Heart rate variability (HRV) in deep breathing tests and 5-min short-term recordings: agreement of ear photoplethysmography with ECG measurements, in 343 subjects

verfasst von: Stefan W. Weinschenk, Reinhard D. Beise, Jürgen Lorenz

Erschienen in: European Journal of Applied Physiology | Ausgabe 8/2016

Einloggen, um Zugang zu erhalten

Abstract

Purpose

We analyzed heart rate variability (HRV) taken by ECG and photoplethysmography (PPG) to assess their agreement. We also analyzed the sensitivity and specificity of PPG to identify subjects with low HRV as an example of its potential use for clinical applications.

Methods

The HRV parameters: mean heart rate (HR), amplitude, and ratio of heart rate oscillation (E–I difference, E/I ratio), RMSSD, SDNN, and Power LF, were measured during 1-min deep breathing tests (DBT) in 343 individuals, followed by a 5-min short-term HRV (s-HRV), where the HRV parameters: HR, SD1, SD2, SDNN, Stress Index, Power HF, Power LF, Power VLF, and Total Power, were determined as well. Parameters were compared through correlation analysis and agreement analysis by Bland–Altman plots.

Results

PPG derived parameters HR and SD2 in s-HRV showed better agreement than SD1, Power HF, and stress index, whereas in DBT HR, E/I ratio and SDNN were superior to Power LF and RMSSD. DBT yielded stronger agreement than s-HRV. A slight overestimation of PPG HRV over HCG HRV was found. HR, Total Power, and SD2 in the s-HRV, HR, Power LF, and SDNN in the DBT showed high sensitivity and specificity to detect individuals with poor HRV. Cutoff percentiles are given for the future development of PPG-based devices.

Conclusion

HRV measured by PPG shows good agreement with ECG HRV when appropriate parameters are used, and PPG-based devices can be employed as an easy screening tool to detect individuals with poor HRV, especially in the 1-min DBT test.
Literatur
Zurück zum Zitat Allen J (2007) Photoplethysmography and its application in clinical physiological measurement. Physiol Meas 28(3):R1–39CrossRefPubMed Allen J (2007) Photoplethysmography and its application in clinical physiological measurement. Physiol Meas 28(3):R1–39CrossRefPubMed
Zurück zum Zitat Bendjelid K, Suter PM, Romand JA (2004) The respiratory change in preejection period: a new method to predict fluid responsiveness. J Appl Physiol 96(1):337–342CrossRefPubMed Bendjelid K, Suter PM, Romand JA (2004) The respiratory change in preejection period: a new method to predict fluid responsiveness. J Appl Physiol 96(1):337–342CrossRefPubMed
Zurück zum Zitat Bland JM, Altman DG (1986) Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1(8476):307–310CrossRefPubMed Bland JM, Altman DG (1986) Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1(8476):307–310CrossRefPubMed
Zurück zum Zitat Charlot K, Cornolo J, Brugniaux JV, Richalet JP, Pichon A (2009) Interchangeability between heart rate and photoplethysmography variabilities during sympathetic stimulations. Physiol Meas 30(12):1357–1369CrossRefPubMed Charlot K, Cornolo J, Brugniaux JV, Richalet JP, Pichon A (2009) Interchangeability between heart rate and photoplethysmography variabilities during sympathetic stimulations. Physiol Meas 30(12):1357–1369CrossRefPubMed
Zurück zum Zitat Chreiteh SS, Belhage B, Hoppe K, Branebjerg J, Thomsen EV (2014) Sternal pulse rate variability compared with heart rate variability on healthy subjects. Conf Proc IEEE Eng Med Biol Soc 2014:3394–3397PubMed Chreiteh SS, Belhage B, Hoppe K, Branebjerg J, Thomsen EV (2014) Sternal pulse rate variability compared with heart rate variability on healthy subjects. Conf Proc IEEE Eng Med Biol Soc 2014:3394–3397PubMed
Zurück zum Zitat Constant I, Laude D, Murat I, Elghozi JL (1999) Pulse rate variability is not a surrogate for heart rate variability. Clin Sci (Lond) 97(4):391–397CrossRef Constant I, Laude D, Murat I, Elghozi JL (1999) Pulse rate variability is not a surrogate for heart rate variability. Clin Sci (Lond) 97(4):391–397CrossRef
Zurück zum Zitat Dehkordi P, Garde A, Karlen W, Wensley D, Ansermino JM, Dumont GA (2013) Pulse rate variability compared with Heart Rate Variability in children with and without sleep disordered breathing. Conf Proc IEEE Eng Med Biol Soc 2013:6563–6566PubMed Dehkordi P, Garde A, Karlen W, Wensley D, Ansermino JM, Dumont GA (2013) Pulse rate variability compared with Heart Rate Variability in children with and without sleep disordered breathing. Conf Proc IEEE Eng Med Biol Soc 2013:6563–6566PubMed
Zurück zum Zitat Giardino ND, Lehrer PM, Edelberg R (2002) Comparison of finger plethysmograph to ECG in the measurement of heart rate variability. Psychophysiology 39(2):246–253CrossRefPubMed Giardino ND, Lehrer PM, Edelberg R (2002) Comparison of finger plethysmograph to ECG in the measurement of heart rate variability. Psychophysiology 39(2):246–253CrossRefPubMed
Zurück zum Zitat Gil E, Orini M, Bailon R, Vergara JM, Mainardi L, Laguna P (2010) Photoplethysmography pulse rate variability as a surrogate measurement of heart rate variability during non-stationary conditions. Physiol Meas 31(9):1271–1290CrossRefPubMed Gil E, Orini M, Bailon R, Vergara JM, Mainardi L, Laguna P (2010) Photoplethysmography pulse rate variability as a surrogate measurement of heart rate variability during non-stationary conditions. Physiol Meas 31(9):1271–1290CrossRefPubMed
Zurück zum Zitat Giles D, Draper N, Neil W (2016) Validity of the Polar V800 heart rate monitor to measure RR intervals at rest. Eur J Appl Physiol 116(3):563–571CrossRefPubMed Giles D, Draper N, Neil W (2016) Validity of the Polar V800 heart rate monitor to measure RR intervals at rest. Eur J Appl Physiol 116(3):563–571CrossRefPubMed
Zurück zum Zitat Heathers JA (2013) Smartphone-enabled pulse rate variability: an alternative methodology for the collection of heart rate variability in psychophysiological research. Int J Psychophysiol 89(3):297–304CrossRefPubMed Heathers JA (2013) Smartphone-enabled pulse rate variability: an alternative methodology for the collection of heart rate variability in psychophysiological research. Int J Psychophysiol 89(3):297–304CrossRefPubMed
Zurück zum Zitat Khandoker AH, Karmakar CK, Palaniswami M (2011) Comparison of pulse rate variability with heart rate variability during obstructive sleep apnea. Med Eng Phys 33(2):204–209CrossRefPubMed Khandoker AH, Karmakar CK, Palaniswami M (2011) Comparison of pulse rate variability with heart rate variability during obstructive sleep apnea. Med Eng Phys 33(2):204–209CrossRefPubMed
Zurück zum Zitat Lollgen D, Mueck-Weymann M, Beise RD (2009) The deep breathing test: median-based expiration-inspiration difference is the measure of choice. Muscle Nerve 39(4):536–544CrossRefPubMed Lollgen D, Mueck-Weymann M, Beise RD (2009) The deep breathing test: median-based expiration-inspiration difference is the measure of choice. Muscle Nerve 39(4):536–544CrossRefPubMed
Zurück zum Zitat Lu S, Zhao H, Ju K, Shin K, Lee M, Shelley K et al (2008) Can photoplethysmography variability serve as an alternative approach to obtain heart rate variability information? J Clin Monit Comput 22(1):23–29CrossRefPubMed Lu S, Zhao H, Ju K, Shin K, Lee M, Shelley K et al (2008) Can photoplethysmography variability serve as an alternative approach to obtain heart rate variability information? J Clin Monit Comput 22(1):23–29CrossRefPubMed
Zurück zum Zitat Radespiel-Troger M, Rauh R, Mahlke C, Gottschalk T, Muck-Weymann M (2003) Agreement of two different methods for measurement of heart rate variability. Clin Auton Res 13(2):99–102CrossRefPubMed Radespiel-Troger M, Rauh R, Mahlke C, Gottschalk T, Muck-Weymann M (2003) Agreement of two different methods for measurement of heart rate variability. Clin Auton Res 13(2):99–102CrossRefPubMed
Zurück zum Zitat Rauh R, Limley R, Bauer RD, Radespiel-Troger M, Mueck-Weymann M (2004) Comparison of heart rate variability and pulse rate variability detected with photoplethysmography. Proc Soc Photo-Opt Instrum Eng 5474:115–126 Rauh R, Limley R, Bauer RD, Radespiel-Troger M, Mueck-Weymann M (2004) Comparison of heart rate variability and pulse rate variability detected with photoplethysmography. Proc Soc Photo-Opt Instrum Eng 5474:115–126
Zurück zum Zitat Schafer A, Vagedes J (2013) How accurate is pulse rate variability as an estimate of heart rate variability? A review on studies comparing photoplethysmographic technology with an electrocardiogram. Int J Cardiol 166(1):15–29CrossRefPubMed Schafer A, Vagedes J (2013) How accurate is pulse rate variability as an estimate of heart rate variability? A review on studies comparing photoplethysmographic technology with an electrocardiogram. Int J Cardiol 166(1):15–29CrossRefPubMed
Zurück zum Zitat Shiyovich A, Jafari J, Blaer Y, Rehby H, Orlov I, Cohen D et al (2010) Respiratory stress response: a novel diagnostic method for detection of significant coronary artery disease from finger pulse wave analysis during brief respiratory exercise. Am J Med Sci 339(5):440–447CrossRefPubMed Shiyovich A, Jafari J, Blaer Y, Rehby H, Orlov I, Cohen D et al (2010) Respiratory stress response: a novel diagnostic method for detection of significant coronary artery disease from finger pulse wave analysis during brief respiratory exercise. Am J Med Sci 339(5):440–447CrossRefPubMed
Zurück zum Zitat TaskForce (1996) Heart rate variability: standards of measurement, physiological interpretation and clinical use. Task Force of the European Society of Cardiology and the North American Society of Pacing and Electrophysiology. Circulation 93(5):1043–1065 TaskForce (1996) Heart rate variability: standards of measurement, physiological interpretation and clinical use. Task Force of the European Society of Cardiology and the North American Society of Pacing and Electrophysiology. Circulation 93(5):1043–1065
Zurück zum Zitat Urbaszek W (1976) Hemodynamic aspects in the evaluation of heart function using systolic time intervals. Z Gesamte Inn Med 31(14):513–520PubMed Urbaszek W (1976) Hemodynamic aspects in the evaluation of heart function using systolic time intervals. Z Gesamte Inn Med 31(14):513–520PubMed
Zurück zum Zitat Waksman R, Sushinsky S, Okubagzi P, Landry P, Torguson R, Bui A et al (2010) An innovative noninvasive respiratory stress test indicates significant coronary artery disease. Cardiovasc Revasc Med 11(1):20–28CrossRefPubMed Waksman R, Sushinsky S, Okubagzi P, Landry P, Torguson R, Bui A et al (2010) An innovative noninvasive respiratory stress test indicates significant coronary artery disease. Cardiovasc Revasc Med 11(1):20–28CrossRefPubMed
Zurück zum Zitat Wong JS, Lu WA, Wu KT, Liu M, Chen GY, Kuo CD (2012) A comparative study of pulse rate variability and heart rate variability in healthy subjects. J Clin Monit Comput 26(2):107–114CrossRefPubMed Wong JS, Lu WA, Wu KT, Liu M, Chen GY, Kuo CD (2012) A comparative study of pulse rate variability and heart rate variability in healthy subjects. J Clin Monit Comput 26(2):107–114CrossRefPubMed
Metadaten
Titel
Heart rate variability (HRV) in deep breathing tests and 5-min short-term recordings: agreement of ear photoplethysmography with ECG measurements, in 343 subjects
verfasst von
Stefan W. Weinschenk
Reinhard D. Beise
Jürgen Lorenz
Publikationsdatum
08.06.2016
Verlag
Springer Berlin Heidelberg
Erschienen in
European Journal of Applied Physiology / Ausgabe 8/2016
Print ISSN: 1439-6319
Elektronische ISSN: 1439-6327
DOI
https://doi.org/10.1007/s00421-016-3401-3

Weitere Artikel der Ausgabe 8/2016

European Journal of Applied Physiology 8/2016 Zur Ausgabe