In-stent restenosis and longitudinal stent deformation: a case report

verfasst von: Daoyuan Si, Yaliang Tong, Bo Yu, Yuquan He, Guohui Liu

Erschienen in: BMC Cardiovascular Disorders | Ausgabe 1/2020

Abstract

Background

Longitudinal stent deformation (LSD) is an infrequent complication of percutaneous coronary intervention (PCI), and it may lead to catastrophic clinical outcomes. However, reports of cardiac adverse events associated with LSD are rare.

Case presentation

A 55-year-old man with chest pain was treated for a severe left anterior descending branch (LAD)-diagonal 1 (D1) bifurcation lesion by PCI with two stents in the proximal LAD. LSD occurred during the withdrawal of the trapped D1 wire. High-pressure balloon dilatation was performed in the deformed stent, and the end-angiographic appearance was acceptable, but no additional corrective measures were implemented. Ten months later, the patient represented with acute coronary syndrome. Severe in-stent restenosis (ISR) had suboccluded the proximal LAD, and optical coherence tomography (OCT) visualized multilayered stent struts protruding into the lumen at the compressed segment of the stent. Following complete apposition with balloon dilation, a drug-coated balloon (DCB) was used to avoid an additional permanent metallic layer. He remained angina free, and the angiographic result demonstrated no residual stenosis at the six-month follow-up. To our knowledge, this case demonstrates the first report of ISR triggered by LSD in patients treated with DCBs under the guidance of OCT.

Conclusions

The current report underscores the importance of awareness of LSD, and OCT seems to be the preferred modality to confirm complete apposition. If left without performing additional corrective measures, LSD may be associated with a risk of ISR. Complete apposition with balloon dilation followed by a DCB is a feasible treatment option.
Hinweise

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Abkürzungen
D1
Diagonal-1
DCB
Drug-coated balloon
DES
Drug-eluting stents
ISR
In-stent restenosis
ISR
In-stent restenosis
IVUS
Intravascular ultrasound
LAD
Left anterior descending
LSD
Longitudinal stent deformation
NC
Noncompliant
OCT
Optical coherence tomography
PCI
Percutaneous coronary intervention
ST
Stent thrombosis

Background

Longitudinal stent deformation (LSD) has gained more attention recently. To date, the clinical implications of LSD have been uncertain [13]. Theoretically, LSD could result in metal overload, malapposition, incomplete plaque coverage and reduced drug delivery, which may lead to a higher risk of stent thrombosis and in-stent restenosis (ISR) [2, 4]. However, reports of adverse events associated with LSD are rare. Here, we present a case of ISR triggered by LSD to raise awareness of this complication. In addition, we also present a treatment using a drug-coated balloon (DCB) under the guidance of optical coherence tomography (OCT).

Case presentation

A 55-year-old man with a 3-year history of chest pain but no history of myocardial infarction or any intervention was admitted to our hospital, after experiencing exacerbation of pain for 5 days. The electrocardiogram (ECG) showed ST-segment depression in lead V2-V5 < 0.1 mV and transthoracic echocardiogram (TTE) revealed left anterior wall hypokinesis. The physical exam and the laboratory findings including cardiac troponin I were unremarkable. The selective coronary angiography revealed a severe bifurcation lesion of Medina type 1,1,1 involving the left anterior descending branch (LAD) and diagonal 1 (D1) (Fig. 1a). Cardiac catheterization was then performed via the right radial artery using a 6 Fr EBU 3.5 guiding catheter (Medtronic, USA). Both LAD and D1 lesions were individually crossed with Balance Middleweight guidewires (Abbott, USA) and sequentially predilated with 2.0 × 15 mm semicompliant balloons (Boston Scientific, USA). Two Element stents (2.75 × 24 mm distally and 3.0 × 16 mm proximally, Boston Scientific, USA) were deployed in the proximal LAD with the crossover technique (Fig. 1b). Both stents were postdilated at high pressure with a 3.0 × 15 mm noncompliant (NC) balloon (Abbott, USA). Despite careful withdrawal of the trapped D1 wire, a deep engagement of the guiding catheter occurred. It was then noted that the proximal edge of the stent exhibited features of longitudinal compression (Fig. 1c and d). Following postdilatation in the deformed stent using a 3.0 × 15 mm TREK NC balloon (Abbott, USA, Fig. 1e), the end-angiographic appearance was acceptable (Fig. 1f). The patient was treated with 75 mg of clopidogrel for 1.5 years and 100 mg of aspirin lifelong.
Ten months later, the patient represented with acute coronary syndrome. The ECG showed ST-segment depression in lead V2-V5 = 0.2 mV and the cardiac troponin I was 0.23 ng/ml. The urgent coronary angiography showed a suboccluded proximal LAD due to the severe ISR (Fig. 2a). Following gentle wiring and predilatation, optical coherence tomography (OCT, St. Jude Medical, USA) showed that the proximal segment of the stent was longitudinally compressed and that there were multilayered stent struts protruding into the lumen (Fig. 2b). The joint decision of our cardiology team was in favor of a drug-coated balloon. Thus, the postdilatation was performed with a 3.0 × 12 mm NC balloon (Yinyi, China), followed by a 3.0 × 26 mm SeQuent Please paclitaxel-eluting balloon (B. Braun Melsungen, Germany) at 16 atm with sustained insufflation for 30 s (Fig. 3a). With the excellent angiographic result (Fig. 3c), subsequent OCT imaging showed the improved apposition of the compressed segment and the reduced strut accumulation (Fig. 3b). The procedure was uneventful, and the patient was discharged after a few days. At the six-month follow-up, he remained angina free, and the angiographic results demonstrated no residual stenosis.

Discussion and conclusion

Although longitudinal stent deformation is still a rare phenomenon, some studies have emphasized the increased rate of LSD in the last few years [2, 4]. The reported cases include all kinds of stent designs, but the Promus Element stent has a relatively high occurrence. Its design with thin struts and fewer connectors are believed to be prone to LSD. On the other hand, its relatively high radio-opacity also provides easier recognition of LSD [1, 5, 6]. The Promus Element stent was also used in the case reported here. The lesion and procedural characteristics are also considered to play a role in the etiology of LSD. Almost all reports cases generally stated that LSD was more frequently observed in calcified lesions, tortuous lesions, ostial and bifurcation lesions. Moreover, its occurrence was also related to some procedural characteristics, such as aggressive guide catheter manipulation, multiple balloons, bifurcation stent techniques, rotational atherectomy, and OCT techniques [2, 79]. In a recent study composed of 750 lesions in 5871 patients, Rhee et al. [1] reported that provisional side-branch stenting or ballooning, additional downstream intervention and IVUS use were independently associated with LSD. In our case, deformation was due to guided catheter compression of the ostial segment of the proximal stent after withdrawing the jailed wire in the D1 branch, and post dilatation before removing the jaled wire might make it worse. Therefore, aggressive catheter manipulation for proximal lesions should be avoided to prevent LSD.
To date, the long-term clinical outcomes due to LSD have been uncertain because of the rarity and variability of LSD. Although some recent studies have shown that LSD has no effect on major adverse cardiac events [8, 10], most previous studies and cases have reported that the identification of LSD is important because it may be associated with mechanical and clinical risks, including underexpansion, malapposition, stent thrombosis (ST) and in-stent restenosis (ISR) [3, 4, 11, 12]. Although rare, ST is a severe complication with a mortality rate of up to 45% and a high relapse rate [13]. Mamas et al. [4] reported that even with successful identification and treatment, LSD might lead to a risk of subsequent stent thrombosis. Similar to that case, LSD was identified and treated appropriately at the first percutaneous coronary intervention (PCI) in our case, but ISR occurred at the LSD segment 10 months later. An explanation for this phenomenon might be that although the apposition was rendered with balloon postdilatation and the angiographic appearance was acceptable at the end of the first PCI, incomplete plaque coverage or stent malapposition, which was difficult to detect by sole angiographic evaluation, could have led to the subsequent ISR. Therefore, when LSD is suspected, IVUS or OCT should be used to confirm full lesion coverage and stent apposition to minimize future cardiac adverse events [14]. Owing to its high resolution, OCT may be better at detailed images of the strut distribution and the lumen interface.
Treating patients with ISR is still challenging. The current guidelines on myocardial revascularization recommend restenting with drug-eluting stents (DESs) and the use of DCBs [15]. In our case, the overlap of multiple stent layers was detected by OCT, so restenting with DES might raise concerns about the addition of a permanent metallic layer. The multiple layers may promote further endothelial growth associated with mechanical complications, including stent thrombosis and recurrent ISR [16]. Following complete apposition with balloon dilation, a DCB was used to avoid the additional permanent metallic layer. To our knowledge, this case demonstrates the first report of ISR triggered by LSD in patients treated with DCBs.
LSD should be treated at the time of the initial stent implantation procedure, and high-resolution OCT seems the preferred modality to confirm the complete apposition. If left without performing additional corrective measures, LSD may be associated with a risk of ISR. Despite second-generation DESs and DCBs, both are currently recommended for ISR treatment, but DCBs might be a better option for ISR related to LSD.

Acknowledgements

Not applicable.
Not applicable.
Written informed consent for publication of the clinical details and images was obtained from the patient. A copy of the consent form is available for review by the Editor of this journal.

Competing interests

The authors declare that they have no competing interests.
Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://​creativecommons.​org/​licenses/​by/​4.​0/​), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://​creativecommons.​org/​publicdomain/​zero/​1.​0/​) applies to the data made available in this article, unless otherwise stated.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Literatur
1.
Zurück zum Zitat Rhee TM, Park KW, Lee JM, Lee MS, Jeon KH, Kang HJ, Koo BK, Rhew JY, Cha KS, Bae JH, et al. Predictors and long-term clinical outcome of longitudinal stent deformation: insights from pooled analysis of Korean Multicenter Drug-Eluting Stent Cohort. Circ Cardiovasc Interv. 2017;10(11):e005518.CrossRef Rhee TM, Park KW, Lee JM, Lee MS, Jeon KH, Kang HJ, Koo BK, Rhew JY, Cha KS, Bae JH, et al. Predictors and long-term clinical outcome of longitudinal stent deformation: insights from pooled analysis of Korean Multicenter Drug-Eluting Stent Cohort. Circ Cardiovasc Interv. 2017;10(11):e005518.CrossRef
2.
Zurück zum Zitat Guler A, Guler Y, Acar E, Aung SM, Efe SC, Kilicgedik A, Karabay CY, Barutcu S, Tigen MK, Pala S, et al. Clinical, angiographic and procedural characteristics of longitudinal stent deformation. Int J Cardiovasc Imaging. 2016;32(8):1163–70.CrossRef Guler A, Guler Y, Acar E, Aung SM, Efe SC, Kilicgedik A, Karabay CY, Barutcu S, Tigen MK, Pala S, et al. Clinical, angiographic and procedural characteristics of longitudinal stent deformation. Int J Cardiovasc Imaging. 2016;32(8):1163–70.CrossRef
3.
Zurück zum Zitat Williams PD, Mamas MA, Morgan KP, El-Omar M, Clarke B, Bainbridge A, Fath-Ordoubadi F, Fraser DG. Longitudinal stent deformation: a retrospective analysis of frequency and mechanisms. EuroIntervention. 2012;8(2):267–74.CrossRef Williams PD, Mamas MA, Morgan KP, El-Omar M, Clarke B, Bainbridge A, Fath-Ordoubadi F, Fraser DG. Longitudinal stent deformation: a retrospective analysis of frequency and mechanisms. EuroIntervention. 2012;8(2):267–74.CrossRef
4.
Zurück zum Zitat Mamas MA, Williams PD. Longitudinal stent deformation: insights on mechanisms, treatments and outcomes from the Food and Drug Administration manufacturer and user facility device experience database. EuroIntervention. 2012;8(2):196–204.CrossRef Mamas MA, Williams PD. Longitudinal stent deformation: insights on mechanisms, treatments and outcomes from the Food and Drug Administration manufacturer and user facility device experience database. EuroIntervention. 2012;8(2):196–204.CrossRef
5.
Zurück zum Zitat Ormiston JA, Webber B, Webster MW. Stent longitudinal integrity bench insights into a clinical problem. JACC Cardiovasc Interv. 2011;4(12):1310–7.CrossRef Ormiston JA, Webber B, Webster MW. Stent longitudinal integrity bench insights into a clinical problem. JACC Cardiovasc Interv. 2011;4(12):1310–7.CrossRef
6.
Zurück zum Zitat Shannon J, Latib A, Takagi K, Chieffo A, Figini F, Sacco F, Ferrarello S, Montorfano M. Colombo a: “procedural trauma risks longitudinal shortening of the Promus Element stent platform”. Catheter Cardiovasc Interv. 2013;81(5):810–7.CrossRef Shannon J, Latib A, Takagi K, Chieffo A, Figini F, Sacco F, Ferrarello S, Montorfano M. Colombo a: “procedural trauma risks longitudinal shortening of the Promus Element stent platform”. Catheter Cardiovasc Interv. 2013;81(5):810–7.CrossRef
7.
Zurück zum Zitat Seth A, Dhall A. Longitudinal stent shortening: the long and short of it!! Catheter Cardiovasc Interv. 2013;81(5):818–9.CrossRef Seth A, Dhall A. Longitudinal stent shortening: the long and short of it!! Catheter Cardiovasc Interv. 2013;81(5):818–9.CrossRef
8.
Zurück zum Zitat Chung MS, Yang DH, Kim YH, Roh JH, Song J, Kang JW, Ahn JM, Park DW, Kang SJ, Lee SW, et al. Stent fracture and longitudinal compression detected on coronary CT angiography in the first- and new-generation drug-eluting stents. Int J Cardiovasc Imaging. 2016;32(4):637–46.CrossRef Chung MS, Yang DH, Kim YH, Roh JH, Song J, Kang JW, Ahn JM, Park DW, Kang SJ, Lee SW, et al. Stent fracture and longitudinal compression detected on coronary CT angiography in the first- and new-generation drug-eluting stents. Int J Cardiovasc Imaging. 2016;32(4):637–46.CrossRef
9.
Zurück zum Zitat Leibundgut G, Gick M, Toma A, Valina C, Loffelhardt N, Buttner HJ, Neumann FJ. Longitudinal compression of the platinum-chromium everolimus-eluting stent during coronary implantation: predisposing mechanical properties, incidence, and predictors in a large patient cohort. Catheter Cardiovasc Interv. 2013;81(5):E206–14.CrossRef Leibundgut G, Gick M, Toma A, Valina C, Loffelhardt N, Buttner HJ, Neumann FJ. Longitudinal compression of the platinum-chromium everolimus-eluting stent during coronary implantation: predisposing mechanical properties, incidence, and predictors in a large patient cohort. Catheter Cardiovasc Interv. 2013;81(5):E206–14.CrossRef
10.
Zurück zum Zitat Sen H, Lam MK, Lowik MM, Danse PW, Jessurun GA, van Houwelingen KG, Anthonio RL, Tjon Joe Gin RM, Hautvast RW, Louwerenburg JH, et al. Clinical events and patient-reported chest pain in all-comers treated with Resolute integrity and Promus Element stents: 2-year follow-up of the DUTCH PEERS (DUrable polymer-based STent CHallenge of Promus ElemEnt versus ReSolute integrity) randomized trial (TWENTE II). JACC Cardiovasc Interv. 2015;8(7):889–99.CrossRef Sen H, Lam MK, Lowik MM, Danse PW, Jessurun GA, van Houwelingen KG, Anthonio RL, Tjon Joe Gin RM, Hautvast RW, Louwerenburg JH, et al. Clinical events and patient-reported chest pain in all-comers treated with Resolute integrity and Promus Element stents: 2-year follow-up of the DUTCH PEERS (DUrable polymer-based STent CHallenge of Promus ElemEnt versus ReSolute integrity) randomized trial (TWENTE II). JACC Cardiovasc Interv. 2015;8(7):889–99.CrossRef
11.
Zurück zum Zitat Buysschaert I, Ughi GJ, Adriaenssens T, Vermeersch P. Very late stent thrombosis and longitudinal stent deformation. Acta Cardiol. 2017;72(2):216–7.CrossRef Buysschaert I, Ughi GJ, Adriaenssens T, Vermeersch P. Very late stent thrombosis and longitudinal stent deformation. Acta Cardiol. 2017;72(2):216–7.CrossRef
12.
Zurück zum Zitat Polimeni A, Weissner M, Schochlow K, Ullrich H, Indolfi C, Dijkstra J, Anadol R, Munzel T, Gori T. Incidence, clinical presentation, and predictors of clinical restenosis in coronary Bioresorbable scaffolds. JACC Cardiovasc Interv. 2017;10(18):1819–27.CrossRef Polimeni A, Weissner M, Schochlow K, Ullrich H, Indolfi C, Dijkstra J, Anadol R, Munzel T, Gori T. Incidence, clinical presentation, and predictors of clinical restenosis in coronary Bioresorbable scaffolds. JACC Cardiovasc Interv. 2017;10(18):1819–27.CrossRef
13.
Zurück zum Zitat Gori T, Polimeni A, Indolfi C, Raber L, Adriaenssens T, Munzel T. Predictors of stent thrombosis and their implications for clinical practice. Nat Rev Cardiol. 2018;16:243–56.CrossRef Gori T, Polimeni A, Indolfi C, Raber L, Adriaenssens T, Munzel T. Predictors of stent thrombosis and their implications for clinical practice. Nat Rev Cardiol. 2018;16:243–56.CrossRef
14.
Zurück zum Zitat Si D, Yang H, Liu G, Tong Y, He Y. Treatment of longitudinal stent compression under intravenous ultrasound guidance: a case report. Medicine. 2017;96(51):e9405.CrossRef Si D, Yang H, Liu G, Tong Y, He Y. Treatment of longitudinal stent compression under intravenous ultrasound guidance: a case report. Medicine. 2017;96(51):e9405.CrossRef
15.
Zurück zum Zitat Authors/Task Force M, Windecker S, Kolh P, Alfonso F, Collet JP, Cremer J, Falk V, Filippatos G, Hamm C, Head SJ, et al. 2014 ESC/EACTS guidelines on myocardial revascularization: The task force on Myocardial Revascularization of the European Society of Cardiology (ESC) and the European Association for Cardio-Thoracic Surgery (EACTS) Developed with the special contribution of the European Association of Percutaneous Cardiovascular Interventions (EAPCI). Eur Heart J. 2014;35(37):2541–619.CrossRef Authors/Task Force M, Windecker S, Kolh P, Alfonso F, Collet JP, Cremer J, Falk V, Filippatos G, Hamm C, Head SJ, et al. 2014 ESC/EACTS guidelines on myocardial revascularization: The task force on Myocardial Revascularization of the European Society of Cardiology (ESC) and the European Association for Cardio-Thoracic Surgery (EACTS) Developed with the special contribution of the European Association of Percutaneous Cardiovascular Interventions (EAPCI). Eur Heart J. 2014;35(37):2541–619.CrossRef
16.
Zurück zum Zitat Waksman R, Steinvil A. In-Stent Restenosis? The Raiders of the Magic Remedy. Circ Cardiovasc Interv. 2016;9(7):e004150.CrossRef Waksman R, Steinvil A. In-Stent Restenosis? The Raiders of the Magic Remedy. Circ Cardiovasc Interv. 2016;9(7):e004150.CrossRef
Metadaten
Titel
In-stent restenosis and longitudinal stent deformation: a case report
verfasst von
Daoyuan Si
Yaliang Tong
Bo Yu
Yuquan He
Guohui Liu
Publikationsdatum
01.12.2020
Verlag
BioMed Central
Erschienen in
BMC Cardiovascular Disorders / Ausgabe 1/2020
Elektronische ISSN: 1471-2261
DOI
https://doi.org/10.1186/s12872-020-01335-1

Weitere Artikel der Ausgabe 1/2020

BMC Cardiovascular Disorders 1/2020 Zur Ausgabe

Update Kardiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.