Evaluation of pressure-controlled mammography compression paddles with respect to force-controlled compression paddles in clinical practice

verfasst von: C. R. L. P. N. Jeukens, T. van Dijk, C. Berben, J. E. Wildberger, M. B. I. Lobbes

Erschienen in: European Radiology | Ausgabe 5/2019

Abstract

Objectives

To reduce pain and discomfort associated with breast compression in mammography, a pressure-controlled compression paddle was recently introduced.
The objective was to evaluate the pressure-controlled paddle by comparing it to the standard force-controlled paddle.

Methods

Differences of compressed breast thickness (CBT), compression force, compression pressure, and average glandular dose (AGD) between annual follow-up full-field digital mammography exams of 3188 patients were retrospectively examined. Two groups were compared: (1) force-force group (FF-group), both examinations were performed with the force-controlled paddle, and (2) force-pressure group (FP-group), only the follow-up examination was performed with the pressure-controlled paddle.
In an additional group of patients, pain scores on a scale of 0 (no pain at all) to 10 (worst pain imaginable) were evaluated prospectively (n = 343) who were randomly assigned to either paddle.

Results

Median differences between follow-up exams in CBT, compression force, compression pressure, and AGD were for the FF- and FP-group respectively − 1.0 vs 0.0 mm (p < 0.001); 0.0 vs − 1.0 daN (p = 0.002); − 1.0 vs − 0.5 kPa (p = 0.005); and 0.05 vs − 0.02 mGy (p < 0.001). These differences were, although statistically significant, clinically non-relevant (defined as ΔCBT > ± 2 mm; Δforce > ± 2 daN; Δpressure > ± 1 kPa and ΔAGD > ± 0.1 mGy). The subanalysis dividing CBT into five categories revealed similar results. The median [interquartile range] pain scores were 6 [3, 7] and 5 [3, 7] for the force-controlled and pressure-controlled paddle, respectively, which was not significantly different (p = 0.113).

Conclusions

We observed no clinically relevant differences in CBT, compression force, compression pressure, AGD, or pain score between the force- and pressure-controlled paddle. As such, we found no basis for preferring one paddle over the other.

Key Points

The pressure-controlled paddle did not show any clinically relevant changes in breast compression parameters compared to the force-controlled paddle.
The pressure-controlled paddle did not lead to significant reduction in pain scores indicated by the patients compared to the force-controlled paddle.
A large variation in compression force and compression pressure was observed in mammography exams for the both the force- and pressure-controlled compression paddle.
Begleitmaterial
Hinweise

Electronic supplementary material

The online version of this article (https://​doi.​org/​10.​1007/​s00330-018-5953-6) contains supplementary material, which is available to authorized users.
Abkürzungen
AGD
Average glandular dose
BI-RADS
Breast imaging-reporting and data system
BRCA
Breast cancer
CBT
Compresed breast thickness
CC-view
Craniocaudal view
DICOM
Digital imaging and communications in medicine
EUREF
The European Network of Reference Centres for Breast Cancer Screening
FF-group
Force-force group
FP-group
Force-pressure group
FFDM
Full-field digital mammography
IQR
Interquartile range
NRS
Numeric Rating Scale
PACS
Picture archiving and communication system
SPSS
Statistical Package for the Social Sciences

Introduction

Mammography is currently the worldwide routine examination for breast cancer detection. During a mammographic exam, the breast is compressed to improve image quality and to decrease radiation dose [1]. However, breast compression comes at the price of discomfort and pain. Hence, efforts have been undertaken to reduce this drawback, such as pain medication prior to the exam, providing patients with information, breast cushions, and curved paddles [2, 3].
Recently, a new method was developed using a pressure-controlled compression paddle [4, 5]. In standard practice, a force-controlled paddle compresses the breast until a target force is reached, irrespective of the breast size. The pressure-controlled paddle compresses the breast until a target pressure is obtained. The pressure is determined as the force exerted divided by the breast-paddle contact area. Thus, compressing until a target pressure leads to a lower compression force for smaller breasts having a smaller contact area than for larger breasts. This would reduce discomfort and pain for the smaller breasts, as it has been shown that the tolerance of compression force decreases with decreasing contact area [5, 6].
Previous studied showed [4, 7] that the pressure-controlled paddle resulted in a more standardized compression pressure, a slightly increased compressed breast thickness (CBT), and lower pain experience without affecting image quality or increasing the average glandular dose (AGD). These studies were performed in an experimental setting using a well-defined compression protocol in which the technician was instructed to compress until a customized display indicated “100% compression,” being either the target force or pressure, for which the technician was blinded.
However, it is known that in everyday clinical practice, being unblinded for the equipment used and less strictly protocolized, compression force varies considerably [8, 9]. The aim of this study was to evaluate the pressure-controlled paddle in the everyday clinical practice of our institution. In case the pressure-controlled paddle would lead to pain reduction, this would be reflected in changes of objective parameters such as CBT, compression force and pressure, and AGD. Therefore, primary study outcomes were differences in these parameters of subsequent follow-up exams using either both force-controlled paddles or first time force-controlled, second time pressure-controlled paddle. Furthermore, a primary study outcome was the subjective pain score of an examination using either the pressure- or force-controlled paddle.

Materials and methods

Study population

FFDM exams performed at our institution for annual follow-up (indications, e.g., family predisposition, BRCA1/2 gene mutation carriers, breast cancer history) were retrospectively retrieved from the Picture Archiving and Communication System (PACS) (period 01/2014–12/2016). All examinations were randomly assigned to either of the two identical mammography units (Senographe Essential, GE Healthcare) undergoing equal quality control testing twice a year. Consecutive follow-up examinations form a so-called follow-up examination pair. CC-views of both breasts, or a single CC-view in case only one breast was examined, were included. Inclusion criteria were time between follow up ≤ 24 months; BI-RADS score was 1 (negative) or 2 (definitely benign), and the same for both follow-up exams (i.e., no significant changes in the breast occurred during follow-up).

Pressure-controlled and force-controlled paddle

The pressure-controlled paddle (Sensitive Sigma Paddle, Sigmascreening) was installed on one mammography unit in February 2016. All 15 mammography technicians received the same application training by the vendor. After an introduction period, the paddle was implemented and exclusively used on one unit, while the second unit used the force-controlled paddle. The protocol for breast compression using a rigid force-controlled paddle was as follows: while carefully positioning the breast, the compression paddle is lowered until a force of 17–18 daN was reached or until the patient indicated intolerable pain. With the pressure-controlled paddle, the breast was compressed until the target pressure of 10 kPa was reached. Eight indicator LEDs on the paddle itself indicate the pressure qualitatively. The target pressure is reached when six LED’s light up.
The technicians were randomly working on both units. Starting from May 2016, exams performed with the pressure-controlled paddle were included.

Data collection

The compression force, CBT, tube potential, tube current, anode and filter material, patient age, exam date, and laterality, were retrieved from the DICOM headers. The AGD was calculated for each exam by an in-house written software script (Python, Python Software Foundation) using the Dance model [10, 11] according to EUREF guidelines [1, 12].
The compression pressure could not be obtained directly, but it was determined by dividing the compression force by the breast-paddle contact area of the breast. The latter was estimated from the images using an in-house written script (Matlab, MathWorks) following the method described by de Groot et al [13].

Data analysis

The follow-up examination pairs were divided in two groups: (1) both examinations being performed with the force-controlled paddle (“force-force group”) and (2) first examination performed with the force-controlled paddle and the second with the pressure-controlled paddle (“force-pressure group”). For each follow-up pair, the differences (exam 2–exam 1) in the CBT (ΔCBT [mm]), compression force (Δforce [daN]), compression pressure (Δpressure [kPa]), and AGD (ΔAGD [mGy]) were calculated. The differences in the parameters for the force-force group were taken as a reference for the intra-group variation when recompressing the same breast. Clinically relevant differences were defined as ΔCBT > ± 2 mm; Δforce > ± 2 daN; Δpressure > ± 1 kPa; and ΔAGD > ± 0.1 mGy (see supplemental material for more details).

Pain score assessment

In an additional group of patients, pain experienced during the mammography exam was scored prospectively (period November 24, 2016–December 29, 2016). All consecutive patients, regardless of their indication, were randomly assigned to a mammography unit with either the pressure-controlled paddle or the force-controlled paddle. Patients were unaware that different paddles were used in our department, nor that a pain score would be assessed. After the complete examination, the technician asked the patient to score the intensity of pain experienced on a ten-point Numeric Rating Scale (NRS; 0 = no pain at all, 10 = worst pain imaginable). NRS pain scores ≥ 2 are considered to be clinically relevant [14, 15]. For this difference, to reach a power of (1-β) > 0.99 for an α = 0.05 and estimated standard deviation of ~ 2.5, a sample size of ≥ 2 × 58 patients would be required.

Medical ethics approval

The local medical ethics committee approved both the retrospective and prospective part of the study. The necessity to obtain written informed consent was waived.

Statistical analysis

Data was tested for normality (Shapiro-Wilk test), and reported as median and interquartile range (IQR), which is represented by [25th percentile, 75th percentile]. Differences between groups were tested using the Mann-Whitney U test. A subanalysis was performed by categorizing the exams according to mean CBT of each pair: ≤ 40 mm, 41–50 mm, 51–60 mm, 61–70 mm, ≥ 71 mm. Correlations were examined using a linear regression model and the correlation coefficient R2 was reported. All statistical analyses were performed using SPSS (IBM SPSS Statistics; version 23, IBM Corporation). Any p < 0.05 was considered statistically significant.

Results

In total, 3188 patients were included having 3848 pairs of follow-up exams as some patients had two or three follow-up exams in the inclusion period. Table 1 shows the characteristics of the force-force and force-pressure groups. No clinically relevant differences were found between the two groups although CBT, time between follow-up, and AGD were significantly different.
Table 1
Details on the patient population in the force-force and force-pressure group. Values are reported as median [IQR] unless specified otherwise. The p values of the Mann-Whithney U test are reported. The median [IQR] values reported for age, CBT, contact area, compression force, compression pressure, and AGD are calculated from the mean value of exams 1 and 2 for each exam pair
 
Force-force group
Force-pressure group
p value
Number of patients
2574
 
614
 
n.a.
Number of exam pairs
3234
 
614
 
n.a.
Age (years)
59
[52–67]
59
[53–68]
0.960
Time between follow-up (months)
12.2
[11.8–12.6]
12.2
[12.0–15.9]
* < 0.001
Mean number of follow-up exams [range]
1.2
[1–3]
1.0
[1]
n.a.
CBT (mm)
59
[50–67]
57
[49–65]
*0.001
Contact area (dm2)
0.62
[0.43–0.87]
0.63
[0.45–0.91]
0.123
Compression force (daN)
17
[16–18]
17
[16–18]
0.513
Compression pressure (kPa)
26
[19–38]
26
[18–37]
0.324
AGD (mGy)
1.64
[1.38–2.05]
1.57
[1.33–1.88]
* < 0.001
* Indicates significance
n.a. not applicable
The median ΔCBT, Δforce, Δpressure, and ΔAGD between follow-up exams were close to zero for both groups (Fig. 1). Although statistically significant, no clinical relevant differences in the median values between the two groups were observed (Table 2).
Table 2
Median [IQR] values of the differences between exam pairs for the force-force and force pressure group for the whole study cohort and for the CBT subgroups. The p values of the Mann-Whithney U test are reported
 
Force-force group
Force-pressure group
p value
All patients
 Number of exam pairs
3234
 
614
  
 ΔCBT (mm)
− 1.0
[− 6.0; 3.0]
0.0
[− 3.0; 4.0]
* < 0.001
 ΔForce (daN)
0.0
[− 2.0; 2.0]
− 1.0
[− 2.0; 1.0]
*0.002
 ΔPressure (kPa)
− 1.0
[− 5.1; 2.5]
− 0.5
[− 3.6; 2.6]
*0.005
 ΔAGD (mGy)
0.05
[− 0.18; 0.30]
− 0.02
[− 0.22; 0.12]
* < 0.001
CBT ≤ 40 mm
 Number of exam pairs
252
 
66
  
 ΔCBT (mm)
− 1.0
[− 5.0; 3.0]
0.0
[− 3.0; 3.0]
*0.030
 ΔForce (daN)
0.0
[− 3.0; 2.0]
− 1.0
[− 3.0; 1.0]
0.237
 ΔPressure (kPa)
− 1.2
[− 6.6; 2.5]
0.0
[− 3.5; 3.1]
0.078
 ΔAGD (mGy)
0.10
[− 0.12; 0.34]
0.01
[− 0.08; 0.11]
*0.010
CBT 41–50 mm
 Number of exam pairs
607
 
117
  
 ΔCBT (mm)
0.0
[− 3.0; 1.0]
0.0
[− 3.0; 3.0]
*0.048
 ΔForce (daN)
− 1.0
[− 2.0; 1.0]
− 1.0
[− 3.0; 1.0]
0.430
 ΔPressure (kPa)
− 1.5
[− 5.7; 2.2]
0.0
[− 4.4; 2.7]
*0.034
 ΔAGD (mGy)
0.07
[− 0.11; 0.28]
− 0.04
[− 0.17; 0.09]
* < 0.001
CBT 51–60 mm
 Number of exam pairs
939
 
192
  
 ΔCBT (mm)
− 1.0
[− 6.0; 3.0]
1.0
[− 3.0; 4.0]
*0.001
 ΔForce (daN)
0.0
[− 2.0; 2.0]
0.0
[− 2.0; 2.0]
0.839
 ΔPressure (kPa)
− 1.1
[− 4.6; 2.3]
− 0.2
[− 3.3; 3.1]
*0.010
 ΔAGD (mGy)
0.05
[− 0.14; 0.25]
− 0.05
[− 0.25; 0.09]
* < 0.001
CBT 61–70 mm
 Number of exam pairs
869
 
161
  
 ΔCBT (mm)
− 1.0
[− 6.0; 3.0]
0.0
[− 2.0; 5.0]
* < 0.001
 ΔForce (daN)
0.0
[− 2.0; 2.0]
− 1.0
[− 2.0; 1.0]
0.159
 ΔPressure (kPa)
− 0.7
[− 4.6; 2.4]
− 0.9
[− 3.3; 2.0]
0.415
 ΔAGD (mGy)
0.02
[− 0.24; 0.30]
0.01
[− 0.27; 0.15]
0.126
CBT > =71 mm
 Number of exam pairs
567
 
78
  
 ΔCBT (mm)
− 1.0
[− 7.0; 4.0]
0.0
[− 4.0; 4.3]
0.186
 ΔForce (daN)
0.0
[− 2.0; 2.0]
− 1.0
[− 3.0; 0.0]
* < 0.001
 ΔPressure (kPa)
− 0.8
[− 5.7; 3.2]
− 1.7
[− 4.0; 0.7]
0.172
 ΔAGD (mGy)
0.02
[− 0.36; 0.45]
0.01
[− 0.35; 0.25]
0.273
* Indicates significance
The subanalysis for CBT categories showed that the median ΔCBT, Δforce, Δpressure, and ΔAGD were close to zero across all CBT categories (Fig. 2). Although statistically significant in 10/20 instances, differences between the median values of both groups were clinically not relevant (Table 2).
Further analysis of the used compression force and pressure showed a large variation for both paddle types (Fig. 3), which may explain the observation that no clinical relevant median differences were found. As protocol dictated breast compression until a target force or pressure was obtained for the force- or pressure-controlled paddle respectively, it would have been expected that the applied force or pressure, respectively, were more or less constant irrespective of the CBT. However, no correlation was found between compression force or pressure and CBT (R2 ≤ 0.008). The large variation observed in Fig. 3 may also explain that for individual patients differences between the follow-up exams may reach clinically relevant levels. However, this is observed both for better or worse and in both the force-force and force-pressure group (Figs. 1 and 2).
The pain scores for exams performed with the force- (n = 165) or pressure-controlled paddle (n = 178) were not statistically different (p = 0.113, Fig. 4). The median [IQR] pain scores were 6 [3, 7] for the force-controlled paddle and 5 [3, 7] for the pressure-controlled paddle. The median [IQR] age and CBT was 57 [49–67] and 56 [49–65] years (p = 0.441) and 55 [46–65] and 57 [48–66] mm (p = 0.075) for the force- and pressure-controlled paddle group respectively.

Discussion

In our study, we evaluated the pressure-controlled compression paddle with respect to the conventional force-controlled compression paddle in the everyday clinical practice of our institution. In contrast to prior studies in controlled experimental settings, we could not observe any clinically relevant added value of the pressure-controlled paddles. As such, we found no basis for preferring one paddle over the other.
Multiple studies were published by a single group showing the potential of the pressure-controlled paddle: a more standardized compression was feasible and discomfort and pain could be reduced, while the AGD was similar and image quality was not affected [4, 7]. However, the studies were performed in a highly-standardized setting, which raises the question of how the pressure-controlled paddle performs in clinical practice, where more variation in breast compression is present. A retrospective study evaluating the pressure-controlled paddle in a clinical setting for n = 39 patients found that the compression pressure of the pressure-controlled paddle is only weakly dependent on the breast contact area as opposed to the compression pressure of the force-controlled paddle [16]. However, whether clinically relevant differences in compression parameters between follow-up exams in favor of the pressure-controlled paddle could be observed was not studied. Our observations show a large spread in compression force, CBT, and AGD between follow-up exams, which was also found by Mercer et al [8, 9]. Moreover, significant differences existed in these parameters between technicians performing the compression.
The EUREF guidelines [1] merely indicate that “The breast should be properly compressed, but no more than is necessary to achieve a good image quality.” A higher compression force would lead to a smaller breast thickness which is believed to be beneficial for image quality, reduction of motion artifacts, and AGD. Although, a study of Holland et al [17] showed in a retrospective analysis of screening mammograms that applying too much or too low pressure lead to a reduced sensitivity and specificity, respectively. In clinical practice, achieving optimal breast compression is a tailor-made process in which technicians consider several factors, such as the patient’s pain tolerance and the risk of motion artifacts, as well as their experience.
There is a complex interplay between the actual discomfort/pain experience of the patient and obtaining a good breast compression. Influencing factors are patient characteristics, such as breast size and density, pain tolerance, and prior existing pain [2, 18], but also proper adjustment of the bucky/detector height [19], number of prior acquisitions within the exam [6], providing information about the examination [2, 20], and psychological factors which can be influenced by the social interaction between patient and technician [20, 21]. This might suggest that repeated training of the technicians, not only in terms of proper positioning but also patient communication, may be an important influential factor in reducing discomfort and pain while maintaining high image quality. As we could not observe any beneficial effects for the pressure-controlled paddles in the clinical setting, we recommend to further investigate the causes for the discomfort/pain experience because this is the basis for seeking solutions. Also analyzing more deeply the observed spread in compression parameters would enable to optimally deploy the potential of pressure-controlled breast compression in clinical practice.
Our study has some limitations. First, the applied pressure was not directly provided by either paddle. Therefore, the pressure was calculated by estimating the contact area from the images, which may deviate from the actual contact area. Despite this limitation, the Δpressure analysis confirms the findings of the ΔCBT and Δforce analysis. Second, there could have been changes in a breast between subsequent exams. To circumvent this, the BI-RADS score was 1 or 2 and equal for both exams. Furthermore, sufficiently large groups were studied to minimize the influence of cases where changes in the breast did occur. Third, we choose not to include an evaluation of the image quality for both paddles. Not finding any differences in compression parameters, we did not expect to observe a difference in image quality, and if so, we would not be able to attribute this to the paddle-type used. Fourth, we performed a single-center study, which might have led to bias in the selection of patients and technicians. It would be valuable to investigate the effect of the pressure-controlled paddle in the clinical practice of other institutions.
In conclusion, the evaluation of the pressure-controlled paddle in clinical practice revealed no clinically relevant differences in CBT, compression force, compression pressure, and AGD in comparison to the force-controlled paddle. The pain score assessment revealed no significant differences between the two paddles. As such, we found no basis for preferring the pressure-controlled paddle over the force-controlled paddle in our clinical practice.

Compliance with ethical standards

Guarantor

The scientific guarantor of this publication is CRLPN Jeukens.

Conflict of interest

The authors of this manuscript declare no relationships with any companies, whose products or services may be related to the subject matter of the article.

Statistics and biometry

No complex statistical methods were necessary for this paper.
Written informed consent was waived by the Institutional Review Board.

Ethical approval

Institutional Review Board approval was obtained.

Methodology

• prospective/retrospective
• case-control study/cross sectional study
• performed at one institution
Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.

Unsere Produktempfehlungen

e.Med Interdisziplinär

Kombi-Abonnement

Für Ihren Erfolg in Klinik und Praxis - Die beste Hilfe in Ihrem Arbeitsalltag

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de.

e.Med Radiologie

Kombi-Abonnement

Mit e.Med Radiologie erhalten Sie Zugang zu CME-Fortbildungen des Fachgebietes Radiologie, den Premium-Inhalten der radiologischen Fachzeitschriften, inklusive einer gedruckten Radiologie-Zeitschrift Ihrer Wahl.

Anhänge

Electronic supplementary material

Literatur
1.
Zurück zum Zitat European Communities (2013) European Guidelines for Quality Assurance in Breast Cancer Screening and Diagnosis, 4th ed., Supp. Luxembourg, Luxembourg: Office for Official Publications of the European Communities. Available via http://www.euref.org/european-guidelines. Accessed 7 Nov 2018 European Communities (2013) European Guidelines for Quality Assurance in Breast Cancer Screening and Diagnosis, 4th ed., Supp. Luxembourg, Luxembourg: Office for Official Publications of the European Communities. Available via http://​www.​euref.​org/​european-guidelines. Accessed 7 Nov 2018
4.
Zurück zum Zitat de Groot JE, Branderhorst W, Grimbergen CA, den Heeten GJ, Broeders MJ (2015) Towards personalized compression in mammography: a comparison study between pressure- and force-standardization. Eur J Radiol 84:384–391 de Groot JE, Branderhorst W, Grimbergen CA, den Heeten GJ, Broeders MJ (2015) Towards personalized compression in mammography: a comparison study between pressure- and force-standardization. Eur J Radiol 84:384–391
5.
Zurück zum Zitat de Groot JE, Broeders MJ, Branderhorst W, den Heeten GJ, Grimbergen CA (2013) A novel approach to mammographic breast compression: improved standardization and reduced discomfort by controlling pressure instead of force. Med Phys. https://doi.org/10.1118/1.4812418 de Groot JE, Broeders MJ, Branderhorst W, den Heeten GJ, Grimbergen CA (2013) A novel approach to mammographic breast compression: improved standardization and reduced discomfort by controlling pressure instead of force. Med Phys. https://​doi.​org/​10.​1118/​1.​4812418
6.
Zurück zum Zitat Feder K, Grunert JH (2017) Is individualizing breast compression during mammography useful? – investigations of pain indications during mammography relating to compression force and surface area of the compressed breast. Rofo 189:39–48 Feder K, Grunert JH (2017) Is individualizing breast compression during mammography useful? – investigations of pain indications during mammography relating to compression force and surface area of the compressed breast. Rofo 189:39–48
7.
Zurück zum Zitat de Groot JE, Hopman IGM, van Lier MGJTB, Branderhorst W, Grimbergen CA, den Heeten GJ (2017) Pressure-standardised mammography does not affect visibility,contrast and sharpness of stable lesions. Eur J Radiol 86:289–295 de Groot JE, Hopman IGM, van Lier MGJTB, Branderhorst W, Grimbergen CA, den Heeten GJ (2017) Pressure-standardised mammography does not affect visibility,contrast and sharpness of stable lesions. Eur J Radiol 86:289–295
8.
Zurück zum Zitat Mercer CE, Hogg P, Szczepura K, Denton ERE (2013) Practitioner compression force variation in mammography: a 6-year study. Radiography 19:200–206CrossRef Mercer CE, Hogg P, Szczepura K, Denton ERE (2013) Practitioner compression force variation in mammography: a 6-year study. Radiography 19:200–206CrossRef
9.
Zurück zum Zitat Mercer CE, Szczepura K, Kelly J et al (2015) A 6-year study of mammographic compression force: practitioner variability within and between screening sites. Radiography 21:68–73CrossRef Mercer CE, Szczepura K, Kelly J et al (2015) A 6-year study of mammographic compression force: practitioner variability within and between screening sites. Radiography 21:68–73CrossRef
10.
Zurück zum Zitat Dance DR, Skinner CL, Young KC, Beckett JR, Kotre CJ (2000) Additional factors for the estimation of mean glandular breast dose using the UK mammography dosimetry protocol. Phys Med Biol 45:3225–3240CrossRefPubMed Dance DR, Skinner CL, Young KC, Beckett JR, Kotre CJ (2000) Additional factors for the estimation of mean glandular breast dose using the UK mammography dosimetry protocol. Phys Med Biol 45:3225–3240CrossRefPubMed
11.
Zurück zum Zitat Dance DR, Young KC, van Engen RE (2009) Further factors for the estimation of mean glandular dose using the United Kingdom, European and IAEA breast dosimetry protocols. Phys Med Biol 54:4361–4372CrossRefPubMed Dance DR, Young KC, van Engen RE (2009) Further factors for the estimation of mean glandular dose using the United Kingdom, European and IAEA breast dosimetry protocols. Phys Med Biol 54:4361–4372CrossRefPubMed
12.
Zurück zum Zitat Lalji UC, Jeukens CR, Houben I et al (2015) Evaluation of low-energy contrast-enhanced spectral mammography images by comparing them to full-field digital mammography using EUREF image quality criteria. Eur Radiol 25:2813–2820 Lalji UC, Jeukens CR, Houben I et al (2015) Evaluation of low-energy contrast-enhanced spectral mammography images by comparing them to full-field digital mammography using EUREF image quality criteria. Eur Radiol 25:2813–2820
14.
Zurück zum Zitat Hawker GA, Mian S, Kendzerska T, French M (2011) Measures of adult pain. Arthritis Care Res (Hoboken) 63:S240–S252 Hawker GA, Mian S, Kendzerska T, French M (2011) Measures of adult pain. Arthritis Care Res (Hoboken) 63:S240–S252
15.
Zurück zum Zitat Ostelo RW, Deyo RA, Stratford P et al (2008) Interpreting Change Scores for Pain and Functional Status in Low Back Pain; Towards International Consensus Regarding Minimal Important Change. Spine (Phila Pa 1976) 33:90–94 Ostelo RW, Deyo RA, Stratford P et al (2008) Interpreting Change Scores for Pain and Functional Status in Low Back Pain; Towards International Consensus Regarding Minimal Important Change. Spine (Phila Pa 1976) 33:90–94
16.
Zurück zum Zitat den Boer D,, Dam-Vervloet LAJ, Boomsma MF, de Boer E, van Dalen JA, Poot L (2018) Clinical validation of a pressure-standardized compression mammography system. Eur J Radiol 105:251–254 den Boer D,, Dam-Vervloet LAJ, Boomsma MF, de Boer E, van Dalen JA, Poot L (2018) Clinical validation of a pressure-standardized compression mammography system. Eur J Radiol 105:251–254
17.
Zurück zum Zitat Holland K, Sechopoulos I, Mann RM, den Heeten GJ, van Gils CH, Karssemeijer N (2017) Influence of breast compression pressure on the performance of population-based mammography-screening. Breast Cancer Res 19:126 Holland K, Sechopoulos I, Mann RM, den Heeten GJ, van Gils CH, Karssemeijer N (2017) Influence of breast compression pressure on the performance of population-based mammography-screening. Breast Cancer Res 19:126
18.
Zurück zum Zitat de Groot JE, Broeders MJ, Grimbergen CA, den Heeten GJ (2015) Pain-preventing strategies in mammography: an observational study of simultaneously recorded pain and breast mechanics throughout the entire breast compression cycle. BMC Womens Health 15:26–34 de Groot JE, Broeders MJ, Grimbergen CA, den Heeten GJ (2015) Pain-preventing strategies in mammography: an observational study of simultaneously recorded pain and breast mechanics throughout the entire breast compression cycle. BMC Womens Health 15:26–34
19.
Zurück zum Zitat Branderhorst W, de Groot JE, Neeter LM et al (2016) Force balancing in mammographic compression. Med Phys 43:518–527 Branderhorst W, de Groot JE, Neeter LM et al (2016) Force balancing in mammographic compression. Med Phys 43:518–527
20.
Zurück zum Zitat Van Goethem M, Mortelmans D, Bruyninckx E et al (2003) Influence of the radiographer on the pain felt during mammography. Eur Radiol 13:2384–2389CrossRefPubMed Van Goethem M, Mortelmans D, Bruyninckx E et al (2003) Influence of the radiographer on the pain felt during mammography. Eur Radiol 13:2384–2389CrossRefPubMed
21.
Zurück zum Zitat Murphy F, Nightingale J, Hogg P, Robinson L, Seddon D, Mackay S (2015) Compression force behaviours: an exploration of the beliefs and values influencing the application of breast compression during screening mammography. Radiography 21:30–35CrossRef Murphy F, Nightingale J, Hogg P, Robinson L, Seddon D, Mackay S (2015) Compression force behaviours: an exploration of the beliefs and values influencing the application of breast compression during screening mammography. Radiography 21:30–35CrossRef
Metadaten
Titel
Evaluation of pressure-controlled mammography compression paddles with respect to force-controlled compression paddles in clinical practice
verfasst von
C. R. L. P. N. Jeukens
T. van Dijk
C. Berben
J. E. Wildberger
M. B. I. Lobbes
Publikationsdatum
07.01.2019
Verlag
Springer Berlin Heidelberg
Erschienen in
European Radiology / Ausgabe 5/2019
Print ISSN: 0938-7994
Elektronische ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-018-5953-6

Weitere Artikel der Ausgabe 5/2019

European Radiology 5/2019 Zur Ausgabe

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.