Influence of environmental temperature, heat-treatment and design on the cyclic fatigue resistance of three generations of a single-file nickel–titanium rotary instrument

verfasst von: Simone Staffoli, Nicola Maria Grande, Gianluca Plotino, Taha Özyürek, Mustafa Gündoğar, Leonzio Fortunato, Antonella Polimeni

Erschienen in: Odontology | Ausgabe 3/2019

Einloggen, um Zugang zu erhalten

Abstract

The aim of the present study was to evaluate the effect of environmental temperature, heat-treatment*** and design on the cyclic fatigue resistance of OneShape (OS), OneShape new generation (OSNG) and One Curve (OC) NiTi rotary single-file instruments. Cyclic fatigue testing of NiTi instruments was performed in an artificial canal with a 5-mm radius of curvature and a 60° angle of curvature located 5 mm from the tip of the instrument. During the experiment 3 different temperatures were used in a thermostatic bath with electronic and infrared controls: 0°, 20° and 35 °C. Sixty OneShape, 60 OneShape new generation and 60 One Curve instruments were rotated in the artificial canal until fracture occurred and the number of cycles to fracture (NCF) and the length of the fractured fragment (FL) was recorded. Data were analyzed by one-way ANOVA test at the 5% significance level. Fatigue resistance was significantly higher as the environmental temperature decreased, with differences ranging from 61 to 791%. OC instruments were significantly more resistant to cyclic fatigue than OS and OSNG instruments at all the temperatures tested (P < 0.05). There was no significant difference with respect to the length of the fractured file fragments for all the instruments tested at all the temperatures tested (P > 0.05). Within the limitation of the present study, fatigue resistance of the files was significantly increased when the environmental temperature decreased. OC instruments were significantly more resistant to cyclic fatigue than OS and OSNG instruments at all the temperatures tested.
Literatur
1.
Zurück zum Zitat McGuigan M, Louca C, Duncan H. Clinical decision-making after endodontic instrument fracture. Br Dent J. 2013;214:395–400.CrossRefPubMed McGuigan M, Louca C, Duncan H. Clinical decision-making after endodontic instrument fracture. Br Dent J. 2013;214:395–400.CrossRefPubMed
2.
Zurück zum Zitat Shen Y, Cheung GSP, Bian Z, Peng B. Comparison of defects in ProFile and ProTaper systems after clinical use. J Endod. 2006;32:61–5.CrossRef Shen Y, Cheung GSP, Bian Z, Peng B. Comparison of defects in ProFile and ProTaper systems after clinical use. J Endod. 2006;32:61–5.CrossRef
3.
Zurück zum Zitat Martin B, Zelada G, Varela P, et al. Factor influening the fracture of nickel-titanium rotary instruments. Int Endod J. 2003;26:262–6.CrossRef Martin B, Zelada G, Varela P, et al. Factor influening the fracture of nickel-titanium rotary instruments. Int Endod J. 2003;26:262–6.CrossRef
4.
Zurück zum Zitat Kitchens GG, Liewehr FR, Moon PC. The effect of operational speed on the fracture of nickel-titanium rotary instruments. J Endod. 2007;33:52–4.CrossRefPubMed Kitchens GG, Liewehr FR, Moon PC. The effect of operational speed on the fracture of nickel-titanium rotary instruments. J Endod. 2007;33:52–4.CrossRefPubMed
5.
Zurück zum Zitat Kim H-C, Kim H-J, Lee C-J, Kim B-M, Park J-K, Versluis A. Mechanical response of nickel–titanium instruments with different cross-sectional designs during shaping of simulated curved canals. Int Endod J. 2009;42:593–602.CrossRefPubMed Kim H-C, Kim H-J, Lee C-J, Kim B-M, Park J-K, Versluis A. Mechanical response of nickel–titanium instruments with different cross-sectional designs during shaping of simulated curved canals. Int Endod J. 2009;42:593–602.CrossRefPubMed
6.
Zurück zum Zitat Shen Y, Zhou H-m, Zheng Y-f, Peng B, Haapasalo M. Current challenges and concepts of the thermomechanical treatment of nickel-titanium instruments. J Endod. 2013;39:163–72.CrossRef Shen Y, Zhou H-m, Zheng Y-f, Peng B, Haapasalo M. Current challenges and concepts of the thermomechanical treatment of nickel-titanium instruments. J Endod. 2013;39:163–72.CrossRef
7.
Zurück zum Zitat Plotino G, Grande NM, Cotti E, Testarelli L, Gambarini G. Blue treatment enhances cyclic fatigue resistance of vortex nickel-titanium rotary files. J Endod. 2014;40:1451–3.CrossRef Plotino G, Grande NM, Cotti E, Testarelli L, Gambarini G. Blue treatment enhances cyclic fatigue resistance of vortex nickel-titanium rotary files. J Endod. 2014;40:1451–3.CrossRef
8.
Zurück zum Zitat Gündoğar M, Özyürek T. Cyclic fatigue resistance of OneShape, HyFlex EDM, WaveOne Gold, and Reciproc blue nickel-titanium instruments. J Endod. 2017;43:1192–6.CrossRef Gündoğar M, Özyürek T. Cyclic fatigue resistance of OneShape, HyFlex EDM, WaveOne Gold, and Reciproc blue nickel-titanium instruments. J Endod. 2017;43:1192–6.CrossRef
9.
Zurück zum Zitat Shen Y, Zhou H-m, Zheng Y-f, Campbell L, Peng B, Haapasalo M. Metallurgical characterization of controlled memory wire nickel-titanium rotary instruments. J Endod. 2011;37:1566–71.CrossRefPubMed Shen Y, Zhou H-m, Zheng Y-f, Campbell L, Peng B, Haapasalo M. Metallurgical characterization of controlled memory wire nickel-titanium rotary instruments. J Endod. 2011;37:1566–71.CrossRefPubMed
10.
Zurück zum Zitat Huang X, Ackland GJ, Rabe KM. Crystal structures and shape-memory behaviour of NiTi. Nat Mater. 2003;2:307–11.CrossRefPubMed Huang X, Ackland GJ, Rabe KM. Crystal structures and shape-memory behaviour of NiTi. Nat Mater. 2003;2:307–11.CrossRefPubMed
11.
Zurück zum Zitat Yoneyama T, Kobayashi C. Endodontic instruments for root canal treatment using Ti–Ni shape memory alloys. Shape memory alloys for biomedical applications. Cambridge: Woodhead Publishing Limited, 2009. 297–305.CrossRef Yoneyama T, Kobayashi C. Endodontic instruments for root canal treatment using Ti–Ni shape memory alloys. Shape memory alloys for biomedical applications. Cambridge: Woodhead Publishing Limited, 2009. 297–305.CrossRef
12.
Zurück zum Zitat Plotino G, Grande NM, Bellido MM, Testarelli L, Gambarini G. Influence of temperature on cyclic fatigue resistance of ProTaper Gold and ProTaper Universal rotary files. J Endod. 2017;43:200–2.CrossRef Plotino G, Grande NM, Bellido MM, Testarelli L, Gambarini G. Influence of temperature on cyclic fatigue resistance of ProTaper Gold and ProTaper Universal rotary files. J Endod. 2017;43:200–2.CrossRef
13.
Zurück zum Zitat Özyürek T. Cyclic fatigue resistance of Reciproc, WaveOne, and WaveOne Gold nickel-titanium instruments. J Endod. 2016;42:1536–9.CrossRef Özyürek T. Cyclic fatigue resistance of Reciproc, WaveOne, and WaveOne Gold nickel-titanium instruments. J Endod. 2016;42:1536–9.CrossRef
14.
Zurück zum Zitat Özyürek T, Yılmaz K, Uslu G. Comparison of cyclic fatigue resistance of old and new generation one shape NiTi instruments. Int J Appl Dent Sci. 2016;2:4–7. Özyürek T, Yılmaz K, Uslu G. Comparison of cyclic fatigue resistance of old and new generation one shape NiTi instruments. Int J Appl Dent Sci. 2016;2:4–7.
15.
Zurück zum Zitat Plotino G, Grande NM, Cordaro M, Testarelli L, Gambarini G. A review of cyclic fatigue testing of nickel-titanium rotary instruments. J Endod. 2009;35:1469–76.CrossRef Plotino G, Grande NM, Cordaro M, Testarelli L, Gambarini G. A review of cyclic fatigue testing of nickel-titanium rotary instruments. J Endod. 2009;35:1469–76.CrossRef
16.
Zurück zum Zitat de Hemptinne F, Slaus G, Vandendael M, Jacquet W, De Moor RJ, Bottenberg P. In vivo intracanal temperature evolution during endodontic treatment after the injection of room temperature or preheated sodium hypochlorite. J Endod. 2015;41:1112–5.CrossRef de Hemptinne F, Slaus G, Vandendael M, Jacquet W, De Moor RJ, Bottenberg P. In vivo intracanal temperature evolution during endodontic treatment after the injection of room temperature or preheated sodium hypochlorite. J Endod. 2015;41:1112–5.CrossRef
17.
Zurück zum Zitat Grande NM, Plotino G, Silla E, et al. Environmental temperature drastically affects flexural fatigue resistance of nickel-titanium rotary files. J Endod. 2017;43:1157–60.CrossRefPubMed Grande NM, Plotino G, Silla E, et al. Environmental temperature drastically affects flexural fatigue resistance of nickel-titanium rotary files. J Endod. 2017;43:1157–60.CrossRefPubMed
18.
Zurück zum Zitat de Vasconcelos RA, Murphy S, Carvalho CAT, Govindjee RG, Govindjee S, Peters OA. Evidence for reduced fatigue resistance of contemporary rotary instruments exposed to body temperature. J Endod. 2016;42:782–7.CrossRef de Vasconcelos RA, Murphy S, Carvalho CAT, Govindjee RG, Govindjee S, Peters OA. Evidence for reduced fatigue resistance of contemporary rotary instruments exposed to body temperature. J Endod. 2016;42:782–7.CrossRef
19.
Zurück zum Zitat Jamleh A, Yahata Y, Ebihara A, Atmeh AR, Bakhsh T, Suda H. Performance of NiTi endodontic instrument under different temperatures. Odontology. 2016;104:324–8.CrossRef Jamleh A, Yahata Y, Ebihara A, Atmeh AR, Bakhsh T, Suda H. Performance of NiTi endodontic instrument under different temperatures. Odontology. 2016;104:324–8.CrossRef
20.
Zurück zum Zitat Dosanjh A, Paurazas S, Askar M. The Effect of temperature on cyclic fatigue of nickel-titanium rotary endodontic instruments. J Endod. 2017;43:823–6.CrossRef Dosanjh A, Paurazas S, Askar M. The Effect of temperature on cyclic fatigue of nickel-titanium rotary endodontic instruments. J Endod. 2017;43:823–6.CrossRef
21.
Zurück zum Zitat Capar ID, Ertas H, Arslan H. Comparison of cyclic fatigue resistance of novel nickel-titanium rotary instruments. Aust Endod J. 2015;41:24–8.CrossRefPubMed Capar ID, Ertas H, Arslan H. Comparison of cyclic fatigue resistance of novel nickel-titanium rotary instruments. Aust Endod J. 2015;41:24–8.CrossRefPubMed
24.
Zurück zum Zitat de Melo MCC, de Azevedo Bahia MG, Buono VTL. Fatigue resistance of engine-driven rotary nickel-titanium endodontic instruments. J Endod. 2002;28:765–9.CrossRef de Melo MCC, de Azevedo Bahia MG, Buono VTL. Fatigue resistance of engine-driven rotary nickel-titanium endodontic instruments. J Endod. 2002;28:765–9.CrossRef
25.
Zurück zum Zitat Cheung G, Darvell B. Low-cycle fatigue of NiTi rotary instruments of various cross-sectional shapes. Int Endod J. 2007;40:626–32.CrossRefPubMed Cheung G, Darvell B. Low-cycle fatigue of NiTi rotary instruments of various cross-sectional shapes. Int Endod J. 2007;40:626–32.CrossRefPubMed
26.
Zurück zum Zitat Haikel Y, Serfaty R, Bateman G, Senger B, Allemann C. Dynamic and cyclic fatigue of engine-driven rotary nickel-titanium endodontic instruments. J Endod. 1999;25:434–40.CrossRef Haikel Y, Serfaty R, Bateman G, Senger B, Allemann C. Dynamic and cyclic fatigue of engine-driven rotary nickel-titanium endodontic instruments. J Endod. 1999;25:434–40.CrossRef
27.
Zurück zum Zitat Grande N, Plotino G, Pecci R, Bedini R, Malagnino V, Somma F. Cyclic fatigue resistance and three-dimensional analysis of instruments from two nickel–titanium rotary systems. Int Endod J. 2006;39:755–63.CrossRefPubMed Grande N, Plotino G, Pecci R, Bedini R, Malagnino V, Somma F. Cyclic fatigue resistance and three-dimensional analysis of instruments from two nickel–titanium rotary systems. Int Endod J. 2006;39:755–63.CrossRefPubMed
28.
Zurück zum Zitat Higuera O, Plotino G, Tocci L, Carrillo G, Gambarini G, Jaramillo DE. Cyclic fatigue resistance of 3 different nickel-titanium reciprocating instruments in artificial canals. J Endod. 2015;41:913–5.CrossRefPubMed Higuera O, Plotino G, Tocci L, Carrillo G, Gambarini G, Jaramillo DE. Cyclic fatigue resistance of 3 different nickel-titanium reciprocating instruments in artificial canals. J Endod. 2015;41:913–5.CrossRefPubMed
29.
Zurück zum Zitat Pedullà E, Plotino G, Scibilia M, et al. Cyclic fatigue comparison among endodontic instruments with similar cross section and different surface coating. Minerva Stomatol. 2016;65:231–4. Pedullà E, Plotino G, Scibilia M, et al. Cyclic fatigue comparison among endodontic instruments with similar cross section and different surface coating. Minerva Stomatol. 2016;65:231–4.
30.
Zurück zum Zitat Pedullà E, Lizio A, Scibilia M, et al. Cyclic fatigue resistance of two nickel–titanium rotary instruments in interrupted rotation. Int Endod J. 2017;50:194–201.CrossRefPubMed Pedullà E, Lizio A, Scibilia M, et al. Cyclic fatigue resistance of two nickel–titanium rotary instruments in interrupted rotation. Int Endod J. 2017;50:194–201.CrossRefPubMed
31.
Zurück zum Zitat Pedullà E, Benites A, La Rosa GM, et al. Cyclic fatigue resistance of heat-treated nickel-titanium instruments after immersion in sodium hypochlorite and/or sterilization. J Endod. 2018;44:648–653.CrossRefPubMed Pedullà E, Benites A, La Rosa GM, et al. Cyclic fatigue resistance of heat-treated nickel-titanium instruments after immersion in sodium hypochlorite and/or sterilization. J Endod. 2018;44:648–653.CrossRefPubMed
32.
Zurück zum Zitat Uslu G, Özyürek T, Yılmaz K, Plotino G. Effect of dynamic immersion in sodium hypochlorite and EDTA solutions on cyclic fatigue resistance of WaveOne and WaveOne gold reciprocating nickel-titanium Files. J Endod. 2018;44:834–7.CrossRefPubMed Uslu G, Özyürek T, Yılmaz K, Plotino G. Effect of dynamic immersion in sodium hypochlorite and EDTA solutions on cyclic fatigue resistance of WaveOne and WaveOne gold reciprocating nickel-titanium Files. J Endod. 2018;44:834–7.CrossRefPubMed
Metadaten
Titel
Influence of environmental temperature, heat-treatment and design on the cyclic fatigue resistance of three generations of a single-file nickel–titanium rotary instrument
verfasst von
Simone Staffoli
Nicola Maria Grande
Gianluca Plotino
Taha Özyürek
Mustafa Gündoğar
Leonzio Fortunato
Antonella Polimeni
Publikationsdatum
27.11.2018
Verlag
Springer Japan
Erschienen in
Odontology / Ausgabe 3/2019
Print ISSN: 1618-1247
Elektronische ISSN: 1618-1255
DOI
https://doi.org/10.1007/s10266-018-0399-5

Weitere Artikel der Ausgabe 3/2019

Odontology 3/2019 Zur Ausgabe

Newsletter

Bestellen Sie unseren kostenlosen Newsletter Update Zahnmedizin und bleiben Sie gut informiert – ganz bequem per eMail.