27.07.2024 | Computed Tomography

Impact of different peak tube voltage settings on adrenal adenomas attenuation at unenhanced CT

verfasst von: Matteo Bonatti, Riccardo Valletta, Valentina Corato, Nadia Oberhofer, Stefano Piffer, Vincenzo Vingiani, Andrea Posteraro, Bernardo Proner, Fabio Lombardo, Giacomo Avesani, Rima Cepurnaite, Giulia A. Zamboni

Erschienen in: European Radiology | Ausgabe 2/2025

Einloggen, um Zugang zu erhalten

Abstract

Objectives

To assess the influence of peak tube voltage peak setting on adrenal adenomas (AA) attenuation on unenhanced abdominal CT.

Materials and methods

IRB-approved retrospective observational cohort study. We included 89 patients with imaging-defined AAs with shortest diameter > 6 mm who underwent two or more unenhanced abdominal CTs using at least two different peak tube voltage settings. Two readers independently measured adenoma attenuation on different CT acquisitions by drawing a round ROI on 3 mm thick axial MPR reconstructions encompassing at least 2/3 of the lesion’s surface. The mean of the values measured by the two readers was used for further analysis. Interobserver variability was assessed (Intraclass Correlation Coefficient). Attenuation values measured on 100, 110 and 140 kVp acquisitions were compared with standard 120 kVp ones (Bland-Altman analysis).

Results

We included 275 unenhanced abdominal CTs (3.1 ± 0.9/patient) in image analysis; 131 acquired at 120 kVp, 65 at 100 kVp, 59 at 110 kVp, and 20 at 140 kVp. 107 lesions were detected in 89 patients (1–4/patient), with a mean maximum diameter of 17 ± 6 mm. Interobserver agreement in attenuation measurement was excellent (ICC: 0.95, CI (92–97)). Median adenoma attenuation was significantly lower on 100 kVp images than on 120 kVp ones (−1 HU, IQR (−5 to 3.6), vs, 2.5 HU, IQR (−1.5 to 8.5); p < 0.001) whereas we didn’t find statistically significant differences in adenoma attenuation between 110 kVp or 140 kVp and 120 kVp ones.

Conclusion

AA attenuation is significantly lower on unenhanced CT scans acquired at 100 kVp than on those acquired at “standard” 120 kVp.

Clinical relevance statement

AA attenuation is significantly lower at 100 kVp in comparison to 120 kVp. This might be exploited to increase unenhanced CT sensitivity in adenoma characterisation, but further studies including non-adenoma lesions are mandatory to confirm this hypothesis.

Key Points

  • CT scans are often acquired using peak tube voltage settings different from the “standard” 120kVp.
  • AA attenuation varies if CT scans are acquired using different tube peak voltage settings.
  • At 100kVp AAs show a significantly lower attenuation than at 120kVp.
Anhänge
Nur mit Berechtigung zugänglich
Literatur
Metadaten
Titel
Impact of different peak tube voltage settings on adrenal adenomas attenuation at unenhanced CT
verfasst von
Matteo Bonatti
Riccardo Valletta
Valentina Corato
Nadia Oberhofer
Stefano Piffer
Vincenzo Vingiani
Andrea Posteraro
Bernardo Proner
Fabio Lombardo
Giacomo Avesani
Rima Cepurnaite
Giulia A. Zamboni
Publikationsdatum
27.07.2024
Verlag
Springer Berlin Heidelberg
Erschienen in
European Radiology / Ausgabe 2/2025
Print ISSN: 0938-7994
Elektronische ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-024-10984-1

Neu im Fachgebiet Radiologie

Röntgen-Thorax oder LDCT fürs Lungenscreening nach HNSCC?

Personen, die an einem Plattenepithelkarzinom im Kopf-Hals-Bereich erkrankt sind, haben ein erhöhtes Risiko für Metastasen oder zweite Primärmalignome der Lunge. Eine Studie hat untersucht, wie die radiologische Überwachung aussehen sollte.

Statine: Was der G-BA-Beschluss für Praxen bedeutet

Nach dem G-BA-Beschluss zur erweiterten Verordnungsfähigkeit von Lipidsenkern rechnet die DEGAM mit 200 bis 300 neuen Dauerpatienten pro Praxis. Im Interview erläutert Präsidiumsmitglied Erika Baum, wie Hausärztinnen und Hausärzte am besten vorgehen.

Brustdichte nicht mit Multivitaminpräparat-Einnahme assoziiert

Der regelmäßige Gebrauch von Nahrungsergänzungsmitteln scheint nicht die mammografische Brustdichte zu erhöhen. In einer US-amerikanischen Studie jedenfalls ließ sich ein derartiger Zusammenhang nicht bestätigen.

Erhöhte Suizidrate unter US-Ärztinnen

Während der Arztberuf Männer eher vor Suizid schützt, erhöht er das Risiko bei Frauen – zumindest in den USA: Die Suizidinzidenz unter Ärztinnen ist um die Hälfte höher als unter Frauen mit anderen Berufen. Männliche Ärzte töten sich dennoch wesentlich häufiger selbst als weibliche.

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.