Implantation of a VDD implantable cardioverter-defibrillator lead via a persistent left superior vena cava

  • Open Access
  • 06.01.2022
  • Case Reports
Erschienen in:
download
DOWNLOAD
print
DRUCKEN
insite
SUCHEN

Abstract

A persistent left superior vena cava (LSVC) represents a challenging congenital abnormality for transvenous cardiac device implantation. In the current case a secondary prophylactic VDD implantable cardioverter-defibrillator (ICD) implantation was planned in a 75-year-old woman presenting with ischemic cardiomyopathy and elevated stroke risk. Since no venous communication to the right side was identified intraoperatively, the lead was placed via the persistent LSVC. The far-field signal on the floating atrial dipole could be successfully blanked out, and appropriate device function with high and stable atrial sensing was demonstrated at follow-up.
Scan QR code & read article online

Medical history

The case of a 75-year-old woman with ventricular tachycardia accompanied by hemodynamic instability due to ischemic cardiomyopathy (left ventricular ejection fraction of 37%) is reported. Since coronary angiography could not identify a target lesion, secondary prophylactic implantable cardioverter-defibrillator (ICD) implantation was indicated. In light of slightly prolonged atrioventricular (AV) conduction (PQ 210 ms), narrow QRS (110 ms) and an elevated stroke risk (CHA2DS2-VASc score: 6), the decision was taken to implant a VDD ICD system (Intica 5 VR‑T DX, Biotronik SE & Co., Berlin, Germany) with the capability of monitoring atrial arrhythmias.

Therapy and its course

After starting the procedure from a typical left infraclavicular side, a persistent left superior vena cava (LSVC) was recognized and no relevant left to right venous communication could be identified by intraoperative venography. Hence, the lead (Plexa ProMRI DF‑1 DX 65/15, Biotronik SE & Co., Berlin, Germany) was successfully placed from the left subclavian approach via the LSVC by using a hand-shaped stylet (Fig. 1). Beside a high atrial sensing amplitude (8.7 mV), a double potential typical for coronary sinus was recorded directly after implantation in the atrial channel, especially during AV-synchronous ventricular pacing, as expected (Fig. 2). The far-field signal could be successfully blanked out with prolongation of the far-field protection window. Appropriate device function with stable and high atrial sensing amplitude was demonstrated at follow-up (4.3–6.5 mV at 1 year) (Fig. 3).
Fig. 1
Left (a) and right (b) anterior oblique fluoroscopy projections of a VDD implantable cardioverter-defibrillator lead inserted via a persistent left superior vena cava
Bild vergrößern
Fig. 2
A double potential typical for coronary sinus was recorded in the atrial channel during atrioventricular-sequential ventricular pacing, which was successfully blanked out with the prolongation of the far-field protection window (FFP)
Bild vergrößern
Fig. 3
Sensing test at 1‑year follow-up showing high and stable atrial and ventricular signal amplitudes without any far-field oversensing
Bild vergrößern

Discussion

To preserve the benefit of atrial sensing without the need to implant an additional lead, a single-lead ICD system with a floating atrial dipole (VDD or DX ICD) has been developed [1]. The VDD ICD system offers an additional atrial intracardiac electrogram, with early detection of atrial arrhythmias, possibly improved supraventricular tachycardia discrimination and AV-sequential pacing in single-lead devices. Moreover, it can be upgraded to a two-lead cardiac resynchronization therapy (CRT)-DX system in the case of stable, long-term atrial sensing and a developing need for CRT. The feasibility of the VDD ICD system in the case of LSCV may be of particular interest.
The incidence of a persistent LSVC has been reported in up to 0.66% of ICD recipients [2]. Lead placement may be technically challenging, especially in the case of absent right superior vena cava. Although most cases can be accomplished with a reliable outcome [2], in some patients implantation of alternative systems should be considered [3, 4].
The current case represents an optimal candidate for a VDD ICD: slightly prolonged AV conduction with a potential need for AV-synchronous pacing and/or upgrade to CRT in the future and a greatly elevated risk of stroke [1]. However, implantation of more complex leads—such as an ICD lead—is often technically challenging in patients with cardiac anatomic variants. To the best of the authorsʼ knowledge, this is the first report demonstrating the feasibility of the VDD ICD system in the case of LSCV.

Declarations

Conflict of interest

M. Vamos reports consulting fees and/or nonfinancial support from Biotronik, Medtronic, Minimal Invasive Technology Ltd., and Pfizer, outside the submitted work. L. Saghy and G. Bencsik declare that they have no competing interests.
For this article no studies with human participants or animals were performed by any of the authors. All studies performed were in accordance with the ethical standards indicated in each case. Additional written informed consent was obtained from all individual participants or their legal representatives for whom identifying information is included in this article.
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
download
DOWNLOAD
print
DRUCKEN
Titel
Implantation of a VDD implantable cardioverter-defibrillator lead via a persistent left superior vena cava
Verfasst von
PD Dr. med. habil. Mate Vamos, MD, PhD
Laszlo Saghy, MD, PhD
Gabor Bencsik, MD, PhD
Publikationsdatum
06.01.2022
Verlag
Springer Medizin
Erschienen in
Herzschrittmachertherapie + Elektrophysiologie / Ausgabe 1/2022
Print ISSN: 0938-7412
Elektronische ISSN: 1435-1544
DOI
https://doi.org/10.1007/s00399-021-00835-7
1.
Zurück zum Zitat Vamos M, Nemeth M, Balazs T, Zima E, Duray GZ (2021) Rationale and feasibility of the atrioventricular single-lead ICD systems with a floating atrial dipole (DX) in clinical practice. Trends Cardiovasc Med. https://doi.org/10.1016/j.tcm.2021.01.003CrossRefPubMed
2.
Zurück zum Zitat Biffi M, Bertini M, Ziacchi M, Martignani C, Valzania C, Diemberger I, Branzi A, Boriani G (2009) Clinical implications of left superior vena cava persistence in candidates for pacemaker or cardioverter-defibrillator implantation. Heart Vessels 24:142–146CrossRef
3.
Zurück zum Zitat Polewczyk A, Kutarski A, Czekajska-Chehab E, Adamczyk P, Boczar K, Polewczyk M, Janion M (2014) Complications of permanent cardiac pacing in patients with persistent left superior vena cava. Cardiol J 21(2):128–137CrossRef
4.
Zurück zum Zitat Erath JW, Sirat AS, Vamos M, Hohnloser SH (2016) Epicardial CRT-P- and S‑ICD implantation in a young patient with persistent left superior vena cava. Herzschrittmacherther Elektrophysiol 27:396–398CrossRef

Neu im Fachgebiet Kardiologie

Beschleunigt leichte nächtliche Hypertonie eine Niereninsuffizienz?

Haben Menschen mit einer chronischen Niereninsuffizienz nachts systolische Blutdruckwerte über 110 mmHg, scheint dies die Nierenfunktion weiter zu verschlechtern. Ab 65 Jahren ist es nach Daten aus China aber genau umgekehrt.

Orales Semaglutid kann Herzinsuffizienz-Komplikationen reduzieren

Die Behandlung eines Typ-2-Diabetes mit oralem Semaglutid scheint sich auch auf eine bestehende Herzinsuffizienz günstig auszuwirken. Laut einer Sekundäranalyse der SOUL-Studie profitieren vor allem Patienten und Patientinnen mit erhaltener Auswurffraktion.

Myokardschäden nach nichtkardialer OP besser kardiologisch abklären lassen

Menschen mit perioperativer Myokardschädigung (PMI) profitieren davon, wenn sie routinemäßig von einer Kardiologin oder einem Kardiologen untersucht werden. Dafür sprechen Ergebnisse einer Schweizer Studie.  

Linksschenkelblock erhöht Mortalität nach TAVI

  • 03.02.2026
  • TAVI
  • Nachrichten

Die Inzidenz eines nach TAVI neu aufgetretenen Linksschenkelblocks ist in den letzten Jahren zurückgegangen. Trotzdem handelt es sich weiterhin um eine häufige Komplikation – die die Aussichten nach dem Eingriff deutlich trübt.

EKG Essentials: EKG befunden mit System

In diesem CME-Kurs können Sie Ihr Wissen zur EKG-Befundung anhand von zwölf Video-Tutorials auffrischen und 10 CME-Punkte sammeln.
Praxisnah, relevant und mit vielen Tipps & Tricks vom Profi.

Update Kardiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.

Bildnachweise
Hand greift nach Skalpell/© Morsa Images / Getty Images / iStock, EKG befunden mit System - EKG Essential/© Springer Medizin Verlag GmbH