In-vivo study of osseointegration in Prestige LP cervical disc prosthesis

verfasst von: Jigang Lou, Beiyu Wang, Tingkui Wu, Wenjie Wu, Huibo Li, Ziyang Liu, Hao Liu

Erschienen in: BMC Musculoskeletal Disorders | Ausgabe 1/2018

Abstract

Background

A study was designed to quantify the extent of porous osseointegration at the prosthesis-bone interface in the Prestige LP prosthesis containing a plasma-sprayed titanium coating.

Methods

Using an anterior surgical approach, cervical disc arthroplasty was performed in 8 mature male goats at the C3-C4 segment, followed by implantation of the Prestige LP prosthesis. The vertebral specimens were examined using microcomputed tomograph for histomorphometric quantification, and proceeded by routine paraffin processing for histological observation. Hence, the porous osseointegration at the prosthesis-bone interface was evaluated based on histologic and histomorphometric analyses.

Results

At 6 months after surgery, there was no evidence of prosthesis migration, loosening, subsidence, or neurologic or vascular complications. Based on gross histologic analysis, there was excellent porous ingrowth at the prosthesis–bone interface, without significant histopathologic changes. Histomorphometric analysis at the prosthesis-bone interface indicated the mean porous ingrowth of 48.5% ± 10.4% and the total ingrowth range of 36.6 to 59.8%.

Conclusions

As the first comprehensive in vivo investigation into the Prestige LP prosthesis, this project established a successful animal model in the evaluation of cervical disc arthroplasty. Moreover, histomorphometric analysis of porous ingrowth at the prosthesis-bone interface was more favorable for cervical disc arthroplasty with the Prestige LP prosthesis compared to historical reports of appendicular total joint arthroplasty.
Hinweise

Electronic supplementary material

The online version of this article (https://​doi.​org/​10.​1186/​s12891-018-1957-2) contains supplementary material, which is available to authorized users.
Abkürzungen
BEI-SEM
Backscattered electron imaging-scanning electron microscopy
CDA
Cervical disc arthroplasty
Micro-CT
Microcomputed tomograph
TiCaPHA
Titanium/calcium phosphate/hydroxyapatite

Background

Cervical disc arthroplasty (CDA) is an exciting new technology to treat symptomatic cervical degenerative disc disease in patients who have failed conservative care. This technology not only allows for the maintenance of normal cervical spinal motion, but also has the potential to prevent or reduce the risk of degeneration at adjacent levels, compared with the conventional fusion techniques [13]. However, the complications associated with CDA prostheses have been reported as heterotopic ossification, prosthesis migration and dislocation, prosthesis subsidence, spontaneous fusion, immune tissue reaction and fracture in the adjacent vertebral body [4, 5]. Among these, the most important complication is the prosthesis migration and dislocation, because it often requires reoperation. In previous studies, the incidence of prosthesis dislocation after CDA has been reported as 0.5%–3.06% [68]. Hence, it is very essential to improve the initial stability and long-term stability of CDA prostheses in order to lower the risk of prosthesis migration and dislocation.
To date, a variety of commercial and experimental disc prostheses with various design concepts are designed to strive for better initial stability and long-term stability in order to obtain long-term survivorship of CDA prostheses. With regard to the current widely used Prestige LP prosthesis (Medtronic Sofamor Danek, Memphis, TN, USA), with two titanium ceramic composite endplates (Fig. 1), its initial device fixation is mainly attributed to the mechanisms of acute fixation achieved via a series of four rails, two on each prosthesis endplate. Besides, its long-term device fixation is acquired by a porous titanium coating on prosthesis endplate. The primary endplate bearing surface contains two layers of pure titanium, with a pore size of 75–300 μm, which serves to increase porosity and surface area and facilitate trabecular ingrowth [9]. The titanium coatings consist of a special adhesive layer (< 90 μm) and a cover layer (< 180 μm). Owing to these features, a porous titanium plasma spray coating on the endplate contacting surfaces facilitates bone ingrowth and promotes the extent of biological osseointegration at the prosthesis-bone interface, ensuring long-term stability. As it is infeasible to quantitatively evaluate the extent of porous osseointegration at the prosthesis-bone interface for patients underwent cervical disc arthroplasty, an animal model is a great alternative.
Serving as the first to demonstrate successful endplate osseointegration in the caprine cervical model, the present study was undertaken to investigate the initial stability and the biologic porous ingrowth characteristics of the Prestige LP artificial cervical disc prosthesis, with success criteria based on radiographic analysis and quantitative histomorphometry.

Methods

Animal research permission

The Institutional Animal Care and Use Committee at the West China Center of Medical Sciences, Sichuan University, Chengdu, Sichuan granted approval for this investigation. Conduct of experimentation on living animals followed the recommendations of the Guide for the Care and Use of Laboratory Animals [10], and under the close supervision of qualified and experienced persons.

Animal model and surgical preparation

Eight mature male goats (2–3 years old, mean weight 30 Kg) from laboratory animal center of Sichuan University were included in this study, and followed for a period of 6 months after surgery. Each animal was sedated with an intravenous injection of anesthetic medications (diazepam 0.2 mg/kg and ketamine HCL 5 mg/kg), followed by endotracheal intubation and general inhalation anesthesia using 1% to 2% isoflurane with continuous intravenous fluids (range 3–6 mL/lb./h) administered for the duration of surgery. In addition, prophylactic intravenous antibiotics (cefazolin sodium, 1 g) and analgesics (butorphanol 0.1 mg/kg) were administered before and after surgery.

Surgical technique and postoperative evaluation

The anterior Smith-Robinson approach to the cervical spine was adapted to the goat model through a right-sided longitudinal incision with length 6- to 8-cm, and a standard anterior cervical discectomy was performed at the C3–C4 intervertebral level. The endplate surfaces were prepared using curettage and a high-speed burr. According to the manufacturer’s recommended tools and procedures, the Prestige LP prosthesis was then implanted at the operative disc level (Fig. 2). Blood loss, operating times, and intra- and perioperative complications were quantified.
Observations of ambulatory activities and wound healing were monitored daily, and all animals received analgesics and prophylactic antibiotics for the first 10 days post-operatively. Lateral X-ray films of the cervical spine were taken intra- and post-operatively to confirm the correct position of all disc prostheses (Fig. 3). Euthanasia was performed for each animal using an overdose (150 mg/kg) of concentrated pentobarbital solution (390 mg/mL) at 6 months after surgery. The spinal column then was carefully dissected, immediately placed in double-wrapped plastic specimen bags and frozen at − 20°C for subsequent radiographic and histological examination.

Histology and histomorphometry

The operative motion segments were examined using microcomputed tomograph (Micro-CT) to histomorphometrically quantify the percentage of trabecular ingrowth at the prosthesis-bone interface. The prosthesis surface was traced manually and expressed as a total endplate area pixel count. The regions of trabecular contact were subsequently traced, quantified in pixels, and expressed as a percentage of the total endplate area (% ingrowth = apparent bone contact area/gross total endplate area). The above quantitative evaluation method has been widely used in several previous studies [1115].
The vertebral specimens were then performed histological evaluation at the Biotechnology Institute Histology Laboratory of Sichuan University. The specimens were fixed in 10% neutral buffered formalin solution, then dehydrated in a series of graded alcohol which was later substituted by dimethylbenzene, and embedded in paraffin. Thereafter, the paraffin embedded sections were cut into 3–5 μm thick using thin-sectioning microtomy, and then treated with two staining techniques: standard Hematoxylin and Eosin and Masson staining.

Statistical analysis

Statistical analysis was performed using SPSS version 19.0 software (SPSS Inc., Chicago, Illinois). Histomorphometric data were presented as the percentage of trabecular bone in contact with the Prestige LP prosthesis (titanium endplates) and statistically compared with historical reports of appendicular total joint arthroplasty using an analysis of variance (ANOVA) with Student–Newman–Keuls test. All data were shown as mean ± standard deviation, and significance was indicated at P < 0.05.

Results

All animals survived the procedures and postoperative time period without incidence of vascular, neurologic or infectious complications. The average operating time required was 72.6 ± 17.5 min (range 58–98 min), with an estimated blood loss of less than 50 ml. All animals had a normal recovery by 1 week after surgery, with clinical assessment indicating normal appearance, ambulation, appetite and wound healing. Based on anteroposterior and lateral plain films, there was no evidence of prosthesis migration, loosening, or subsidence at the prosthesis-bone interface.

Histomorphometry

Micro-CT of the operative motion segments showed excellent osseointegration at the prosthesis–bone interface (Fig. 4a and b). Histomorphometric analysis at the prosthesis-bone interface (apparent bone contact area/gross total endplate area) indicated that the mean porous ingrowth was 48.5% ± 10.4% (total range: 36.6% to 59.8%) at 6 months after surgery, which was higher than that reported for acetabular components, tibial plateaus, and femoral stem components found in the appendicular skeleton (Fig. 5).

Bone histology

As an overall statement, gross histologic analysis of the Prestige LP prosthesis demonstrated excellent ingrowth at the prosthesis–bone interface, without evidence of particulate wear debris or significant histopathologic changes. All the vertebral specimens were fixed and underwent routine paraffin processing and slide preparation. The paraffin embedded sections were cut, slide mounted, and stained using standard Hematoxylin and Eosin and Masson staining. As a result, there were both plenty of proliferated osteoblasts and regenerated osseous tissues in some regions of the prosthesis-bone interface. Moreover, there was a distinct interface between regenerated osseous tissues and mature bone tissues. (Fig. 6a and b).

Discussion

To date, a wide variety of commercial and experimental artificial prosthetic discs are now available on the market or are under clinical trial. Despite with different design concepts, all artificial cervical discs are designed to replace the diseased intervertebral discs to perform their functions. The successful outcome of cervical disc replacement, on one hand, rests upon the mechanisms of acute fixation, which are accomplished by endplate modifications involving the use of keels, teeth, rails, and serrations providing initial device fixation [16], on the other hand, depends on the extent of biological osseointegration at the prosthesis-bone interface, ensuring long-term device fixation. Therefore, as a current widely used artificial disc, it is very essential to evaluate the initial stability and the biological porous ingrowth characteristics of the Prestige LP prosthesis.
In the current study, radiographic analysis showed no evidence of prosthesis migration, loosening, or subsidence. Most importantly and most challenging is for device to encourage osseointegration at the prosthesis-bone interface in order to minimize the incidence of clinical device loosening or migration after CDA. Based on the histomorphometry data, the mean porous ingrowth was 48.5% ± 10.4%. This demonstrated excellent porous osseointegration at the prosthesis–bone interface for the Prestige LP prosthesis, similar to that reported in several previous studies [1315]. Moreover, the mean porous ingrowth was much higher than that reported for porous ingrowth found in appendicular total joint arthroplasty (only 20%–30% ingrowth). Harvey et al. [17] found the values of bone ingrowth to be 9.7 ± 5.38% for a composite stem compared with 28.1 ± 5.31% for a titanium alloy stem in a canine total hip arthroplasty model. Jasty et al. [18] retrieved five porous-coated femoral components from patients underwent revision arthroplasty, and found the ingrowth was 4 to 44% (mean 24%). Sumner et al. [19, 20] investigated the uncemented femoral components at 2 years in a canine total hip arthroplasty model, and found the mean ingrowth to be 32.7 ± 4.7% (range 19.7–47.5%) with fiber metal coatings compared with 24.1 ± 1.8% (range 19.0–31.2%) using bead coatings. Pidhorz et al. [21] retrieved 11 cementless acetabular components at autopsy at an average of 41 months (5 weeks to 75 months) after surgery, with mean ingrowth of 12.1 ± 8.2%. As for the improved extent of porous ingrowth in the Prestige LP prosthesis, one reason may be that the rails provide a better primary press-fit fixation, facing toward the bony endplates to resist pull-out, and the porous titanium plasma spray coating facilitates bone ingrowth and significantly encourages osseointegration at the prosthesis-bone interface in the long run. In addition, we postulate that the more favorable porous ingrowth may be the result of acute ligamentotaxis causing long-term and sustained compression across the prosthesis-bone interface.
There is controversy regarding the most accurate method of measuring the porous ingrowth of cementless prostheses [2224]. The three most widely used methods are microradiography, stained histology, and backscattered electron imaging-scanning electron microscopy (BEI-SEM). In this study, we adopted microradiography and stained histology to evaluate the porous ingrowth at the prosthesis-bone interface. It turned out that there was evidence of excellent osseointegration at the prosthesis–bone interface. Sumner et al. [25] compared the porous ingrowth of acetabular cups using the three methods, finding that BEI-SEM and histologic sections possessed comparable results, whereas microradiography underestimated the porosity of the porous coating by a mean of 17% and simultaneously overestimated the amount of bony ingrowth by a mean of 0.8%.
There are still some aspects need to be improved in the current study. First, the caprine model used only a 6-month follow-up period and a small sample size. It is expected that longer follow-up evaluation with greater numbers of subjects would be required to obtain a more reliable measure of the biological osseointegration of CDA with the Prestige LP prosthesis. In addition, it is a relatively simplified design for the Prestige LP prosthesis with a porous titanium plasma spray coating. A unique bioactive TiCaPHA (titanium/calcium phosphate/hydroxyapatite) coating is more favorable to encourage osseointegration at the prosthesis-bone interface [13, 14]. Moreover, the prosthesis may possess stronger surface and more favorable porous osseointegration through the procedure for an electrochemically bonded coating than a plasma-sprayed coating [12].

Conclusions

In summary, as the first comprehensive in vivo investigation into the Prestige LP prosthesis, the present project established a successful animal model for CDA. Moreover, histomorphometric analysis of porous ingrowth at the prosthesis-bone interface was more favorable for CDA with the Prestige LP prosthesis, compared to historical reports of appendicular total joint arthroplasty.

Acknowledgements

We would like to thank Dr. Li Tian for her highly valuable participation in animal care and operations.

Funding

This study was supported by the foundation of Science & Technology Department of Sichuan Province (grant No. 0040205301C35), PR China.

Availability of data and materials

The dataset supporting the conclusions of this article is included within the (Additional file 1).

Ethics approval

This study was approved by the Institutional Animal Care and Use Committee at the West China Center of Medical Sciences, Sichuan University, Chengdu, Sichuan, China.
Not applicable.

Competing interests

The authors declare that they have no competing interests.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://​creativecommons.​org/​licenses/​by/​4.​0/​), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://​creativecommons.​org/​publicdomain/​zero/​1.​0/​) applies to the data made available in this article, unless otherwise stated.
Literatur
1.
Zurück zum Zitat Sasso RC, Smucker JD, Hacker RJ, Heller JG. Artificial disc versus fusion: a prospective, randomized study with 2-year follow-up on 99 patients. Spine (Phila Pa 1976). 2007;32:2933–40.CrossRef Sasso RC, Smucker JD, Hacker RJ, Heller JG. Artificial disc versus fusion: a prospective, randomized study with 2-year follow-up on 99 patients. Spine (Phila Pa 1976). 2007;32:2933–40.CrossRef
2.
Zurück zum Zitat Phillips FM, Allen TR, Regan JJ, Albert TJ, Cappuccino A, Devine JG, et al. Cervical disc replacement in patients with and without previous adjacent level fusion surgery: a prospective study. Spine (Phila Pa 1976). 2009;34:556–65.CrossRef Phillips FM, Allen TR, Regan JJ, Albert TJ, Cappuccino A, Devine JG, et al. Cervical disc replacement in patients with and without previous adjacent level fusion surgery: a prospective study. Spine (Phila Pa 1976). 2009;34:556–65.CrossRef
3.
Zurück zum Zitat Maldonado CV, Paz RD, Martin CB. Adjacent-level degeneration after cervical disc arthroplasty versus fusion. Eur Spine J. 2011;20(Suppl 3):403–7.CrossRefPubMedPubMedCentral Maldonado CV, Paz RD, Martin CB. Adjacent-level degeneration after cervical disc arthroplasty versus fusion. Eur Spine J. 2011;20(Suppl 3):403–7.CrossRefPubMedPubMedCentral
4.
Zurück zum Zitat Cavanaugh DA, Nunley PD, Kerr EJ 3rd, Werner DJ, Jawahar A. Delayed hyper-reactivity to metal ions after cervical disc arthroplasty: a case report and literature review. Spine (Phila Pa 1976). 2009;34:E262-5.CrossRef Cavanaugh DA, Nunley PD, Kerr EJ 3rd, Werner DJ, Jawahar A. Delayed hyper-reactivity to metal ions after cervical disc arthroplasty: a case report and literature review. Spine (Phila Pa 1976). 2009;34:E262-5.CrossRef
5.
Zurück zum Zitat Thaler M, Hartmann S, Gstöttner M, Lechner R, Gabl M, Bach C. Footprint mismatch in total cervical disc arthroplasty. Eur Spine J. 2013;22:759–65.CrossRefPubMed Thaler M, Hartmann S, Gstöttner M, Lechner R, Gabl M, Bach C. Footprint mismatch in total cervical disc arthroplasty. Eur Spine J. 2013;22:759–65.CrossRefPubMed
6.
Zurück zum Zitat Goffin J, Van Calenbergh F, van Loon J, Casey A, Kehr P, Liebig K, et al. Intermediate follow-up after treatment of degenerative disc disease with the Bryan cervical disc prosthesis: single-level and bi-level. Spine (Phila Pa 1976). 2003;28:2673–8.CrossRef Goffin J, Van Calenbergh F, van Loon J, Casey A, Kehr P, Liebig K, et al. Intermediate follow-up after treatment of degenerative disc disease with the Bryan cervical disc prosthesis: single-level and bi-level. Spine (Phila Pa 1976). 2003;28:2673–8.CrossRef
7.
Zurück zum Zitat Anderson PA, Sasso RC, Riew KD. Comparison of adverse events between the Bryan artificial cervical disc and anterior cervical arthrodesis. Spine (Phila Pa 1976). 2008;33:1305–12.CrossRef Anderson PA, Sasso RC, Riew KD. Comparison of adverse events between the Bryan artificial cervical disc and anterior cervical arthrodesis. Spine (Phila Pa 1976). 2008;33:1305–12.CrossRef
8.
Zurück zum Zitat Ozbek Z, Ozkara E, Arslantas A. Implant migration in cervical disk Arthroplasty. World Neurosurg. 2017;97:390–7.CrossRefPubMed Ozbek Z, Ozkara E, Arslantas A. Implant migration in cervical disk Arthroplasty. World Neurosurg. 2017;97:390–7.CrossRefPubMed
9.
Zurück zum Zitat Rahbek O, Overgaard S, Jensen TB, Bendix K, Søballe K. Sealing effect of hydroxyapatite coating: a 12-month study in canines. Acta Orthop Scand. 2000;71:563–73.CrossRefPubMed Rahbek O, Overgaard S, Jensen TB, Bendix K, Søballe K. Sealing effect of hydroxyapatite coating: a 12-month study in canines. Acta Orthop Scand. 2000;71:563–73.CrossRefPubMed
10.
Zurück zum Zitat Mason TJ, Matthews M. Aquatic environment, housing, and management in the eighth edition of the guide for the care and use of laboratory animals: additional considerations and recommendations. J Am Assoc Lab Anim Sci. 2012;51:329–32.PubMedPubMedCentral Mason TJ, Matthews M. Aquatic environment, housing, and management in the eighth edition of the guide for the care and use of laboratory animals: additional considerations and recommendations. J Am Assoc Lab Anim Sci. 2012;51:329–32.PubMedPubMedCentral
11.
Zurück zum Zitat Cunningham BW, Lowery GL, Serhan HA, Dmitriev AE, Orbegoso CM, McAfee PC, et al. Total disc replacement arthroplasty using the AcroFlex lumbar disc: a non-human primate model. Eur Spine J. 2002;11(Suppl 2):S115–23.PubMedPubMedCentral Cunningham BW, Lowery GL, Serhan HA, Dmitriev AE, Orbegoso CM, McAfee PC, et al. Total disc replacement arthroplasty using the AcroFlex lumbar disc: a non-human primate model. Eur Spine J. 2002;11(Suppl 2):S115–23.PubMedPubMedCentral
12.
Zurück zum Zitat McAfee PC, Cunningham BW, Orbegoso CM, Sefter JC, Dmitriev AE, Fedder IL. Analysis of porous ingrowth in intervertebral disc prostheses: a nonhuman primate model. Spine (Phila Pa 1976). 2003;28:332–40. McAfee PC, Cunningham BW, Orbegoso CM, Sefter JC, Dmitriev AE, Fedder IL. Analysis of porous ingrowth in intervertebral disc prostheses: a nonhuman primate model. Spine (Phila Pa 1976). 2003;28:332–40.
13.
Zurück zum Zitat Hu N, Cunningham BW, McAfee PC, Kim SW, Sefter JC, Cappuccino A, et al. Porous coated motion cervical disc replacement: a biomechanical, histomorphometric, and biologic wear analysis in a caprine model. Spine (Phila Pa 1976). 2006;31:1666–73.CrossRef Hu N, Cunningham BW, McAfee PC, Kim SW, Sefter JC, Cappuccino A, et al. Porous coated motion cervical disc replacement: a biomechanical, histomorphometric, and biologic wear analysis in a caprine model. Spine (Phila Pa 1976). 2006;31:1666–73.CrossRef
14.
Zurück zum Zitat Cunningham BW, Hu N, Zorn CM, McAfee PC. Bioactive titanium calcium phosphate coating for disc arthroplasty: analysis of 58 vertebral end plates after 6- to 12-month implantation. Spine J. 2009;9:836–45.CrossRefPubMed Cunningham BW, Hu N, Zorn CM, McAfee PC. Bioactive titanium calcium phosphate coating for disc arthroplasty: analysis of 58 vertebral end plates after 6- to 12-month implantation. Spine J. 2009;9:836–45.CrossRefPubMed
15.
Zurück zum Zitat Lou J, Wu W, Li H, Wang B, Liu H. Analysis of bony ingrowth in novel cervical disc prosthesis. Int J Clin Exp Med. 2017;10:10196–201. Lou J, Wu W, Li H, Wang B, Liu H. Analysis of bony ingrowth in novel cervical disc prosthesis. Int J Clin Exp Med. 2017;10:10196–201.
16.
Zurück zum Zitat Cunningham BW, Hu N, Zorn CM, McAfee PC. Comparative fixation methods of cervical disc arthroplasty versus conventional methods of anterior cervical arthrodesis: serration, teeth, keels, or screws? J Neurosurg Spine. 2010;12:214–20.CrossRefPubMed Cunningham BW, Hu N, Zorn CM, McAfee PC. Comparative fixation methods of cervical disc arthroplasty versus conventional methods of anterior cervical arthrodesis: serration, teeth, keels, or screws? J Neurosurg Spine. 2010;12:214–20.CrossRefPubMed
17.
Zurück zum Zitat Harvey EJ, Bobyn JD, Tanzer M, Stackpool GJ, Krygier JJ, Hacking SA. Effect of flexibility of the femoral stem on bone-remodeling and fixation of the stem in a canine total hip arthroplasty model without cement. J Bone Joint Surg Am. 1999;81:93–107.CrossRefPubMed Harvey EJ, Bobyn JD, Tanzer M, Stackpool GJ, Krygier JJ, Hacking SA. Effect of flexibility of the femoral stem on bone-remodeling and fixation of the stem in a canine total hip arthroplasty model without cement. J Bone Joint Surg Am. 1999;81:93–107.CrossRefPubMed
18.
Zurück zum Zitat Jasty M, Bragdon CR, Maloney WJ, Haire T, Harris WH. Ingrowth of bone in failed fixation of porous-coated femoral components. J Bone Joint Surg Am. 1991;73:1331–7.CrossRefPubMed Jasty M, Bragdon CR, Maloney WJ, Haire T, Harris WH. Ingrowth of bone in failed fixation of porous-coated femoral components. J Bone Joint Surg Am. 1991;73:1331–7.CrossRefPubMed
19.
Zurück zum Zitat Sumner DR, Turner TM, Urban RM, Galante JO. Remodeling and ingrowth of bone at two years in a canine cementless total hip-arthroplasty model. J Bone Joint Surg Am. 1992;74:239–50.CrossRefPubMed Sumner DR, Turner TM, Urban RM, Galante JO. Remodeling and ingrowth of bone at two years in a canine cementless total hip-arthroplasty model. J Bone Joint Surg Am. 1992;74:239–50.CrossRefPubMed
20.
Zurück zum Zitat Sumner DR, Kienapfel H, Jacobs JJ, Urban RM, Turner TM, Galante JO. Bone ingrowth and wear debris in well-fixed cementless porous-coated tibial components removed from patients. J Arthroplast. 1995;10:157–67.CrossRef Sumner DR, Kienapfel H, Jacobs JJ, Urban RM, Turner TM, Galante JO. Bone ingrowth and wear debris in well-fixed cementless porous-coated tibial components removed from patients. J Arthroplast. 1995;10:157–67.CrossRef
21.
Zurück zum Zitat Pidhorz LE, Urban RM, Jacobs JJ, Sumner DR, Galante JO. A quantitative study of bone and soft tissues in cementless porous-coated acetabular components retrieved at autopsy. J Arthroplast. 1993;8:213–25.CrossRef Pidhorz LE, Urban RM, Jacobs JJ, Sumner DR, Galante JO. A quantitative study of bone and soft tissues in cementless porous-coated acetabular components retrieved at autopsy. J Arthroplast. 1993;8:213–25.CrossRef
22.
Zurück zum Zitat Engh CA, Zettl-Schaffer KF, Kukita Y, Sweet D, Jasty M, Bragdon C. Histological and radiographic assessment of well functioning porous-coated acetabular components. A human postmortem retrieval study. J Bone Joint Surg Am. 1993;75:814–24.CrossRefPubMed Engh CA, Zettl-Schaffer KF, Kukita Y, Sweet D, Jasty M, Bragdon C. Histological and radiographic assessment of well functioning porous-coated acetabular components. A human postmortem retrieval study. J Bone Joint Surg Am. 1993;75:814–24.CrossRefPubMed
23.
Zurück zum Zitat Urban RM, Jacobs JJ, Sumner DR, Peters CL, Voss FR, Galante JO. The bone-implant interface of femoral stems with non-circumferential porous coating. J Bone Joint Surg Am. 1996;78:1068–81.CrossRefPubMed Urban RM, Jacobs JJ, Sumner DR, Peters CL, Voss FR, Galante JO. The bone-implant interface of femoral stems with non-circumferential porous coating. J Bone Joint Surg Am. 1996;78:1068–81.CrossRefPubMed
24.
Zurück zum Zitat Bloebaum RD, Rhodes DM, Rubman MH, Hofmann AA. Bilateral tibial components of different cementless designs and materials. Microradiographic, backscattered imaging, and histologic analysis. Clin Orthop Relat Res. 1991;268:179–87. Bloebaum RD, Rhodes DM, Rubman MH, Hofmann AA. Bilateral tibial components of different cementless designs and materials. Microradiographic, backscattered imaging, and histologic analysis. Clin Orthop Relat Res. 1991;268:179–87.
25.
Zurück zum Zitat Sumner DR, Bryan JM, Urban RM, Kuszak JR. Measuring the volume fraction of bone ingrowth: a comparison of three techniques. J Orthop Res. 1990;8:448–52.CrossRefPubMed Sumner DR, Bryan JM, Urban RM, Kuszak JR. Measuring the volume fraction of bone ingrowth: a comparison of three techniques. J Orthop Res. 1990;8:448–52.CrossRefPubMed
Metadaten
Titel
In-vivo study of osseointegration in Prestige LP cervical disc prosthesis
verfasst von
Jigang Lou
Beiyu Wang
Tingkui Wu
Wenjie Wu
Huibo Li
Ziyang Liu
Hao Liu
Publikationsdatum
01.12.2018
Verlag
BioMed Central
Erschienen in
BMC Musculoskeletal Disorders / Ausgabe 1/2018
Elektronische ISSN: 1471-2474
DOI
https://doi.org/10.1186/s12891-018-1957-2

Weitere Artikel der Ausgabe 1/2018

BMC Musculoskeletal Disorders 1/2018 Zur Ausgabe

Arthropedia

Grundlagenwissen der Arthroskopie und Gelenkchirurgie. Erweitert durch Fallbeispiele, Videos und Abbildungen. 
» Jetzt entdecken

Update Orthopädie und Unfallchirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.