Abstract

Background

Temporal window failure (TWF) is found in 8–20% of subjects. There are still insufficient studies about the factors affecting TWF. We aimed to elucidate the underlying causes of TWF.

Methods

We analyzed 376 patients who underwent both transcranial Doppler sonography and cerebral angiographic imaging. They were divided into two groups: with and without TWF. Demographics, cardiovascular factors, degree of stenosis from the proximal intracranial artery to the middle cerebral artery (MCA), MCA diameter, and skull features were examined.

Results

The subjects were 314 TWF-negative patients and 62 TWF-positive patients. The TWF-negative group was younger than that of the TWF-positive group (67.0 ± 12.1 vs. 75.2 ± 9.4, p < 0.001). The proportion of men in the TWF-negative group was higher than in the TWF-positive group (71% vs. 29%; p < 0.001). The TWF-negative group had a higher smoking rate than the TWF-positive group (34.4% vs. 12.9%; p = 0.001). In multivariate logistic regression analysis, age (odds ratio (OR), 1.05; p = 0.019), sex (OR, 4.64; p = 0.002), temporal bone thickness (OR, 6.03; p < 0.001), temporal bone density (OR, 0.996; p = 0.002), and soft tissue thickness (OR, 1.31; p = 0.004) significantly affected TWF.

Conclusions

In addition to age, sex, temporal bone thickness, and temporal bone density which were previously reported as variables associated with TWF, we confirmed that soft tissue thickness of the temporal area is a new associated factor of TWF. Measuring soft tissue thickness of the temporal area for patients with suspected TWF could be useful in identifying measurement error due to technical problems.
Anhänge
Nur mit Berechtigung zugänglich
Literatur
1.
Zurück zum Zitat Marinoni M, Ginanneschi A, Forleo P, Amaducci L (1997) Technical limits in transcranial Doppler recording: inadequate acoustic windows. Ultrasound Med Biol 23:1275–1277CrossRef Marinoni M, Ginanneschi A, Forleo P, Amaducci L (1997) Technical limits in transcranial Doppler recording: inadequate acoustic windows. Ultrasound Med Biol 23:1275–1277CrossRef
2.
Zurück zum Zitat Postert T, Federlein J, Przuntek H, Büttner T (1997) Insufficient and absent acoustic temporal bone window: potential and limitations of transcranial contrast-enhanced color-coded sonography and contrast-enhanced power-based sonography. Ultrasound Med Biol 23:857–862CrossRef Postert T, Federlein J, Przuntek H, Büttner T (1997) Insufficient and absent acoustic temporal bone window: potential and limitations of transcranial contrast-enhanced color-coded sonography and contrast-enhanced power-based sonography. Ultrasound Med Biol 23:857–862CrossRef
3.
Zurück zum Zitat Bazan R, Braga GP, Luvizutto GJ, Hueb JC, Hokama NK, Zanati Bazan SG, de Carvalho Nunes HR, Leite JP, Pontes-Neto OM (2015) Evaluation of the temporal acoustic window for transcranial Doppler in a multi-ethnic population in Brazil. Ultrasound Med Biol 41:2131–2134CrossRef Bazan R, Braga GP, Luvizutto GJ, Hueb JC, Hokama NK, Zanati Bazan SG, de Carvalho Nunes HR, Leite JP, Pontes-Neto OM (2015) Evaluation of the temporal acoustic window for transcranial Doppler in a multi-ethnic population in Brazil. Ultrasound Med Biol 41:2131–2134CrossRef
4.
Zurück zum Zitat Krejza J, Swiat M, Pawlak MA, Oszkinis G, Weigele J, Hurst RW, Kasner S (2007) Suitability of temporal bone acoustic window: conventional TCD versus transcranial color-coded duplex sonography. J Neuroimaging 17:311–314CrossRef Krejza J, Swiat M, Pawlak MA, Oszkinis G, Weigele J, Hurst RW, Kasner S (2007) Suitability of temporal bone acoustic window: conventional TCD versus transcranial color-coded duplex sonography. J Neuroimaging 17:311–314CrossRef
5.
Zurück zum Zitat Lindsey BD, Nicoletto HA, Bennett ER, Laskowitz DT, Smith SW (2013) Simultaneous bilateral real-time 3-D transcranial ultrasound imaging at 1 MHz through poor acoustic windows. Ultrasound Med Biol 39:721–734CrossRef Lindsey BD, Nicoletto HA, Bennett ER, Laskowitz DT, Smith SW (2013) Simultaneous bilateral real-time 3-D transcranial ultrasound imaging at 1 MHz through poor acoustic windows. Ultrasound Med Biol 39:721–734CrossRef
6.
Zurück zum Zitat Brunser AM, Silva C, Carcamo D, Muñoz P, Hoppe A, Olavarría VV et al (2012) Transcranial Doppler in a Hispanic-Mestizo population with neurological diseases: a study of sonographic window and its determinants. Brain Behav 2:231–236CrossRef Brunser AM, Silva C, Carcamo D, Muñoz P, Hoppe A, Olavarría VV et al (2012) Transcranial Doppler in a Hispanic-Mestizo population with neurological diseases: a study of sonographic window and its determinants. Brain Behav 2:231–236CrossRef
7.
Zurück zum Zitat Kwon JH, Kim JS, Kang DW, Bae KS, Kwon SU (2006) The thickness and texture of temporal bone in brain CT predict acoustic window failure of transcranial Doppler. J Neuroimaging 16:347–352CrossRef Kwon JH, Kim JS, Kang DW, Bae KS, Kwon SU (2006) The thickness and texture of temporal bone in brain CT predict acoustic window failure of transcranial Doppler. J Neuroimaging 16:347–352CrossRef
8.
Zurück zum Zitat Laurent A, Mistretta F, Bottigioli D, Dahel K, Goujon C, Nicolas JF, Hennino A, Laurent PE (2007) Echographic measurement of skin thickness in adults by high frequency ultrasound to assess the appropriate microneedle length for intradermal delivery of vaccines. Vaccine 25:6423–6430CrossRef Laurent A, Mistretta F, Bottigioli D, Dahel K, Goujon C, Nicolas JF, Hennino A, Laurent PE (2007) Echographic measurement of skin thickness in adults by high frequency ultrasound to assess the appropriate microneedle length for intradermal delivery of vaccines. Vaccine 25:6423–6430CrossRef
9.
Zurück zum Zitat North American Symptomatic Carotid Endarterectomy Trial C, HJM B, Taylor DW, Haynes RB, Sackett DL, Peerless SJ et al (1991) Beneficial effect of carotid endarterectomy in symptomatic patients with high-grade carotid stenosis. N Engl J Med 325:445–453CrossRef North American Symptomatic Carotid Endarterectomy Trial C, HJM B, Taylor DW, Haynes RB, Sackett DL, Peerless SJ et al (1991) Beneficial effect of carotid endarterectomy in symptomatic patients with high-grade carotid stenosis. N Engl J Med 325:445–453CrossRef
10.
Zurück zum Zitat Ryu WS, Park SS, Kim YS, Lee SH, Kang K, Kim C, Sohn CH, Lee SH, Yoon BW (2014) Long-term natural history of intracranial arterial stenosis: an MRA follow-up study. Cerebrovasc Dis 38:290–296CrossRef Ryu WS, Park SS, Kim YS, Lee SH, Kang K, Kim C, Sohn CH, Lee SH, Yoon BW (2014) Long-term natural history of intracranial arterial stenosis: an MRA follow-up study. Cerebrovasc Dis 38:290–296CrossRef
11.
Zurück zum Zitat Lin YP, Fu MH, Tan (2015) Factors associated with no or insufficient temporal bone window using transcranial color-coded sonography. J Med Ultrasound 23:129–132CrossRef Lin YP, Fu MH, Tan (2015) Factors associated with no or insufficient temporal bone window using transcranial color-coded sonography. J Med Ultrasound 23:129–132CrossRef
12.
Zurück zum Zitat Kollar J, Schulte-Altedorneburg G, Sikula J, Fülesdi B, Ringelstein EB, Mehta V et al (2004) Image quality of the temporal bone window examined by transcranial Doppler sonography and correlation with postmortem computed tomography measurements. Cerebrovasc Dis 17:61–65CrossRef Kollar J, Schulte-Altedorneburg G, Sikula J, Fülesdi B, Ringelstein EB, Mehta V et al (2004) Image quality of the temporal bone window examined by transcranial Doppler sonography and correlation with postmortem computed tomography measurements. Cerebrovasc Dis 17:61–65CrossRef
13.
Zurück zum Zitat Jarquin-Valdivia AA, McCartney J, Palestrant D, Johnston SC, Gress D (2004) The thickness of the temporal squama and its implication for transcranial sonography. J Neuroimaging 14:139–142CrossRef Jarquin-Valdivia AA, McCartney J, Palestrant D, Johnston SC, Gress D (2004) The thickness of the temporal squama and its implication for transcranial sonography. J Neuroimaging 14:139–142CrossRef
14.
Zurück zum Zitat Wijnhoud AD, Franckena M, van der Lugt A, Koudstaal PJ, Dippel ED (2008) Inadequate acoustical temporal bone window in patients with a transient ischemic attack or minor stroke: role of skull thickness and bone density. Ultrasound Med Biol 34:923–929CrossRef Wijnhoud AD, Franckena M, van der Lugt A, Koudstaal PJ, Dippel ED (2008) Inadequate acoustical temporal bone window in patients with a transient ischemic attack or minor stroke: role of skull thickness and bone density. Ultrasound Med Biol 34:923–929CrossRef
15.
Zurück zum Zitat Ovel S (2014) Sonography exam review: physics, abdomen, obstetrics and gynecology. Elsevier Health Sciences, St. Louis Ovel S (2014) Sonography exam review: physics, abdomen, obstetrics and gynecology. Elsevier Health Sciences, St. Louis
16.
Zurück zum Zitat Shmulewitz A, Teefey SA, Robinson BS (1993) Factors affecting image quality and diagnostic efficacy in abdominal sonography: a prospective study of 140 patients. J Clin Ultrasound 21:623–630CrossRef Shmulewitz A, Teefey SA, Robinson BS (1993) Factors affecting image quality and diagnostic efficacy in abdominal sonography: a prospective study of 140 patients. J Clin Ultrasound 21:623–630CrossRef
17.
Zurück zum Zitat Siadecki SD, Frasure SE, Lewiss RE, Saul T (2016) High body mass index is strongly correlated with decreased image quality in focused bedside echocardiography. J Emerg Med 50:295–301CrossRef Siadecki SD, Frasure SE, Lewiss RE, Saul T (2016) High body mass index is strongly correlated with decreased image quality in focused bedside echocardiography. J Emerg Med 50:295–301CrossRef
18.
Zurück zum Zitat Zimmerman AA, Burrows FA, Jonas RA, Hickey PR (1997) The limits of detectable cerebral perfusion by transcranial doppler sonography in neonates undergoing deep hypothermic low-flow cardiopulmonary bypass. J Thorac Cardiovasc Surg 114:594–600CrossRef Zimmerman AA, Burrows FA, Jonas RA, Hickey PR (1997) The limits of detectable cerebral perfusion by transcranial doppler sonography in neonates undergoing deep hypothermic low-flow cardiopulmonary bypass. J Thorac Cardiovasc Surg 114:594–600CrossRef
Metadaten
Titel
Factors associated with temporal window failure in transcranial Doppler sonography
verfasst von
Chan-Hyuk Lee
Seung-Ho Jeon
Su-Jung Wang
Byoung-Soo Shin
Hyun Goo Kang
Publikationsdatum
13.05.2020
Verlag
Springer International Publishing
Erschienen in
Neurological Sciences / Ausgabe 11/2020
Print ISSN: 1590-1874
Elektronische ISSN: 1590-3478
DOI
https://doi.org/10.1007/s10072-020-04459-6

Weitere Artikel der Ausgabe 11/2020

Neurological Sciences 11/2020 Zur Ausgabe

Leitlinien kompakt für die Neurologie

Mit medbee Pocketcards sicher entscheiden.

Seit 2022 gehört die medbee GmbH zum Springer Medizin Verlag

Update Neurologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.