Expert consensus regarding standardization of sample preparation for clotting time assays

verfasst von: Masahiro Ieko, Yutaka Komiyama, Satoshi Yamazaki, Hisako Katagiri, Chisato Shimazu, Sumiyoshi Naito, Masato Matsuda, Yoichi Yuki, Kazuyoshi Miyata, Rika Arai, Osamu Kumano, Mari Emmi, Tatsuya Yoshida, Hideaki Tanaka, Toshihiko Okuhara, Tetsuji Yamashita, Yasuji Sekine, Masaru Homma, Yohko Kawai, Kaoru Tohyama, Masahiro Ieko, Yutaka Komiyama, Satoshi Yamazaki, Hisako Katagiri, Chisato Shimazu, Sumiyoshi Naito, Masato Matsuda, Yoichi Yuki, Kazuyoshi Miyata, Rika Arai, Osamu Kumano, Mari Emmi, Tatsuya Yoshida, Hideaki Tanaka, Toshihiko Okuhara, Tetsuji Yamashita, Yasuji Sekine, Masaru Homma, Yohko Kawai, Kaoru Tohyama, The working group for standardization of sample preparation for clotting time assays, Japanese society for laboratory hematology.

Erschienen in: International Journal of Hematology | Ausgabe 5/2020

Einloggen, um Zugang zu erhalten

Abstract

Accurate clotting time assay results are vital, as the test is employed to indicate the amount of oral anticoagulant to be prescribed, while it is also used for screening the hemorrhagic and thrombotic diseases. The procedure chosen for preparation of a patient blood sample including centrifugation can contribute to significant differences in the results obtained. Thus, for the purpose of proposing a standardized method to appropriately prepare blood samples prior to assay, the Japanese Society of Laboratory Hematology organized the Working Group for Standardization of Sample Preparation for Clotting Time Assays (WG). Following reviews of previously announced guidelines and original experimental results, consensus was obtained by the WG, with the main findings as follows. (1) The recommended anticoagulant in the blood collection tube is sodium citrate solution at 0.105–0.109 M (3.13–3.2%). (2) Whole blood samples should be stored at room temperature (18–25 ˚C) within 1 h of collection from the patient. (3) For plasma preparation, centrifugation at 1500 × g should be performed for at least 15 min or at 2000 × g for at least 10 min at room temperature. (4) After the plasma sample is prepared, it should be stored at room temperature and assayed within 4 h.
Literatur
1.
Zurück zum Zitat Yoshida M, Ieko M, Naito S, Morikawa C, Kumano O, Murakami M, et al. Is plasma separated from blood by double centrifugation useful for determination of lupus anticoagulant? J Thromb Haemost. 2011;9:879 (Supple.). Yoshida M, Ieko M, Naito S, Morikawa C, Kumano O, Murakami M, et al. Is plasma separated from blood by double centrifugation useful for determination of lupus anticoagulant? J Thromb Haemost. 2011;9:879 (Supple.).
2.
Zurück zum Zitat Matsuda M, Komiyama Y, Suzuki T, Moriyama M. Impact of sample storage and centrifugation conditions on hemostasis testing. J Japanese Soc Lab Hematol. 2020;21:136–44. Matsuda M, Komiyama Y, Suzuki T, Moriyama M. Impact of sample storage and centrifugation conditions on hemostasis testing. J Japanese Soc Lab Hematol. 2020;21:136–44.
3.
Zurück zum Zitat Adcock DM, Hoefner DM, Kottke-Marchant K, Marlar RA, Szamosi DI, Warunek DJ. Collection, transport, and processing of blood specimens for testing of plasma-based coagulation assays and molecular hemostasis assays; approved guideline. Fifth Edition. Wayne PA: Clinical and Laboratory Standards Institute; 2008: 28. CLSI Document H21-A5. Adcock DM, Hoefner DM, Kottke-Marchant K, Marlar RA, Szamosi DI, Warunek DJ. Collection, transport, and processing of blood specimens for testing of plasma-based coagulation assays and molecular hemostasis assays; approved guideline. Fifth Edition. Wayne PA: Clinical and Laboratory Standards Institute; 2008: 28. CLSI Document H21-A5.
4.
Zurück zum Zitat Mackie I, Cooper P, Lawrie A, Kitchen S, Gray E, Laffan M, et al. on behalf of British committee for standards in haematology: British committee for standards in haematology: guidelines on the laboratory aspects of assays used in haemostasis and thrombosis. Int J Lab Hematol. 2013;35:1–13.CrossRef Mackie I, Cooper P, Lawrie A, Kitchen S, Gray E, Laffan M, et al. on behalf of British committee for standards in haematology: British committee for standards in haematology: guidelines on the laboratory aspects of assays used in haemostasis and thrombosis. Int J Lab Hematol. 2013;35:1–13.CrossRef
5.
Zurück zum Zitat Adcock DM, Kressin DC, Marlar R. Effect of 3.2% vs. 3.8% sodium citrate concentration on routine coagulation testing. Am J Clin Pathol. 1997;107:105–10.CrossRef Adcock DM, Kressin DC, Marlar R. Effect of 3.2% vs. 3.8% sodium citrate concentration on routine coagulation testing. Am J Clin Pathol. 1997;107:105–10.CrossRef
6.
Zurück zum Zitat Kobayashi K, Yoshida M, Kanda S, Hisa K, Matsumoto K. Effects of the differences in sodium citrate concentration on laboratory tests. Tokyo J Med Tech. 2010;38:154 (author’s translate from Japanese). Kobayashi K, Yoshida M, Kanda S, Hisa K, Matsumoto K. Effects of the differences in sodium citrate concentration on laboratory tests. Tokyo J Med Tech. 2010;38:154 (author’s translate from Japanese).
7.
Zurück zum Zitat Ando H. Proposal for unification of blood collection tubes for blood coagulation tests II-Recommendation on use of blood collection tubes with a concentration of trisodium citrate of 3.2%-. Med Technol. 2012;40:350–1 (author’s translate from Japanese). Ando H. Proposal for unification of blood collection tubes for blood coagulation tests II-Recommendation on use of blood collection tubes with a concentration of trisodium citrate of 3.2%-. Med Technol. 2012;40:350–1 (author’s translate from Japanese).
8.
Zurück zum Zitat Japanese committee for clinical laboratory standards. In: Watanabe T (eds) JCCLS standard phlebotomy guideline (GP4-A3). Tokyo: Japanese committee for clinical laboratory standards. 2019. p. 13–35 (author’s translate from Japanese). Japanese committee for clinical laboratory standards. In: Watanabe T (eds) JCCLS standard phlebotomy guideline (GP4-A3). Tokyo: Japanese committee for clinical laboratory standards. 2019. p. 13–35 (author’s translate from Japanese).
10.
Zurück zum Zitat Komiyama Y, Yoshika M. Inadequate technique of phlebotomy affects several data of laboratory tests. Rinsho Kensa. 2015;59:20–5. Komiyama Y, Yoshika M. Inadequate technique of phlebotomy affects several data of laboratory tests. Rinsho Kensa. 2015;59:20–5.
11.
Zurück zum Zitat Lippi G, Salvagno GL, Montagnana M, Poli G, Guidi GC. Influence of the needle bore size on platelet count and routine coagulation testing. Blood Coagul Fibrinolysis. 2006;17:557–61.CrossRef Lippi G, Salvagno GL, Montagnana M, Poli G, Guidi GC. Influence of the needle bore size on platelet count and routine coagulation testing. Blood Coagul Fibrinolysis. 2006;17:557–61.CrossRef
12.
Zurück zum Zitat Yoshida M, Ieko M, Naito S, Suzuki T, Takahashi N, Shibata T. Influence of whole blood preservation on prothrombin time measurement. J Japanese Soc Lab Hematol. 2013;14:267–73. Yoshida M, Ieko M, Naito S, Suzuki T, Takahashi N, Shibata T. Influence of whole blood preservation on prothrombin time measurement. J Japanese Soc Lab Hematol. 2013;14:267–73.
13.
Zurück zum Zitat Favaloro E, Soltani S, McDonald J. Potential laboratory misdiagnosis of hemophilia and von Willebrand disorder owing to cold activation of blood samples for testing. Am J Clin Pathol. 2004;122:686–92.CrossRef Favaloro E, Soltani S, McDonald J. Potential laboratory misdiagnosis of hemophilia and von Willebrand disorder owing to cold activation of blood samples for testing. Am J Clin Pathol. 2004;122:686–92.CrossRef
14.
Zurück zum Zitat Ieko M, Komiyama Y, Yamazaki S, Katagiri H, Shomazu T, Naito S, et al. Expert consensus on standardization of sample preparation for clotting time assays. J Japanese Soc Lab Hematol. 2016;17:140–68. Ieko M, Komiyama Y, Yamazaki S, Katagiri H, Shomazu T, Naito S, et al. Expert consensus on standardization of sample preparation for clotting time assays. J Japanese Soc Lab Hematol. 2016;17:140–68.
15.
Zurück zum Zitat Daves M, Giacomuzzi K, Tagnin E, Jani E, Adcock Funk DM, Favaloro EJ, et al. Influence of centrifuge brake on residual platelet count and routine coagulation tests in citrated plasma. Blood Coagul Fibrinolysis. 2014;25:292–5.CrossRef Daves M, Giacomuzzi K, Tagnin E, Jani E, Adcock Funk DM, Favaloro EJ, et al. Influence of centrifuge brake on residual platelet count and routine coagulation tests in citrated plasma. Blood Coagul Fibrinolysis. 2014;25:292–5.CrossRef
16.
Zurück zum Zitat Woodhams B, Girardot O, Blanco MJ, Colesse G, Gourmelin Y. Stability of coagulation proteins in frozen plasma. Blood Coagul Fibrinolysis. 2001;12:229–36.CrossRef Woodhams B, Girardot O, Blanco MJ, Colesse G, Gourmelin Y. Stability of coagulation proteins in frozen plasma. Blood Coagul Fibrinolysis. 2001;12:229–36.CrossRef
17.
Zurück zum Zitat Lippi G, Rossi R, Ippolito L, Zobbi V, Azzi D, Pipitone S, et al. Influence of residual platelet count on routine coagulation, factor VIII, and factor IX testing in postfreeze-thaw samples. Semin Thromb Hemost. 2013;39:834–9.CrossRef Lippi G, Rossi R, Ippolito L, Zobbi V, Azzi D, Pipitone S, et al. Influence of residual platelet count on routine coagulation, factor VIII, and factor IX testing in postfreeze-thaw samples. Semin Thromb Hemost. 2013;39:834–9.CrossRef
Metadaten
Titel
Expert consensus regarding standardization of sample preparation for clotting time assays
verfasst von
Masahiro Ieko
Yutaka Komiyama
Satoshi Yamazaki
Hisako Katagiri
Chisato Shimazu
Sumiyoshi Naito
Masato Matsuda
Yoichi Yuki
Kazuyoshi Miyata
Rika Arai
Osamu Kumano
Mari Emmi
Tatsuya Yoshida
Hideaki Tanaka
Toshihiko Okuhara
Tetsuji Yamashita
Yasuji Sekine
Masaru Homma
Yohko Kawai
Kaoru Tohyama
Masahiro Ieko
Yutaka Komiyama
Satoshi Yamazaki
Hisako Katagiri
Chisato Shimazu
Sumiyoshi Naito
Masato Matsuda
Yoichi Yuki
Kazuyoshi Miyata
Rika Arai
Osamu Kumano
Mari Emmi
Tatsuya Yoshida
Hideaki Tanaka
Toshihiko Okuhara
Tetsuji Yamashita
Yasuji Sekine
Masaru Homma
Yohko Kawai
Kaoru Tohyama
The working group for standardization of sample preparation for clotting time assays, Japanese society for laboratory hematology.
Publikationsdatum
14.09.2020
Verlag
Springer Singapore
Erschienen in
International Journal of Hematology / Ausgabe 5/2020
Print ISSN: 0925-5710
Elektronische ISSN: 1865-3774
DOI
https://doi.org/10.1007/s12185-020-02983-x

Weitere Artikel der Ausgabe 5/2020

International Journal of Hematology 5/2020 Zur Ausgabe

Update Onkologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.