Evaluation of low-contrast detectability for iterative reconstruction in pediatric abdominal computed tomography: a phantom study

verfasst von: Nicholas Rubert, Richard Southard, Susan M. Hamman, Ryan Robison

Erschienen in: Pediatric Radiology | Ausgabe 3/2020

Einloggen, um Zugang zu erhalten

Abstract

Background

Iterative reconstruction is offered by all vendors to achieve similar or better CT image quality at lower doses than images reconstructed with filtered back-projection.

Objective

The purpose of this study was to investigate the dose-reduction potential for pediatric abdominal CT imaging when using either a commercially available hybrid or a commercially available model-based iterative reconstruction algorithm from a single manufacturer.

Materials and methods

A phantom containing four low-contrast inserts and a uniform background with total attenuation equivalent to the abdomen of an average 8-year-old child was imaged on a CT scanner (IQon; Philips Healthcare, Cleveland, OH). We reconstructed images using both hybrid iterative reconstruction (iDose4) and model-based iterative reconstruction (Iterative Model Reconstruction). The four low-contrast inserts had circular cross-section with diameters of 3 mm, 5 mm, 7 mm and 10 mm and contrasts of 14 Hounsfield units (HU), 7 HU, 5 HU and 3 HU, respectively. Helical scans with identical kilovoltage (kV), pitch, rotation time, and collimation were repeated on the phantom at volume CT dose index (CTDIvol) of 2.0 milligrays (mGy), 3.0 mGy, 4.5 mGy and 6.0 mGy. We measured the contrast-to-noise ratio (CNR) in each rod across scans. Additionally, we collected sub-images containing each rod and sub-images containing the background and used them in two-alternative forced choice observer experiments with four observers (two radiologists and two physicists). We calculated the dose-reduction potential of both iterative reconstruction algorithms relative to a scan performed at 6 mGy and reconstructed with filtered back-projection.

Results

We calculated dose-reduction potential by either matching average equal observer performance in the two-alternative forced choice experiments or matching CNR. When matching CNR, the dose-reduction potential was 34% to 45% for hybrid iterative reconstruction and 89% to 95% for model-based iterative reconstruction. When matching average observer performance, the dose-reduction potential was 9% to 30% for hybrid iterative reconstruction and 57% to 74% for model-based iterative reconstruction. The range in dose-reduction potential depended on the rod size and contrast level.

Conclusion

Observer performance in this phantom study indicates that the dose-reduction potential indicated by an observer study is less than that of CNR; extrapolating the results to clinical studies suggests that the dose-reduction potential would also be less.
Anhänge
Nur mit Berechtigung zugänglich
Literatur
1.
Zurück zum Zitat Brink JA, Amis ES Jr (2010) Image Wisely: a campaign to increase awareness about adult radiation protection. Radiology 257:601–602CrossRef Brink JA, Amis ES Jr (2010) Image Wisely: a campaign to increase awareness about adult radiation protection. Radiology 257:601–602CrossRef
2.
Zurück zum Zitat Goske MJ, Applegate KE, Boylan J et al (2008) The Image Gently campaign: working together to change practice. AJR Am J Roentgenol 190:273–274CrossRef Goske MJ, Applegate KE, Boylan J et al (2008) The Image Gently campaign: working together to change practice. AJR Am J Roentgenol 190:273–274CrossRef
3.
Zurück zum Zitat United Nations Scientific Committee on the Effects of Atomic Radiation (2013) UNSCEAR 2013 report Vol. II. Sources, effects and risks of ionizing radiation. United Nations, New York United Nations Scientific Committee on the Effects of Atomic Radiation (2013) UNSCEAR 2013 report Vol. II. Sources, effects and risks of ionizing radiation. United Nations, New York
4.
Zurück zum Zitat Hsieh J (2008) Adaptive statistical iterative reconstruction (white paper). GE Healthcare, Waukesha Hsieh J (2008) Adaptive statistical iterative reconstruction (white paper). GE Healthcare, Waukesha
5.
Zurück zum Zitat Beister M, Kolditz D, Kalender W (2012) Iterative reconstruction methods in X-ray CT. Phys Med 28:94–108CrossRef Beister M, Kolditz D, Kalender W (2012) Iterative reconstruction methods in X-ray CT. Phys Med 28:94–108CrossRef
6.
Zurück zum Zitat Khawaja RDA, Singh S, Otrakji A et al (2015) Dose reduction in pediatric abdominal CT: use of iterative reconstruction techniques across different CT platforms. Pediatr Radiol 45:1046–1055CrossRef Khawaja RDA, Singh S, Otrakji A et al (2015) Dose reduction in pediatric abdominal CT: use of iterative reconstruction techniques across different CT platforms. Pediatr Radiol 45:1046–1055CrossRef
7.
Zurück zum Zitat Hsieh J (2015) Computed tomography: principles, design, artifacts, and recent advances, 3rd edn. SPIE Press, BellinghamCrossRef Hsieh J (2015) Computed tomography: principles, design, artifacts, and recent advances, 3rd edn. SPIE Press, BellinghamCrossRef
8.
Zurück zum Zitat Smith TB, Solomon J, Samei E (2018) Estimating detectability index in vivo: development and validation of an automated methodology. J Med Imaging 5:031403CrossRef Smith TB, Solomon J, Samei E (2018) Estimating detectability index in vivo: development and validation of an automated methodology. J Med Imaging 5:031403CrossRef
9.
Zurück zum Zitat Samei E, Richard S (2015) Assessment of the dose reduction potential of a model-based iterative reconstruction algorithm using a task-based performance metrology. Med Phys 42:314–323CrossRef Samei E, Richard S (2015) Assessment of the dose reduction potential of a model-based iterative reconstruction algorithm using a task-based performance metrology. Med Phys 42:314–323CrossRef
10.
Zurück zum Zitat McCollough CH, Yu L, Kofler JM et al (2015) Degradation of CT low-contrast spatial resolution due to the use of iterative reconstruction and reduced dose levels. Radiology 276:499–506CrossRef McCollough CH, Yu L, Kofler JM et al (2015) Degradation of CT low-contrast spatial resolution due to the use of iterative reconstruction and reduced dose levels. Radiology 276:499–506CrossRef
11.
Zurück zum Zitat Vaishnav JY, Jung WC, Popescu LM et al (2014) Objective assessment of image quality and dose reduction in CT iterative reconstruction. Med Phys 41:071904CrossRef Vaishnav JY, Jung WC, Popescu LM et al (2014) Objective assessment of image quality and dose reduction in CT iterative reconstruction. Med Phys 41:071904CrossRef
12.
Zurück zum Zitat Hernandez-Giron I, Calzado A, Geleijns J et al (2014) Comparison between human and model observer performance in low-contrast detection tasks in CT images: application to images reconstructed with filtered back projection and iterative algorithms. Br J Radiol 87:20140014CrossRef Hernandez-Giron I, Calzado A, Geleijns J et al (2014) Comparison between human and model observer performance in low-contrast detection tasks in CT images: application to images reconstructed with filtered back projection and iterative algorithms. Br J Radiol 87:20140014CrossRef
13.
Zurück zum Zitat Yu L, Leng S, Chen L, Kofler JM et al (2013) Prediction of human observer performance in a 2-alternative forced choice low-contrast detection task using channelized Hotelling observer: impact of radiation dose and reconstruction algorithms. Med Phys 40:041908CrossRef Yu L, Leng S, Chen L, Kofler JM et al (2013) Prediction of human observer performance in a 2-alternative forced choice low-contrast detection task using channelized Hotelling observer: impact of radiation dose and reconstruction algorithms. Med Phys 40:041908CrossRef
14.
Zurück zum Zitat Geyer LL, Schoepf UJ, Meinel FG et al (2015) State of the art: iterative CT reconstruction techniques. Radiology 276:339–357CrossRef Geyer LL, Schoepf UJ, Meinel FG et al (2015) State of the art: iterative CT reconstruction techniques. Radiology 276:339–357CrossRef
15.
Zurück zum Zitat Noël P, Fingerle A, Renger B et al (2011) Initial performance characterization of a clinical noise-suppressing reconstruction algorithm for MDCT. AJR Am J Roentgenol 197:1404–1409CrossRef Noël P, Fingerle A, Renger B et al (2011) Initial performance characterization of a clinical noise-suppressing reconstruction algorithm for MDCT. AJR Am J Roentgenol 197:1404–1409CrossRef
16.
Zurück zum Zitat Mehta D, Thomspon R, Morton T et al (2013) Iterative model reconstruction: simultaneously lowered computed tomography radiation dose and improved image quality. Med Phys Int 1:147–155 Mehta D, Thomspon R, Morton T et al (2013) Iterative model reconstruction: simultaneously lowered computed tomography radiation dose and improved image quality. Med Phys Int 1:147–155
17.
Zurück zum Zitat Boone JM, Strauss KJ, Cody DD et al (2011) Report of AAPM Task Group 204: size-specific dose estimates (SSDE) in pediatric and adult body CT examinations. American Association of Physicists in Medicine, AlexandriaCrossRef Boone JM, Strauss KJ, Cody DD et al (2011) Report of AAPM Task Group 204: size-specific dose estimates (SSDE) in pediatric and adult body CT examinations. American Association of Physicists in Medicine, AlexandriaCrossRef
18.
Zurück zum Zitat Green D, Swets J (1966) Signal detection theory and psychophysics. Wiley, New York Green D, Swets J (1966) Signal detection theory and psychophysics. Wiley, New York
19.
Zurück zum Zitat Christianson O, Chen JJ, Yang Z et al (2015) An improved index of image quality for task-based performance of CT iterative reconstruction across three commercial implementations. Radiology 275:725–734CrossRef Christianson O, Chen JJ, Yang Z et al (2015) An improved index of image quality for task-based performance of CT iterative reconstruction across three commercial implementations. Radiology 275:725–734CrossRef
20.
Zurück zum Zitat American College of Radiology (2018) Executive summary report, Jan-Jun 2018, National Radiology Data Registry, Dose Index Registry. ACR, Reston American College of Radiology (2018) Executive summary report, Jan-Jun 2018, National Radiology Data Registry, Dose Index Registry. ACR, Reston
21.
Zurück zum Zitat Ryu YJ, Choi YH, Cheon JE et al (2016) Knowledge-based iterative model reconstruction: comparative image quality and radiation dose with a pediatric computed tomography phantom. Pediatr Radiol 46:303–315CrossRef Ryu YJ, Choi YH, Cheon JE et al (2016) Knowledge-based iterative model reconstruction: comparative image quality and radiation dose with a pediatric computed tomography phantom. Pediatr Radiol 46:303–315CrossRef
22.
Zurück zum Zitat Eck BL, Fahmi R, Brown KM et al (2015) Computational and human observer image quality evaluation of low dose, knowledge-based CT iterative reconstruction. Med Phys 42:6098–6111CrossRef Eck BL, Fahmi R, Brown KM et al (2015) Computational and human observer image quality evaluation of low dose, knowledge-based CT iterative reconstruction. Med Phys 42:6098–6111CrossRef
23.
Zurück zum Zitat Singh S, Kaira MK, Shenoy-Bhangle AS et al (2012) Radiation dose reduction with hybrid iterative reconstruction for pediatric CT. Radiology 263:537–546CrossRef Singh S, Kaira MK, Shenoy-Bhangle AS et al (2012) Radiation dose reduction with hybrid iterative reconstruction for pediatric CT. Radiology 263:537–546CrossRef
24.
Zurück zum Zitat Gay F, Pavia Y, Pierrat N et al (2014) Dose reduction with adaptive statistical iterative reconstruction for paediatric CT: phantom study and clinical experience on chest and abdomen CT. Eur Radiol 24:102–111CrossRef Gay F, Pavia Y, Pierrat N et al (2014) Dose reduction with adaptive statistical iterative reconstruction for paediatric CT: phantom study and clinical experience on chest and abdomen CT. Eur Radiol 24:102–111CrossRef
25.
Zurück zum Zitat Smith EA, Dillman JR, Goodsitt MM et al (2014) Model-based iterative reconstruction: effect on patient radiation dose and image quality in pediatric body CT. Radiology 270:526–534CrossRef Smith EA, Dillman JR, Goodsitt MM et al (2014) Model-based iterative reconstruction: effect on patient radiation dose and image quality in pediatric body CT. Radiology 270:526–534CrossRef
26.
Zurück zum Zitat Lee M, Kim MJ, Han KH et al (2015) Half-dose abdominal CT with sinogram-affirmed iterative reconstruction technique in children — comparison with full-dose CT with filtered back projection. Pediatr Radiol 45:188–193CrossRef Lee M, Kim MJ, Han KH et al (2015) Half-dose abdominal CT with sinogram-affirmed iterative reconstruction technique in children — comparison with full-dose CT with filtered back projection. Pediatr Radiol 45:188–193CrossRef
27.
Zurück zum Zitat Solomon J, Samei E (2014) Quantum noise properties of CT images with anatomical textured backgrounds across reconstruction algorithms: FBP and SAFIRE. Med Phys 41:091908CrossRef Solomon J, Samei E (2014) Quantum noise properties of CT images with anatomical textured backgrounds across reconstruction algorithms: FBP and SAFIRE. Med Phys 41:091908CrossRef
28.
Zurück zum Zitat Racine D, Ba AH, Ott JG et al (2016) Objective assessment of low contrast detectability in computed tomography with channelized Hotelling observer. Phys Med 32:76–83CrossRef Racine D, Ba AH, Ott JG et al (2016) Objective assessment of low contrast detectability in computed tomography with channelized Hotelling observer. Phys Med 32:76–83CrossRef
29.
Zurück zum Zitat Tseng HW, Fan J, Kupinski MA et al (2014) Assessing image quality and dose reduction of a new X-ray computed tomography iterative reconstruction algorithm using model observers. Med Phys 41:07190 Tseng HW, Fan J, Kupinski MA et al (2014) Assessing image quality and dose reduction of a new X-ray computed tomography iterative reconstruction algorithm using model observers. Med Phys 41:07190
Metadaten
Titel
Evaluation of low-contrast detectability for iterative reconstruction in pediatric abdominal computed tomography: a phantom study
verfasst von
Nicholas Rubert
Richard Southard
Susan M. Hamman
Ryan Robison
Publikationsdatum
09.11.2019
Verlag
Springer Berlin Heidelberg
Erschienen in
Pediatric Radiology / Ausgabe 3/2020
Print ISSN: 0301-0449
Elektronische ISSN: 1432-1998
DOI
https://doi.org/10.1007/s00247-019-04561-8

Weitere Artikel der Ausgabe 3/2020

Pediatric Radiology 3/2020 Zur Ausgabe

Hermes

Hermes

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.