15.09.2015 | Computed Tomography

Impact of an advanced image-based monoenergetic reconstruction algorithm on coronary stent visualization using third generation dual-source dual-energy CT: a phantom study

verfasst von: Stefanie Mangold, Paola M. Cannaó, U. Joseph Schoepf, Julian L. Wichmann, Christian Canstein, Stephen R. Fuller, Giuseppe Muscogiuri, Akos Varga-Szemes, Konstantin Nikolaou, Carlo N. De Cecco

Erschienen in: European Radiology | Ausgabe 6/2016

Einloggen, um Zugang zu erhalten

Abstract

Purpose

To evaluate the impact of an advanced monoenergetic (ME) reconstruction algorithm on CT coronary stent imaging in a phantom model.

Materials and methods

Three stents with lumen diameters of 2.25, 3.0 and 3.5 mm were examined with a third-generation dual-source dual-energy CT (DECT). Tube potential was set at 90/Sn150 kV for DE and 70, 90 or 120 kV for single-energy (SE) acquisitions and advanced modelled iterative reconstruction was used. Overall, 23 reconstructions were evaluated for each stent including three SE acquisitions and ten advanced and standard ME images with virtual photon energies from 40 to 130 keV, respectively. In-stent luminal diameter was measured and compared to nominal lumen diameter to determine stent lumen visibility. Contrast-to-noise ratio was calculated.

Results

Advanced ME reconstructions substantially increased lumen visibility in comparison to SE for stents ≤3 mm. 130 keV images produced the best mean lumen visibility: 86 % for the 2.25 mm stent (82 % for standard ME and 64 % for SE) and 82 % for the 3.0 mm stent (77 % for standard ME and 69 % for SE). Mean DLP for SE 120 kV and DE acquisitions were 114.4 ± 9.8 and 58.9 ± 2.2 mGy × cm, respectively.

Conclusion

DECT with advanced ME reconstructions improves the in-lumen visibility of small stents in comparison with standard ME and SE imaging.

Key Points

An advanced image-based monoenergetic reconstruction algorithm improves lumen visualization in stents ≤3.0 mm.
Application of high keV reconstructions significantly improves in-stent lumen visualization.
DECT acquisition resulted in 49 % radiation dose reduction compared with 120 kV SE.
Literatur
1.
Zurück zum Zitat Mahnken AH (2012) CT imaging of coronary stents: past, present, and future. ISRN Cardiol 139823 Mahnken AH (2012) CT imaging of coronary stents: past, present, and future. ISRN Cardiol 139823
2.
Zurück zum Zitat Min JK, Swaminathan RV, Vass M, Gallagher S, Weinsaft JW (2009) High-definition multidetector computed tomography for evaluation of coronary artery stents: comparison to standard-definition 64-detector row computed tomography. J Cardiovasc Comput Tomogr 3:246–251CrossRefPubMed Min JK, Swaminathan RV, Vass M, Gallagher S, Weinsaft JW (2009) High-definition multidetector computed tomography for evaluation of coronary artery stents: comparison to standard-definition 64-detector row computed tomography. J Cardiovasc Comput Tomogr 3:246–251CrossRefPubMed
3.
Zurück zum Zitat Gassenmaier T, Petri N, Allmendinger T et al (2014) Next generation coronary CT angiography: in vitro evaluation of 27 coronary stents. Eur Radiol 24:2953–2961CrossRefPubMed Gassenmaier T, Petri N, Allmendinger T et al (2014) Next generation coronary CT angiography: in vitro evaluation of 27 coronary stents. Eur Radiol 24:2953–2961CrossRefPubMed
4.
Zurück zum Zitat Ulrich A, Burg MC, Raupach R et al (2015) Coronary stent imaging with dual-source CT: assessment of lumen visibility using different convolution kernels and postprocessing filters. Acta Radiol 56:42–50CrossRefPubMed Ulrich A, Burg MC, Raupach R et al (2015) Coronary stent imaging with dual-source CT: assessment of lumen visibility using different convolution kernels and postprocessing filters. Acta Radiol 56:42–50CrossRefPubMed
5.
Zurück zum Zitat von Spiczak J, Morsbach F, Winklhofer S et al (2013) Coronary artery stent imaging with CT using an integrated electronics detector and iterative reconstructions: first in vitro experience. J Cardiovasc Comput Tomogr 7:215–222CrossRef von Spiczak J, Morsbach F, Winklhofer S et al (2013) Coronary artery stent imaging with CT using an integrated electronics detector and iterative reconstructions: first in vitro experience. J Cardiovasc Comput Tomogr 7:215–222CrossRef
6.
Zurück zum Zitat Eisentopf J, Achenbach S, Ulzheimer S et al (2013) Low-dose dual-source CT angiography with iterative reconstruction for coronary artery stent evaluation. JACC Cardiovasc Imaging 6:458–465CrossRefPubMed Eisentopf J, Achenbach S, Ulzheimer S et al (2013) Low-dose dual-source CT angiography with iterative reconstruction for coronary artery stent evaluation. JACC Cardiovasc Imaging 6:458–465CrossRefPubMed
7.
Zurück zum Zitat Zhou Q, Jiang B, Dong F, Huang P, Liu H, Zhang M (2014) Computed tomography coronary stent imaging with iterative reconstruction: a trade-off study between medium kernel and sharp kernel. J Comput Assist Tomogr 38:604–612CrossRefPubMed Zhou Q, Jiang B, Dong F, Huang P, Liu H, Zhang M (2014) Computed tomography coronary stent imaging with iterative reconstruction: a trade-off study between medium kernel and sharp kernel. J Comput Assist Tomogr 38:604–612CrossRefPubMed
8.
Zurück zum Zitat Geyer LL, Glenn GR, De Cecco CN et al (2015) CT evaluation of small-diameter coronary artery stents: effect of an integrated circuit detector with iterative reconstruction. Radiology 276:706–714 Geyer LL, Glenn GR, De Cecco CN et al (2015) CT evaluation of small-diameter coronary artery stents: effect of an integrated circuit detector with iterative reconstruction. Radiology 276:706–714
9.
Zurück zum Zitat Wichmann JL, Arbaciauskaite R, Kerl JM et al (2014) Evaluation of monoenergetic late iodine enhancement dual-energy computed tomography for imaging of chronic myocardial infarction. Eur Radiol 24:1211–1218CrossRefPubMed Wichmann JL, Arbaciauskaite R, Kerl JM et al (2014) Evaluation of monoenergetic late iodine enhancement dual-energy computed tomography for imaging of chronic myocardial infarction. Eur Radiol 24:1211–1218CrossRefPubMed
10.
Zurück zum Zitat Bamberg F, Dierks A, Nikolaou K, Reiser MF, Becker CR, Johnson TR (2011) Metal artifact reduction by dual energy computed tomography using monoenergetic extrapolation. Eur Radiol 21:1424–1429CrossRefPubMed Bamberg F, Dierks A, Nikolaou K, Reiser MF, Becker CR, Johnson TR (2011) Metal artifact reduction by dual energy computed tomography using monoenergetic extrapolation. Eur Radiol 21:1424–1429CrossRefPubMed
11.
Zurück zum Zitat Bongers MN, Schabel C, Krauss B et al (2014) Noise-optimized virtual monoenergetic images and iodine maps for the detection of venous thrombosis in second-generation dual-energy CT (DECT): an ex vivo phantom study. Eur Radiol 25:1655–1664 Bongers MN, Schabel C, Krauss B et al (2014) Noise-optimized virtual monoenergetic images and iodine maps for the detection of venous thrombosis in second-generation dual-energy CT (DECT): an ex vivo phantom study. Eur Radiol 25:1655–1664
12.
Zurück zum Zitat Schabel C, Bongers M, Sedlmair M et al (2014) Assessment of the hepatic veins in poor contrast conditions using dual energy CT: evaluation of a novel monoenergetic extrapolation software algorithm. Röfo 186:591–597PubMed Schabel C, Bongers M, Sedlmair M et al (2014) Assessment of the hepatic veins in poor contrast conditions using dual energy CT: evaluation of a novel monoenergetic extrapolation software algorithm. Röfo 186:591–597PubMed
13.
Zurück zum Zitat Grant KL, Flohr TG, Krauss B, Sedlmair M, Thomas C, Schmidt B (2014) Assessment of an advanced image-based technique to calculate virtual monoenergetic computed tomographic images from a dual-energy examination to improve contrast-to-noise ratio in examinations using iodinated contrast media. Investig Radiol 49:586–592CrossRef Grant KL, Flohr TG, Krauss B, Sedlmair M, Thomas C, Schmidt B (2014) Assessment of an advanced image-based technique to calculate virtual monoenergetic computed tomographic images from a dual-energy examination to improve contrast-to-noise ratio in examinations using iodinated contrast media. Investig Radiol 49:586–592CrossRef
14.
Zurück zum Zitat Stehli J, Fuchs TA, Singer A et al (2014) First experience with single-source, dual-energy CCTA for monochromatic stent imaging. Eur Heart J Cardiovasc Imaging 16:507–512 Stehli J, Fuchs TA, Singer A et al (2014) First experience with single-source, dual-energy CCTA for monochromatic stent imaging. Eur Heart J Cardiovasc Imaging 16:507–512
15.
Zurück zum Zitat Winklehner A, Goetti R, Baumueller S et al (2011) Automated attenuation-based tube potential selection for thoracoabdominal computed tomography angiography: improved dose effectiveness. Investig Radiol 46:767–773CrossRef Winklehner A, Goetti R, Baumueller S et al (2011) Automated attenuation-based tube potential selection for thoracoabdominal computed tomography angiography: improved dose effectiveness. Investig Radiol 46:767–773CrossRef
17.
Zurück zum Zitat Carbone I, Francone M, Algeri E et al (2008) Non-invasive evaluation of coronary artery stent patency with retrospectively ECG-gated 64-slice CT angiography. Eur Radiol 18:234–243CrossRefPubMed Carbone I, Francone M, Algeri E et al (2008) Non-invasive evaluation of coronary artery stent patency with retrospectively ECG-gated 64-slice CT angiography. Eur Radiol 18:234–243CrossRefPubMed
18.
Zurück zum Zitat Oncel D, Oncel G, Tastan A, Tamci B (2008) Evaluation of coronary stent patency and in-stent restenosis with dual-source CT coronary angiography without heart rate control. AJR Am J Roentgenol 191:56–63CrossRefPubMed Oncel D, Oncel G, Tastan A, Tamci B (2008) Evaluation of coronary stent patency and in-stent restenosis with dual-source CT coronary angiography without heart rate control. AJR Am J Roentgenol 191:56–63CrossRefPubMed
19.
Zurück zum Zitat Pugliese F, Weustink AC, Van Mieghem C et al (2008) Dual source coronary computed tomography angiography for detecting in-stent restenosis. Heart 94:848–854CrossRefPubMed Pugliese F, Weustink AC, Van Mieghem C et al (2008) Dual source coronary computed tomography angiography for detecting in-stent restenosis. Heart 94:848–854CrossRefPubMed
20.
Zurück zum Zitat Taylor AJ, Cerqueira M, Hodgson JM et al (2010) ACCF/SCCT/ACR/AHA/ASE/ASNC/NASCI/SCAI/SCMR 2010 appropriate Use criteria for cardiac computed tomography. A report of the AMERICAN college of Cardiology Foundation appropriate use criteria task force, the society of cardiovascular computed tomography, the American College of Radiology, the American Heart Association, the American Society of Echocardiography, the American Society of Nuclear Cardiology, the North American Society for Cardiovascular Imaging, the Society for Cardiovascular Angiography and Interventions, and the Society for Cardiovascular Magnetic Resonance. Circulation 122:e525–e555CrossRefPubMed Taylor AJ, Cerqueira M, Hodgson JM et al (2010) ACCF/SCCT/ACR/AHA/ASE/ASNC/NASCI/SCAI/SCMR 2010 appropriate Use criteria for cardiac computed tomography. A report of the AMERICAN college of Cardiology Foundation appropriate use criteria task force, the society of cardiovascular computed tomography, the American College of Radiology, the American Heart Association, the American Society of Echocardiography, the American Society of Nuclear Cardiology, the North American Society for Cardiovascular Imaging, the Society for Cardiovascular Angiography and Interventions, and the Society for Cardiovascular Magnetic Resonance. Circulation 122:e525–e555CrossRefPubMed
21.
Zurück zum Zitat Ebersberger U, Tricarico F, Schoepf UJ et al (2013) CT evaluation of coronary artery stents with iterative image reconstruction: improvements in image quality and potential for radiation dose reduction. Eur Radiol 23:125–132CrossRefPubMed Ebersberger U, Tricarico F, Schoepf UJ et al (2013) CT evaluation of coronary artery stents with iterative image reconstruction: improvements in image quality and potential for radiation dose reduction. Eur Radiol 23:125–132CrossRefPubMed
22.
Zurück zum Zitat Leipsic J, Heilbron BG, Hague C (2012) Iterative reconstruction for coronary CT angiography: finding its way. Int J Cardiovasc Imaging 28:613–620CrossRefPubMed Leipsic J, Heilbron BG, Hague C (2012) Iterative reconstruction for coronary CT angiography: finding its way. Int J Cardiovasc Imaging 28:613–620CrossRefPubMed
23.
Zurück zum Zitat Renker M, Geyer LL, Krazinski AW, Silverman JR, Ebersberger U, Schoepf UJ (2013) Iterative image reconstruction: a realistic dose-saving method in cardiac CT imaging? Expert Rev Cardiovasc Ther 11:403–409CrossRefPubMed Renker M, Geyer LL, Krazinski AW, Silverman JR, Ebersberger U, Schoepf UJ (2013) Iterative image reconstruction: a realistic dose-saving method in cardiac CT imaging? Expert Rev Cardiovasc Ther 11:403–409CrossRefPubMed
24.
Zurück zum Zitat Maintz D, Burg MC, Seifarth H et al (2009) Update on multidetector coronary CT angiography of coronary stents: in vitro evaluation of 29 different stent types with dual-source CT. Eur Radiol 19:42–49CrossRefPubMed Maintz D, Burg MC, Seifarth H et al (2009) Update on multidetector coronary CT angiography of coronary stents: in vitro evaluation of 29 different stent types with dual-source CT. Eur Radiol 19:42–49CrossRefPubMed
25.
Zurück zum Zitat Maintz D, Seifarth H, Raupach R et al (2006) 64-slice multidetector coronary CT angiography: in vitro evaluation of 68 different stents. Eur Radiol 16:818–826CrossRefPubMed Maintz D, Seifarth H, Raupach R et al (2006) 64-slice multidetector coronary CT angiography: in vitro evaluation of 68 different stents. Eur Radiol 16:818–826CrossRefPubMed
Metadaten
Titel
Impact of an advanced image-based monoenergetic reconstruction algorithm on coronary stent visualization using third generation dual-source dual-energy CT: a phantom study
verfasst von
Stefanie Mangold
Paola M. Cannaó
U. Joseph Schoepf
Julian L. Wichmann
Christian Canstein
Stephen R. Fuller
Giuseppe Muscogiuri
Akos Varga-Szemes
Konstantin Nikolaou
Carlo N. De Cecco
Publikationsdatum
15.09.2015
Verlag
Springer Berlin Heidelberg
Erschienen in
European Radiology / Ausgabe 6/2016
Print ISSN: 0938-7994
Elektronische ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-015-3997-4

Weitere Artikel der Ausgabe 6/2016

European Radiology 6/2016 Zur Ausgabe

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.