25.01.2019 | Original Paper

Image quality and radiation dose of ECG-triggered High-Pitch Dual-Source cardiac computed tomography angiography in children for the evaluation of central vascular stents

verfasst von: Christian A. Barrera, Hansel J. Otero, Ammie M. White, David Saul, David M. Biko

Erschienen in: The International Journal of Cardiovascular Imaging | Ausgabe 2/2019

Einloggen, um Zugang zu erhalten

Abstract

Assess image quality and radiation dose of ECG-triggered High-Pitch Dual-Source CTA for the evaluation central vascular stents in children. We included all children ≤ 21 years old with one or more central vascular stents and available prospective ECG-triggered High-Pitch Dual-Source CTA performed at our institution between January 2015 and August 2017. Demographic and scanner information was retrieved. Two board-certified pediatric radiologists blinded to the clinical data, independently reviewed and scored each case using a four-point quality score. Scores 1, 2 and 3 were considered of diagnostic image quality. Inter-observer agreement and non-parametric test were used. 18 patients (10 girls, 8 boys) with a mean age of 9.47 ± 7.38 years (mean ± SD) met inclusion criteria. Thirty-two central vascular stents were evaluated. Mean quality score was 2.07 ± 0.94 with 12.5% (4/32) of the cases classified as unevaluable. Interobserver agreement was excellent (k = 0.86). There is no significant difference between quality score and stent location (p = 0.07). There is a significant difference with stent material as all non-diagnostic scores were only seen in covered stents made of platinum-iridium (p < 0.001). There was no association between image quality and age, height, weight, BSA, heart rate, radiation dose or stent lumen size (p > 0.05). ECG-triggered high-pitch spiral DS-CTA offers appropriate image quality for assessment of central vascular stents in children.
Literatur
2.
Zurück zum Zitat Temel MT, Coskun ME, Baspinar O, Demiryurek AT (2017) Prevalence and characteristics of coronary artery anomalies in children with congenital heart disease diagnosed with coronary angiography. Turk Kardiyoloji Dernegi arsivi: Turk Kardiyoloji Derneginin yayin organidir 45(6):527–532. https://doi.org/10.5543/tkda.2017.24162 CrossRef Temel MT, Coskun ME, Baspinar O, Demiryurek AT (2017) Prevalence and characteristics of coronary artery anomalies in children with congenital heart disease diagnosed with coronary angiography. Turk Kardiyoloji Dernegi arsivi: Turk Kardiyoloji Derneginin yayin organidir 45(6):527–532. https://​doi.​org/​10.​5543/​tkda.​2017.​24162 CrossRef
5.
Zurück zum Zitat O’Laughlin MP, Perry SB, Lock JE, Mullins CE (1991) Use of endovascular stents in congenital heart disease. Circulation 83(6):1923–1939CrossRefPubMed O’Laughlin MP, Perry SB, Lock JE, Mullins CE (1991) Use of endovascular stents in congenital heart disease. Circulation 83(6):1923–1939CrossRefPubMed
7.
Zurück zum Zitat Eichhorn JG, Long FR, Hill SL, O’Donovan J, Chisolm JL, Fernandez SA, Cheatham JP (2006) Assessment of in-stent stenosis in small children with congenital heart disease using multi-detector computed tomography: a validation study. Catheter Cardiovasc Interv 68(1):11–20. https://doi.org/10.1002/ccd.20760 CrossRefPubMed Eichhorn JG, Long FR, Hill SL, O’Donovan J, Chisolm JL, Fernandez SA, Cheatham JP (2006) Assessment of in-stent stenosis in small children with congenital heart disease using multi-detector computed tomography: a validation study. Catheter Cardiovasc Interv 68(1):11–20. https://​doi.​org/​10.​1002/​ccd.​20760 CrossRefPubMed
8.
Zurück zum Zitat Ehara M, Kawai M, Surmely JF, Matsubara T, Terashima M, Tsuchikane E, Kinoshita Y, Ito T, Takeda Y, Nasu K, Tanaka N, Murata A, Fujita H, Sato K, Kodama A, Katoh O, Suzuki T (2007) Diagnostic accuracy of coronary in-stent restenosis using 64-slice computed tomography: comparison with invasive coronary angiography. J Am Coll Cardiol 49(9):951–959. https://doi.org/10.1016/j.jacc.2006.10.065 CrossRefPubMed Ehara M, Kawai M, Surmely JF, Matsubara T, Terashima M, Tsuchikane E, Kinoshita Y, Ito T, Takeda Y, Nasu K, Tanaka N, Murata A, Fujita H, Sato K, Kodama A, Katoh O, Suzuki T (2007) Diagnostic accuracy of coronary in-stent restenosis using 64-slice computed tomography: comparison with invasive coronary angiography. J Am Coll Cardiol 49(9):951–959. https://​doi.​org/​10.​1016/​j.​jacc.​2006.​10.​065 CrossRefPubMed
10.
Zurück zum Zitat Eichhorn JG, Long FR, Jourdan C, Heverhagen JT, Hill SL, Raman SV, Cheatham JP (2008) Usefulness of multidetector CT imaging to assess vascular stents in children with congenital heart disease: an in vivo and in vitro study. Catheter Cardiovasc Interv 72(4):544–551. https://doi.org/10.1002/ccd.21680 CrossRefPubMed Eichhorn JG, Long FR, Jourdan C, Heverhagen JT, Hill SL, Raman SV, Cheatham JP (2008) Usefulness of multidetector CT imaging to assess vascular stents in children with congenital heart disease: an in vivo and in vitro study. Catheter Cardiovasc Interv 72(4):544–551. https://​doi.​org/​10.​1002/​ccd.​21680 CrossRefPubMed
11.
Zurück zum Zitat Sun Z, Ng KH, Sarji SA (2010) Is utilisation of computed tomography justified in clinical practice? Part IV: applications of paediatric computed tomography. Singapore Med J 51(6):457–463PubMed Sun Z, Ng KH, Sarji SA (2010) Is utilisation of computed tomography justified in clinical practice? Part IV: applications of paediatric computed tomography. Singapore Med J 51(6):457–463PubMed
19.
Zurück zum Zitat American Association of Physicists in Medicine (2008) The measure, reporting and management of radiation dose in CT. Report # 96 of AAPM task Group 23 of the Diagnostic Imaging Council CT Committee American Association of Physicists in Medicine (2008) The measure, reporting and management of radiation dose in CT. Report # 96 of AAPM task Group 23 of the Diagnostic Imaging Council CT Committee
20.
Zurück zum Zitat Layritz C, Schmid J, Achenbach S, Ulzheimer S, Wuest W, May M, Ropers D, Klinghammer L, Daniel WG, Pflederer T, Lell M (2014) Accuracy of prospectively ECG-triggered very low-dose coronary dual-source CT angiography using iterative reconstruction for the detection of coronary artery stenosis: comparison with invasive catheterization. Eur Heart J Cardiovasc Imag 15(11):1238–1245. https://doi.org/10.1093/ehjci/jeu113 CrossRef Layritz C, Schmid J, Achenbach S, Ulzheimer S, Wuest W, May M, Ropers D, Klinghammer L, Daniel WG, Pflederer T, Lell M (2014) Accuracy of prospectively ECG-triggered very low-dose coronary dual-source CT angiography using iterative reconstruction for the detection of coronary artery stenosis: comparison with invasive catheterization. Eur Heart J Cardiovasc Imag 15(11):1238–1245. https://​doi.​org/​10.​1093/​ehjci/​jeu113 CrossRef
21.
Zurück zum Zitat Rybicki FJ, Otero HJ, Steigner ML, Vorobiof G, Nallamshetty L, Mitsouras D, Ersoy H, Mather RT, Judy PF, Cai T, Coyner K, Schultz K, Whitmore AG, Di Carli MF (2008) Initial evaluation of coronary images from 320-detector row computed tomography. Int J Cardiovasc Imag 24(5):535–546. https://doi.org/10.1007/s10554-008-9308-2 CrossRef Rybicki FJ, Otero HJ, Steigner ML, Vorobiof G, Nallamshetty L, Mitsouras D, Ersoy H, Mather RT, Judy PF, Cai T, Coyner K, Schultz K, Whitmore AG, Di Carli MF (2008) Initial evaluation of coronary images from 320-detector row computed tomography. Int J Cardiovasc Imag 24(5):535–546. https://​doi.​org/​10.​1007/​s10554-008-9308-2 CrossRef
22.
Zurück zum Zitat Viera AJ, Garrett JM (2005) Understanding interobserver agreement: the kappa statistic. Fam Med 37(5):360–363PubMed Viera AJ, Garrett JM (2005) Understanding interobserver agreement: the kappa statistic. Fam Med 37(5):360–363PubMed
23.
Zurück zum Zitat Nordmeyer J, Gaudin R, Tann OR, Lurz PC, Bonhoeffer P, Taylor AM, Muthurangu V (2010) MRI may be sufficient for noninvasive assessment of great vessel stents: an in vitro comparison of MRI, CT, and conventional angiography. Am J Roentgenol 195(4):865–871. https://doi.org/10.2214/ajr.09.4166 CrossRef Nordmeyer J, Gaudin R, Tann OR, Lurz PC, Bonhoeffer P, Taylor AM, Muthurangu V (2010) MRI may be sufficient for noninvasive assessment of great vessel stents: an in vitro comparison of MRI, CT, and conventional angiography. Am J Roentgenol 195(4):865–871. https://​doi.​org/​10.​2214/​ajr.​09.​4166 CrossRef
25.
Zurück zum Zitat Achenbach S, Marwan M, Ropers D, Schepis T, Pflederer T, Anders K, Kuettner A, Daniel WG, Uder M, Lell MM (2010) Coronary computed tomography angiography with a consistent dose below 1 mSv using prospectively electrocardiogram-triggered high-pitch spiral acquisition. Eur Heart J 31(3):340–346. https://doi.org/10.1093/eurheartj/ehp470 CrossRefPubMed Achenbach S, Marwan M, Ropers D, Schepis T, Pflederer T, Anders K, Kuettner A, Daniel WG, Uder M, Lell MM (2010) Coronary computed tomography angiography with a consistent dose below 1 mSv using prospectively electrocardiogram-triggered high-pitch spiral acquisition. Eur Heart J 31(3):340–346. https://​doi.​org/​10.​1093/​eurheartj/​ehp470 CrossRefPubMed
31.
Zurück zum Zitat Marrocco-Trischitta MM, van Bakel TM, Romarowski RM, de Beaufort HW, Conti M, van Herwaarden JA, Moll FL, Auricchio F, Trimarchi S (2018) The modified arch landing areas nomenclature (MALAN) improves prediction of stent graft displacement forces: proof of concept by computational fluid dynamics modelling. Eur J Vasc Endovasc Surg 55(4):584–592. https://doi.org/10.1016/j.ejvs.2017.12.019 CrossRefPubMed Marrocco-Trischitta MM, van Bakel TM, Romarowski RM, de Beaufort HW, Conti M, van Herwaarden JA, Moll FL, Auricchio F, Trimarchi S (2018) The modified arch landing areas nomenclature (MALAN) improves prediction of stent graft displacement forces: proof of concept by computational fluid dynamics modelling. Eur J Vasc Endovasc Surg 55(4):584–592. https://​doi.​org/​10.​1016/​j.​ejvs.​2017.​12.​019 CrossRefPubMed
32.
Zurück zum Zitat Karmonik C, Muller-Eschner M, Partovi S, Geisbusch P, Ganten MK, Bismuth J, Davies MG, Bockler D, Loebe M, Lumsden AB, von Tengg-Kobligk H (2013) Computational fluid dynamics investigation of chronic aortic dissection hemodynamics versus normal aorta. Vasc Endovasc Surg 47(8):625–631. https://doi.org/10.1177/1538574413503561 CrossRef Karmonik C, Muller-Eschner M, Partovi S, Geisbusch P, Ganten MK, Bismuth J, Davies MG, Bockler D, Loebe M, Lumsden AB, von Tengg-Kobligk H (2013) Computational fluid dynamics investigation of chronic aortic dissection hemodynamics versus normal aorta. Vasc Endovasc Surg 47(8):625–631. https://​doi.​org/​10.​1177/​1538574413503561​ CrossRef
35.
Zurück zum Zitat Oberhoffer R, Lang D, Feilen K (1989) The diameter of coronary arteries in infants and children without heart disease. Eur J Pediatrics 148(5):389–392CrossRef Oberhoffer R, Lang D, Feilen K (1989) The diameter of coronary arteries in infants and children without heart disease. Eur J Pediatrics 148(5):389–392CrossRef
37.
Zurück zum Zitat Eichhorn JG, Long FR, Hill SL, Cheatham JP (2006) Multislice computed tomography as an adjunct to the management of an in-stent stenosis in an infant with congenital heart disease: imaging for the future. Catheter Cardiovasc Intervent 67(3):477–481. https://doi.org/10.1002/ccd.20644 CrossRef Eichhorn JG, Long FR, Hill SL, Cheatham JP (2006) Multislice computed tomography as an adjunct to the management of an in-stent stenosis in an infant with congenital heart disease: imaging for the future. Catheter Cardiovasc Intervent 67(3):477–481. https://​doi.​org/​10.​1002/​ccd.​20644 CrossRef
39.
44.
Metadaten
Titel
Image quality and radiation dose of ECG-triggered High-Pitch Dual-Source cardiac computed tomography angiography in children for the evaluation of central vascular stents
verfasst von
Christian A. Barrera
Hansel J. Otero
Ammie M. White
David Saul
David M. Biko
Publikationsdatum
25.01.2019
Verlag
Springer Netherlands
Erschienen in
The International Journal of Cardiovascular Imaging / Ausgabe 2/2019
Print ISSN: 1569-5794
Elektronische ISSN: 1875-8312
DOI
https://doi.org/10.1007/s10554-019-01539-3

Weitere Artikel der Ausgabe 2/2019

The International Journal of Cardiovascular Imaging 2/2019 Zur Ausgabe

Update Kardiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.