Feasibility study of a novel portable digital radiography system modified for fluoroscopy in the neonatal intensive care unit

verfasst von: Mark C. Liszewski, Samuel Richard, Jordana N. Gross, Alison Schonberger, Terry L. Levin, Einat Blumfield, Suhas M. Nafday, Benjamin H. Taragin

Erschienen in: Pediatric Radiology | Ausgabe 9/2021

Einloggen, um Zugang zu erhalten

Abstract

Background

A portable radiographic system capable of fluoroscopic imaging in the neonatal intensive care unit (NICU) potentially benefits critically ill neonates by eliminating the need to transport them to a fluoroscopy suite.

Objective

To evaluate whether a portable bedside fluoroscopy system in the NICU can deliver comparable image quality at a similar dose rate to a standard system in a fluoroscopy suite.

Materials and methods

In phase A, 20 patients <3 years of age and scheduled to undergo upper gastrointestinal series (upper GI) or voiding cystourethrograms (VCUG) in the radiology fluoroscopy suite were recruited to evaluate a portable fluoroscopic unit. A modified portable radiographic system with a cassette-sized detector and an in-room fluoroscopy system were sequentially used in the same examination. Four radiologists compared the image quality of 20 images from each system using the Radlex score (1–4) for five image quality attributes. The radiation dose rates for the portable and in-suite systems were collected. In phase B, fluoroscopy studies were performed in 5 neonates in the NICU and compared to the 20 previous neonatal studies performed in the department. Clinical workflow, examination time, fluoroscopy time, scattered radiation dose and patient radiation dose were evaluated.

Results

In phase A, average dose rates for in-room and portable systems were equivalent, (0.322 mGy/min and 0.320 mGy/min, respectively). Reader-averaged Radlex scores for in-room and portable systems were statistically significantly greater (P<0.05) for all attributes on the portable system except for image contrast. In phase B, scattered radiation from the average fluoroscopy time (26 s) was equivalent to the scattered radiation of 2.6 portable neonatal chest radiographs. Procedure time and diagnostic quality were deemed equivalent. The average dose rate in the NICU with the portable system was 0.21 mGy/min compared to 0.29 mGy/min for the in-room system.

Conclusion

The portable fluoroscopy unit is capable of providing comparable image quality at equivalent dose levels to an in-room system for neonates with minimal risks to the staff and other patients in the NICU.
Literatur
1.
Zurück zum Zitat Roberts KB, Subcommittee on Urinary Tract Infection, Steering Committee on Quality Improvement and Management (2011) Urinary tract infection: clinical practice guideline for the diagnosis and management of the initial UTI in febrile infants and children 2 to 24 months. Pediatrics 128:595–610CrossRef Roberts KB, Subcommittee on Urinary Tract Infection, Steering Committee on Quality Improvement and Management (2011) Urinary tract infection: clinical practice guideline for the diagnosis and management of the initial UTI in febrile infants and children 2 to 24 months. Pediatrics 128:595–610CrossRef
2.
Zurück zum Zitat Strouse PJ (2004) Disorders of intestinal rotation and fixation (“malrotation”). Pediatr Radiol 34:837–851CrossRef Strouse PJ (2004) Disorders of intestinal rotation and fixation (“malrotation”). Pediatr Radiol 34:837–851CrossRef
3.
Zurück zum Zitat Vieira ALP, dos Santos AMN, Okuyama MK et al (2011) Factors associated with clinical complications during intra-hospital transports in a neonatal unit in Brazil. J Trop Pediatr 57:368–374CrossRef Vieira ALP, dos Santos AMN, Okuyama MK et al (2011) Factors associated with clinical complications during intra-hospital transports in a neonatal unit in Brazil. J Trop Pediatr 57:368–374CrossRef
4.
Zurück zum Zitat Fenton AC, Leslie A, Skeoch CH (2004) Optimizing neonatal transfer. Arch Dis Child Fetal Neonatal Ed 89:F215–F219CrossRef Fenton AC, Leslie A, Skeoch CH (2004) Optimizing neonatal transfer. Arch Dis Child Fetal Neonatal Ed 89:F215–F219CrossRef
5.
Zurück zum Zitat Lim MTC, Ratnavel N (2008) A prospective review of adverse events during interhospital transfers of neonates by dedicated neonatal transfer service. Pediatr Crit Care Med 9:289–293CrossRef Lim MTC, Ratnavel N (2008) A prospective review of adverse events during interhospital transfers of neonates by dedicated neonatal transfer service. Pediatr Crit Care Med 9:289–293CrossRef
6.
Zurück zum Zitat Nayak GK, Levin TL, Kurian J et al (2014) Bedside upper gastrointestinal series in critically ill low birth weight infants. Pediatr Radiol 44:1252–1257CrossRef Nayak GK, Levin TL, Kurian J et al (2014) Bedside upper gastrointestinal series in critically ill low birth weight infants. Pediatr Radiol 44:1252–1257CrossRef
7.
Zurück zum Zitat Abdullah AC, Adnan JS, Rahman NAA, Palur R (2017) Limited evaluation of image quality produced by a portable head CT scanner (CereTom) in a neurosurgery centre. Malays J Med Sci 24:104–112PubMedPubMedCentral Abdullah AC, Adnan JS, Rahman NAA, Palur R (2017) Limited evaluation of image quality produced by a portable head CT scanner (CereTom) in a neurosurgery centre. Malays J Med Sci 24:104–112PubMedPubMedCentral
8.
Zurück zum Zitat Masaryk T, Kolonick R, Painter T, Weinreb DB (2008) The economic and clinical benefits of portable head/neck CT imaging in the intensive care unit. Radiol Manage 30:50–54PubMed Masaryk T, Kolonick R, Painter T, Weinreb DB (2008) The economic and clinical benefits of portable head/neck CT imaging in the intensive care unit. Radiol Manage 30:50–54PubMed
9.
Zurück zum Zitat Chacko J, Brar G (2014) Bedside ultrasonography: applications in critical care: part I. Indian J Crit Care Med 18:301–309CrossRef Chacko J, Brar G (2014) Bedside ultrasonography: applications in critical care: part I. Indian J Crit Care Med 18:301–309CrossRef
Metadaten
Titel
Feasibility study of a novel portable digital radiography system modified for fluoroscopy in the neonatal intensive care unit
verfasst von
Mark C. Liszewski
Samuel Richard
Jordana N. Gross
Alison Schonberger
Terry L. Levin
Einat Blumfield
Suhas M. Nafday
Benjamin H. Taragin
Publikationsdatum
18.03.2021
Verlag
Springer Berlin Heidelberg
Erschienen in
Pediatric Radiology / Ausgabe 9/2021
Print ISSN: 0301-0449
Elektronische ISSN: 1432-1998
DOI
https://doi.org/10.1007/s00247-021-05022-x

Neu im Fachgebiet Radiologie

KI-gestütztes Mammografiescreening überzeugt im Praxistest

Mit dem Einsatz künstlicher Intelligenz lässt sich die Detektionsrate im Mammografiescreening offenbar deutlich steigern. Mehr unnötige Zusatzuntersuchungen sind laut der Studie aus Deutschland nicht zu befürchten.

Stumme Schlaganfälle − ein häufiger Nebenbefund im Kopf-CT?

In 4% der in der Notfallambulanz initiierten zerebralen Bildgebung sind „alte“ Schlaganfälle zu erkennen. Gar nicht so selten handelt es sich laut einer aktuellen Studie dabei um unbemerkte Insulte. Bietet sich hier womöglich die Chance auf ein effektives opportunistisches Screening?

Die elektronische Patientenakte kommt: Das sollten Sie jetzt wissen

Am 15. Januar geht die „ePA für alle“ zunächst in den Modellregionen an den Start. Doch schon bald soll sie in allen Praxen zum Einsatz kommen. Was ist jetzt zu tun? Was müssen Sie wissen? Wir geben in einem FAQ Antworten auf 21 Fragen.

Stören weiße Wände und viel Licht die Bildqualitätskontrolle?

Wenn es darum geht, die technische Qualität eines Mammogramms zu beurteilen, könnten graue Wandfarbe und reduzierte Beleuchtung im Bildgebungsraum von Vorteil sein. Darauf deuten zumindest Ergebnisse einer kleinen Studie hin. 

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.