Skip to main content
Erschienen in:

07.10.2022 | Original Article

Gas leakages from gastrointestinal endoscopy system—its visualization and semi-quantification utilizing schlieren optical system in the swine models

verfasst von: Tomo Ishida, Yoshinori Hayashi, Takuro Saito, Kazuyoshi Yamamoto, Kotarou Yamashita, Koji Tanaka, Tomoki Makino, Tsuyoshi Takahashi, Yukinori Kurokawa, Hidetoshi Eguchi, Yuichiro Doki, Kiyokazu Nakajima

Erschienen in: Surgical Endoscopy | Ausgabe 3/2023

Einloggen, um Zugang zu erhalten

Abstract

Background and aims

With the global epidemic of SARS-CoV-2, there has been a growing concern regarding the risk of aerosol exposure to healthcare workers and patients during medical/surgical interventions. The Schlieren device is capable of visualizing fine gas-flows by using refractive index differences in the medium. We aimed to reveal the existence of gas leakage from gastro-intestinal endoscopy system by utilizing Schlieren device and to clarify the factors which relates to the amount of gas leakage.

Methods

The experiments were performed on the excised swine stomach while maintaining a constant pressure environment in the stomach. The System Schlieren (SS100,KatoKoken) was used to visualize possible gas leakages from forceps plugs of endoscopy. We attempted to semi-quantify the leakage by referring to the image of the gas from the forceps plug and by measuring the initial velocity and diffusion area of the leakage.

Results

Regardless of the type of forceps plugs, a certain amount of leakage was detected during both insertion and removal of forceps. The initial velocity and the diffusion area of the leakage increased with the increase in intragastric pressure. Semi-quantitative comparison showed that there was a difference in the amount of gas leakage among various forceps plugs. Furthermore, the amount of gas leakage was significantly greater in the forceps plugs that were used repeatedly.

Conclusion

It was possible to visualize gas leakages from the gastrointestinal endoscope system using the Schlieren optical device. Avoiding too high intragastric pressure and not using deteriorated plugs may reduce the risk of aerosol exposure.
Anhänge
Nur mit Berechtigung zugänglich
Literatur
1.
Zurück zum Zitat Liu Y, Ning Z, Chen Y, Guo M, Liu Y, Gali NK, Sun L, Duan Y, Cai J, Westerdahl D, Liu X, Xu K, Ho KF, Kan H, Fu Q, Lan K (2020) Aerodynamic analysis of SARS-CoV-2 in two Wuhan hospitals. Nature 582:557–560CrossRefPubMed Liu Y, Ning Z, Chen Y, Guo M, Liu Y, Gali NK, Sun L, Duan Y, Cai J, Westerdahl D, Liu X, Xu K, Ho KF, Kan H, Fu Q, Lan K (2020) Aerodynamic analysis of SARS-CoV-2 in two Wuhan hospitals. Nature 582:557–560CrossRefPubMed
2.
Zurück zum Zitat Wang C, Horby PW, Hayden FG, Gao GF (2020) A novel coronavirus outbreak of global health concern. Lancet (London, England) 395:470–473CrossRefPubMed Wang C, Horby PW, Hayden FG, Gao GF (2020) A novel coronavirus outbreak of global health concern. Lancet (London, England) 395:470–473CrossRefPubMed
3.
Zurück zum Zitat Nguyen LH, Drew DA, Graham MS, Joshi AD, Guo CG, Ma W, Mehta RS, Warner ET, Sikavi DR, Lo CH, Kwon S, Song M, Mucci LA, Stampfer MJ, Willett WC, Eliassen AH, Hart JE, Chavarro JE, Rich-Edwards JW, Davies R, Capdevila J, Lee KA, Lochlainn MN, Varsavsky T, Sudre CH, Cardoso MJ, Wolf J, Spector TD, Ourselin S, Steves CJ, Chan AT (2020) Risk of COVID-19 among front-line health-care workers and the general community: a prospective cohort study. Lancet Public Health 5:e475–e483CrossRefPubMedPubMedCentral Nguyen LH, Drew DA, Graham MS, Joshi AD, Guo CG, Ma W, Mehta RS, Warner ET, Sikavi DR, Lo CH, Kwon S, Song M, Mucci LA, Stampfer MJ, Willett WC, Eliassen AH, Hart JE, Chavarro JE, Rich-Edwards JW, Davies R, Capdevila J, Lee KA, Lochlainn MN, Varsavsky T, Sudre CH, Cardoso MJ, Wolf J, Spector TD, Ourselin S, Steves CJ, Chan AT (2020) Risk of COVID-19 among front-line health-care workers and the general community: a prospective cohort study. Lancet Public Health 5:e475–e483CrossRefPubMedPubMedCentral
4.
Zurück zum Zitat Chiu PWY, Ng SC, Inoue H, Reddy DN, Ling HuE, Cho JY, Ho LK, Hewett DG, Chiu HM, Rerknimitr R, Wang HP, Ho SH, Seo DW, Goh KL, Tajiri H, Kitano S, Chan FKL (2020) Practice of endoscopy during COVID-19 pandemic: position statements of the Asian Pacific Society for Digestive Endoscopy (APSDE-COVID statements). Gut 69:991–996CrossRefPubMed Chiu PWY, Ng SC, Inoue H, Reddy DN, Ling HuE, Cho JY, Ho LK, Hewett DG, Chiu HM, Rerknimitr R, Wang HP, Ho SH, Seo DW, Goh KL, Tajiri H, Kitano S, Chan FKL (2020) Practice of endoscopy during COVID-19 pandemic: position statements of the Asian Pacific Society for Digestive Endoscopy (APSDE-COVID statements). Gut 69:991–996CrossRefPubMed
5.
Zurück zum Zitat Sinonquel P, Roelandt P, Demedts I, Van Gerven L, Vandenbriele C, Wilmer A, Van Wijngaerden E, Bisschops R (2020) COVID-19 and gastrointestinal endoscopy: what should be taken into account? Dig Endosc 32:723–731CrossRefPubMedPubMedCentral Sinonquel P, Roelandt P, Demedts I, Van Gerven L, Vandenbriele C, Wilmer A, Van Wijngaerden E, Bisschops R (2020) COVID-19 and gastrointestinal endoscopy: what should be taken into account? Dig Endosc 32:723–731CrossRefPubMedPubMedCentral
6.
Zurück zum Zitat Mazumdar.M (2013) Principles and Techniques of Schlieren Imaging Systems. Columbia University Computer Science Technical Reports Mazumdar.M (2013) Principles and Techniques of Schlieren Imaging Systems. Columbia University Computer Science Technical Reports
7.
Zurück zum Zitat Miyake S, Ashikari K, Kato S, Takatsu T, Kuwashima H, Kaneko H, Nagai K, Watari I, Sato T, Yamaoka Y, Yamamoto T, Ryo A, Maeda S, Nakajima A, Higurashi T (2021) Severe acute respiratory syndrome coronavirus 2 prevalence in saliva and gastric and intestinal fluid in patients undergoing gastrointestinal endoscopy in coronavirus disease 2019 endemic areas: prospective cross-sectional study in Japan. Dig Endosc. https://doi.org/10.1111/den.13945CrossRefPubMedPubMedCentral Miyake S, Ashikari K, Kato S, Takatsu T, Kuwashima H, Kaneko H, Nagai K, Watari I, Sato T, Yamaoka Y, Yamamoto T, Ryo A, Maeda S, Nakajima A, Higurashi T (2021) Severe acute respiratory syndrome coronavirus 2 prevalence in saliva and gastric and intestinal fluid in patients undergoing gastrointestinal endoscopy in coronavirus disease 2019 endemic areas: prospective cross-sectional study in Japan. Dig Endosc. https://​doi.​org/​10.​1111/​den.​13945CrossRefPubMedPubMedCentral
8.
Zurück zum Zitat Takahashi H, Hirota M, Takahashi T, Yamasaki M, Miyazaki Y, Makino T, Kurokawa Y, Takiguchi S, Mori M, Doki Y, Nakajima K (2016) Simultaneous automatic insufflation and smoke-evacuation system in flexible gastrointestinal endoscopy. Endoscopy 48:579–583CrossRefPubMed Takahashi H, Hirota M, Takahashi T, Yamasaki M, Miyazaki Y, Makino T, Kurokawa Y, Takiguchi S, Mori M, Doki Y, Nakajima K (2016) Simultaneous automatic insufflation and smoke-evacuation system in flexible gastrointestinal endoscopy. Endoscopy 48:579–583CrossRefPubMed
9.
Zurück zum Zitat Yamada T, Hirota M, Tsutsui S, Kato M, Takahashi T, Yasuda K, Sumiyama K, Tsujii M, Takehara T, Mori M, Doki Y, Nakajima K (2015) Gastric endoscopic submucosal dissection under steady pressure automatically controlled endoscopy (SPACE); a multicenter randomized preclinical trial. Surg Endosc 29:2748–2755CrossRefPubMed Yamada T, Hirota M, Tsutsui S, Kato M, Takahashi T, Yasuda K, Sumiyama K, Tsujii M, Takehara T, Mori M, Doki Y, Nakajima K (2015) Gastric endoscopic submucosal dissection under steady pressure automatically controlled endoscopy (SPACE); a multicenter randomized preclinical trial. Surg Endosc 29:2748–2755CrossRefPubMed
10.
Zurück zum Zitat Hirota M, Miyazaki Y, Takahashi T, Yamasaki M, Takiguchi S, Mori M, Doki Y, Nakajima K (2014) Dynamic article: steady pressure CO2 colonoscopy; its feasibility and underlying mechanism. Dis Colon Rectum 57:1120–1128CrossRefPubMed Hirota M, Miyazaki Y, Takahashi T, Yamasaki M, Takiguchi S, Mori M, Doki Y, Nakajima K (2014) Dynamic article: steady pressure CO2 colonoscopy; its feasibility and underlying mechanism. Dis Colon Rectum 57:1120–1128CrossRefPubMed
11.
Zurück zum Zitat Kato M, Nakajima K, Yamada T, Hirota M, Miyazaki Y, Yamasaki M, Nishida T, Mori M, Doki Y, Tsujii M, Takehara T (2014) Esophageal submucosal dissection under steady pressure automatically controlled endoscopy (SPACE): a clinical feasibility study. Endoscopy 46:680–684CrossRefPubMed Kato M, Nakajima K, Yamada T, Hirota M, Miyazaki Y, Yamasaki M, Nishida T, Mori M, Doki Y, Tsujii M, Takehara T (2014) Esophageal submucosal dissection under steady pressure automatically controlled endoscopy (SPACE): a clinical feasibility study. Endoscopy 46:680–684CrossRefPubMed
12.
Zurück zum Zitat Ushimaru Y, Nakajima K, Hirota M, Miyazaki Y, Yamashita K, Saito T, Tanaka K, Makino T, Takahashi T, Kurokawa Y, Yamasaki M, Mori M, Doki Y (2020) The endoluminal pressures during flexible gastrointestinal endoscopy. Sci Rep 10:18169CrossRefPubMedPubMedCentral Ushimaru Y, Nakajima K, Hirota M, Miyazaki Y, Yamashita K, Saito T, Tanaka K, Makino T, Takahashi T, Kurokawa Y, Yamasaki M, Mori M, Doki Y (2020) The endoluminal pressures during flexible gastrointestinal endoscopy. Sci Rep 10:18169CrossRefPubMedPubMedCentral
13.
Zurück zum Zitat Black JRM, Bailey C, Przewrocka J, Dijkstra KK, Swanton C (2020) COVID-19: the case for health-care worker screening to prevent hospital transmission. Lancet (London, England) 395:1418–1420CrossRefPubMed Black JRM, Bailey C, Przewrocka J, Dijkstra KK, Swanton C (2020) COVID-19: the case for health-care worker screening to prevent hospital transmission. Lancet (London, England) 395:1418–1420CrossRefPubMed
14.
Zurück zum Zitat Verbeek JH, Rajamaki B, Ijaz S, Sauni R, Toomey E, Blackwood B, Tikka C, Ruotsalainen JH, Kilinc Balci FS (2021) Personal protective equipment for preventing highly infectious diseases due to exposure to contaminated body fluids in healthcare staff. Emergencias 33:59–61PubMed Verbeek JH, Rajamaki B, Ijaz S, Sauni R, Toomey E, Blackwood B, Tikka C, Ruotsalainen JH, Kilinc Balci FS (2021) Personal protective equipment for preventing highly infectious diseases due to exposure to contaminated body fluids in healthcare staff. Emergencias 33:59–61PubMed
15.
Zurück zum Zitat Urakawa S, Hirashita T, Momose K, Nishimura M, Nakajima K, Milsom JW (2021) Intraluminal gas escape from biopsy valves and endoscopic devices during endoscopy: caution advised during the COVID-19 era. Endosc Int Open 9:E443-e449CrossRefPubMedPubMedCentral Urakawa S, Hirashita T, Momose K, Nishimura M, Nakajima K, Milsom JW (2021) Intraluminal gas escape from biopsy valves and endoscopic devices during endoscopy: caution advised during the COVID-19 era. Endosc Int Open 9:E443-e449CrossRefPubMedPubMedCentral
16.
Zurück zum Zitat Flemming S, Hankir M, Hering I, Meybohm P, Krone M, Weissbrich B, Germer CT, Wiegering A (2020) Abdominal fluid samples (negative for SARS-CoV-2) from a critically unwell patient with respiratory COVID-19. Br J Surg 107:e259–e260CrossRefPubMedPubMedCentral Flemming S, Hankir M, Hering I, Meybohm P, Krone M, Weissbrich B, Germer CT, Wiegering A (2020) Abdominal fluid samples (negative for SARS-CoV-2) from a critically unwell patient with respiratory COVID-19. Br J Surg 107:e259–e260CrossRefPubMedPubMedCentral
17.
Zurück zum Zitat Mina MJ, Parker R, Larremore DB (2020) Rethinking Covid-19 test sensitivity—a strategy for containment. N Engl J Med 383:e120CrossRefPubMed Mina MJ, Parker R, Larremore DB (2020) Rethinking Covid-19 test sensitivity—a strategy for containment. N Engl J Med 383:e120CrossRefPubMed
Metadaten
Titel
Gas leakages from gastrointestinal endoscopy system—its visualization and semi-quantification utilizing schlieren optical system in the swine models
verfasst von
Tomo Ishida
Yoshinori Hayashi
Takuro Saito
Kazuyoshi Yamamoto
Kotarou Yamashita
Koji Tanaka
Tomoki Makino
Tsuyoshi Takahashi
Yukinori Kurokawa
Hidetoshi Eguchi
Yuichiro Doki
Kiyokazu Nakajima
Publikationsdatum
07.10.2022
Verlag
Springer US
Erschienen in
Surgical Endoscopy / Ausgabe 3/2023
Print ISSN: 0930-2794
Elektronische ISSN: 1432-2218
DOI
https://doi.org/10.1007/s00464-022-09668-y

Neu im Fachgebiet Chirurgie

Krebs-Op. besser erst nach mehrwöchigem Rauchverzicht?

Aktive Raucher haben nach onkologischen Operationen ein höheres Komplikationsrisiko als Nieraucher, aber auch als Exraucher. Damit der Rauchstopp einen Nutzen hat, darf er aber nicht zu kurz vor der Op. erfolgen, wie eine Metaanalyse nahelegt.

Periprozeduraler Myokardschaden nach NSTEMI prognostisch bedeutsam

Erleiden NSTEMI-Patienten und -Patientinnen infolge einer PCI Myokardschädigungen mit oder ohne Infarkt, erhöht das die Gesamtmortalität und das Risiko für weitere schwere Herzereignisse. Dafür sprechen zumindest Daten aus zwei italienischen Zentren.

Höhere Sterberate nach Freitags-Op.

Wer direkt vor dem Wochenende operiert wird, läuft eher Gefahr, in den folgenden Wochen und Monaten zu sterben, als Menschen mit einer Op. unmittelbar nach dem Wochenende. Der Unterschied bei der Sterberate beläuft sich nach Daten einer US-Analyse auf etwa 10%.

Ärztinnen überholen Ärzte bei Praxisgründungen

Bei Praxisgründungen haben inzwischen die Frauen deutlich die Nase vorn: Seit zehn Jahren wagen laut apoBank mehr Ärztinnen als Ärzte den Schritt in die Selbstständigkeit. In puncto Finanzierung sind sie aber vorsichtiger als die männlichen Kollegen.

Update Chirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.