Open Access 01.12.2018 | Research

Image-guided study of inter-fraction and intra-fraction set-up variability and margins in reverse semi-decubitus breast radiotherapy

verfasst von: Jie Lee, Shih-Hua Liu, Jhen-Bin Lin, Meng-Hao Wu, Chieh-Ju Wu, Hung-Chi Tai, Shih-Ming Hsu, Yin-Ju Chen, Jo-Chiao Tai, Yu-Jen Chen

Erschienen in: Radiation Oncology | Ausgabe 1/2018

Abstract

Background

This study aimed to evaluate the inter-fraction set-up error and intra-fraction motion during reverse semi-decubitus (RSD) breast radiotherapy, and to determine a planning target volume (PTV) margin.

Material and methods

Pre- and post-treatment cone-beam computed tomography (CBCT) scans were prospectively acquired at fractions 1, 4, 7, 8, 11, and 14 for 30 patients who underwent RSD breast radiotherapy. Online correction for initial set-up error greater than 5 mm or 2° was performed and post-correction CBCT was acquired. An off-line analysis was performed to quantify initial and residual inter-fraction set-up errors and intra-fraction motion in three-dimensions. Patient inter-fraction errors were analysed for time trends during the course of radiotherapy. PTV margins were calculated from the systematic and random errors.

Results

The initial inter-fraction population systematic errors were 1.8–3.3 mm (translation) and 0.5° (rotation); random errors were 1.8–2.1 mm (translation) and 0.3–0.5° (rotation). After online correction, the residual inter-fraction population systematic errors were 1.2–1.8 mm (translation) and 0.3–0.4° (rotation); random errors were 1.4–1.6 mm (translation) and 0.3–0.4° (rotation). Intra-fraction population systematic and random errors were ≤ 1.3 mm (translation) and ≤ 0.2° (rotation). The magnitude of inter-fraction set-up errors in the anterior-posterior direction, roll, and yaw were significantly correlated with higher body weight and body mass index (BMI). The inter-fraction set-up error did not change significantly as a function of time during the course of radiotherapy. The magnitude of intra-fraction motion was not correlated with patient characteristics and treatment time. The total PTV margins accounting for pre-correction and intra-fraction errors were 6.5–10.2 mm; those accounting for post-correction and intra-fraction errors were 4.7–6.3 mm.

Conclusions

CBCT is an effective modality to evaluate and improve the inter-fraction set-up reproducibility in RSD breast radiotherapy, particularly for patients with higher BMI. Intra-fraction motion was minimal during RSD breast radiotherapy.
Abkürzungen
ANOVA
Analysis of variance
AP
Anterior–posterior
BMI
Body mass index
CBCT
Cone-beam computed tomography
CTV
Clinical target volume
DIBH
Deep inspiration breath-hold
LAD
Left anterior descending coronary artery
PTV
Planning target volume
RL
Right–left
RSD
Reverse semi-decubitus
SD
Standard deviation
SI
Superior–inferior

Introduction

Several studies have shown an increase in the rate of ischaemic heart disease after adjuvant radiotherapy for left-sided breast cancer [14]. The occurrences of acute coronary events exhibit dose-effect relationships [35]. To improve the therapeutic ratio of breast radiotherapy, efforts have been made to develop simulation and treatment techniques in order to reduce the volume of cardiac irradiation.
Deep inspiration breath-hold (DIBH) and prone are widely used to improve cardiac dosimetry [615]. However, certain factors may preclude their use, such as intolerance of the technique by the patient or requirement for specialized equipment. The reasons why patients may not be able to tolerate breath-hold include medical co-morbidity, anxiety, inability to tolerate specialized equipment, or language barriers. The benefits of the prone technique in cardiac sparing might be limited to large breasted women, and may in fact be detrimental in women with small breasts [1518]. A novel free-breathing technique for left breast irradiation in the reverse semi-decubitus (RSD) position could benefit patients unable to tolerate breath-hold, by reducing the cardiac dose [19, 20]. The RSD simulation was performed by rotating the patient into a semi-lateral decubitus position, with the right side towards the treatment couch and the left side elevated. To ensure that radiotherapy is given in a safe and consistent manner, the reproducibility of the RSD, particularly that of rotation, is of great importance. However, the translational and rotational inter-fraction set-up error and intra-fraction motion are unknown. Cone-beam computed tomography (CBCT) can help reduce the setup error and random deviation, by quantifying the three-dimensional translational and rotational errors [21, 22]. This study aimed to evaluate the translational and rotational inter-fraction set-up error and intra-fraction motion using CBCT and calculate an appropriate clinical target volume (CTV) to planning target volume (PTV) margin for RSD breast radiotherapy.

Materials and methods

Patients

This study was approved by the Institutional Review Board and all procedures were performed according to the Declaration of Helsinki. Thirty patients with early-stage left-sided breast cancer were enrolled at our institutions. All patients had undergone breast-conserving surgery, followed by left breast radiotherapy, between September 2017 and January 2018.

Patient positioning and image acquisition

All patients undergoing breast radiotherapy were scanned in the RSD position on the custom-made Alpha Cradle (Fig. 1). RSD simulation was performed by rotating the patient into a semi-lateral decubitus position, with the right side towards the treatment couch and the left side elevated [19, 20]. The patient’s arms were abducted above the head and immobilized using an Alpha Cradle. Markers were placed ipsilaterally, 2 cm lateral to all palpable breast tissue along the midaxillary line and midsternal line. Surgical scars and all visible breast tissue were circled with wires. Patients were scanned using the Philips Brilliance CT Big Bore scanner (Philips Healthcare, Amsterdam, Netherlands). The CT images were acquired from the C6 vertebral body to the diaphragm at 3-mm slice intervals. The simulation and treatment were performed while free-breathing.

Treatment

Figure 1d shows the axial RSD CT planning images. The CTV encompassed the breast tissue visualized on CT. The CTV was limited by the pectoral fascia. The PTV was generated by adding a 7-mm margin around the CTV, except at the skin surface. Mammary chain or supraclavicular fossa irradiation was not performed. The organs at risk included the heart, left anterior descending coronary artery (LAD), and LAD region, which were delineated based on the previously published cardiac atlas [2325]. All plans were implemented using a three-dimensional treatment planning system (Pinnacle, version 9.8; Phillips Medical Systems, Fitchburg, WI, USA). Patients were treated using 6–10 MV photons with 40 Gy in 15 fractions over 3 weeks. The tangential wedge-based field plans were applied, and ≥ 90% of the whole breast CTV was encompassed by the 95% isodose.

CBCT analysis

The RSD position was reproduced during treatment by aligning skin tattoos to lasers, with photographs used for assistance where necessary. The CBCT images of the chest were acquired before (pre-treatment CBCT) and after (post-treatment CBCT) treatment at fractions 1, 4, 7, 8, 11, and 14, by using the Elekta Versa-HD X-ray Volume Imaging system (Elekta, Crawley, UK). The pre-treatment CBCT data were registered online to the treatment planning CT. If there was rotational error greater than 2°, the patient was manually repositioned and another CBCT scan was acquired. Translational displacements of greater than 5 mm were corrected online by using automated treatment table movement; then, a second CBCT was performed to confirm the accuracy of correction.
All CBCT data were analysed offline. Automated rigid registration was implemented by using the Pinnacle, with priority given to accurate registration of the sternum and ribs on the left side to achieve a position that would provide the most accurate bony reproducibility. The displacement of the isocentre was calculated in the three axes (RL: right–left; AP: anterior–posterior; and SI: superior–inferior). The + and – signs indicate right/anterior/superior and left/posterior/inferior directions, respectively. The nomenclature used in this study was as follows: roll is a rotation about the SI axis, yaw is a rotation about the AP axis, and pitch is a rotation about the RL axis. The pre-correction CBCT scan acquired after the in-room setup was used to calculate the initial inter-fraction error. The post-correction CBCT scan acquired after any corrections for translational or rotational set-up error, or the pre-correction CBCT scan for fractions where the initial setup was within ±5 mm set-up error, was used to calculate the residual inter-fraction error. The difference between the pre-correction and post-treatment CBCT was used to calculate intra-fraction motion for patients with initial set-up without correction; the post-correction and post-treatment CBCT was used for intra-fraction motion analysis for patients with correction for translational or rotational set-up error, as previously mentioned. The data are presented as the mean ± standard deviation (SD).

Predictive factors for the inter-fraction set-up error and intra-fraction motion

Patient characteristics, such as age, body weight, and body mass index (BMI) were recorded to examine predictive factors for inter-fraction set-up error and intra-fraction motion. Correlations between set-up errors and patient body characteristics were analysed by using Pearson’s product-moment correlation coefficient (R), and the t-test was used to test the significance of the correlation. The treatment time was defined as the time interval between pre-correction and post-treatment CBCT acquisition time for patients with initial set-up without correction; the time interval between post-correction and post-treatment CBCT acquisition time was used for patients having correction for translational or rotational set-up error. The correlation between treatment time and intra-fraction motion was also analysed. One-way repeated measures analysis of variance (ANOVA) was used to analyse the time dependency of inter-fraction set-up errors. The significance level was set to < 0.05. The patients were divided into two groups on the basis of the median values of factors with significance, and the Student’s t-test was used to analyse differences between the groups.

Calculation of PTV margin

Mean displacements and SD were calculated in three-dimensions for each patient. Population systematic errors (Σ) were calculated from the SD of all mean displacements, and population random errors (σ) from the root mean square of all SD values [26]. The CTV–PTV margins were calculated based on the van Herk formula: 2.5Σ + 0.7σ [27]. The overall Σ and σ were defined as the square root of the quadratic sum of the inter-fraction and intra-fraction Σ and σ, respectively [26].

Results

The median (range) age, body weight, and BMI were 52 (38–75) years, 63.0 (47.0–88.2) kg, and 25.9 (18.1–33.6) kg/m2, respectively. A total of 435 CBCT images were acquired from 30 left breast cancer patients, including 180 pre-correction images, 75 post-correction images, and 180 post-treatment images. Among all patients, the median number of fractions with set-up errors requiring correction and re-imaging was two (range, 0–5). When dividing patients into two groups on the basis of the median value of BMI, the median numbers of fractions with set-up errors requiring correction and re-imaging were zero (range, 0–3) and four (range, 1–5) for patients with ≤25.9 kg/m2 and > 25.9 kg/m2, respectively (p < 0.001).

Inter-fraction set-up error

The distributions of inter-fraction setup errors in each of the three orthogonal directions were calculated by using the 180 pre-correction images and 75 post-correction images (Fig. 2). For initial inter-fraction shifts, the numbers of fractions exceeding ±5 mm in the RL, AP, and SI directions were 5 (2.7%), 51 (28.3%) and 25 (13.9%), respectively; 16 (8.9%), 14 (7.8%) and 8 (4.4%) fractions exceeded 2° for roll, yaw, and pitch, respectively. In all directions, the distributions obtained from the post-correction scans were narrower than those obtained from the pre-treatment scans, all lying within the 5 mm and 2° tolerance level. Table 1 displays the population mean, systematic, and random inter-fraction translational and rotational errors. The residual inter-fraction Σ and σ values were smaller than the initial inter-fraction Σ and σ values.
Table 1
Population mean (M), systematic (Σ), and random (σ) translational and rotational errors in 3-dimensions in reverse semi-decubitus positions
  
RL (mm)
AP (mm)
SI (mm)
Roll (°)
Yaw (°)
Pitch (°)
Initial inter-fraction error
M
0.2
−2.5
0.7
1.1
1.0
0.7
Σ
1.8
3.3
2.7
0.5
0.5
0.5
σ
1.8
1.8
2.1
0.5
0.4
0.3
Residual inter-fraction error
M
0.2
−0.9
0.4
1.0
0.9
0.6
Σ
1.2
1.8
1.4
0.4
0.4
0.3
σ
1.4
1.5
1.6
0.4
0.4
0.3
Intra-fraction motion
M
−0.1
−0.4
0.2
0.2
0.2
0.2
Σ
0.9
1.0
1.0
0.1
0.1
0.1
σ
1.0
1.3
1.2
0.2
0.1
0.1
RL right-left direction, AP anterior–posterior direction, SI superior-inferior direction
The magnitude of the inter-fraction set-up errors in the AP direction, roll, and yaw were correlated with high body weight and BMI (Table 2). The inter-fraction translational and rotational set-up errors did not change significantly as a function of time (Fig. 3). When dividing the patients into two groups based on the median BMI, the mean set-up errors were − 1.1 ± 3.0 and − 4.0 ± 2.9 mm in the AP direction (p = 0.01), 0.7° ± 0.3° and 1.5° ± 0.5° in the roll (p < 0.001), and 0.8° ± 0.4° and 1.1° ± 0.6° in the yaw (p = 0.08), for patients with BMI ≤ 25.9 kg/m2 and > 25.9 kg/m2, respectively. The inter-fraction set-up errors were − 1.6 ± 2.6 and − 3.4 ± 3.7 mm in the AP direction (p = 0.14), 0.9° ± 0.5° and 1.4° ± 0.5° in the roll (p = 0.004), and 0.8° ± 0.4° and 1.2° ± 0.6° in the yaw (p = 0.06), for patients with weight ≤ 63 kg and > 63 kg, respectively.
Table 2
Correlation coefficients between patient characteristics and translational and rotational set-up errors
Characteristics
Median (range)
Inter-fractiona
Intra-fraction
RL
AP
SI
Roll
Yaw
Pitch
RL
AP
SI
Roll
Yaw
Pitch
Age
52 (38–75)
0.17
0.05
0.09
0.28
0.02
0.05
0.29
0.07
0.01
0.22
0.07
0.13
Body weight (kg)
63.0 (47.0–88.2)
0.33
0.54†
0.13
0.54†
0.50†
0.30
0.02
0.33
0.30
0.24
0.21
0.22
BMI (kg/m2)
25.9 (18.1–33.6)
0.24
0.55†
0.12
0.68†
0.53†
0.33
0.03
0.23
0.32
0.28
0.24
0.13
Treatment time (min)
4.1 (3.0–5.6)
0.08
0.32
0.10
0.14
0.03
0.14
RL right-left direction, AP anterior–posterior direction, SI superior-inferior direction, BMI body mass index
†Correlation was significant at p < 0.05
aThe pre-correction cone-beam computerized tomography images were used for these analyses

Intra-fraction motion

The mean time interval of intra-fraction motion assessment was 4.1 ± 0.8 min (range, 3.0–5.6 min). Figure 4 shows a histogram distribution of intra-fraction motion. The numbers of fractions exceeding ±5 mm in the RL, AP, and SI directions were 1 (0.6%), 1 (0.6%) and 3 (1.7%), respectively. The intra-fraction rotational errors were all within 1°. Table 1 displays the population mean systematic and random intra-fraction translational and rotational errors. The magnitude of intra-fraction translational and rotational motion was not correlated with age, body weight, BMI, or treatment time (Table 2).

PTV margins for reverse semi-decubitus breast radiotherapy

For all patients, a comparison of the PTV margins obtained from pre-correction, post-correction, and post-treatment CBCT images showed that PTV margins can be reduced with the online correction (Table 3). The total margins accounting for initial inter-fraction set-up error and intra-fraction motion were 6.5, 10.2, and 8.9 mm in the RL, AP, and SI directions, respectively; in contrast, the corresponding required target margins were only 4.7, 6.3, and 5.4 mm in each of the three directions, if the residual and intra-fraction errors were considered after online correction.
Table 3
Calculated CTV–PTV margins in 3-dimensions
Margin
RL (mm)
AP (mm)
SI (mm)
Initial inter-fraction
 Overall
5.8
9.4
8.3
 BMI > 25.9 kg/m2
5.9
9.8
8.3
 BMI ≤25.9 kg/m2
5.4
7.4
8.2
Residual inter-fraction
 Overall
4.0
5.5
4.7
 BMI > 25.9 kg/m2
3.1
4.4
3.7
 BMI ≤25.9 kg/m2
4.8
6.4
5.6
Intra-fraction
 Overall
2.9
3.3
3.4
 BMI > 25.9 kg/m2
2.9
3.3
3.1
 BMI ≤25.9 kg/m2
3.0
3.2
3.7
Total without CBCT correction
 Overall
6.5
10.2
8.9
 BMI > 25.9 kg/m2
6.6
10.3
8.8
 BMI ≤25.9 kg/m2
6.3
8.3
9.1
Total with CBCT correction
 Overall
4.7
6.3
5.4
 BMI > 25.9 kg/m2
4.3
5.4
4.7
 BMI ≤25.9 kg/m2
5.6
7.2
6.8
CTV clinical target volume, PTV planning target volume, RL right-left direction, AP anterior–posterior direction, SI superior-inferior direction, BMI body mass index, CBCT cone-beam computerized tomography
For patients with higher BMI, the calculated PTV margins for initial inter-fraction set-up error and intra-fraction motion were larger than those for patients with lower BMI (Table 3). With online correction, the PTV margins could be reduced both for patients with BMI ≤ 25.9 kg/m2 and > 25.9 kg/m2.

Discussion

This is the first study to analyse inter-fraction set-up error and intra-fraction motion by CBCT scans for RSD breast radiotherapy. The CBCT allows a reduction in inter-fraction set-up errors, particularly for heavier patients. The magnitude of inter-fraction set-up errors in the AP direction, roll, and yaw were significantly correlated with higher body weight and BMI. Intra-fraction motion was small, suggesting that patients were able to perform a stable RSD within the treatment fraction.
In the present study, the RSD position was reproduced by aligning skin tattoos to lasers; the set-up variability was evaluated by analysing CBCT results. Pre-treatment CBCT could effectively detect the translational and rotational set-up errors that required adjustment and improved the accuracy of RSD breast radiotherapy, particularly for patients with higher BMI. By using image-guided set-up, the inter-fraction systemic errors were reduced from 1.8–3.3 mm to 1.2–1.8 mm. Most of the patients had inter-fraction set-up error toward the posterior direction (Fig. 2b); the magnitude of set-up errors in the AP direction was significantly correlated with high weight and BMI. The inter-fraction rotational errors in the roll and yaw were also significantly larger for heavier patients. These findings might be attributed to differences in the magnitude of compression of the subcutaneous back fat in each fraction [28], although set-up included matching to skin tattoos. In addition, the inter-fraction set-up error did not change significantly as a function of time in this study. Image-guided set-up would be helpful to ensure favourable reproducibility in RSD breast radiotherapy.
Intra-fraction motions were evaluated by comparing pre-treatment and post-treatment CBCT without considering respiratory-induced motion in this study. This implied that only intra-fraction baseline drift, which may depend on the deformation or movement of fat or muscle relaxation during treatment, could be evaluated. We found that intra-fraction baseline drift, in terms of translation and rotation, was small during RSD breast radiotherapy (Table 1); this suggested that patients could maintain the RSD position with high stability during radiotherapy. The clinical and dosimetric impact of such minimal intra-fraction motion might be limited. In addition, baseline drift that occurs during prolonged treatment might induce notable uncertainties [29]. The patients in the present study received tangential radiotherapy with a mean treatment time of 4.1 ± 0.8 min; however, intra-fraction motion was not correlated with treatment time. The main intra-fraction baseline drifts were in the posterior direction; this finding was similar to the findings in two recent studies, where intra-fraction motion was evaluated in patients receiving supine breast radiotherapy [28, 29]. It might also be attributed to the compression of back subcutaneous fat or muscle relaxation during treatment [28]. However, the magnitude of intra-fraction translational and rotational motion was not correlated with body weight and BMI in this study.
The isotropic 7-mm CTV-PTV margin was used to treat patients in this study; a total of 19 out of 180 fractions (10.6%) showed an inter-fraction set-up error exceeding 7 mm. The frequencies of inter-fraction set-up errors exceeding 7 mm along the RL, AP, and SI axes were 1.1, 7.8, and 2.8%, respectively. Furthermore, most of the inter-fraction set-up errors exceeding 7 mm occurred in patients with a BMI > 25.9 kg/m2 (73.7%). An increase in PTV margins based on the van Herk formula using these data would ensure greater coverage of the target, but would also increase the radiation to nearby organs at risk. According to this study, the clinical benefit of increased PTV margins might be limited especially when tangential field radiotherapy was used. If the more conformal technique like intensity-modulated radiotherapy or volumetric arc therapy was used, substantial set-up errors would potentially have an impact on the total dose to the target and adjacent organs at risk, and these calculated margins might be used [30, 31]. Compared with the calculated margins for supine or prone position in the previous studies [10, 32], the calculated margins for RSD in this study were smaller. A possible explanation might be the differences in body composition between Western and Asian populations; the BMI of most Asian populations are lower than those of the North American population and possibly some European populations [33]. Online image guidance could help minimize the occurrence of substantial set-up error and potentially permit reduction in PTV margins [21]. The use of CBCT might be of greater benefit to patients with higher body weight or BMI. In the present study, the calculated PTV margins were reduced after image-guided set-up. However, we found that patients with BMI > 25.9 kg/m2 had greater reduction in PTV margins; this may have been related to the greater number of fractions requiring correction for set-up errors greater than 5 mm and small residual errors in these patients. In contrast, we found that most initial set-up errors were within the 5-mm tolerance setting in patients with BMI ≤ 25.9 kg/m2. Hence, the calculated PTV margin with CBCT correction was smaller for heavier patients in this study. In addition to CBCT, three-dimensional surface imaging or an optical tracking system might also be useful to compensate for the inter-fraction set-up errors and intra-fraction motion in breast radiotherapy, without the use of additional radiation [28, 29, 34]. However, the evaluation of set-up variability during RSD breast radiotherapy by three-dimensional surface imaging is beyond the scope of this study and must be addressed in future studies.
There are some limitations in this study. First, this study only evaluated six fractions among all 15 fractions of radiotherapy, rather than in the setting of daily CBCT. For the other nine fractions of radiotherapy, the RSD position was reproduced by aligning skin tattoos to lasers alone. Hence, substantial set-up errors might have occurred during these fractions, particularly for the heavier patients. A study with daily image-guided set-up would provide more comprehensive views of set-up errors and robust calculations of PTV margins in RSD breast radiotherapy. Second, this study only analysed patient set-up error; notably, changes in breast shape and reproducibility of the heart may play roles in RSD breast radiotherapy [31, 35]. Future studies evaluating the volumetric variation of the breast, as well as the reproducibility of the heart and LAD, are needed. Third, the respiratory-induced motion during RSD breast radiotherapy could not be evaluated in this study. Although previous studies have evaluated respiratory-induced motion in the supine or prone positions [10, 28], the optimal margin (including respiratory-induced motion in the RSD position) must be investigated in future studies. Despite these limitations, the present study is the first to evaluate inter-fraction set-up error and intra-fraction motion by using CBCT in RSD breast radiotherapy. RSD is an alternative technique for patients who are not able to tolerate breath-holding [19, 20]. At our institutions, RSD positioning is now one of the standard techniques because patients can receive breast radiotherapy with high comfort and compliance. Generally, the overall set-up and treatment time of RSD breast radiotherapy is similar to that in the supine position in our clinical practice, although the set-up time might be longer for patients with higher BMI. As robust results of daily image-guided set-up are not yet available, this study provides the first evidence of inter-fraction and intra-fraction set-up errors and margins for RSD breast radiotherapy.

Conclusions

This study determined the inter-fraction set-up error and intra-fraction motion in RSD breast radiotherapy with free breathing, supporting the use of CBCT as an effective modality to evaluate and improve the accuracy of RSD breast radiotherapy. Inter-fraction set-up errors did not increase as a function of time during the course of radiotherapy. The magnitude of inter-fraction set-up error was correlated with high body weight and BMI. Our results suggest that CBCT might be beneficial to these patients for the detection of substantial set-up error, which could not have been detected if these patients were aligned by using skin marks alone. In addition, patients could maintain the RSD position with minimal baseline drift during treatment. The anisotropic PTV margins of 6.5, 10.2, and 8.9 mm in the RL, AP, and SI directions might be used for RSD breast radiotherapy without CBCT; margins of 4.7, 6.3, and 5.4 mm in the RL, AP, and SI directions might be used for those treated with rigorous daily CBCT scans.

Acknowledgements

Not applicable.

Funding

No funding provided.

Availability of data and materials

The datasets used and/or analysed during the current study are available from the corresponding author on reasonable request.
This present study was approved by the Institutional Review Board in MacKay Memorial Hospital [17MMHSO51e].
Written informed consent was obtained from the patient for the publication of this report and any accompanying images.

Competing interests

The authors declare that they have no competing interests.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://​creativecommons.​org/​licenses/​by/​4.​0/​), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://​creativecommons.​org/​publicdomain/​zero/​1.​0/​) applies to the data made available in this article, unless otherwise stated.
Literatur
1.
Zurück zum Zitat Nilsson G, Holmberg L, Garmo H, et al. Distribution of coronary artery stenosis after radiation for breast cancer. J Clin Oncol. 2012;30:380–6.CrossRef Nilsson G, Holmberg L, Garmo H, et al. Distribution of coronary artery stenosis after radiation for breast cancer. J Clin Oncol. 2012;30:380–6.CrossRef
2.
Zurück zum Zitat Taylor C, Correa C, Duane FK, et al. Estimating the risks of breast cancer radiotherapy: evidence from modern radiation doses to the lungs and heart and from previous randomized trials. J Clin Oncol. 2017;35:1641–9.CrossRef Taylor C, Correa C, Duane FK, et al. Estimating the risks of breast cancer radiotherapy: evidence from modern radiation doses to the lungs and heart and from previous randomized trials. J Clin Oncol. 2017;35:1641–9.CrossRef
3.
Zurück zum Zitat Darby SC, Ewertz M, McGale P, et al. Risk of ischemic heart disease in women after radiotherapy for breast cancer. N Engl J Med. 2013;368:987–98.CrossRef Darby SC, Ewertz M, McGale P, et al. Risk of ischemic heart disease in women after radiotherapy for breast cancer. N Engl J Med. 2013;368:987–98.CrossRef
4.
Zurück zum Zitat van den Bogaard VA, Ta BD, van der Schaaf A, et al. Validation and modification of a prediction model for acute cardiac events in patients with breast cancer treated with radiotherapy based on three-dimensional dose distributions to cardiac substructures. J Clin Oncol. 2017;35:1171–8.CrossRef van den Bogaard VA, Ta BD, van der Schaaf A, et al. Validation and modification of a prediction model for acute cardiac events in patients with breast cancer treated with radiotherapy based on three-dimensional dose distributions to cardiac substructures. J Clin Oncol. 2017;35:1171–8.CrossRef
5.
Zurück zum Zitat Mendez LC, Louie AV, Moreno C, et al. Evaluation of a new predictor of heart and left anterior descending artery dose in patients treated with adjuvant radiotherapy to the left breast. Radiat Oncol. 2018;13:124.CrossRef Mendez LC, Louie AV, Moreno C, et al. Evaluation of a new predictor of heart and left anterior descending artery dose in patients treated with adjuvant radiotherapy to the left breast. Radiat Oncol. 2018;13:124.CrossRef
7.
Zurück zum Zitat Nissen HD, Appelt AL. Improved heart, lung and target dose with deep inspiration breath hold in a large clinical series of breast cancer patients. Radiother Oncol. 2013;106:28–32.CrossRef Nissen HD, Appelt AL. Improved heart, lung and target dose with deep inspiration breath hold in a large clinical series of breast cancer patients. Radiother Oncol. 2013;106:28–32.CrossRef
8.
Zurück zum Zitat Zagar TM, Kaidar-Person O, Tang X, et al. Utility of deep inspiration breath hold for left-sided breast radiation therapy in preventing early cardiac perfusion defects: a prospective study. Int J Radiat Oncol Biol Phys. 2017;97:903–9.CrossRef Zagar TM, Kaidar-Person O, Tang X, et al. Utility of deep inspiration breath hold for left-sided breast radiation therapy in preventing early cardiac perfusion defects: a prospective study. Int J Radiat Oncol Biol Phys. 2017;97:903–9.CrossRef
9.
Zurück zum Zitat Bartlett FR, Colgan RM, Donovan EM, et al. The UK HeartSpare Study (Stage IB): randomised comparison of a voluntary breath-hold technique and prone radiotherapy after breast conserving surgery. Radiother Oncol. 2015;114:66–72.CrossRef Bartlett FR, Colgan RM, Donovan EM, et al. The UK HeartSpare Study (Stage IB): randomised comparison of a voluntary breath-hold technique and prone radiotherapy after breast conserving surgery. Radiother Oncol. 2015;114:66–72.CrossRef
10.
Zurück zum Zitat Kirby AM, Evans PM, Helyer SJ, et al. A randomised trial of supine versus prone breast radiotherapy (SuPr study): comparing set-up errors and respiratory motion. Radiother Oncol. 2011;100:221–6.CrossRef Kirby AM, Evans PM, Helyer SJ, et al. A randomised trial of supine versus prone breast radiotherapy (SuPr study): comparing set-up errors and respiratory motion. Radiother Oncol. 2011;100:221–6.CrossRef
11.
Zurück zum Zitat Dunkerley N, Bartlett FR, Kirby AM, Evans PM, Donovan EM. Mean heart dose variation over a course of breath-holding breast cancer radiotherapy. Br J Radiol. 2016;89:20160536.CrossRef Dunkerley N, Bartlett FR, Kirby AM, Evans PM, Donovan EM. Mean heart dose variation over a course of breath-holding breast cancer radiotherapy. Br J Radiol. 2016;89:20160536.CrossRef
12.
Zurück zum Zitat Schonecker S, Walter F, Freislederer P, et al. Treatment planning and evaluation of gated radiotherapy in left-sided breast cancer patients using the catalyst(TM)/sentinel(TM) system for deep inspiration breath-hold (DIBH). Radiat Oncol. 2016;11:143.CrossRef Schonecker S, Walter F, Freislederer P, et al. Treatment planning and evaluation of gated radiotherapy in left-sided breast cancer patients using the catalyst(TM)/sentinel(TM) system for deep inspiration breath-hold (DIBH). Radiat Oncol. 2016;11:143.CrossRef
13.
Zurück zum Zitat Zhao F, Shen J, Lu Z, et al. Abdominal DIBH reduces the cardiac dose even further: a prospective analysis. Radiat Oncol. 2018;13:116.CrossRef Zhao F, Shen J, Lu Z, et al. Abdominal DIBH reduces the cardiac dose even further: a prospective analysis. Radiat Oncol. 2018;13:116.CrossRef
14.
Zurück zum Zitat Deseyne P, Speleers B, De Neve W, et al. Whole breast and regional nodal irradiation in prone versus supine position in left sided breast cancer. Radiat Oncol. 2017;12:89.CrossRef Deseyne P, Speleers B, De Neve W, et al. Whole breast and regional nodal irradiation in prone versus supine position in left sided breast cancer. Radiat Oncol. 2017;12:89.CrossRef
15.
Zurück zum Zitat Kahan Z, Rarosi F, Gaal S, et al. A simple clinical method for predicting the benefit of prone vs. supine positioning in reducing heart exposure during left breast radiotherapy. Radiother Oncol. 2018;126:487–92.CrossRef Kahan Z, Rarosi F, Gaal S, et al. A simple clinical method for predicting the benefit of prone vs. supine positioning in reducing heart exposure during left breast radiotherapy. Radiother Oncol. 2018;126:487–92.CrossRef
16.
Zurück zum Zitat Kirby AM, Evans PM, Donovan EM, et al. Prone versus supine positioning for whole and partial-breast radiotherapy: a comparison of non-target tissue dosimetry. Radiother Oncol. 2010;96:178–84.CrossRef Kirby AM, Evans PM, Donovan EM, et al. Prone versus supine positioning for whole and partial-breast radiotherapy: a comparison of non-target tissue dosimetry. Radiother Oncol. 2010;96:178–84.CrossRef
17.
Zurück zum Zitat Formenti SC, DeWyngaert JK, Jozsef G, Goldberg JD. Prone vs supine positioning for breast cancer radiotherapy. JAMA. 2012;308:861–3.CrossRef Formenti SC, DeWyngaert JK, Jozsef G, Goldberg JD. Prone vs supine positioning for breast cancer radiotherapy. JAMA. 2012;308:861–3.CrossRef
18.
Zurück zum Zitat Lymberis SC, deWyngaert JK, Parhar P, et al. Prospective assessment of optimal individual position (prone versus supine) for breast radiotherapy: volumetric and dosimetric correlations in 100 patients. Int J Radiat Oncol Biol Phys. 2012;84:902–9.CrossRef Lymberis SC, deWyngaert JK, Parhar P, et al. Prospective assessment of optimal individual position (prone versus supine) for breast radiotherapy: volumetric and dosimetric correlations in 100 patients. Int J Radiat Oncol Biol Phys. 2012;84:902–9.CrossRef
19.
Zurück zum Zitat Niglas M, McCann C, Keller BM, et al. A dosimetric study of cardiac dose sparing using the reverse semi-decubitus technique for left breast and internal mammary chain irradiation. Radiother Oncol. 2016;118:187–93.CrossRef Niglas M, McCann C, Keller BM, et al. A dosimetric study of cardiac dose sparing using the reverse semi-decubitus technique for left breast and internal mammary chain irradiation. Radiother Oncol. 2016;118:187–93.CrossRef
20.
Zurück zum Zitat Helal A, Farouk M, Gamie SH, et al. The impact of semilateral decubitus position on the dose-volume parameters of the heart and lung for left sided breast cancer patients: a comparative dosimetric study. Pract Radiat Oncol. 2018;8:71–80.CrossRef Helal A, Farouk M, Gamie SH, et al. The impact of semilateral decubitus position on the dose-volume parameters of the heart and lung for left sided breast cancer patients: a comparative dosimetric study. Pract Radiat Oncol. 2018;8:71–80.CrossRef
21.
Zurück zum Zitat White EA, Cho J, Vallis KA, et al. Cone beam computed tomography guidance for setup of patients receiving accelerated partial breast irradiation. Int J Radiat Oncol Biol Phys. 2007;68:547–54.CrossRef White EA, Cho J, Vallis KA, et al. Cone beam computed tomography guidance for setup of patients receiving accelerated partial breast irradiation. Int J Radiat Oncol Biol Phys. 2007;68:547–54.CrossRef
22.
Zurück zum Zitat Topolnjak R, Sonke JJ, Nijkamp J, et al. Breast patient setup error assessment: comparison of electronic portal image devices and cone-beam computed tomography matching results. Int J Radiat Oncol Biol Phys. 2010;78:1235–43.CrossRef Topolnjak R, Sonke JJ, Nijkamp J, et al. Breast patient setup error assessment: comparison of electronic portal image devices and cone-beam computed tomography matching results. Int J Radiat Oncol Biol Phys. 2010;78:1235–43.CrossRef
23.
Zurück zum Zitat Feng M, Moran JM, Koelling T, et al. Development and validation of a heart atlas to study cardiac exposure to radiation following treatment for breast cancer. Int J Radiat Oncol Biol Phys. 2011;79:10–8.CrossRef Feng M, Moran JM, Koelling T, et al. Development and validation of a heart atlas to study cardiac exposure to radiation following treatment for breast cancer. Int J Radiat Oncol Biol Phys. 2011;79:10–8.CrossRef
24.
Zurück zum Zitat Lee J, Hua KL, Hsu SM, et al. Development of delineation for the left anterior descending coronary artery region in left breast cancer radiotherapy: an optimized organ at risk. Radiother Oncol. 2017;122:423–30.CrossRef Lee J, Hua KL, Hsu SM, et al. Development of delineation for the left anterior descending coronary artery region in left breast cancer radiotherapy: an optimized organ at risk. Radiother Oncol. 2017;122:423–30.CrossRef
25.
Zurück zum Zitat Duane F, Aznar MC, Bartlett F, et al. A cardiac contouring atlas for radiotherapy. Radiother Oncol. 2017;122:416–22.CrossRef Duane F, Aznar MC, Bartlett F, et al. A cardiac contouring atlas for radiotherapy. Radiother Oncol. 2017;122:416–22.CrossRef
26.
Zurück zum Zitat van Herk M. Errors and margins in radiotherapy. Semin Radiat Oncol. 2004;14:52–64.CrossRef van Herk M. Errors and margins in radiotherapy. Semin Radiat Oncol. 2004;14:52–64.CrossRef
27.
Zurück zum Zitat van Herk M, Remeijer P, Rasch C, Lebesque JV. The probability of correct target dosage: dose-population histograms for deriving treatment margins in radiotherapy. Int J Radiat Oncol Biol Phys. 2000;47:1121–35.CrossRef van Herk M, Remeijer P, Rasch C, Lebesque JV. The probability of correct target dosage: dose-population histograms for deriving treatment margins in radiotherapy. Int J Radiat Oncol Biol Phys. 2000;47:1121–35.CrossRef
28.
Zurück zum Zitat Hirata K, Yoshimura M, Mukumoto N, et al. Three-dimensional intrafractional internal target motions in accelerated partial breast irradiation using three-dimensional conformal external beam radiotherapy. Radiother Oncol. 2017;124:118–23.CrossRef Hirata K, Yoshimura M, Mukumoto N, et al. Three-dimensional intrafractional internal target motions in accelerated partial breast irradiation using three-dimensional conformal external beam radiotherapy. Radiother Oncol. 2017;124:118–23.CrossRef
29.
Zurück zum Zitat Ricotti R, Ciardo D, Fattori G, et al. Intra-fraction respiratory motion and baseline drift during breast Helical Tomotherapy. Radiother Oncol. 2017;122:79–86.CrossRef Ricotti R, Ciardo D, Fattori G, et al. Intra-fraction respiratory motion and baseline drift during breast Helical Tomotherapy. Radiother Oncol. 2017;122:79–86.CrossRef
30.
Zurück zum Zitat Feng CH, Gerry E, Chmura SJ, Hasan Y, Al-Hallaq HA. An image-guided study of setup reproducibility of postmastectomy breast cancer patients treated with inverse-planned intensity modulated radiation therapy. Int J Radiat Oncol Biol Phys. 2015;91:58–64.CrossRef Feng CH, Gerry E, Chmura SJ, Hasan Y, Al-Hallaq HA. An image-guided study of setup reproducibility of postmastectomy breast cancer patients treated with inverse-planned intensity modulated radiation therapy. Int J Radiat Oncol Biol Phys. 2015;91:58–64.CrossRef
31.
Zurück zum Zitat van Mourik A, van Kranen S, den Hollander S, et al. Effects of setup errors and shape changes on breast radiotherapy. Int J Radiat Oncol Biol Phys. 2011;79:1557–64.CrossRef van Mourik A, van Kranen S, den Hollander S, et al. Effects of setup errors and shape changes on breast radiotherapy. Int J Radiat Oncol Biol Phys. 2011;79:1557–64.CrossRef
32.
Zurück zum Zitat Jozsef G, DeWyngaert JK, Becker SJ, Lymberis S, Formenti SC. Prospective study of cone-beam computed tomography image-guided radiotherapy for prone accelerated partial breast irradiation. Int J Radiat Oncol Biol Phys. 2011;81:568–74.CrossRef Jozsef G, DeWyngaert JK, Becker SJ, Lymberis S, Formenti SC. Prospective study of cone-beam computed tomography image-guided radiotherapy for prone accelerated partial breast irradiation. Int J Radiat Oncol Biol Phys. 2011;81:568–74.CrossRef
33.
Zurück zum Zitat NCD Risk Factor Collaboration (NCD-RisC). Trends in adult body-mass index in 200 countries from 1975 to 2014: a pooled analysis of 1698 population-based measurement studies with 19.2 million participants. Lancet. 2016;387:1377–96. NCD Risk Factor Collaboration (NCD-RisC). Trends in adult body-mass index in 200 countries from 1975 to 2014: a pooled analysis of 1698 population-based measurement studies with 19.2 million participants. Lancet. 2016;387:1377–96.
34.
Zurück zum Zitat Betgen A, Alderliesten T, Sonke JJ, et al. Assessment of set-up variability during deep inspiration breath hold radiotherapy for breast cancer patients by 3D-surface imaging. Radiother Oncol. 2013;106:225–30.CrossRef Betgen A, Alderliesten T, Sonke JJ, et al. Assessment of set-up variability during deep inspiration breath hold radiotherapy for breast cancer patients by 3D-surface imaging. Radiother Oncol. 2013;106:225–30.CrossRef
35.
Zurück zum Zitat Koivumaki T, Tujunen J, Viren T, Heikkila J, Seppala J. Geometrical uncertainty of heart position in deep-inspiration breath-hold radiotherapy of left-sided breast cancer patients. Acta Oncol. 2017;56:879–83.CrossRef Koivumaki T, Tujunen J, Viren T, Heikkila J, Seppala J. Geometrical uncertainty of heart position in deep-inspiration breath-hold radiotherapy of left-sided breast cancer patients. Acta Oncol. 2017;56:879–83.CrossRef
Metadaten
Titel
Image-guided study of inter-fraction and intra-fraction set-up variability and margins in reverse semi-decubitus breast radiotherapy
verfasst von
Jie Lee
Shih-Hua Liu
Jhen-Bin Lin
Meng-Hao Wu
Chieh-Ju Wu
Hung-Chi Tai
Shih-Ming Hsu
Yin-Ju Chen
Jo-Chiao Tai
Yu-Jen Chen
Publikationsdatum
01.12.2018
Verlag
BioMed Central
Erschienen in
Radiation Oncology / Ausgabe 1/2018
Elektronische ISSN: 1748-717X
DOI
https://doi.org/10.1186/s13014-018-1200-1

Weitere Artikel der Ausgabe 1/2018

Radiation Oncology 1/2018 Zur Ausgabe

Update Onkologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.