Skip to main content
Erschienen in: Surgical Endoscopy 1/2019

Open Access 14.08.2018 | New Technology

Fecal diverting device for the substitution of defunctioning stoma: preliminary clinical study

verfasst von: Jae Hwang Kim, Sohyun Kim, Sang Hun Jung

Erschienen in: Surgical Endoscopy | Ausgabe 1/2019

Abstract

Background

A novel fecal diverting device (FDD) made for the prevention of sepsis resulting from anastomotic leakage (AL) was tested successfully in an animal study. This study was undertaken to evaluate the clinical safety and effectiveness of the FDD.

Methods

A prospective observation trial was implemented in a tertiary referral university hospital. The study enrolled patients who needed a defunctioning stoma to preserve low-lying rectal anastomosis. The FDD was fixed to the proximal colon 15 cm from the anastomosis and scheduled to divert feces for 3 weeks. The duration could be extended for more than 3 weeks if AL was noted. Postoperative evaluations of AL were performed by obtaining a computed tomography (CT) scan after 1 week and a contrast study after 3 weeks. The outcomes were FDD-related complications, and the capacity of the FDD to preserve the anastomosis. The median follow-up period was 10 (range 5–40) months.

Results

Thirty-one patients, including 5 benign cases, were evaluated. There was no case of stoma conversion or surgical re-intervention. Evidence of AL was identified in 10 (32%) patients using the CT scan at 1 week after surgery. However, in the contrast study at 3 weeks after surgery, only 5 cases of AL sinus were noted. Conservative treatments including 1–3 weeks prolongation of FDD maintenance were enough to preserve the anastomosis. There were 3 cases of partial colonic wall erosions at the FDD attachment area. All of these patients showed improvement with conservative treatment. The limitations were that the study was performed in a single institute and without a control group.

Conclusions

The FDD showed a sufficient capacity of fecal diversion and maintenance duration that prevented aggravation of sepsis in the case of AL without significant complications.
Hinweise
Early results of the contents of this article were presented at the New Technology Symposium during the meeting of The American Society of Colon and Rectal Surgeons, Los Angeles, CA, 30 April–4 May 2016.
Fecal diversion with stoma is considered to be the best option to prevent the catastrophic cascade after anastomotic leakage (AL) associated with lower rectal resection [13]. However, this procedure cannot, by itself, prevent leakage [4]. The morbidity and mortality associated with stoma remain as the important clinical consideration [5]. An impaired quality of life and the economic burden for patients represent other disadvantages associated with stoma [6].
The basic mechanism of defunctioning stoma is diverting feces from the anastomotic wound. By doing this, inflammation at the anastomotic wound can be decreased due to a reduction of infection sources. Several tube-structured devices have been attempted to replace the temporary stoma [711]; however, they have not been successful. Their limitations, such as limited maintenance duration, incontinence, and questionable safety, may be the main cause of this failure. Anastomotic wound healing usually occurs in a week [12]; however, some conditions need more time for healing. Delayed wound healing may occur in patients with preoperative chemoradiation therapy, long-term use of steroids, AL, or various other conditions [1315]. If these factors are present, the time for anastomotic wound healing may be prolonged.
A novel tube device, the fecal diverting device (FDD, Yushin Med. Co. Seoul, Korea), was developed recently (Fig. 1). It showed a successful fecal diversion and long maintenance duration of up to 82 days in an animal study [16]. Thus, the FDD could prevent development of serious septicemia in conditions of AL.
Before the current study, we performed a pilot study of the FDD in 8 patients who required fecal diversion. We noted that the results were similar to those of the animal study.
In the present clinical study, we attempted to identify the safety and effectiveness of the using the FDD system in patients with low-lying colorectal anastomosis.

Materials and methods

The study was designed as a prospective observation trial. All the procedures were performed by experienced colorectal surgeons in a tertiary referral university hospital. The study was approved by the ethical review board at Yeungnam University and performed in accordance with the ethical standards of the Declaration of Helsinki.
The participants were patients who needed a defunctioning stoma after low-lying colorectal anastomosis, regardless of whether the cause was benign or malignant. A standard stapled or hand-sewn extraperitoneal anastomosis was created within 8 cm from the anal verge. Patients who were aged < 18 years, had an American Society of Anesthesiologists score > 3, or who were pregnant were excluded. All the patients and caregivers were informed in detail about the procedure, including the risks and benefits, and were invited to participate. Written consent was obtained before surgery or during surgery.
Any mortality and morbidity data of the enrolled patients that were not associated with the FDD were excluded from analysis.

The FDD system

The FDD system consists of 2 main devices: the FDD and a band (BAND) for fixing the FDD as shown in Fig. 2.
The FDD is a silicone tube device [16], designed for the protection from fecal contamination of an anastomotic wound using fecal diversion. The FDD is composed of 2 parts, a solid tubular head portion and a thin tubular tail portion. At the head portion, there are 2 outer balloons and 1 inner balloon. The outer balloons are used to attach the FDD at the colon. The inner balloon is used for stop-flow in the tube. A channel for the infusion of irrigation water is located parallel to the catheters for ballooning. The function of the balloons during the FDD procedure is illustrated in Fig. 3. The tail, i.e., the thin tube below the head portion, is long enough to hang out from the anus.
A BAND is an absorbable poly lactic-co-glycolic acid (PGLA) mesh band (NEOSORB MESH®, Samyang Biopham. Co. Daejeon, Korea) with a half-life of 6 weeks. The extraluminal BAND wraps the waist made between the 2 outer balloons to attach the FDD at a certain location of the bowel (Fig. 2). For the safe application of the BAND, an automatic tension measuring instrument (ATMI, JSR medical Inc, Daegu, Korea) was newly developed and applied in the current study (Fig. 4).

The FDD procedure

Intraoperative FDD installation procedure

Following removal of the pathological specimen, an ATMI was inserted in the mesocolon to measure an adequate length of the BAND. The resulting length was marked on the BAND. The BAND was then introduced and hung at the ATMI-penetrated mesocolic opening with loose suturing using 2 silk sutures (2–0) at the marked point of the BAND. Anvil insertion and circular stapling were performed by a standard stapling technique after setting up the BAND. After anastomosis, an FDD was inserted from the anus to the BAND site with the inner balloon inflated and outer balloons deflated. Two sutures at the marked points of the BAND were tightened after the inflation of the outer balloons (Fig. 5). The procedure was applied in open or laparoscopic conditions with little difference.

Postoperative maintenance period

There was no restriction of postoperative oral intake. Daily bowel movement with twice daily irrigation of tepid water via the FDD started a day after intake of a full liquid diet. The patients and caregivers were educated about the mechanism (Fig. 3) of the irrigation procedure and encouraged to perform the procedures.
Patients or guardians were ordered to record the volume of infused irrigation water for the evaluation of adequate irrigation water volume. We recommended to infuse water until the patient felt abdominal discomfort. The volume was counted in 100 mL increments for convenience.
Patients or guardians were also ordered to record the evacuation time. Time was counted in 5-min increments.
The scheduled maintenance duration of the FDD was 3 weeks. If there was AL present, the duration was prolonged. Removal of the FDD was performed with fluoroscopic guidance without anesthesia.

Postoperative study

One week after surgery, an abdominal computed tomography (CT) scan was scheduled to detect any hidden AL. If there was any evidence of AL clinically before 1 week, a CT scan was performed without delay. The findings were evaluated by experienced radiologists and participating surgeons.
Three weeks after surgery a water-soluble contrast study was performed. If there was no evidence of AL, the FDD was removed. If there was evidence of sinus, the result of AL, the patients were recommended to maintain the FDD for an additional 1–3 weeks. The actual duration was determined by the surgeons. The maximum duration of the FDD maintenance was 6 weeks due to the half-life of the BAND. Sigmoidoscopic examination for the evaluation of the anastomotic site and possible colonic wall injury at the FDD attachment area was performed immediately after the FDD was removed.

Follow-up evaluation

The Clavien–Dindo classification (Dindo) was applied for the evaluation of possible surgical complications of the FDD and anastomosis. The definition of AL in this study was a defect in the intestinal wall integrity at the colorectal or coloanal anastomotic site, which led to communication between the intraluminal and extraluminal compartments. A pelvic abscess close to the anastomosis is also considered as AL [17]. Additionally, peri-anastomotic air and peri-anastomotic fluid collection were included [18].

Endpoints

The primary purpose of the study was to determine the safety and effectiveness of the FDD system. Safety was assessed by identification of significant intra-abdominal complications caused by bowel wall injury at the FDD attachment area. Inadvertent adverse events caused by FDD installation, maintenance, removal and anodermal injury were also included. Effectiveness was assessed by the capability of complete fecal diversion for a sufficient duration in the case of AL to avoid defunctioning stoma.
The secondary purpose was to identify the adequate volume of irrigation water and adequate evacuation time of diluted feces.

Results

Thirty-two patients (19 males) with benign or malignant colorectal disease were enrolled between October 2014 and September 2017. One patient was excluded. The remaining 31 cases were included in the analysis for the current study (Fig. 6). The median follow-up period was 10 (range 5–40) months. Baseline clinical characteristics are listed in Table 1.
Table 1
Clinical characteristics
Characteristic
All patients (n = 31)
Age (year)
70.5 (range 39–86)
Sex (male/female)
19:12
ASA score
 1
10 (32.3)
 2
18 (58.1)
 3
3 (9.7)
Anastomotic height (cm)
5.9 ± 2.8
Lesion of colon
 Malignancy
26 (83.9)
 Benign
5 (16.1)
PCRT
7/26 (26.9)
Maintenance of FDD (day)
23.6 ± 6.1
Hospital stay after operation (day)
13.9 ± 5.8
Values are presented as mean ± standard deviation or number (%)
ASA score physical status classification system of American Society of Anesthesiologists, PCRT preoperative chemoradiation therapy, FDD fecal diverting device

Exclusion

Permanent ileostomy conversion was performed in 1 patient (female, 78 years) because of a long segment of remaining colonic necrosis. There was no evidence of AL or BAND site abnormality. The patient experienced low anterior resection with high ligation. However, she had a history of right hemicolectomy for ascending colon cancer 12 years prior. The cause of necrosis was ischemia caused by the poor blood supply from the only remaining midcolic artery.

Endpoints

Safety of FDD

The time of FDD removal for all participating patients was decided by the surgeons. The average duration of FDD use was 23.6 ± 6.1 days. The average duration of hospital stay was 13.9 ± 5.8 days. There were 3 partial colonic wall erosions (3/31, 9.7%) noted at the BAND area on days 22, 26, and 36 after surgery. Detection was possible by using a water-soluble contrast study and sigmoidoscopy. Two cases were asymptomatic, and no other treatment was needed (Dindo Grade I). For the other patient, the erosion caused a pericolic abscess; however, it improved with antibiotic treatment (Dindo Grade II).
The follow-up of these patients was uneventful, and there was no evidence of wall defect or stricture in follow-up colonoscopic examination up to 9 months afterward.
Intraoperative installation of the FDD was composed of an intraoperative BAND application and FDD insertion via the anus. There was no failure of the insertion procedure and no traumatic disruption of the anastomotic wound.
During the maintenance period, all patients experienced twice daily irrigation with the FDD for their bowel control. The majority of the irrigation procedures were performed by the patients themselves without significant difficulty. If it was difficult for elderly or handicapped patients to perform the procedure by themselves, healthy caregivers were able to help them easily.
There was no failure of FDD removal and removal of the FDDs was uneventful. In a few instances, there was mild resistance when the solid portion of FDD was passing the anastomotic area. However, it was overcome by slow and gentle pulling of the device.
There was no significant anodermal or perianal skin problem that required medication or treatment.

Effectiveness of the FDD in case of AL

Ten cases (32%) of AL were detected by the CT scan examination at 1-week postoperative evaluation. Eight (33%) and 2 (29%) ALs were noted in patients without (n = 24) and with (n = 7) preoperative chemoradiation, respectively. These were classified as Dindo Grade I (3), Grade II (6), and Grade IIIa (1) surgical complications. The Grade III complication was noted in a patient with local sepsis because of a large peri-anastomotic fluid collection on the 4th postoperative day. Septic symptoms improved dramatically with ultrasonogram guided percutaneous drainage. There were no patients with stoma conversion due to the failure of fecal diversion with the FDD. The contrast study after 3 weeks revealed 5 (16%) cases of remaining sinus in the previous 10 AL cases. Prolonged FDD maintenance was decided in these 5 cases and resulted in a median of 36 days (range 27–43 days) of maintenance duration. The final contrast study before the removal of the FDD revealed a persistent chronic sinus in 4 (13%) cases. No additional defunctioning stoma was provided to these patients.
In the follow-up, after more than 6 months and up to 18 months, there was no recurrence of an anastomotic wound problem in the patients with chronic sinus.

Secondary endpoints

The median volume of tepid irrigation water was 800 (range 400–1100) mL. The median evacuation time was 20 (range 15–35) minutes. After evacuation, only minimal soilage occurred before the next irrigation.

Discussion

The advantage of FDD

The FDD in the current study is a kind of bypass tube device. The FDD, unlike previously introduced devices, overcame most of the existing limitations, such as insufficient and uncontrollable maintenance duration, and incontinence. The maintenance duration of FDD can be controlled by surgeons and can be prolonged for up to 6 weeks. A twice daily irrigation procedure can help most patients to achieve continence.

Safety

The most serious anticipated adverse effect in this FDD concept is bowel injury, such as necrotic perforation or erosion due to colonic wall ischemia. We were able to estimate the safe pressure range from the previous animal study [16] and could make a device ATMI. This worked well and resulted in no loss of device and no acute intra-abdominal surgical complication.
There was 1 case of pericolic abscess at the erosion site, meaning erosion perforation of the colonic wall. We suppose that the cause of local sepsis might delay wound healing between irradiated adjacent structures [13]. The other 2 patients with erosion required no admission and no additional medication. The sigmoidoscopic examination revealed colonic edemas above the anastomosis in 1 patient. The second patient with erosion had a history of frequent strong pulling of the FDD catheter complex. The patient fixed the catheters at her abdomen tightly. Even though erosion developed at a rate of 10%, the absence of treatment or conservative treatment was enough to solve the problem. In the previous animal study, there was a 15% erosion rate; however, there were no serious conditions that required re-intervention [16]. Nevertheless, it would be preferable if no erosion developed, even if it did not require a surgical re-intervention. Future improvement of the FDD should lead to increased safety of the device.

Effectiveness

In the current study protocol, the scheduled maintenance duration was 3 weeks. It is known that the collagen deposition at an anastomotic wound is highest 6–12 days after surgery [12]. However, this mechanism might be disturbed due to various causes and results of delayed wound healing [1315]. In our current study, maintenance duration of 10 patients with AL ranged from 27 to 43 days, which is 1–3 weeks longer than that of the 21 patients without AL. However, there was no increased incidence of complication.
The AL detection rate in our study is higher at 32% than other reports in the literature. We suggest that the reason might be routine evaluation with a CT scan at 1 week after surgery. This high rate is similar to other studies that included early routine evaluation of AL. Pakkastie et al. reported 39% of AL incidence with a routine contrast study approximately 1 week after surgery in their randomized controlled trial [19]. Tagart reported 36% of AL with routine contrast approximately 2 weeks after surgery [20]. If we include the 4 cases (12.9%) of persistent sinus after 26–43 days of fecal diversion, the incidence rate of AL would be similar to the other results evaluated before stoma closing [21, 22].
In this study, it is notable that there were no complications that required surgical re-intervention even though there was a high rate of AL. We suppose that the capacity of fecal diversion of FDD is comparable to the conventional defunctioning stoma.
During the maintenance period of the FDD, we controlled bowel movement with irrigation twice a day. Fecal contents cannot pass easily through non-peristaltic tube structures without modification of the solid characteristics; therefore, dilution of feces is inevitable to avoid passage disturbance. We targeted dilution before the feces solidified. From the pilot study, we found that once daily dilution sometimes resulted in passage disturbance. The time for the meals to reach the cecum, known as the orocecal transit time, is about 90 min [23], and the normal transit time of the ascending colon is 24 h [24]. Therefore, we elected for twice daily irrigation with tepid water in order to dilute the feces in the ascending colon for easy evacuation.
A limitation of this study is that all procedures were performed by experienced colorectal surgeons from a single institution with experience in animal studies. Additionally, the population of patients with preoperative chemoradiation was relatively small for the analysis of safety and effectiveness of the current procedure. Finally, we did not perform quantitative analysis regarding patient tolerance.
In conclusion, the FDD procedure showed no serious complications associated with the attachment of the FDD that required re-intervention. The capacity of fecal diversion in a situation of anastomotic wound disruption was sufficient to prevent aggravation of septicemia and to avoid re-laparotomy or temporary stoma. The FDD could be maintained for up to 6 weeks without serious complication.

Compliance with ethical standards

Disclosures

Jae Hwang Kim, Sohyun Kim, and Sang Hun Jung have no conflicts of interest or financial ties to disclose.
Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://​creativecommons.​org/​licenses/​by/​4.​0/​), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.

Unsere Produktempfehlungen

Die Chirurgie

Print-Titel

Das Abo mit mehr Tiefe

Mit der Zeitschrift Die Chirurgie erhalten Sie zusätzlich Online-Zugriff auf weitere 43 chirurgische Fachzeitschriften, CME-Fortbildungen, Webinare, Vorbereitungskursen zur Facharztprüfung und die digitale Enzyklopädie e.Medpedia.

Bis 30. April 2024 bestellen und im ersten Jahr nur 199 € zahlen!

e.Med Interdisziplinär

Kombi-Abonnement

Für Ihren Erfolg in Klinik und Praxis - Die beste Hilfe in Ihrem Arbeitsalltag

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de.

Literatur
1.
Zurück zum Zitat Poon RT, Chu KW, Ho JW, Chan CW, Law WL, Wong J (1999) Prospective evaluation of selective defunctioning stoma for low anterior resection with total mesorectal excision. World J Surg 23:463–467CrossRefPubMed Poon RT, Chu KW, Ho JW, Chan CW, Law WL, Wong J (1999) Prospective evaluation of selective defunctioning stoma for low anterior resection with total mesorectal excision. World J Surg 23:463–467CrossRefPubMed
2.
Zurück zum Zitat Huser N, Michalski CW, Erkan M, Schuster T, Rosenberg R, Kleeff J, Friess H (2008) Systematic review and meta-analysis of the role of defunctioning stoma in low rectal cancer surgery. Ann Surg 248:52–60CrossRefPubMed Huser N, Michalski CW, Erkan M, Schuster T, Rosenberg R, Kleeff J, Friess H (2008) Systematic review and meta-analysis of the role of defunctioning stoma in low rectal cancer surgery. Ann Surg 248:52–60CrossRefPubMed
3.
Zurück zum Zitat Matthiessen P, Hallbook O, Rutegard J, Simert G, Sjodahl R (2007) Defunctioning stoma reduces symptomatic anastomotic leakage after low anterior resection of the rectum for cancer: a randomized multicenter trial. Ann Surg 246:207–214CrossRefPubMedPubMedCentral Matthiessen P, Hallbook O, Rutegard J, Simert G, Sjodahl R (2007) Defunctioning stoma reduces symptomatic anastomotic leakage after low anterior resection of the rectum for cancer: a randomized multicenter trial. Ann Surg 246:207–214CrossRefPubMedPubMedCentral
4.
Zurück zum Zitat Wong NY, Eu KW (2005) A defunctioning ileostomy does not prevent clinical anastomotic leak after a low anterior resection: a prospective, comparative study. Dis Colon Rectum 48:2076–2079CrossRefPubMed Wong NY, Eu KW (2005) A defunctioning ileostomy does not prevent clinical anastomotic leak after a low anterior resection: a prospective, comparative study. Dis Colon Rectum 48:2076–2079CrossRefPubMed
5.
Zurück zum Zitat Leong AP, Londono-Schimmer EE, Phillips RK (1994) Life-table analysis of stomal complications following ileostomy. Br J Surg 81:727–729CrossRefPubMed Leong AP, Londono-Schimmer EE, Phillips RK (1994) Life-table analysis of stomal complications following ileostomy. Br J Surg 81:727–729CrossRefPubMed
6.
Zurück zum Zitat Nugent KP, Daniels P, Stewart B, Patankar R, Johnson CD (1999) Quality of life in stoma patients. Dis Colon Rectum 42:1569–1574CrossRefPubMed Nugent KP, Daniels P, Stewart B, Patankar R, Johnson CD (1999) Quality of life in stoma patients. Dis Colon Rectum 42:1569–1574CrossRefPubMed
7.
Zurück zum Zitat Yoon WH, Song IS, Chang ES (1994) Intraluminal bypass technique using a condom for protection of coloanal anastomosis. Dis Colon Rectum 37:1046–1047CrossRefPubMed Yoon WH, Song IS, Chang ES (1994) Intraluminal bypass technique using a condom for protection of coloanal anastomosis. Dis Colon Rectum 37:1046–1047CrossRefPubMed
8.
Zurück zum Zitat Ruiz PL, Facciuto EM, Facciuto ME, Rodriguez Otero JC, Pigatto J, Cominelli H (1995) New intraluminal bypass tube for management of acutely obstructed left colon. Dis Colon Rectum 38:1108–1109CrossRefPubMed Ruiz PL, Facciuto EM, Facciuto ME, Rodriguez Otero JC, Pigatto J, Cominelli H (1995) New intraluminal bypass tube for management of acutely obstructed left colon. Dis Colon Rectum 38:1108–1109CrossRefPubMed
9.
Zurück zum Zitat Ye F, Wang D, Xu X, Liu F, Lin J (2008) Use of intracolonic bypass secured by a biodegradable anastomotic ring to protect the low rectal anastomosis. Dis Colon Rectum 51:109–115CrossRefPubMed Ye F, Wang D, Xu X, Liu F, Lin J (2008) Use of intracolonic bypass secured by a biodegradable anastomotic ring to protect the low rectal anastomosis. Dis Colon Rectum 51:109–115CrossRefPubMed
11.
Zurück zum Zitat Amin AI, Ramalingam T, Sexton R, Heald RJ, Leppington-Clarke A, Moran BJ (2003) Comparison of transanal stent with defunctioning stoma in low anterior resection for rectal cancer. Br J Surg 90:581–582CrossRefPubMed Amin AI, Ramalingam T, Sexton R, Heald RJ, Leppington-Clarke A, Moran BJ (2003) Comparison of transanal stent with defunctioning stoma in low anterior resection for rectal cancer. Br J Surg 90:581–582CrossRefPubMed
12.
Zurück zum Zitat Oxlund H, Christensen H, Seyer-Hansen M, Andreassen TT (1996) Collagen deposition and mechanical strength of colon anastomoses and skin incisional wounds of rats. J Surg Res 66:25–30CrossRefPubMed Oxlund H, Christensen H, Seyer-Hansen M, Andreassen TT (1996) Collagen deposition and mechanical strength of colon anastomoses and skin incisional wounds of rats. J Surg Res 66:25–30CrossRefPubMed
13.
Zurück zum Zitat Seifert WF, Wobbes T, Hoogenhout J, de Man BM, Hendriks T (1997) Intra-operative irradiation prolongs the presence of matrix metalloproteinase activity in large bowel anastomoses of the rat. Radiat Res 147:354–361CrossRefPubMed Seifert WF, Wobbes T, Hoogenhout J, de Man BM, Hendriks T (1997) Intra-operative irradiation prolongs the presence of matrix metalloproteinase activity in large bowel anastomoses of the rat. Radiat Res 147:354–361CrossRefPubMed
14.
Zurück zum Zitat Konishi T, Watanabe T, Kishimoto J, Nagawa H (2006) Risk factors for anastomotic leakage after surgery for colorectal cancer: results of prospective surveillance. J Am Coll Surg 202:439–444CrossRefPubMed Konishi T, Watanabe T, Kishimoto J, Nagawa H (2006) Risk factors for anastomotic leakage after surgery for colorectal cancer: results of prospective surveillance. J Am Coll Surg 202:439–444CrossRefPubMed
15.
Zurück zum Zitat Frasson M, Granero-Castro P, Ramos Rodriguez JL, Flor-Lorente B, Braithwaite M, Marti Martinez E, Alvarez Perez JA, Codina Cazador A, Espi A, Garcia-Granero E, Group AS (2016) Risk factors for anastomotic leak and postoperative morbidity and mortality after elective right colectomy for cancer: results from a prospective, multicentric study of 1102 patients. Int J Colorectal Dis 31:105–114CrossRefPubMed Frasson M, Granero-Castro P, Ramos Rodriguez JL, Flor-Lorente B, Braithwaite M, Marti Martinez E, Alvarez Perez JA, Codina Cazador A, Espi A, Garcia-Granero E, Group AS (2016) Risk factors for anastomotic leak and postoperative morbidity and mortality after elective right colectomy for cancer: results from a prospective, multicentric study of 1102 patients. Int J Colorectal Dis 31:105–114CrossRefPubMed
16.
Zurück zum Zitat Kim JH, Jung SH, Kim YJ, Park SL, Kim DH (2013) Long-term fecal diverting device for the prevention of sepsis in case of colorectal anastomotic leakage: an animal experiment. Int J Colorectal Dis 28:477–484CrossRefPubMed Kim JH, Jung SH, Kim YJ, Park SL, Kim DH (2013) Long-term fecal diverting device for the prevention of sepsis in case of colorectal anastomotic leakage: an animal experiment. Int J Colorectal Dis 28:477–484CrossRefPubMed
17.
Zurück zum Zitat Rahbari NN, Weitz J, Hohenberger W, Heald RJ, Moran B, Ulrich A, Holm T, Wong WD, Tiret E, Moriya Y, Laurberg S, den Dulk M, van de Velde C, Buchler MW (2010) Definition and grading of anastomotic leakage following anterior resection of the rectum: a proposal by the International Study Group of Rectal Cancer. Surgery 147:339–351CrossRefPubMed Rahbari NN, Weitz J, Hohenberger W, Heald RJ, Moran B, Ulrich A, Holm T, Wong WD, Tiret E, Moriya Y, Laurberg S, den Dulk M, van de Velde C, Buchler MW (2010) Definition and grading of anastomotic leakage following anterior resection of the rectum: a proposal by the International Study Group of Rectal Cancer. Surgery 147:339–351CrossRefPubMed
18.
Zurück zum Zitat Kaur P, Karandikar SS, Roy-Choudhury S (2014) Accuracy of multidetector CT in detecting anastomotic leaks following stapled left-sided colonic anastomosis. Clin Radiol 69:59–62CrossRefPubMed Kaur P, Karandikar SS, Roy-Choudhury S (2014) Accuracy of multidetector CT in detecting anastomotic leaks following stapled left-sided colonic anastomosis. Clin Radiol 69:59–62CrossRefPubMed
19.
Zurück zum Zitat Pakkastie TE, Ovaska JT, Pekkala ES, Luukkonen PE, Jarvinen HJ (1997) A randomised study of colostomies in low colorectal anastomoses. Eur J Surg 163:929–933PubMed Pakkastie TE, Ovaska JT, Pekkala ES, Luukkonen PE, Jarvinen HJ (1997) A randomised study of colostomies in low colorectal anastomoses. Eur J Surg 163:929–933PubMed
20.
21.
Zurück zum Zitat Dehni N, Schlegel RD, Cunningham C, Guiguet M, Tiret E, Parc R (1998) Influence of a defunctioning stoma on leakage rates after low colorectal anastomosis and colonic J pouch-anal anastomosis. Br J Surg 85:1114–1117CrossRefPubMed Dehni N, Schlegel RD, Cunningham C, Guiguet M, Tiret E, Parc R (1998) Influence of a defunctioning stoma on leakage rates after low colorectal anastomosis and colonic J pouch-anal anastomosis. Br J Surg 85:1114–1117CrossRefPubMed
22.
Zurück zum Zitat Hallbook O, Pahlman L, Krog M, Wexner SD, Sjodahl R (1996) Randomized comparison of straight and colonic J pouch anastomosis after low anterior resection. Ann Surg 224:58–65CrossRefPubMedPubMedCentral Hallbook O, Pahlman L, Krog M, Wexner SD, Sjodahl R (1996) Randomized comparison of straight and colonic J pouch anastomosis after low anterior resection. Ann Surg 224:58–65CrossRefPubMedPubMedCentral
23.
Zurück zum Zitat Ning Y, Lou C, Huang Z, Chen D, Huang H, Chen L, Zhang B, Dai N, Zhao J, Zhen X (2016) Clinical value of radionuclide small intestine transit time measurement combined with lactulose hydrogen breath test for the diagnosis of bacterial overgrowth in irritable bowel syndrome. Hell J Nucl Med 19:124–129PubMed Ning Y, Lou C, Huang Z, Chen D, Huang H, Chen L, Zhang B, Dai N, Zhao J, Zhen X (2016) Clinical value of radionuclide small intestine transit time measurement combined with lactulose hydrogen breath test for the diagnosis of bacterial overgrowth in irritable bowel syndrome. Hell J Nucl Med 19:124–129PubMed
24.
Zurück zum Zitat Herman R, Gregorczyk A, Walega P, Kawiorski W (1994) Radiologic methods of evaluating colonic transit time in functional disorders of the large intestine. Przegl Lek 51:343–346PubMed Herman R, Gregorczyk A, Walega P, Kawiorski W (1994) Radiologic methods of evaluating colonic transit time in functional disorders of the large intestine. Przegl Lek 51:343–346PubMed
Metadaten
Titel
Fecal diverting device for the substitution of defunctioning stoma: preliminary clinical study
verfasst von
Jae Hwang Kim
Sohyun Kim
Sang Hun Jung
Publikationsdatum
14.08.2018
Verlag
Springer US
Erschienen in
Surgical Endoscopy / Ausgabe 1/2019
Print ISSN: 0930-2794
Elektronische ISSN: 1432-2218
DOI
https://doi.org/10.1007/s00464-018-6389-4

Weitere Artikel der Ausgabe 1/2019

Surgical Endoscopy 1/2019 Zur Ausgabe

Update Chirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.

S3-Leitlinie „Diagnostik und Therapie des Karpaltunnelsyndroms“

Karpaltunnelsyndrom BDC Leitlinien Webinare
CME: 2 Punkte

Das Karpaltunnelsyndrom ist die häufigste Kompressionsneuropathie peripherer Nerven. Obwohl die Anamnese mit dem nächtlichen Einschlafen der Hand (Brachialgia parästhetica nocturna) sehr typisch ist, ist eine klinisch-neurologische Untersuchung und Elektroneurografie in manchen Fällen auch eine Neurosonografie erforderlich. Im Anfangsstadium sind konservative Maßnahmen (Handgelenksschiene, Ergotherapie) empfehlenswert. Bei nicht Ansprechen der konservativen Therapie oder Auftreten von neurologischen Ausfällen ist eine Dekompression des N. medianus am Karpaltunnel indiziert.

Prof. Dr. med. Gregor Antoniadis
Berufsverband der Deutschen Chirurgie e.V.

S2e-Leitlinie „Distale Radiusfraktur“

Radiusfraktur BDC Leitlinien Webinare
CME: 2 Punkte

Das Webinar beschäftigt sich mit Fragen und Antworten zu Diagnostik und Klassifikation sowie Möglichkeiten des Ausschlusses von Zusatzverletzungen. Die Referenten erläutern, welche Frakturen konservativ behandelt werden können und wie. Das Webinar beantwortet die Frage nach aktuellen operativen Therapiekonzepten: Welcher Zugang, welches Osteosynthesematerial? Auf was muss bei der Nachbehandlung der distalen Radiusfraktur geachtet werden?

PD Dr. med. Oliver Pieske
Dr. med. Benjamin Meyknecht
Berufsverband der Deutschen Chirurgie e.V.

S1-Leitlinie „Empfehlungen zur Therapie der akuten Appendizitis bei Erwachsenen“

Appendizitis BDC Leitlinien Webinare
CME: 2 Punkte

Inhalte des Webinars zur S1-Leitlinie „Empfehlungen zur Therapie der akuten Appendizitis bei Erwachsenen“ sind die Darstellung des Projektes und des Erstellungswegs zur S1-Leitlinie, die Erläuterung der klinischen Relevanz der Klassifikation EAES 2015, die wissenschaftliche Begründung der wichtigsten Empfehlungen und die Darstellung stadiengerechter Therapieoptionen.

Dr. med. Mihailo Andric
Berufsverband der Deutschen Chirurgie e.V.