In vitro surface characteristics and impurity analysis of five different commercially available dental zirconia implants

verfasst von: B. Beger, H. Goetz, M. Morlock, E. Schiegnitz, B. Al-Nawas

Erschienen in: International Journal of Implant Dentistry | Ausgabe 1/2018

Abstract

Background

The aim of this study was to assess surface characteristics, element composition, and surface roughness of five different commercially available dental zirconia implants.
Five zirconia implants (Bredent whiteSKY™ (I1), Straumann® PURE Ceramic (I2), ceramic.implant vitaclinical (I3), Zeramex® (I4), Ceralog Monobloc M10 (I5)) were evaluated.

Methods

The evaluation was performed by means of scanning electron microscopy (SEM), energy-dispersive X-ray spectroscopy (EDX), and confocal laser scanning microscopy (CLSM).

Results

The semi-quantitative element composition showed no significant impurity of any implant tested. Both the machined and the rough areas of the investigated implants were predominated by zirconium, oxygen, and carbon. Roughness values (Sa) showed highest values for I2 and I5.

Conclusions

The investigated zirconia implants showed surface characteristics and roughness values close to those of conventionally produced titanium implants, making them a promising alternative. However, zirconia implants have yet to prove themselves in clinical practice and clinical controlled trials.
Abkürzungen
AL2O3
Aluminum oxide
CIM
Ceramic injection molding
CLSM
Confocal laser scanning microscopy
EDX
Energy-dispersive X-ray spectroscopy
HIP
Hot isostatic pressing
I
Implant
kV
Kilovolt
mbar
Millibar
MPa
Megapascal
nm
Nanometer
S a
Area roughness parameter
SEM
Scanning electron microscopy
SLA™
Sandblasted, Large-grit, Acid-etched
Y-TZP
Yttrium-stabilized tetragonal zirconium polycrystalline
μm
Micrometer

Background

Dental implants have become a well-established treatment method for oral rehabilitation after tooth loss. Pure titanium is still the material of choice when it comes to dental intraosseous implants and has been used for decades. However, titanium implants have esthetic limitations, especially in the front aspect of the maxillary jaw. The recession of the gingiva can lead to visible implant necks. Furthermore, titanium may cause immunological reactions with early local infection and possible risk for implant loss [1]. Ceramic implants are proclaimed as a new alternative to titanium implants. The first tooth-colored ceramic implants were inferior to titanium-based implants due to their biomechanical characteristics such as low fracture toughness [2]. In the 1980s, the Tübinger immediate implant was introduced, fully made of aluminum oxide (AL2O3), but was withdrawn from the market because of high fracture rates [3]. Other investigations on different AL2O3 implants found less bone-implant contact compared to titanium [4] as well as reduced survival rates [2, 5]. Since the introduction of yttrium-stabilized tetragonal zirconia polycrystalline (Y-TZP)-based implants, it could be shown that these implants show high similarity in osseointegration compared to titanium implants [2].
Titanium implants with smooth or roughened surfaces have shown high success rates in various indications [2, 6, 7]. Surface characteristics of dental implants, as a new development over the last decades, are seen as an important factor that affects osseointegration, especially in compromised patients (e.g., following radiation therapy, bone augmentation, class D4 bone) [8]. By improving the implant design, implant material, and implant surface characteristics as well as surgical techniques and implant loading conditions, osseointegration can be affected [9]. Several new techniques are performed nowadays to speed up the osseointegration process by altering the surface of the implant chemically (incorporating inorganic phases onto the titanium oxide layer) or physically (increasing the level of roughness) [10, 11]. Advantages of surface-modified implants include (a) establishing a greater contact area followed by better primary stability, (b) providing surface-retaining blood clots, and (c) stimulating bone formation [10, 12]. In vitro tests of surface roughness showed higher proliferation, cytokine, and growth factor production of osteoblast-like cells. Those factors are known to affect proliferation, differentiation, and matrix synthesis of chondrocytes [1316]. Many studies on surface characteristics of titanium implants were performed over the last years. Due to the renaissance and new development of zirconia implants, it is now necessary to study their behavior and surface characteristics and to compare them to titanium implants. However, data regarding the surface characteristics of these zirconia implants are very rare. Therefore, the aim of this study was to examine the surface characteristics, element composition, and surface roughness of the five different commercially available dental zirconia implants.

Methods

Investigated implants

The following five commercially available dental zirconia implants were used in this study (Table 1). Bredent whiteSKY™ implant (I1) is made from unground Brezirkon™, an yttrium oxide (Y2O3)-stabilized tetragonal polycrystalline zirconium oxide and is sandblasted. Zirconium oxide is endowed with 3 mol% yttrium oxide to gain a rectangle and room temperature stable structure [17]. Straumann® PURE Ceramic Implant (I2) is generally made from yttrium oxide-stabilized tetragonal polycrystalline zirconium oxide. The surface due to the manufacturer is coated with a special process called ZLA™ which shall be similar to the SLA™ process (Sandblasted, Large-grit, Acid-etched) of titanium implants. Ceramic.implant vitaclinical (I3) is made from zirconium oxide. The Zeramex® implant (I4) is made from zirconium and has a sandblasted and etched surface structure with their so-called ZERAFIL™ technology. Camlog’s Ceralog Monobloc M10 ceramic implant (I5) is also made from yttrium-stabilized zirconium dioxide. Unlike the other ceramic implants, it is produced with ceramic injection molding (CIM) technique. This technique requires no sandblasting or etching. The implants’ geometrical design and the surface structure are already molded via CIM before the sintering and hot isostatic pressing (HIP) process.
Table 1
Five commercially available ceramic implants and surface characteristics
https://static-content.springer.com/image/art%3A10.1186%2Fs40729-018-0124-8/MediaObjects/40729_2018_124_Tab1_HTML.gif
*Due to the processes CIM and HIP, see the “Methods” section

Scanning electron microscopy

For a more detailed illustration of the implant surface topology, the technique of scanning electron microscopy (SEM) was used. A Quanta 200 FEG (FEI Company, Netherlands) field emission SEM equipped with environmental low vacuum mode makes it possible to avoid the typical surface charging-up problems of uncoated highly insulating ceramic implants without the need for sample preparation. Therefore, high-resolution SEM images with magnifications up to 25,000 are possible to demonstrate the micro-structured appearances at different locations. Comparable areas for all implants under investigation are selected by splitting up the cylindrical shape of the implant into sections (Fig. 1). For the comparison of surface structures between the tested implants, two regions of interest were selected: machined and rough area (compare Figs. 1 and 3). Each section was observed under different degrees of magnifications (× 2000, × 10,000, × 25,000) with the same microscope parameters (HV 20 kV, Det LFD, pressure 0.90 mbar). The low vacuum pressure in the sample chamber was reduced until charging levels on the sample surface were reduced to the level at which electron imaging of the sample surface was possible.

Energy-dispersive X-ray spectroscopy

Analysis of the element composition of the implant surfaces by means of energy-dispersive X-ray spectroscopy (EDX) was performed with an INCA Energy 350 system (Oxford Instruments, Wiesbaden, Germany) coupled with the SEM Quanta 200 FEG (Fig. 2). Similar to the micro-morphological presentation, each implant was divided into comparable sites of interest. Typical areas were selected and evaluated (Fig. 3). With the “Point &ID” mode of the INCA Energy software, both points of interest and the areas of interest are selected for the EDX analysis. Microscopic conditions (magnification × 2000) and excitation energy (HV 20 kV) are kept constant for all types of implants. For a semi-quantitative approach, the main components identified on all of the sample surfaces are evaluated as shown in Table 2. Intervals of minimum and maximum values are presented to demonstrate the high inhomogeneous situation found at most of the selected areas.
Table 2
EDX
Element composition/semi-quantitative evaluation
Location
Type
Zr at %min–at %max
Hf at %
Y at %min–at %max
Al at %min–at %max
O at %min–at %max
C at %min–at %max
N at %min–at %max
Machined area
WhiteSKY
16.0–19.5
< 0.25
1.47–1.67
< 0.5
55–58
17.6–24.0
< 1.0
Straumann ZLA
19.4–22.4
< 0.35
1.6–1.8
< 0.12
48.5–52.1
20.6–22.3
5.2–7.2
Vitaclinical
23.7
< 0.30
<  1.5
< 0.13
56
9.8
None
ZERAMEX
17.7
< 0.23
< Det. limit
< 9.6
57.5
7.7
7.7
Monobloc M10
4.0–11.0
< 0.09
< Det. limit
0.4–2.3
12.0–21.0
63.0–80.0
0–11.0
Rough area
WhiteSKY
15.6–19.3
< 0.23
0–2.8
1.1–3.8
49.8–80.7
0–20.7
0–6.3
Straumann ZLA
17.4–28.9
< 0.25
1.7–3.4
< 0.13
48.8–63.7
7.4–15.4
8.2–14.7
Vitaclinical
17.2–23.4
< 0.26
1.3–2.6
< 0.24
48.6–64.5
11.5–18.9
3.8–8.2
ZERAMEX
6.9–18.3
< 0.23
< 1.7
7.8–18.7
67.1–71.5
3.0–6.7
6.1–7.8
Monobloc M10
4.6–28.0
< 0.40
< Det. limit
2.9–13.9
12.0–69.0
28.0–79.0
None

Confocal laser scanning microscopy

Evaluation of the zirconia implant surface roughness as well as their surface texture parameters is carried out by means of confocal laser scanning microscopic technique. A Leica TCS SP2 (Leica Microsystems, Wetzlar, Germany) upright microscope with a red He-laser (633 nm) and a high-performance objective (HC PL FLUOTAR × 50/0.80) was used to acquire high spatial resolution images (1024 × 1024 pixels). Image stacks are created by capturing all the light reflected from the deepest to the highest point of the selected sample surface area. The image stacks are created in defined steps and acquired for five uniformly distributed points at the circumferences of representative-treated and none-treated locations on each type of implant (compare Fig. 4). The step size was calculated for optically optimized values by the LCS Leica confocal software. Because of the cylindrically shaped surface character, a zoom factor of 2 which generates an image size of 150 × 150 μm was used to avoid artificial height values.
Maximum projections and height distribution images (depth map) are calculated by LCS software from the image stacks and viewed exemplary in Fig. 5.
Subsequently, the depth map images are imported in the SPIP™ 4.2.6 (Image Metrology) software for roughness and texture evaluation. According to the ISO 25178-2 reference, all surface roughness parameters implemented in SPIP™ are evaluated and classified as amplitude, hybrid, functional, and spatial parameters. Selected values are shown in Table 3.
Table 3
Roughness analysis
  
Amplitude parameters
Group
Name
Sa (μm)
Machined area
WhiteSKY
0.24 ± 0.04
Straumann ZLA
0.36 ± 0.03
Vitaclinical
0.20 ± 0.06
ZERAMEX
0.30 ± 0.05
Monobloc M10
0.61 ± 0.03
Rough area
WhiteSKY (Impl1)
0.91 ± 0.13
Straumann ZLA (Impl2)
1.27 ± 0.24
Vitaclinical (Impl3)
1.05 ± 0.17
Zeramex (Impl4)
0.73 ± 0.95
Monobloc M10 (impl5)
1.22 ± 0.36

Results

SEM

SEM micrographs presented in Fig. 2 demonstrate the dissimilarity of the sample surface microstructure. Implant 1 shows an overall smoother surface and a slaty-like surface without evidence of a typical etching process. The surface shows sparse roughness. Implants 2–4 show deep markings from their brand’s specific etching and sandblasting processes. In × 10,000 magnification, immersions can be found that look like little craters. Implant 2 shows the biggest immersions, and implant 4 shows the smallest. In a × 25,000 magnification, implants 2–5 show droplet-like-shaped particles on the outer surface as a basic structure of the immersions under × 10,000 magnification. The finest droplets can be found on implant 2, and the biggest droplets can be found on implants 4 and 5. Implant 5 stands out from the other implants. It shows very evenly spread droplets on the surface in every magnification (Fig. 2).

EDX analysis

The semi-quantitative element composition showed no significant impurity of any implant tested (Table 2). Both the machined and the rough areas (Fig. 3) were predominated by zirconium, oxygen, and carbon. Yttrium could be found in implants 1–3. Implants 4 and 5 showed yttrium under the detection limit and just less than 1.7 atomic % in the apical aspect of implant 4. Minor traces of hafnium could be shown in all implants 1–5. Implants 1, 4, and 5 showed traces of aluminum on the surface. The highest amount of aluminum could be found on the surface of implant 4.

Confocal laser scanning microscopy (CLSM)

CLSM images including the topological information of all five implants are shown in Fig. 4.
Untreated areas (machined areas) of implants 1–4 showed parallel grooves of the machining process in the interface area of the neck (Fig. 4). Treated areas (rough areas) show roughened surfaces due to special treatment with acid and sandblasting. Implant 5 showed roughened surface in both areas and no sign for a machined neck part.

Roughness analysis (SPIP)

Implant 2 (Sa 1.27 μm ± 0.24) and implant 5 (Sa 1.22 μm ± 0.36) show the highest roughness values (Sa) of all tested implants: Straumann’s pure ceramic implant was blasted and etched and shows the overall highest Sa value in the rough area. Implant 3 (vitaclinical) shows correspondingly lower Sa around 1.05 μm (± 0.17) (Table 3). The lowest Sa value could be found in implant 4, which was only sandblasted due to manufacturer’s specifications. However, the Zeramex implant despite being sandblasted and etched shows the lowest roughness value around 0.73 μm (± 0.95). Nevertheless, Zeramex shows a fine distribution of small pores all over the surface in the SEM sample images. Camlog’s Ceralog shows the highest roughness in the untreated area with 0.61 μm (± 0.03). Figure 6 shows the box plot of the roughness analysis with implant 5 having the widest distribution of measured values.

Discussion

Implant surface characteristics are of ongoing scientific interest. Implants made from titanium are still the most common to be used. Titanium implants are made from alpha-beta alloy which consists of 6% aluminum and 4% vanadium (Ti-6Al-4V). These materials have low density, high strength, and resistance to fatigue and corrosion, and their modulus of elasticity is closer to the bone than any other implant material [18]. However, titanium implants are discussed to trigger hypersensitivity reactions due to surface corrosion [1, 19]. Titanium implant surfaces are machined, etched, sandblasted, and sometimes coated with special (company-specific) coatings. For titanium implants, roughness values (Ra) around 1.5 μm are known to provide successful osseointegration [20].
Ceramic implants experienced a renaissance since their reentry into the market. New ceramic implants with yttria (Y2O3)-stabilized tetragonal zirconium polycrystalline (Y-TZP) material have superior corrosion and wear resistance in comparison to titanium implants as well as high flexural strength (800 to 1000 MPa) [18]. However, due to manufacturing imperfections or flaws created during zirconia implant fabrication and because of special surface treatments, their strength can be compromised [18, 21]. Due to their brittle nature, ceramic implants tend to fracture. Especially sharp, deep, and thin threads can easily lead to implant failures [18, 21]. The surface treatment on ceramic is developed due to a process of sandblasting, etching, and heat treatment [22]. Sandblasting is usually done with alumina particles that lead to sharp edges and scratches on the surface. The treatment with hydrofluoric acid as the following procedure may smoothen the surface again [2224]. However, in zirconia implants, due to stress caused by sandblasting, a tetragonal to monoclinic phase transition may be caused [22, 25]. This monoclinic volume fraction can be seen in 10–15% of the cases [26] and initially leads to a surface compression of the zirconia material [22]. According to Fischer et al., the long-term effects and the implant stability after this procedure are not yet proven [22]. However, it can be reversed by a thermal treatment that is higher than the transition temperature [22, 27].
The surface shape (droplet-like surface), which was observed in the SEM samples, can be caused due to the sintering process in which ceramic powder was melted and then formed. Different particle, immersion, and droplet sizes can also change due to possible reasons like usage of various types and dosages of acid for the etching process and change of exposure time to acid effect. A longer exposure time to etching process could also be responsible for lowering aluminum corundum from sandblasting processes. However, despite a very fine surface microstructure, implant 4 shows the highest amount of aluminum on the outer surface. This could be explained by sandblasting with aluminum-containing corundum particles followed by a shorter etching process. The higher amount of aluminum in implants 1, 4, and 5 might be due to the individual material composition while sintering the material mixture or to corundum particles of the machining and sandblasting process. Implants with aluminum under the detection limit could be caused by a final etching process. Implants 1 and 5 are not advertised with a special etching process. However, implant 4 is supposed to be etched. The etching could have happened before sandblasting, or the acid used was not strong enough to eliminate all aluminum particles.
All implants excluding the Ceralog Monobloc (implant 5) show typical parallel grooves of the machining process in the confocal laser scan and rougher surfaces in the treated areas. Ceralog is the only implant with a rough surface that can also be found in the machined area. Zirconia implants which are treated with a process of sandblasting, etching, and heat treatment are showing a micro-structured surface resulting in a surface roughness in the range of 1.2 μm [22]. In this study, implants 2 and 5 showed roughness values in the range of 1.2 μm. The other implants showed different roughness values. The surface porosity of titanium implants after sandblasting and etching processes is much more rigorous than that of the ceramic implants that were investigated. In this study, implants 2 and 5 can approximately be compared to titanium surface characteristics in the SEM samples. However, implant 5 was not sandblasted and etched because of a special “injection molding technique” and shows a wide distribution of roughness values. A similarity to the surface structure of titanium implants cannot be proven yet.
The semi-quantitative energy-dispersive X-ray spectroscopy (EDX) can be used to further analyze the components of the implant surface. None of the implants showed any impurity or unexpected results. Implants 4 and 5 showed yttrium under the detection limit in the EDX analysis. This could be caused by the lower dosage of yttrium endowment in the stabilization processing in comparison to other implants [17].
This investigation shows results on a sample basis with one implant tested and shall not be used for generalization.

Conclusions

New ceramic implants are showing a variety of surface characteristics due to different manufacturing processes as shown by other groups [2, 28]. The surface structures of the investigated implants are close to titanium implants. If the surface characteristics really have a high influence on osseointegration, ceramic implants cannot yet compare to the long experience with titanium. However, there are several indications for using ceramic implants. In the future, ceramic implants have to prove themselves in clinical practice and clinical controlled trials.
Not applicable.
Not applicable.

Competing interests

Beger B, Goetz H, Morlock M, Schiegnitz E, and Al-Nawas B declare that they have no competing interests.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://​creativecommons.​org/​licenses/​by/​4.​0/​), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.
Literatur
1.
Zurück zum Zitat Jacobi-Gresser E, Huesker K, Schutt S. Genetic and immunological markers predict titanium implant failure: a retrospective study. Int J Oral Maxillofac Surg. 2013;42(4):537–43.CrossRefPubMed Jacobi-Gresser E, Huesker K, Schutt S. Genetic and immunological markers predict titanium implant failure: a retrospective study. Int J Oral Maxillofac Surg. 2013;42(4):537–43.CrossRefPubMed
2.
Zurück zum Zitat Wenz HJ, Bartsch J, Wolfart S, Kern M. Osseointegration and clinical success of zirconia dental implants: a systematic review. Int J Prosthodont. 2008;21(1):27–36.PubMed Wenz HJ, Bartsch J, Wolfart S, Kern M. Osseointegration and clinical success of zirconia dental implants: a systematic review. Int J Prosthodont. 2008;21(1):27–36.PubMed
3.
Zurück zum Zitat Shulte W. The intra-osseous Al2O3 (Frialit) Tuebingen implant. Developmental status after eight years (I). Quintessence Int. 1984;15(1):9. Shulte W. The intra-osseous Al2O3 (Frialit) Tuebingen implant. Developmental status after eight years (I). Quintessence Int. 1984;15(1):9.
4.
Zurück zum Zitat Steflik DE, Lake FT, Sisk AL, Parr GR, Hanes PJ, Davis HC, et al. A comparative investigation in dogs: 2-year morphometric results of the dental implant—bone interface. Int J Oral Maxillofac Implants. 1996;11(1):15–25.PubMed Steflik DE, Lake FT, Sisk AL, Parr GR, Hanes PJ, Davis HC, et al. A comparative investigation in dogs: 2-year morphometric results of the dental implant—bone interface. Int J Oral Maxillofac Implants. 1996;11(1):15–25.PubMed
5.
Zurück zum Zitat Berge TI, Gronningsaeter AG. Survival of single crystal sapphire implants supporting mandibular overdentures. Clin Oral Implants Res. 2000;11(2):154–62.CrossRefPubMed Berge TI, Gronningsaeter AG. Survival of single crystal sapphire implants supporting mandibular overdentures. Clin Oral Implants Res. 2000;11(2):154–62.CrossRefPubMed
6.
Zurück zum Zitat Branemark PI, Adell R, Breine U, Hansson BO, Lindstrom J, Ohlsson A. Intra-osseous anchorage of dental prostheses. I. Experimental studies. Scand J Plast Reconstr Surg. 1969;3(2):81–100.CrossRefPubMed Branemark PI, Adell R, Breine U, Hansson BO, Lindstrom J, Ohlsson A. Intra-osseous anchorage of dental prostheses. I. Experimental studies. Scand J Plast Reconstr Surg. 1969;3(2):81–100.CrossRefPubMed
7.
Zurück zum Zitat Adell R, Eriksson B, Lekholm U, Branemark PI, Jemt T. Long-term follow-up study of osseointegrated implants in the treatment of totally edentulous jaws. Int J Oral Maxillofac Implants. 1990;5(4):347–59.PubMed Adell R, Eriksson B, Lekholm U, Branemark PI, Jemt T. Long-term follow-up study of osseointegrated implants in the treatment of totally edentulous jaws. Int J Oral Maxillofac Implants. 1990;5(4):347–59.PubMed
8.
Zurück zum Zitat Grötz KA, Wahlmann UW, Krummenauer F, Wegener J, Al-Nawas B, Kuffner H-D, et al. Prognosis and factors affecting prognosis for enossal implants in the irradiated jaw. Mund Kiefer Gesichtschir. 1999;3(1):S117–S24.CrossRefPubMed Grötz KA, Wahlmann UW, Krummenauer F, Wegener J, Al-Nawas B, Kuffner H-D, et al. Prognosis and factors affecting prognosis for enossal implants in the irradiated jaw. Mund Kiefer Gesichtschir. 1999;3(1):S117–S24.CrossRefPubMed
9.
Zurück zum Zitat Albrektsson T, Branemark PI, Hansson HA, Lindstrom J. Osseointegrated titanium implants. Requirements for ensuring a long-lasting, direct bone-to-implant anchorage in man. Acta Orthop Scand. 1981;52(2):155–70.CrossRefPubMed Albrektsson T, Branemark PI, Hansson HA, Lindstrom J. Osseointegrated titanium implants. Requirements for ensuring a long-lasting, direct bone-to-implant anchorage in man. Acta Orthop Scand. 1981;52(2):155–70.CrossRefPubMed
10.
Zurück zum Zitat Karamian E, Khandan A, Motamedi MR, Mirmohammadi H. Surface characteristics and bioactivity of a novel natural HA/zircon nanocomposite coated on dental implants. Biomed Res Int. 2014;2014:410627.CrossRefPubMedPubMedCentral Karamian E, Khandan A, Motamedi MR, Mirmohammadi H. Surface characteristics and bioactivity of a novel natural HA/zircon nanocomposite coated on dental implants. Biomed Res Int. 2014;2014:410627.CrossRefPubMedPubMedCentral
11.
Zurück zum Zitat Dohan Ehrenfest DM, Coelho PG, Kang BS, Sul YT, Albrektsson T. Classification of osseointegrated implant surfaces: materials, chemistry and topography. Trends Biotechnol. 2010;28(4):198–206.CrossRefPubMed Dohan Ehrenfest DM, Coelho PG, Kang BS, Sul YT, Albrektsson T. Classification of osseointegrated implant surfaces: materials, chemistry and topography. Trends Biotechnol. 2010;28(4):198–206.CrossRefPubMed
12.
Zurück zum Zitat Wennerberg A, Albrektsson T. Effects of titanium surface topography on bone integration: a systematic review. Clin Oral Implants Res. 2009;20(Suppl 4):172–84.CrossRefPubMed Wennerberg A, Albrektsson T. Effects of titanium surface topography on bone integration: a systematic review. Clin Oral Implants Res. 2009;20(Suppl 4):172–84.CrossRefPubMed
13.
Zurück zum Zitat Kieswetter K, Schwartz Z, Dean DD, Boyan BD. The role of implant surface characteristics in the healing of bone. Crit Rev Oral Biol Med. 1996;7(4):329–45.CrossRefPubMed Kieswetter K, Schwartz Z, Dean DD, Boyan BD. The role of implant surface characteristics in the healing of bone. Crit Rev Oral Biol Med. 1996;7(4):329–45.CrossRefPubMed
14.
Zurück zum Zitat Ong JL, Carnes DL, Cardenas HL, Cavin R. Surface roughness of titanium on bone morphogenetic protein-2 treated osteoblast cells in vitro. Implant Dent. 1997;6(1):19–24.CrossRefPubMed Ong JL, Carnes DL, Cardenas HL, Cavin R. Surface roughness of titanium on bone morphogenetic protein-2 treated osteoblast cells in vitro. Implant Dent. 1997;6(1):19–24.CrossRefPubMed
15.
Zurück zum Zitat Schwartz Z, Kieswetter K, Dean DD, Boyan BD. Underlying mechanisms at the bone-surface interface during regeneration. J Periodontal Res. 1997;32(1 Pt 2):166–71.CrossRefPubMed Schwartz Z, Kieswetter K, Dean DD, Boyan BD. Underlying mechanisms at the bone-surface interface during regeneration. J Periodontal Res. 1997;32(1 Pt 2):166–71.CrossRefPubMed
16.
Zurück zum Zitat Al-Nawas B, Gotz H. Three-dimensional topographic and metrologic evaluation of dental implants by confocal laser scanning microscopy. Clin Implant Dent Relat Res. 2003;5(3):176–83.CrossRefPubMed Al-Nawas B, Gotz H. Three-dimensional topographic and metrologic evaluation of dental implants by confocal laser scanning microscopy. Clin Implant Dent Relat Res. 2003;5(3):176–83.CrossRefPubMed
17.
Zurück zum Zitat Schweiger M. Zirkoniumdioxid–hochfeste und bruchzähe Strukturkeramik. Ästhetische Zahnmedizin. 2004;5:248–57. Schweiger M. Zirkoniumdioxid–hochfeste und bruchzähe Strukturkeramik. Ästhetische Zahnmedizin. 2004;5:248–57.
18.
19.
Zurück zum Zitat Javed F, Al-Hezaimi K, Almas K, Romanos GE. Is titanium sensitivity associated with allergic reactions in patients with dental implants? A systematic review. Clin Implant Dent Relat Res. 2013;15(1):47–52.CrossRefPubMed Javed F, Al-Hezaimi K, Almas K, Romanos GE. Is titanium sensitivity associated with allergic reactions in patients with dental implants? A systematic review. Clin Implant Dent Relat Res. 2013;15(1):47–52.CrossRefPubMed
20.
Zurück zum Zitat Wennerberg A, Hallgren C, Johansson C, Danelli S. A histomorphometric evaluation of screw-shaped implants each prepared with two surface roughnesses. Clin Oral Implants Res. 1998;9(1):11–9.CrossRefPubMed Wennerberg A, Hallgren C, Johansson C, Danelli S. A histomorphometric evaluation of screw-shaped implants each prepared with two surface roughnesses. Clin Oral Implants Res. 1998;9(1):11–9.CrossRefPubMed
21.
Zurück zum Zitat Osman RB, Ma S, Duncan W, De Silva RK, Siddiqi A, Swain MV. Fractured zirconia implants and related implant designs: scanning electron microscopy analysis. Clin Oral Implants Res. 2013;24(5):592–7.CrossRefPubMed Osman RB, Ma S, Duncan W, De Silva RK, Siddiqi A, Swain MV. Fractured zirconia implants and related implant designs: scanning electron microscopy analysis. Clin Oral Implants Res. 2013;24(5):592–7.CrossRefPubMed
22.
Zurück zum Zitat Fischer J, Schott A, Martin S. Surface micro-structuring of zirconia dental implants. Clin Oral Implants Res. 2016;27(2):162–6.CrossRefPubMed Fischer J, Schott A, Martin S. Surface micro-structuring of zirconia dental implants. Clin Oral Implants Res. 2016;27(2):162–6.CrossRefPubMed
23.
Zurück zum Zitat Bachle M, Butz F, Hubner U, Bakalinis E, Kohal RJ. Behavior of CAL72 osteoblast-like cells cultured on zirconia ceramics with different surface topographies. Clin Oral Implants Res. 2007;18(1):53–9.CrossRefPubMed Bachle M, Butz F, Hubner U, Bakalinis E, Kohal RJ. Behavior of CAL72 osteoblast-like cells cultured on zirconia ceramics with different surface topographies. Clin Oral Implants Res. 2007;18(1):53–9.CrossRefPubMed
24.
Zurück zum Zitat Gahlert M, Rohling S, Wieland M, Eichhorn S, Kuchenhoff H, Kniha H. A comparison study of the osseointegration of zirconia and titanium dental implants. A biomechanical evaluation in the maxilla of pigs. Clin Implant Dent Relat Res. 2010;12(4):297–305.CrossRefPubMed Gahlert M, Rohling S, Wieland M, Eichhorn S, Kuchenhoff H, Kniha H. A comparison study of the osseointegration of zirconia and titanium dental implants. A biomechanical evaluation in the maxilla of pigs. Clin Implant Dent Relat Res. 2010;12(4):297–305.CrossRefPubMed
25.
Zurück zum Zitat Chevalier J, Grandjean S, Kuntz M, Pezzotti G. On the kinetics and impact of tetragonal to monoclinic transformation in an alumina/zirconia composite for arthroplasty applications. Biomaterials. 2009;30(29):5279–82.CrossRefPubMed Chevalier J, Grandjean S, Kuntz M, Pezzotti G. On the kinetics and impact of tetragonal to monoclinic transformation in an alumina/zirconia composite for arthroplasty applications. Biomaterials. 2009;30(29):5279–82.CrossRefPubMed
26.
Zurück zum Zitat Guazzato M, Quach L, Albakry M, Swain MV. Influence of surface and heat treatments on the flexural strength of Y-TZP dental ceramic. J Dent. 2005;33(1):9–18.CrossRefPubMed Guazzato M, Quach L, Albakry M, Swain MV. Influence of surface and heat treatments on the flexural strength of Y-TZP dental ceramic. J Dent. 2005;33(1):9–18.CrossRefPubMed
27.
Zurück zum Zitat Papanagiotou HP, Morgano SM, Giordano RA, Pober R. In vitro evaluation of low-temperature aging effects and finishing procedures on the flexural strength and structural stability of Y-TZP dental ceramics. J Prosthet Dent. 2006;96(3):154–64.CrossRefPubMed Papanagiotou HP, Morgano SM, Giordano RA, Pober R. In vitro evaluation of low-temperature aging effects and finishing procedures on the flexural strength and structural stability of Y-TZP dental ceramics. J Prosthet Dent. 2006;96(3):154–64.CrossRefPubMed
28.
Zurück zum Zitat Ewais OH, Al Abbassy F, Ghoneim MM, Aboushelib MN. Novel zirconia surface treatments for enhanced osseointegration: laboratory characterization. Int J Dent. 2014;2014:203940.CrossRefPubMedPubMedCentral Ewais OH, Al Abbassy F, Ghoneim MM, Aboushelib MN. Novel zirconia surface treatments for enhanced osseointegration: laboratory characterization. Int J Dent. 2014;2014:203940.CrossRefPubMedPubMedCentral
Metadaten
Titel
In vitro surface characteristics and impurity analysis of five different commercially available dental zirconia implants
verfasst von
B. Beger
H. Goetz
M. Morlock
E. Schiegnitz
B. Al-Nawas
Publikationsdatum
01.12.2018
Verlag
Springer Berlin Heidelberg
Erschienen in
International Journal of Implant Dentistry / Ausgabe 1/2018
Elektronische ISSN: 2198-4034
DOI
https://doi.org/10.1186/s40729-018-0124-8

Weitere Artikel der Ausgabe 1/2018

International Journal of Implant Dentistry 1/2018 Zur Ausgabe

Newsletter

Bestellen Sie unseren kostenlosen Newsletter Update Zahnmedizin und bleiben Sie gut informiert – ganz bequem per eMail.