High-pitch low-voltage CT coronary artery calcium scoring with tin filtration: accuracy and radiation dose reduction

Zeitschrift:
European Radiology > Ausgabe 7/2018
Autoren:
Georg Apfaltrer, Moritz H. Albrecht, U. Joseph Schoepf, Taylor M. Duguay, Carlo N. De Cecco, John W. Nance, Domenico De Santis, Paul Apfaltrer, Marwen H. Eid, Chelsea D. Eason, Zachary M. Thompson, Maximilian J. Bauer, Akos Varga-Szemes, Brian E. Jacobs, Erich Sorantin, Christian Tesche

Abstract

Objectives

To investigate diagnostic accuracy and radiation dose of high-pitch CT coronary artery calcium scoring (CACS) with tin filtration (Sn100kVp) versus standard 120kVp high-pitch acquisition.

Methods

78 patients (58% male, 61.5±9.1 years) were prospectively enrolled. Subjects underwent clinical 120kVp high-pitch CACS using third-generation dual-source CT followed by additional high-pitch Sn100kVp acquisition. Agatston scores, calcium volume scores, Agatston score categories, percentile-based risk categorization and radiation metrics were compared.

Results

61/78 patients showed coronary calcifications. Median Agatston scores were 34.9 [0.7–197.1] and 41.7 [0.7–207.2] and calcium volume scores were 34.1 [0.7–218.0] for Sn100kVp and 35.7 [1.1–221.0] for 120kVp acquisitions, respectively (both p<0.0001). Bland-Altman analysis revealed underestimated Agatston scores and calcium volume scores with Sn100kVp versus 120kVp acquisitions (mean difference: 16.4 and 11.5). However, Agatston score categories and percentile-based risk categories showed excellent agreement (ĸ=0.98 and ĸ=0.99). Image noise was 25.8±4.4HU and 16.6±2.9HU in Sn100kVp and 120kVp scans, respectively (p<0.0001). Dose-length-product was 9.9±4.8mGy*cm and 40.9±14.4mGy*cm with Sn100kVp and 120kVp scans, respectively (p<0.0001). This resulted in significant effective radiation dose reduction (0.13±0.07mSv vs. 0.57±0.2mSv, p<0.0001) for Sn100kVp acquisitions.

Conclusion

CACS using high-pitch low-voltage tin-filtered acquisitions demonstrates excellent agreement in Agatston score and percentile-based cardiac risk categorization with standard 120kVp high-pitch acquisitions. Furthermore, radiation dose was significantly reduced by 78% while maintaining accurate risk prediction.

Key points

• Coronary artery calcium scoring with tin filtration reduces radiation dose by 78%.
• There is excellent correlation between high-pitch Sn100kVp and standard 120kVp acquisitions.
• Excellent agreement regarding Agatston score categories and percentile-based risk categorization was achieved.
• No cardiac risk reclassifications were observed using Sn100kVp coronary artery calcium scoring.

Bitte loggen Sie sich ein, um Zugang zu diesem Inhalt zu erhalten

★ PREMIUM-INHALT
e.Med Interdisziplinär

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de. Zusätzlich können Sie eine Zeitschrift Ihrer Wahl in gedruckter Form beziehen – ohne Aufpreis.

Weitere Produktempfehlungen anzeigen
Literatur
Über diesen Artikel

Weitere Artikel der Ausgabe 7/2018

European Radiology 7/2018 Zur Ausgabe
  1. Sie können e.Med Radiologie 14 Tage kostenlos testen (keine Print-Zeitschrift enthalten). Der Test läuft automatisch und formlos aus. Es kann nur einmal getestet werden.

Neu im Fachgebiet Radiologie

Meistgelesene Bücher aus der Radiologie

2016 | Buch

Medizinische Fremdkörper in der Bildgebung

Thorax, Abdomen, Gefäße und Kinder

Dieses einzigartige Buch enthält ca. 1.600 hochwertige radiologische Abbildungen und Fotos iatrogen eingebrachter Fremdmaterialien im Röntgenbild und CT.

Herausgeber:
Dr. med. Daniela Kildal

2011 | Buch

Atlas Klinische Neuroradiologie des Gehirns

Radiologie lebt von Bildern! Der vorliegende Atlas trägt dieser Tatsache Rechnung. Sie finden zu jedem Krankheitsbild des Gehirns Referenzbilder zum Abgleichen mit eigenen Befunden.

Autoren:
Priv.-Doz. Dr. med. Jennifer Linn, Prof. Dr. med. Martin Wiesmann, Prof. Dr. med. Hartmut Brückmann

Mail Icon II Newsletter

Bestellen Sie unseren kostenlosen Newsletter Update Radiologie und bleiben Sie gut informiert – ganz bequem per eMail.

Bildnachweise