Image quality in dual-source multiphasic dynamic computed tomography of the abdomen: evaluating the effects of a low tube voltage (70 kVp) in combination with contrast dose reduction

verfasst von: Keisuke Miyoshi, Hideko Onoda, Mashiro Tanabe, Sei Nakao, Mayumi Higashi, Etsushi Iida, Munemasa Okada, Matakazu Furukawa, Katsuyoshi Ito

Erschienen in: Abdominal Radiology | Ausgabe 11/2020

Einloggen, um Zugang zu erhalten

Abstract

Purpose

To compare the image quality of multiphasic (arterial, portal, and equilibrium phases) dynamic computed tomography (CT) of the abdomen obtained by a low tube voltage (70kVp) in combination with a half-dose iodine load using low-concentration contrast agent in high tube output dual-source CT with a standard tube voltage (120kVp) and full-dose iodine load using the same group of adult patients.

Methods

Fifty-five patients who underwent both low-tube-voltage (70kVp) abdominal CT with a half-dose iodine load and standard-tube-voltage (120kVp) CT with a full-dose iodine load were analyzed. The mean CT values and signal-to-noise ratio (SNR) of the liver, aorta and portal veins were quantitatively assessed. In addition, the contrast enhancement of the abdominal organs and overall image quality were qualitatively evaluated.

Results

The mean CT values and SNR of the liver parenchyma were significantly higher in 70-kVp protocol than in 120-kVp protocol in all 3 phases (p = 0.018 ~  < 0.001). Regarding the qualitative analysis, the overall image quality in the 70-kVp protocol was significantly better than in the 120-kVp protocol in all 3 phases (p < 0.001). In addition, the contrast enhancement scores of the liver parenchyma and hepatic vein in the equilibrium phase were also significantly higher in the 70-kVp protocol than in the 120-kVp protocol (p < 0.001).

Conclusion

A low tube voltage (70kVp) in combination with a half-dose iodine load using a low-concentration contrast agent and an iterative reconstruction algorithm in high tube output dual-source CT may improve the contrast enhancement and image quality in multiphasic dynamic CT of the abdomen in patients under 71 kg of body weight.
Literatur
1.
Zurück zum Zitat McDonald JS, McDonald RJ, Carter RE, Katzberg RW, Kallmes DF, Williamson EE. Risk of intravenous contrast material-mediated acute kidney injury: a propensity score-matched study stratified by baseline-estimated glomerular filtration rate. Radiology. 2014;271(1):65-73.CrossRef McDonald JS, McDonald RJ, Carter RE, Katzberg RW, Kallmes DF, Williamson EE. Risk of intravenous contrast material-mediated acute kidney injury: a propensity score-matched study stratified by baseline-estimated glomerular filtration rate. Radiology. 2014;271(1):65-73.CrossRef
2.
Zurück zum Zitat Launay-Vacher V, Oudard S, Janus N, et al. Prevalence of Renal Insufficiency in cancer patients and implications for anticancer drug management: the renal insufficiency and anticancer medications (IRMA) study. Cancer. 2007;110(6):1376-84.CrossRef Launay-Vacher V, Oudard S, Janus N, et al. Prevalence of Renal Insufficiency in cancer patients and implications for anticancer drug management: the renal insufficiency and anticancer medications (IRMA) study. Cancer. 2007;110(6):1376-84.CrossRef
3.
Zurück zum Zitat Garcia-Compean D, Jaquez-Quintana JO, Maldonado-Garza H. Hepatogenous diabetes. Current views of an ancient problem. Annals of Hepatology. 2009;8(1):13–20. Garcia-Compean D, Jaquez-Quintana JO, Maldonado-Garza H. Hepatogenous diabetes. Current views of an ancient problem. Annals of Hepatology. 2009;8(1):13–20.
4.
Zurück zum Zitat Stacul F, van der Molen AJ, Reimer P, et al. Contrast induced nephropathy: updated ESUR Contrast Media Safety Committee guidelines. European radiology. 2011;21(12):2527-41.CrossRef Stacul F, van der Molen AJ, Reimer P, et al. Contrast induced nephropathy: updated ESUR Contrast Media Safety Committee guidelines. European radiology. 2011;21(12):2527-41.CrossRef
5.
Zurück zum Zitat Cicin I, Erdogan B, Gulsen E, et al. Incidence of contrast-induced nephropathy in hospitalised patients with cancer. European radiology. 2014;24(1):184-90.CrossRef Cicin I, Erdogan B, Gulsen E, et al. Incidence of contrast-induced nephropathy in hospitalised patients with cancer. European radiology. 2014;24(1):184-90.CrossRef
6.
Zurück zum Zitat Hunsaker AR, Oliva IB, Cai T, et al. Contrast opacification using a reduced volume of iodinated contrast material and low peak kilovoltage in pulmonary CT angiography: Objective and subjective evaluation. AJR American journal of roentgenology. 2010;195(2):W118-24.CrossRef Hunsaker AR, Oliva IB, Cai T, et al. Contrast opacification using a reduced volume of iodinated contrast material and low peak kilovoltage in pulmonary CT angiography: Objective and subjective evaluation. AJR American journal of roentgenology. 2010;195(2):W118-24.CrossRef
7.
Zurück zum Zitat Huda W. Dose and image quality in CT. Pediatric radiology. 2002;32(10):709–13; discussion 51–4. Huda W. Dose and image quality in CT. Pediatric radiology. 2002;32(10):709–13; discussion 51–4.
8.
Zurück zum Zitat Thor D, Brismar TB, Fischer MA. Low tube voltage dual source computed tomography to reduce contrast media doses in adult abdomen examinations: A phantom study. Medical physics. 2015;42(9):5100-9.CrossRef Thor D, Brismar TB, Fischer MA. Low tube voltage dual source computed tomography to reduce contrast media doses in adult abdomen examinations: A phantom study. Medical physics. 2015;42(9):5100-9.CrossRef
9.
Zurück zum Zitat Schindera ST, Nelson RC, Mukundan S, Jr., et al. Hypervascular liver tumors: low tube voltage, high tube current multi-detector row CT for enhanced detection--phantom study. Radiology. 2008;246(1):125-32.CrossRef Schindera ST, Nelson RC, Mukundan S, Jr., et al. Hypervascular liver tumors: low tube voltage, high tube current multi-detector row CT for enhanced detection--phantom study. Radiology. 2008;246(1):125-32.CrossRef
10.
Zurück zum Zitat Noda Y, Kanematsu M, Goshima S, et al. Reducing iodine load in hepatic CT for patients with chronic liver disease with a combination of low-tube-voltage and adaptive statistical iterative reconstruction. European journal of radiology. 2015;84(1):11-8.CrossRef Noda Y, Kanematsu M, Goshima S, et al. Reducing iodine load in hepatic CT for patients with chronic liver disease with a combination of low-tube-voltage and adaptive statistical iterative reconstruction. European journal of radiology. 2015;84(1):11-8.CrossRef
11.
Zurück zum Zitat Botsikas D, Barnaure I, Terraz S, Becker CD, Kalovidouri A, Montet X. Value of liver computed tomography with iodixanol 270, 80 kVp and iterative reconstruction. World journal of radiology. 2016;8(7):693-9.CrossRef Botsikas D, Barnaure I, Terraz S, Becker CD, Kalovidouri A, Montet X. Value of liver computed tomography with iodixanol 270, 80 kVp and iterative reconstruction. World journal of radiology. 2016;8(7):693-9.CrossRef
12.
Zurück zum Zitat Araki K, Yoshizako T, Yoshida R, Tada K, Kitagaki H. Low-voltage (80-kVp) abdominopelvic computed tomography allows 60% contrast dose reduction in patients at risk of contrast-induced nephropathy. Clinical imaging. 2018;51:352-5.CrossRef Araki K, Yoshizako T, Yoshida R, Tada K, Kitagaki H. Low-voltage (80-kVp) abdominopelvic computed tomography allows 60% contrast dose reduction in patients at risk of contrast-induced nephropathy. Clinical imaging. 2018;51:352-5.CrossRef
13.
Zurück zum Zitat Onoda H, Ueno H, Hashimoto M, Kuwahara H, Sobajima M, Kinugawa K. Clinical Advantages of Using Low Tube Voltage in Third-Generation 192-Slice Dual-Source Computed Tomographic Angiography Before Transcatheter Aortic Valve Implantation. International heart journal. 2019;60(5):1091-7.CrossRef Onoda H, Ueno H, Hashimoto M, Kuwahara H, Sobajima M, Kinugawa K. Clinical Advantages of Using Low Tube Voltage in Third-Generation 192-Slice Dual-Source Computed Tomographic Angiography Before Transcatheter Aortic Valve Implantation. International heart journal. 2019;60(5):1091-7.CrossRef
14.
Zurück zum Zitat Svensson A, Thor D, Fischer MA, Brismar T. Dual source abdominal computed tomography: the effect of reduced X-ray tube voltage and intravenous contrast media dosage in patients with reduced renal function. Acta radiologica (Stockholm, Sweden : 1987). 2019;60(3):293–300. Svensson A, Thor D, Fischer MA, Brismar T. Dual source abdominal computed tomography: the effect of reduced X-ray tube voltage and intravenous contrast media dosage in patients with reduced renal function. Acta radiologica (Stockholm, Sweden : 1987). 2019;60(3):293–300.
15.
Zurück zum Zitat Peer A, Averbukh Z, Berman S, Modai D, Averbukh M, Weissgarten J. Contrast media augmented apoptosis of cultured renal mesangial, tubular, epithelial, endothelial, and hepatic cells. Invest Radiol. 2003;38(3):177-82.PubMed Peer A, Averbukh Z, Berman S, Modai D, Averbukh M, Weissgarten J. Contrast media augmented apoptosis of cultured renal mesangial, tubular, epithelial, endothelial, and hepatic cells. Invest Radiol. 2003;38(3):177-82.PubMed
16.
Zurück zum Zitat Romano G, Briguori C, Quintavalle C, et al. Contrast agents and renal cell apoptosis. Eur Heart J. 2008;29(20):2569-76.CrossRef Romano G, Briguori C, Quintavalle C, et al. Contrast agents and renal cell apoptosis. Eur Heart J. 2008;29(20):2569-76.CrossRef
17.
Zurück zum Zitat Persson PB, Tepel M. Contrast medium-induced nephropathy: the pathophysiology. Kidney Int Suppl. 2006(100):S8-10.CrossRef Persson PB, Tepel M. Contrast medium-induced nephropathy: the pathophysiology. Kidney Int Suppl. 2006(100):S8-10.CrossRef
18.
Zurück zum Zitat Rudnick MR, Leonberg-Yoo AK, Litt HI, Cohen RM, Hilton S, Reese PP. The Controversy of Contrast-Induced Nephropathy With Intravenous Contrast: What Is the Risk? Am J Kidney Dis. 2020;75(1):105–13. Rudnick MR, Leonberg-Yoo AK, Litt HI, Cohen RM, Hilton S, Reese PP. The Controversy of Contrast-Induced Nephropathy With Intravenous Contrast: What Is the Risk? Am J Kidney Dis. 2020;75(1):105–13.
19.
Zurück zum Zitat Luk L, Steinman J, Newhouse JH. Intravenous Contrast-Induced Nephropathy-The Rise and Fall of a Threatening Idea. Adv Chronic Kidney Dis. 2017;24(3):169-75.CrossRef Luk L, Steinman J, Newhouse JH. Intravenous Contrast-Induced Nephropathy-The Rise and Fall of a Threatening Idea. Adv Chronic Kidney Dis. 2017;24(3):169-75.CrossRef
20.
Zurück zum Zitat McDonald RJ, McDonald JS, Newhouse JH, Davenport MS. Controversies in Contrast Material-induced Acute Kidney Injury: Closing in on the Truth? Radiology. 2015;277(3):627-32.CrossRef McDonald RJ, McDonald JS, Newhouse JH, Davenport MS. Controversies in Contrast Material-induced Acute Kidney Injury: Closing in on the Truth? Radiology. 2015;277(3):627-32.CrossRef
21.
Zurück zum Zitat Onishi H, Murakami T, Kim T, et al. Abdominal multi-detector row CT: effectiveness of determining contrast medium dose on basis of body surface area. European journal of radiology. 2011;80(3):643-7.CrossRef Onishi H, Murakami T, Kim T, et al. Abdominal multi-detector row CT: effectiveness of determining contrast medium dose on basis of body surface area. European journal of radiology. 2011;80(3):643-7.CrossRef
22.
Zurück zum Zitat Kalra MK, Woisetschlager M, Dahlstrom N, et al. Radiation dose reduction with Sinogram Affirmed Iterative Reconstruction technique for abdominal computed tomography. J Comput Assist Tomogr. 2012;36(3):339-46.CrossRef Kalra MK, Woisetschlager M, Dahlstrom N, et al. Radiation dose reduction with Sinogram Affirmed Iterative Reconstruction technique for abdominal computed tomography. J Comput Assist Tomogr. 2012;36(3):339-46.CrossRef
23.
Zurück zum Zitat Joemai RM, Geleijns J, Veldkamp WJ. Development and validation of a low dose simulator for computed tomography. European radiology. 2010;20(4):958-66.CrossRef Joemai RM, Geleijns J, Veldkamp WJ. Development and validation of a low dose simulator for computed tomography. European radiology. 2010;20(4):958-66.CrossRef
24.
Zurück zum Zitat Higashigaito K, Schmid T, Puippe G, et al. CT Angiography of the Aorta: Prospective Evaluation of Individualized Low-Volume Contrast Media Protocols. Radiology. 2016;280(3):960-8.CrossRef Higashigaito K, Schmid T, Puippe G, et al. CT Angiography of the Aorta: Prospective Evaluation of Individualized Low-Volume Contrast Media Protocols. Radiology. 2016;280(3):960-8.CrossRef
25.
Zurück zum Zitat Schoenhagen P, Hausleiter J, Achenbach S, Desai MY, Tuzcu EM. Computed tomography in the evaluation for transcatheter aortic valve implantation (TAVI). Cardiovasc Diagn Ther. 2011;1(1):44-56.PubMedPubMedCentral Schoenhagen P, Hausleiter J, Achenbach S, Desai MY, Tuzcu EM. Computed tomography in the evaluation for transcatheter aortic valve implantation (TAVI). Cardiovasc Diagn Ther. 2011;1(1):44-56.PubMedPubMedCentral
26.
Zurück zum Zitat Sharma SK, Kini A. Effect of nonionic radiocontrast agents on the occurrence of contrast-induced nephropathy in patients with mild-moderate chronic renal insufficiency: pooled analysis of the randomized trials. Catheter Cardiovasc Interv. 2005;65(3):386-93.CrossRef Sharma SK, Kini A. Effect of nonionic radiocontrast agents on the occurrence of contrast-induced nephropathy in patients with mild-moderate chronic renal insufficiency: pooled analysis of the randomized trials. Catheter Cardiovasc Interv. 2005;65(3):386-93.CrossRef
27.
Zurück zum Zitat Marin D, Nelson RC, Schindera ST, et al. Low-tube-voltage, high-tube-current multidetector abdominal CT: improved image quality and decreased radiation dose with adaptive statistical iterative reconstruction algorithm--initial clinical experience. Radiology. 2010;254(1):145-53.CrossRef Marin D, Nelson RC, Schindera ST, et al. Low-tube-voltage, high-tube-current multidetector abdominal CT: improved image quality and decreased radiation dose with adaptive statistical iterative reconstruction algorithm--initial clinical experience. Radiology. 2010;254(1):145-53.CrossRef
28.
Zurück zum Zitat Yonekura Y. Diagnostic reference levels based on latest surveys in Japan. Report from Japan network for research and information on medical exposure (J-RIME). 2015:1–11. Yonekura Y. Diagnostic reference levels based on latest surveys in Japan. Report from Japan network for research and information on medical exposure (J-RIME). 2015:1–11.
Metadaten
Titel
Image quality in dual-source multiphasic dynamic computed tomography of the abdomen: evaluating the effects of a low tube voltage (70 kVp) in combination with contrast dose reduction
verfasst von
Keisuke Miyoshi
Hideko Onoda
Mashiro Tanabe
Sei Nakao
Mayumi Higashi
Etsushi Iida
Munemasa Okada
Matakazu Furukawa
Katsuyoshi Ito
Publikationsdatum
07.05.2020
Verlag
Springer US
Erschienen in
Abdominal Radiology / Ausgabe 11/2020
Print ISSN: 2366-004X
Elektronische ISSN: 2366-0058
DOI
https://doi.org/10.1007/s00261-020-02565-9

Weitere Artikel der Ausgabe 11/2020

Abdominal Radiology 11/2020 Zur Ausgabe

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.