Impact of transducer frequency setting on speckle tracking measures

verfasst von: Flemming Javier Olsen, Jesper Hastrup Svendsen, Lars Køber, Søren Højberg, Ketil Haugan, Jan Skov Jensen, Tor Biering-Sørensen

Erschienen in: The International Journal of Cardiovascular Imaging | Ausgabe 3/2018

Einloggen, um Zugang zu erhalten

Abstract

Speckle tracking echocardiography is an emerging technique, which is currently being included in clinical guidelines. We sought to investigate the impact of transducer frequency settings on speckle tracking derived measures. The study comprised of 22 subjects prospectively enrolled for a randomized controlled trial (LOOP-study, Clinicaltrials.gov:NCT02036450). Patients were above 70 years of age with increased risk of stroke, and had an echocardiogram performed, which included focused images of the left ventricle. Focused images were obtained with the transducer frequency set at both 1.7/3.3 and 1.5/3.0 MHz. The images were obtained immediately after each other at the exact same position for the two settings. Speckle tracking was performed in three apical projections, allowing for acquisition of layered global longitudinal strain (GLS) and strain rate measures. Concordance between the frequency settings was tested for endo-, mid-, and epicardial GLS and strain rates by coefficients of variation, bias coefficients and visually displayed by Bland–Altman plots. Bland–Altman plots did not reveal any significant over- or underestimation of any speckle tracking measure. Bias coefficients showed that none of the measurements differed significantly between the two settings (bias for GLSendo = − 0.07 ± 2.94, p = 0.91; GLSmid = 0.02 ± 2.70, p = 0.98, GLSepi = 0.07 ± 2.53, p = 0.90). Coefficients of variation were as follows: GLSendo = 15.11%, GLSmid = 15.28%, GLSepi = 17.26%, systolic strain rate = 15.66%, early diastolic strain rate = 38.46%, late diastolic strain rate = 11%. Changing between transducer frequency settings does not systematically derange speckle tracking measures. One can safely reduce the transducer frequency without compromising the validity of speckle tracking derived measures.
Literatur
1.
Zurück zum Zitat Biering-Sørensen T, Biering-Sørensen SR, Olsen FJ et al (2017) Global longitudinal strain by echocardiography predicts long-term risk of cardiovascular Morbidity and mortality in a low-risk general population: the Copenhagen city heart study. Circ Cardiovasc Imaging. doi:10.1161/CIRCIMAGING.116.005521 Biering-Sørensen T, Biering-Sørensen SR, Olsen FJ et al (2017) Global longitudinal strain by echocardiography predicts long-term risk of cardiovascular Morbidity and mortality in a low-risk general population: the Copenhagen city heart study. Circ Cardiovasc Imaging. doi:10.​1161/​CIRCIMAGING.​116.​005521
2.
Zurück zum Zitat Bax JJ, Delgado V, Sogaard P et al (2017) Prognostic implications of left ventricular global longitudinal strain in heart failure patients with narrow QRS complex treated with cardiac resynchronization therapy: a subanalysis of the randomized EchoCRT trial. Eur Heart J 38:720–726. doi:10.1093/eurheartj/ehw506 CrossRefPubMed Bax JJ, Delgado V, Sogaard P et al (2017) Prognostic implications of left ventricular global longitudinal strain in heart failure patients with narrow QRS complex treated with cardiac resynchronization therapy: a subanalysis of the randomized EchoCRT trial. Eur Heart J 38:720–726. doi:10.​1093/​eurheartj/​ehw506 CrossRefPubMed
4.
Zurück zum Zitat Plana JC, Galderisi M, Barac A et al (2014) Expert consensus for multimodality imaging evaluation of adult patients during and after cancer therapy: a report from the American Society of Echocardiography and the European Association of Cardiovascular Imaging. Eur Heart J Cardiovasc Imaging 15:1063–1093. doi:10.1093/ehjci/jeu192 CrossRefPubMedPubMedCentral Plana JC, Galderisi M, Barac A et al (2014) Expert consensus for multimodality imaging evaluation of adult patients during and after cancer therapy: a report from the American Society of Echocardiography and the European Association of Cardiovascular Imaging. Eur Heart J Cardiovasc Imaging 15:1063–1093. doi:10.​1093/​ehjci/​jeu192 CrossRefPubMedPubMedCentral
5.
Zurück zum Zitat Voigt J-U, Pedrizzetti G, Lysyansky P et al (2015) Definitions for a common standard for 2D speckle tracking echocardiography: consensus document of the EACVI/ASE/Industry Task Force to standardize deformation imaging. Eur Heart J Cardiovasc Imaging 16:1–11. doi:10.1093/ehjci/jeu184 CrossRefPubMed Voigt J-U, Pedrizzetti G, Lysyansky P et al (2015) Definitions for a common standard for 2D speckle tracking echocardiography: consensus document of the EACVI/ASE/Industry Task Force to standardize deformation imaging. Eur Heart J Cardiovasc Imaging 16:1–11. doi:10.​1093/​ehjci/​jeu184 CrossRefPubMed
7.
Zurück zum Zitat Armstrong WF, Ryan T (2009) Feigenbaum’s echocardiography, 7th edn. Lippincott Williams & Wilkins, Philadelphia Armstrong WF, Ryan T (2009) Feigenbaum’s echocardiography, 7th edn. Lippincott Williams & Wilkins, Philadelphia
8.
Zurück zum Zitat Diederichsen SZ, Haugan KJ, Køber L et al (2017) Atrial fibrillation detected by continuous electrocardiographic monitoring using implantable loop recorder to prevent stroke in individuals at risk (the LOOP study): rationale and design of a large randomized controlled trial. Am Heart J 187:122–132. doi:10.1016/j.ahj.2017.02.017 CrossRefPubMed Diederichsen SZ, Haugan KJ, Køber L et al (2017) Atrial fibrillation detected by continuous electrocardiographic monitoring using implantable loop recorder to prevent stroke in individuals at risk (the LOOP study): rationale and design of a large randomized controlled trial. Am Heart J 187:122–132. doi:10.​1016/​j.​ahj.​2017.​02.​017 CrossRefPubMed
10.
Zurück zum Zitat Sarvari SI, Haugaa KH, Zahid W et al (2013) Layer-specific quantification of myocardial deformation by strain echocardiography may reveal significant CAD in patients with non-ST-segment elevation acute coronary syndrome. JACC Cardiovasc Imaging 6:535–544. doi:10.1016/j.jcmg.2013.01.009 CrossRefPubMed Sarvari SI, Haugaa KH, Zahid W et al (2013) Layer-specific quantification of myocardial deformation by strain echocardiography may reveal significant CAD in patients with non-ST-segment elevation acute coronary syndrome. JACC Cardiovasc Imaging 6:535–544. doi:10.​1016/​j.​jcmg.​2013.​01.​009 CrossRefPubMed
12.
13.
Zurück zum Zitat Risum N, Ali S, Olsen NT et al (2012) Variability of global left ventricular deformation analysis using vendor dependent and independent two-dimensional speckle-tracking software in adults. J Am Soc Echocardiogr 25:1195–1203. doi:10.1016/j.echo.2012.08.007 CrossRefPubMed Risum N, Ali S, Olsen NT et al (2012) Variability of global left ventricular deformation analysis using vendor dependent and independent two-dimensional speckle-tracking software in adults. J Am Soc Echocardiogr 25:1195–1203. doi:10.​1016/​j.​echo.​2012.​08.​007 CrossRefPubMed
Metadaten
Titel
Impact of transducer frequency setting on speckle tracking measures
verfasst von
Flemming Javier Olsen
Jesper Hastrup Svendsen
Lars Køber
Søren Højberg
Ketil Haugan
Jan Skov Jensen
Tor Biering-Sørensen
Publikationsdatum
03.11.2017
Verlag
Springer Netherlands
Erschienen in
The International Journal of Cardiovascular Imaging / Ausgabe 3/2018
Print ISSN: 1569-5794
Elektronische ISSN: 1875-8312
DOI
https://doi.org/10.1007/s10554-017-1254-4

Weitere Artikel der Ausgabe 3/2018

The International Journal of Cardiovascular Imaging 3/2018 Zur Ausgabe

Update Kardiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.