26.05.2021 | Original Article

Identification of various orthodontic materials as foreign bodies via panoramic radiography, cone beam computed tomography, magnetic resonance imaging, and ultrasonography: an in vitro study

verfasst von: Ozlem Isman, Eren Isman

Erschienen in: Oral Radiology | Ausgabe 3/2021

Einloggen, um Zugang zu erhalten

Abstract

Objectives

We evaluated the in vitro detection sensitivity of orthodontic materials (serving as foreign bodies) using panoramic radiography, cone beam computed tomography (CBCT), magnetic resonance imaging (MRI), and ultrasonography.

Methods

Five different orthodontic materials served as foreign bodies: titanium–molybdenum alloy wire (TMA; ORMCO, Orange, CA, USA; 0.017 × 0.025 in in cross-sectional dimensions and 1 cm long); stainless steel bracket tooth #34 (American Orthodontics, Sheboygan, WI, USA); a monocrystalline, sapphire ceramic bracket tooth #34 (Skyortho Dental Supplies Medical, China); a polycrystalline alumina clear bracket, Damon clear bracket tooth #34 (ORMCO); and a 1 × 1 × 0.1 cm polyurethane-based thermoplastic material, Invisalign clear aligner (Align Technology, San Jose, CA, USA). Panoramic radiography, CBCT, MRI, and ultrasonography were used, and four observers scored all findings independently.

Results

The TMA and stainless steel bracket were visualised in all fields by panoramic radiography and CBCT. The sapphire and Damon brackets were very clear on CBCT. The Invisalign in air was evident only on CBCT. MRI was unable to identify any material in muscle. Ultrasonography detected the TMA, sapphire bracket, and the Invisalign in muscle but only the TMA on bone.

Conclusions

Panoramic radiography does not reveal nonmetallic orthodontic equipment in air and reveals them only poorly in muscle. CBCT was the optimal imaging modality for all materials in all fields except for the Invisalign in muscle and bone. CBCT was the only method that revealed the Invisalign in air. MRI and ultrasonography should be used to detect orthodontic materials in muscle.
Literatur
1.
Zurück zum Zitat Aras MH, Miloglu O, Barutcugil C, Kantarci M, Ozcan E, Harorli A. Comparison of the sensitivity for detecting foreign bodies among conventional plain radiography, computed tomography and ultrasonography. Dentomaxillofac Radiol. 2010;39:72–8.CrossRef Aras MH, Miloglu O, Barutcugil C, Kantarci M, Ozcan E, Harorli A. Comparison of the sensitivity for detecting foreign bodies among conventional plain radiography, computed tomography and ultrasonography. Dentomaxillofac Radiol. 2010;39:72–8.CrossRef
2.
Zurück zum Zitat Anderson MA, Newmeyer WL 3rd, Kilgore ES Jr. Diagnosis and treatment of retained foreign bodies in the hand. Am J Surg. 1982;144:63–7.CrossRef Anderson MA, Newmeyer WL 3rd, Kilgore ES Jr. Diagnosis and treatment of retained foreign bodies in the hand. Am J Surg. 1982;144:63–7.CrossRef
3.
Zurück zum Zitat Javadrashid R, Fouladi DF, Golamian M, Hajalioghli P, Daghighi MH, Shahmorady Z, et al. Visibility of different foreign bodies in the maxillofacial region using plain radiography, CT, MRI and ultrasonography: an in vitro study. Dentomaxillofac Radiol. 2015;44:20140229.CrossRef Javadrashid R, Fouladi DF, Golamian M, Hajalioghli P, Daghighi MH, Shahmorady Z, et al. Visibility of different foreign bodies in the maxillofacial region using plain radiography, CT, MRI and ultrasonography: an in vitro study. Dentomaxillofac Radiol. 2015;44:20140229.CrossRef
4.
Zurück zum Zitat Atala MH, Atala N, Yegin E, Bayrak S. Comparison of radiopacity of current restorative CAD/CAM blocks with digital radiography. J Esthet Restor Dent. 2019;31:88–92.CrossRef Atala MH, Atala N, Yegin E, Bayrak S. Comparison of radiopacity of current restorative CAD/CAM blocks with digital radiography. J Esthet Restor Dent. 2019;31:88–92.CrossRef
5.
Zurück zum Zitat Beuf O, Lissac M, Cremillieux Y, Briguet A. Correlation between magnetic resonance imaging disturbances and the magnetic susceptibility of dental materials. Dent Mater. 1994;10:265–8.CrossRef Beuf O, Lissac M, Cremillieux Y, Briguet A. Correlation between magnetic resonance imaging disturbances and the magnetic susceptibility of dental materials. Dent Mater. 1994;10:265–8.CrossRef
6.
Zurück zum Zitat Klinke T, Daboul A, Maron J, Gredes T, Puls R, Jaghsi A, et al. Artifacts in magnetic resonance imaging and computed tomography caused by dental materials. PLoS ONE. 2012;7:e31766.CrossRef Klinke T, Daboul A, Maron J, Gredes T, Puls R, Jaghsi A, et al. Artifacts in magnetic resonance imaging and computed tomography caused by dental materials. PLoS ONE. 2012;7:e31766.CrossRef
7.
Zurück zum Zitat Stockmann P, Vairaktaris E, Fenner M, Tudor C, Neukam FW, Nkenke E. Conventional radiographs: are they still the standard in localization of projectiles? Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2007;104:e71–5.CrossRef Stockmann P, Vairaktaris E, Fenner M, Tudor C, Neukam FW, Nkenke E. Conventional radiographs: are they still the standard in localization of projectiles? Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2007;104:e71–5.CrossRef
8.
Zurück zum Zitat Eggers G, Welzel T, Mukhamadiev D, Wortche R, Hassfeld S, Muhling J. X-ray-based volumetric imaging of foreign bodies: a comparison of computed tomography and digital volume tomography. J Oral Maxillofac Surg. 2007;65:1880–5.CrossRef Eggers G, Welzel T, Mukhamadiev D, Wortche R, Hassfeld S, Muhling J. X-ray-based volumetric imaging of foreign bodies: a comparison of computed tomography and digital volume tomography. J Oral Maxillofac Surg. 2007;65:1880–5.CrossRef
9.
Zurück zum Zitat Orlinsky M, Knittel P, Feit T, Chan L, Mandavia D. The comparative accuracy of radiolucent foreign body detection using ultrasonography. Am J Emerg Med. 2000;18:401–3.CrossRef Orlinsky M, Knittel P, Feit T, Chan L, Mandavia D. The comparative accuracy of radiolucent foreign body detection using ultrasonography. Am J Emerg Med. 2000;18:401–3.CrossRef
10.
Zurück zum Zitat Oikarinen KS, Nieminen TM, Makarainen H, Pyhtinen J. Visibility of foreign bodies in soft tissue in plain radiographs, computed tomography, magnetic resonance imaging, and ultrasound. An in vitro study. Int J Oral Maxillofac Surg. 1993;22:119–24.CrossRef Oikarinen KS, Nieminen TM, Makarainen H, Pyhtinen J. Visibility of foreign bodies in soft tissue in plain radiographs, computed tomography, magnetic resonance imaging, and ultrasound. An in vitro study. Int J Oral Maxillofac Surg. 1993;22:119–24.CrossRef
11.
Zurück zum Zitat Schulze R, Heil U, Gross D, Bruellmann DD, Dranischnikow E, Schwanecke U, et al. Artefacts in CBCT: a review. Dentomaxillofac Radiol. 2011;40:265–73.CrossRef Schulze R, Heil U, Gross D, Bruellmann DD, Dranischnikow E, Schwanecke U, et al. Artefacts in CBCT: a review. Dentomaxillofac Radiol. 2011;40:265–73.CrossRef
12.
Zurück zum Zitat Isman O, Aktan AM, Ertas ET. Evaluating the effects of orthodontic materials, field of view, and artifact reduction mode on accuracy of CBCT-based caries detection. Clin Oral Investig. 2020;24:2487–96.CrossRef Isman O, Aktan AM, Ertas ET. Evaluating the effects of orthodontic materials, field of view, and artifact reduction mode on accuracy of CBCT-based caries detection. Clin Oral Investig. 2020;24:2487–96.CrossRef
14.
Zurück zum Zitat Papadopoulou AK, Cantele A, Polychronis G, Zinelis S, Eliades T. Changes in roughness and mechanical properties of invisalign((R)) appliances after one- and two-weeks use. Materials (Basel). 2019;12:2406.CrossRef Papadopoulou AK, Cantele A, Polychronis G, Zinelis S, Eliades T. Changes in roughness and mechanical properties of invisalign((R)) appliances after one- and two-weeks use. Materials (Basel). 2019;12:2406.CrossRef
15.
Zurück zum Zitat Qi HJ, Boyce MC. Stress-strain behavior of thermoplastic polyurethanes. Mech Mater. 2005;37:817–39.CrossRef Qi HJ, Boyce MC. Stress-strain behavior of thermoplastic polyurethanes. Mech Mater. 2005;37:817–39.CrossRef
16.
Zurück zum Zitat Bittner RC, Felix R. Magnetic resonance (MR) imaging of the chest: state-of-the-art. Eur Respir J. 1998;11:1392–404.CrossRef Bittner RC, Felix R. Magnetic resonance (MR) imaging of the chest: state-of-the-art. Eur Respir J. 1998;11:1392–404.CrossRef
17.
Zurück zum Zitat Destine D, Mizutani H, Igarashi Y. Metallic artifacts in MRI caused by dental alloys and magnetic keeper. Nihon Hotetsu Shika Gakkai Zasshi. 2008;52:205–10.CrossRef Destine D, Mizutani H, Igarashi Y. Metallic artifacts in MRI caused by dental alloys and magnetic keeper. Nihon Hotetsu Shika Gakkai Zasshi. 2008;52:205–10.CrossRef
18.
Zurück zum Zitat Wylezinska M, Pinkstone M, Hay N, Scott AD, Birch MJ, Miquel ME. Impact of orthodontic appliances on the quality of craniofacial anatomical magnetic resonance imaging and real-time speech imaging. Eur J Orthod. 2015;37:610–7.CrossRef Wylezinska M, Pinkstone M, Hay N, Scott AD, Birch MJ, Miquel ME. Impact of orthodontic appliances on the quality of craniofacial anatomical magnetic resonance imaging and real-time speech imaging. Eur J Orthod. 2015;37:610–7.CrossRef
19.
Zurück zum Zitat Cassetta M, Pranno N, Stasolla A, Orsogna N, Fierro D, Cavallini C, et al. The effects of a common stainless steel orthodontic bracket on the diagnostic quality of cranial and cervical 3T- MR images: a prospective, case-control study. Dentomaxillofac Radiol. 2017;46:20170051.CrossRef Cassetta M, Pranno N, Stasolla A, Orsogna N, Fierro D, Cavallini C, et al. The effects of a common stainless steel orthodontic bracket on the diagnostic quality of cranial and cervical 3T- MR images: a prospective, case-control study. Dentomaxillofac Radiol. 2017;46:20170051.CrossRef
20.
Zurück zum Zitat Manthey DE, Storrow AB, Milbourn JM, Wagner BJ. Ultrasound versus radiography in the detection of soft-tissue foreign bodies. Ann Emerg Med. 1996;28:7–9.CrossRef Manthey DE, Storrow AB, Milbourn JM, Wagner BJ. Ultrasound versus radiography in the detection of soft-tissue foreign bodies. Ann Emerg Med. 1996;28:7–9.CrossRef
21.
Zurück zum Zitat Schlager D, Sanders AB, Wiggins D, Boren W. Ultrasound for the detection of foreign bodies. Ann Emerg Med. 1991;20:189–91.CrossRef Schlager D, Sanders AB, Wiggins D, Boren W. Ultrasound for the detection of foreign bodies. Ann Emerg Med. 1991;20:189–91.CrossRef
Metadaten
Titel
Identification of various orthodontic materials as foreign bodies via panoramic radiography, cone beam computed tomography, magnetic resonance imaging, and ultrasonography: an in vitro study
verfasst von
Ozlem Isman
Eren Isman
Publikationsdatum
26.05.2021
Verlag
Springer Singapore
Erschienen in
Oral Radiology / Ausgabe 3/2021
Print ISSN: 0911-6028
Elektronische ISSN: 1613-9674
DOI
https://doi.org/10.1007/s11282-021-00537-3

Weitere Artikel der Ausgabe 3/2021

Oral Radiology 3/2021 Zur Ausgabe

Newsletter

Bestellen Sie unseren kostenlosen Newsletter Update Zahnmedizin und bleiben Sie gut informiert – ganz bequem per eMail.