Ex vivo orbital volumetry using stereology and CT imaging: A comparison with manual planimetry

verfasst von: Georgios Bontzos, Michael Mazonakis, Efrosini Papadaki, Thomas G. Maris, Styliani Blazaki, Eleni E. Drakonaki, Efstathios T. Detorakis

Erschienen in: European Radiology | Ausgabe 3/2019

Einloggen, um Zugang zu erhalten

Abstract

Objectives

To evaluate the applicability of stereology and planimetry in orbital volume measurements using computed tomography (CT) and to compare the results between the two measurements.

Methods

Experimental study using sheep craniums for CT imaging. Water filling measurements were performed, as the validation technique. Quantification techniques were also evaluated in five human subjects. To examine the proportion of agreement among measurements, we tested intra- and inter-observer agreement.

Results

For stereology customization, a 1/8 systematic sampling scheme was considered as optimal; this resulted in a low coefficient of error (2.59 %) and low measurement time (1.9 mins). In sheep craniums, mean volume measured by water displacement, planimetry and stereology was 17.81 ± 0.59 cm3, 17.87 ± 0.68 cm3 and 17.54 ± 0.49 cm3, respectively. Total volumes, obtained by stereology, were highly correlated with the water-filling method (r=0.893; p = 0.001) and a paired t-test showed significant difference between methods (t=3.047; p = 0.014). Planimetry results displayed a high correlation with the water-filling method (r=0.957; p ≈ 0.001) but no statistically significant difference was found (p = 0.154). Mean difference using planimetry and stereology was 0.332 ± 0.322 cm3. In human subjects, using stereology, the estimated volume ranged between 18.57 cm3 and 19.27 cm3, and the mean orbital volume was 19.05 ± 0.50 cm3 with CE=3.75 ± 0.16 %. Mean measure time was 2.1 ± 0.1 mins.

Conclusions

Stereological measurements were superior to manual planimetry in terms of user effort and time spent. Stereology sampling of 1/8 was successfully applied in human subjects and yielded a strong correlation with manual planimetry.

Key Points

• Stereology can be applied to measure the orbital volume using computed tomography.
• Stereological measurements display high correlation with gold standard planimetry and combine low coefficient of error (2.59%) with low measurement time (1.9 min).
• Stereology is superior in terms of user effort and time spent.
Literatur
1.
Zurück zum Zitat Alinasab B, Beckman MO, Pansell T, Abdi S, Westermark AH, Stjärne P (2011) Relative difference in orbital volume as an indication for surgical reconstruction in isolated orbital floor fractures. Craniomaxillofac Trauma Reconstr. 4:203–212 Alinasab B, Beckman MO, Pansell T, Abdi S, Westermark AH, Stjärne P (2011) Relative difference in orbital volume as an indication for surgical reconstruction in isolated orbital floor fractures. Craniomaxillofac Trauma Reconstr. 4:203–212
2.
Zurück zum Zitat Imai K, Fujimoto T, Takahashi M, Maruyama Y, Yamaguchi K (2013) Preoperative andpostoperative orbital volume in patients with Crouzon and Apert syndrome. J Craniofac Surg 24:191–194 Imai K, Fujimoto T, Takahashi M, Maruyama Y, Yamaguchi K (2013) Preoperative andpostoperative orbital volume in patients with Crouzon and Apert syndrome. J Craniofac Surg 24:191–194
3.
Zurück zum Zitat Schiff BA, McMullen CP, Farinhas J et al (2015) Use of computed tomography to assess volume change after endoscopic orbital decompression for Graves' ophthalmopathy. Am J Otolaryngol 36:729–735CrossRefPubMed Schiff BA, McMullen CP, Farinhas J et al (2015) Use of computed tomography to assess volume change after endoscopic orbital decompression for Graves' ophthalmopathy. Am J Otolaryngol 36:729–735CrossRefPubMed
4.
Zurück zum Zitat Sung YS, Chung CM, Hong IP (2013) The correlation between the degree of enophthalmos and the extent of fracture in medial orbital wall fracture left untreated for over six months: a retrospective analysis of 81 cases at a single institution. Arch Plast Surg. 40:335–340CrossRefPubMedPubMedCentral Sung YS, Chung CM, Hong IP (2013) The correlation between the degree of enophthalmos and the extent of fracture in medial orbital wall fracture left untreated for over six months: a retrospective analysis of 81 cases at a single institution. Arch Plast Surg. 40:335–340CrossRefPubMedPubMedCentral
5.
Zurück zum Zitat Oh TS, Jeong WS, Chang TJ, Koh KS, Choi JW (2016) Customized orbital wall reconstruction using three-dimensionally printed rapid prototype model in patients with orbital wall fracture. J Craniofac Surg. 27:2020–2024 Oh TS, Jeong WS, Chang TJ, Koh KS, Choi JW (2016) Customized orbital wall reconstruction using three-dimensionally printed rapid prototype model in patients with orbital wall fracture. J Craniofac Surg. 27:2020–2024
6.
Zurück zum Zitat Forbes G, Gorman CA, Gehring D, Baker HL Jr (1983) Computer analysis of orbital fat and muscle volumes in Graves ophthalmopathy. AJNR Am J Neuroradiol. 4:737–740 Forbes G, Gorman CA, Gehring D, Baker HL Jr (1983) Computer analysis of orbital fat and muscle volumes in Graves ophthalmopathy. AJNR Am J Neuroradiol. 4:737–740
7.
Zurück zum Zitat Forbes G, Gehring DG, Gorman CA, Brennan MD, Jackson IT (1985) Volume measurements of normal orbital structures by computed tomographic analysis. AJR Am J Roentgenol. 145:149–154 Forbes G, Gehring DG, Gorman CA, Brennan MD, Jackson IT (1985) Volume measurements of normal orbital structures by computed tomographic analysis. AJR Am J Roentgenol. 145:149–154
8.
Zurück zum Zitat McGurk M, Whitehouse RW, Taylor PM, Swinson B (1992) Orbital volume measured by a low-dose CT scanning technique. Dentomaxillofac Radiol. 21:70–72 McGurk M, Whitehouse RW, Taylor PM, Swinson B (1992) Orbital volume measured by a low-dose CT scanning technique. Dentomaxillofac Radiol. 21:70–72
9.
Zurück zum Zitat Lutzemberger L, Salvetti O (1998) Volumetric analysis of CT orbital images. Med Biol Eng Comput. 36661-666 Lutzemberger L, Salvetti O (1998) Volumetric analysis of CT orbital images. Med Biol Eng Comput. 36661-666
10.
Zurück zum Zitat Deveci M, Oztürk S, Sengezer M, Pabuşcu Y (2000) Measurement of orbital volume by a 3-dimensional software program: an experimental study. J Oral Maxillofac Surg. 58:645–648 Deveci M, Oztürk S, Sengezer M, Pabuşcu Y (2000) Measurement of orbital volume by a 3-dimensional software program: an experimental study. J Oral Maxillofac Surg. 58:645–648
11.
Zurück zum Zitat Warfield SK, Kaus M, Jolesz FA, Kikinis R (2000) Adaptive, template moderated, spatially varying statistical classification. Med Image Anal. 4:43–55 Warfield SK, Kaus M, Jolesz FA, Kikinis R (2000) Adaptive, template moderated, spatially varying statistical classification. Med Image Anal. 4:43–55
12.
Zurück zum Zitat Bartling SH, Majdani O, Gupta R et al (2007) Large scan field, high spatial resolution flat-panel detector based volumetric CT of the whole human skull base and for maxillofacial imaging. Dentomaxillofac Radiol. 36:317–327CrossRefPubMed Bartling SH, Majdani O, Gupta R et al (2007) Large scan field, high spatial resolution flat-panel detector based volumetric CT of the whole human skull base and for maxillofacial imaging. Dentomaxillofac Radiol. 36:317–327CrossRefPubMed
13.
Zurück zum Zitat Bijlsma WR, Mourits MP (2006) Radiologic measurement of extraocular muscle volumes in patients with Graves' orbitopathy: a review and guideline. Orbit. 25:83–91CrossRefPubMed Bijlsma WR, Mourits MP (2006) Radiologic measurement of extraocular muscle volumes in patients with Graves' orbitopathy: a review and guideline. Orbit. 25:83–91CrossRefPubMed
14.
Zurück zum Zitat Cooper WC (1985) A method for volume determination of the orbit and its contents by high resolution axial tomography and quantitative digital image analysis. Trans Am Ophthalmol Soc. 83:546–609PubMedPubMedCentral Cooper WC (1985) A method for volume determination of the orbit and its contents by high resolution axial tomography and quantitative digital image analysis. Trans Am Ophthalmol Soc. 83:546–609PubMedPubMedCentral
15.
Zurück zum Zitat Acer N, Sahin B, Ergür H, Basaloglu H, Ceri NG (2009) Stereological estimation of the orbital volume: a criterion standard study. J Craniofac Surg. 20:921–925 Acer N, Sahin B, Ergür H, Basaloglu H, Ceri NG (2009) Stereological estimation of the orbital volume: a criterion standard study. J Craniofac Surg. 20:921–925
16.
Zurück zum Zitat Mazonakis M, Karampekios S, Damilakis J, Voloudaki A, Gourtsoyiannis N (2004) Stereological estimation of total intracranial volume on CT images. Eur Radiol. 14:1285–1290 Mazonakis M, Karampekios S, Damilakis J, Voloudaki A, Gourtsoyiannis N (2004) Stereological estimation of total intracranial volume on CT images. Eur Radiol. 14:1285–1290
17.
Zurück zum Zitat Unal B, Kara A, Aksak S, Unal D (2010) A stereological assessment method for estimating the surface area of cycloids. Eurasian J Med 42:66–73 Unal B, Kara A, Aksak S, Unal D (2010) A stereological assessment method for estimating the surface area of cycloids. Eurasian J Med 42:66–73
18.
Zurück zum Zitat Bilgic S, Sahin B, Sonmez OF et al (2005) A new approach for the estimation of intervertebral disc volume using the Cavalieri principle and computed tomography images. Clin Neurol Neurosurg 107:282–288CrossRefPubMed Bilgic S, Sahin B, Sonmez OF et al (2005) A new approach for the estimation of intervertebral disc volume using the Cavalieri principle and computed tomography images. Clin Neurol Neurosurg 107:282–288CrossRefPubMed
19.
Zurück zum Zitat Gundersen HJ, Jensen EB (1987) The efficiency of systematic sampling in stereology and its prediction. J Microsc. 147:229–263CrossRefPubMed Gundersen HJ, Jensen EB (1987) The efficiency of systematic sampling in stereology and its prediction. J Microsc. 147:229–263CrossRefPubMed
20.
Zurück zum Zitat Roberts N, Puddephat MJ, McNulty V (2000) The benefit of stereology for quantitative radiology. Br J Radiol. 73:679–697CrossRefPubMed Roberts N, Puddephat MJ, McNulty V (2000) The benefit of stereology for quantitative radiology. Br J Radiol. 73:679–697CrossRefPubMed
21.
Zurück zum Zitat Mazonakis M, Damilakis J, Varveris H (1998) Bladder and rectum volume estimations using CT and stereology. Comput Med Imaging Graph. 22:195–201CrossRefPubMed Mazonakis M, Damilakis J, Varveris H (1998) Bladder and rectum volume estimations using CT and stereology. Comput Med Imaging Graph. 22:195–201CrossRefPubMed
22.
Zurück zum Zitat Furuta M (2001) Measurement of orbital volume by computed tomography: especially on the growth of the orbit. Jpn J Ophthalmol. 45:600–606CrossRefPubMed Furuta M (2001) Measurement of orbital volume by computed tomography: especially on the growth of the orbit. Jpn J Ophthalmol. 45:600–606CrossRefPubMed
23.
Zurück zum Zitat Sugamata A, Yoshizawa N (2010) Clinical analysis of orbital blowout fractures caused by a globe-to-wall contact mechanism. J Plast Surg Hand Surg. 44:278–281CrossRefPubMed Sugamata A, Yoshizawa N (2010) Clinical analysis of orbital blowout fractures caused by a globe-to-wall contact mechanism. J Plast Surg Hand Surg. 44:278–281CrossRefPubMed
24.
Zurück zum Zitat Sahin B, Emirzeoglu M, Uzun A et al (2003) Unbiased estimation of the liver volume by the Cavalieri principle using magnetic resonance images. Eur J Radiol. 47:164–170CrossRefPubMed Sahin B, Emirzeoglu M, Uzun A et al (2003) Unbiased estimation of the liver volume by the Cavalieri principle using magnetic resonance images. Eur J Radiol. 47:164–170CrossRefPubMed
25.
Zurück zum Zitat Mazonakis M, Pagonidis K, Damilakis J (2011) Right ventricular volumes and ejection fraction by MR imaging and stereology: comparison with standard image analysis method. Clin Anat. 24:868–873CrossRefPubMed Mazonakis M, Pagonidis K, Damilakis J (2011) Right ventricular volumes and ejection fraction by MR imaging and stereology: comparison with standard image analysis method. Clin Anat. 24:868–873CrossRefPubMed
26.
Zurück zum Zitat Mazonakis M, Stratakis J, Damilakis J (2015) Efficient stereological approaches for the volumetry of a normal or enlarged spleen from MDCT images. Eur Radiol. 25:1761–1767CrossRefPubMed Mazonakis M, Stratakis J, Damilakis J (2015) Efficient stereological approaches for the volumetry of a normal or enlarged spleen from MDCT images. Eur Radiol. 25:1761–1767CrossRefPubMed
27.
Zurück zum Zitat Emirzeoglu M, Sahin B, Selcuk MB, Kaplan S (2005) The effects of section thickness on the estimation of liver volume by the Cavalieri principle using computed tomography images. Eur J Radiol. 56:391–397 Emirzeoglu M, Sahin B, Selcuk MB, Kaplan S (2005) The effects of section thickness on the estimation of liver volume by the Cavalieri principle using computed tomography images. Eur J Radiol. 56:391–397
Metadaten
Titel
Ex vivo orbital volumetry using stereology and CT imaging: A comparison with manual planimetry
verfasst von
Georgios Bontzos
Michael Mazonakis
Efrosini Papadaki
Thomas G. Maris
Styliani Blazaki
Eleni E. Drakonaki
Efstathios T. Detorakis
Publikationsdatum
22.08.2018
Verlag
Springer Berlin Heidelberg
Erschienen in
European Radiology / Ausgabe 3/2019
Print ISSN: 0938-7994
Elektronische ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-018-5691-9

Weitere Artikel der Ausgabe 3/2019

European Radiology 3/2019 Zur Ausgabe

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.