Skip to main content
Erschienen in: Clinical Orthopaedics and Related Research® 8/2016

21.01.2016 | Clinical Research

Inaccuracies in the Use of Magnification Markers in Digital Hip Radiographs

verfasst von: Michael J. Archibeck, MD, Tamara Cummins, RT, Krishna R. Tripuraneni, MD, Joshua T. Carothers, MD, Cristina Murray-Krezan, MS, Mohammad Hattab, PhD, Richard E. White Jr, MD

Erschienen in: Clinical Orthopaedics and Related Research® | Ausgabe 8/2016

Einloggen, um Zugang zu erhalten

Abstract

Background

With the ubiquity of digital radiographs, the use of digital templating for arthroplasty has become commonplace. Although improved accuracy with digital radiographs and magnification markers is assumed, it has not been shown.

Questions/Purposes

We wanted to (1) evaluate the accuracy of magnification markers in estimating the magnification of the true hip and (2) determine if the use of magnification markers improves on older techniques of assuming a magnification of 20% for all patients.

Methods

Between April 2013 and September 2013 we collected 100 AP pelvis radiographs of patients who had a THA prosthesis in situ and a magnification marker placed per the manufacturer’s instructions. Radiographs seen during our standard radiographic review process, which met our inclusion criteria (AP pelvic view that included a well-positioned and observed magnification marker, and a prior total hip replacement with a known femoral head size), were included in the analysis. We then used OrthoViewTM software program to calculate magnification of the radiograph using the magnification marker (measured magnification) and the femoral head of known size (true magnification).

Results

The mean true magnification using the femoral head was 21% (SD, 2%). The mean magnification using the marker was 15% (SD, 5%). The 95% CI for the mean difference between the two measurements was 6% to 7% (p < 0.001). The use of a magnification marker to estimate magnification at the level of the hip using standard radiographic techniques was shown in this study to routinely underestimate the magnification of the radiograph using an arthroplasty femoral head of known diameter as the reference. If we assume a magnification of 20%, this more closely approximated the true magnification routinely. With this assumption, we were within 2% magnification in 64 of the 100 hips and off by 4% or more in only four hips. In contrast, using the magnification marker we were within 2% of true magnification in only 20 hips and were off by 4% or more in 59 hips.

Conclusion

We found the use of a magnification marker with digital radiographs for preoperative templating to be generally inaccurate, with a mean error of 6% and range from −5% to 15%. Additionally, these data suggest that the use of a magnification marker while taking preoperative radiographs of the hip may be unnecessary, as simply setting the software to assume a 20% magnification actually was more accurate.

Level of Evidence

Level III, diagnostic study.
Literatur
1.
Zurück zum Zitat Gamble P, de Beer J, Petruccelli D, Winemaker M. The accuracy of digital templating in uncemented total hip arthroplasty. J Arthroplasty. 2010;25:529–532CrossRefPubMed Gamble P, de Beer J, Petruccelli D, Winemaker M. The accuracy of digital templating in uncemented total hip arthroplasty. J Arthroplasty. 2010;25:529–532CrossRefPubMed
2.
Zurück zum Zitat Iorio R, Siegel J, Specht LM, Tilzey JF, Hartman A, Healy WL. A comparison of acetate vs digital templating for preoperative planning of total hip arthroplasty: is digital templating accurate and safe? J Arthroplasty. 2009;24:175–179.CrossRefPubMed Iorio R, Siegel J, Specht LM, Tilzey JF, Hartman A, Healy WL. A comparison of acetate vs digital templating for preoperative planning of total hip arthroplasty: is digital templating accurate and safe? J Arthroplasty. 2009;24:175–179.CrossRefPubMed
3.
Zurück zum Zitat Shaarani SR, McHugh G, Collins DA. Accuracy of digital preoperative templating in 100 consecutive uncemented total hip arthroplasties: a single surgeon series. J Arthroplasty. 2013;28:331–337.CrossRefPubMed Shaarani SR, McHugh G, Collins DA. Accuracy of digital preoperative templating in 100 consecutive uncemented total hip arthroplasties: a single surgeon series. J Arthroplasty. 2013;28:331–337.CrossRefPubMed
4.
Zurück zum Zitat Sinclair VF, Wilson J, Jain NP, Knowles D. Assessment of accuracy of marker ball placement in pre-operative templating for total hip arthroplasty. J Arthroplasty. 2014;29:1658–1660.CrossRefPubMed Sinclair VF, Wilson J, Jain NP, Knowles D. Assessment of accuracy of marker ball placement in pre-operative templating for total hip arthroplasty. J Arthroplasty. 2014;29:1658–1660.CrossRefPubMed
5.
Zurück zum Zitat Stanish WM, Taylor N. Estimation of the intraclass correlation coefficient for the analysis of covariance model. Am Stat. 1983; 37:221–224. Stanish WM, Taylor N. Estimation of the intraclass correlation coefficient for the analysis of covariance model. Am Stat. 1983; 37:221–224.
6.
Zurück zum Zitat Steinberg EL, Shasha N, Menahem A, Dekel S. Preoperative planning of total hip replacement using the TraumaCadTM system. Arch Orthop Trauma Surg. 2010;130:1429–1432.CrossRefPubMed Steinberg EL, Shasha N, Menahem A, Dekel S. Preoperative planning of total hip replacement using the TraumaCadTM system. Arch Orthop Trauma Surg. 2010;130:1429–1432.CrossRefPubMed
7.
Zurück zum Zitat Whiddon DR, Bono JV. Digital templating in total hip arthroplasty. Instr Course Lect. 2008;57:273–279.PubMed Whiddon DR, Bono JV. Digital templating in total hip arthroplasty. Instr Course Lect. 2008;57:273–279.PubMed
8.
Zurück zum Zitat Whiddon DR, Bono JV, Lang JE, Smith EL, Salyapongse AK. Accuracy of digital templating in total hip arthroplasty. Am J Orthop (Belle Mead NJ). 2011;40:395–398.PubMed Whiddon DR, Bono JV, Lang JE, Smith EL, Salyapongse AK. Accuracy of digital templating in total hip arthroplasty. Am J Orthop (Belle Mead NJ). 2011;40:395–398.PubMed
Metadaten
Titel
Inaccuracies in the Use of Magnification Markers in Digital Hip Radiographs
verfasst von
Michael J. Archibeck, MD
Tamara Cummins, RT
Krishna R. Tripuraneni, MD
Joshua T. Carothers, MD
Cristina Murray-Krezan, MS
Mohammad Hattab, PhD
Richard E. White Jr, MD
Publikationsdatum
21.01.2016
Verlag
Springer US
Erschienen in
Clinical Orthopaedics and Related Research® / Ausgabe 8/2016
Print ISSN: 0009-921X
Elektronische ISSN: 1528-1132
DOI
https://doi.org/10.1007/s11999-016-4704-8

Weitere Artikel der Ausgabe 8/2016

Clinical Orthopaedics and Related Research® 8/2016 Zur Ausgabe

Arthropedia

Grundlagenwissen der Arthroskopie und Gelenkchirurgie. Erweitert durch Fallbeispiele, Videos und Abbildungen. 
» Jetzt entdecken

Update Orthopädie und Unfallchirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.