, usefulness and cost of RETeval system for diabetic retinopathy screening. Transl Vis Sci Technol 6(3):3CrossRef
17.
Zurück zum Zitat Golshani C, Brodie SE (2018) Quantitative calibration of sensor strip ERG electrodes. Poster presented at the 2018 annual meeting of the association for research in vision and ophthalmology, Honolulu, HI Golshani C, Brodie SE (2018) Quantitative calibration of sensor strip ERG electrodes. Poster presented at the 2018 annual meeting of the association for research in vision and ophthalmology, Honolulu, HI
18.
Zurück zum Zitat Leguire LE, Rogers GL (1985) Pattern electroretinogram: use of noncorneal skin electrodes. Vis Res 25(6):867–870CrossRef Leguire LE, Rogers GL (1985) Pattern electroretinogram: use of noncorneal skin electrodes. Vis Res 25(6):867–870CrossRef
19.
Zurück zum Zitat Miura G, Nakamura Y, Sato E, Yamamoto S (2016) Effects of cataracts on flicker electroretinograms recorded with RETeval™ system: new mydriasis-free ERG device. BMC Ophthalmol 16(1):22CrossRef Miura G, Nakamura Y, Sato E, Yamamoto S (2016) Effects of cataracts on flicker electroretinograms recorded with RETeval™ system: new mydriasis-free ERG device. BMC Ophthalmol 16(1):22CrossRef
Metadaten
Titel
Evaluation of light- and dark-adapted ERGs using a mydriasis-free, portable system: clinical classifications and normative data
verfasst von
Henry Liu
Xiang Ji
Sabrina Dhaliwal
Syeda Naima Rahman
Michelle McFarlane
Anupreet Tumber
Jeff Locke
Tom Wright
Ajoy Vincent
Carol Westall
Publikationsdatum
24.10.2018
Verlag
Springer Berlin Heidelberg
Erschienen in
Documenta Ophthalmologica / Ausgabe 3/2018
Print ISSN: 0012-4486
Elektronische ISSN: 1573-2622
DOI
https://doi.org/10.1007/s10633-018-9660-z

Weitere Artikel der Ausgabe 3/2018

Documenta Ophthalmologica 3/2018 Zur Ausgabe

Neu im Fachgebiet Augenheilkunde

Update Augenheilkunde

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.