01.06.2013 | Original Article

Influence of implant design on the method of failure for three implants designed for use in the treatment of intertrochanteric fractures: the dynamic hip screw (DHS), DHS blade and X-BOLT

verfasst von: F. O’Neill, T. McGloughlin, B. Lenehan, F. Condon, J. C. Coffey, M. Walsh

Erschienen in: European Journal of Trauma and Emergency Surgery | Ausgabe 3/2013

Einloggen, um Zugang zu erhalten

Abstract

Background

The dynamic hip screw (DHS) has been widely adopted as the implant of choice in the treatment of intertrochanteric fractures. There have been attempts over the years to improve on the DHS lag screw design in order to reduce failure in the form of “cut out”. The purpose of this study was to investigate how two new design variations of the DHS, the DHS blade and the X-BOLT, behave within bone, and if these design modifications do indeed improve the fixation achieved and lead to a reduction in failure due to cut out.

Methods

“Pushout” tests were chosen as the means of investigating the failure modes and patterns for these implants that lead to cut out. These pushout studies were performed in artificial bone substrate in the form of polyurethane foam blocks and in cadaveric femoral heads.

Results

The results demonstrated that each individual implant produces its own specific distinct force–displacement curve or pattern of failure, and that despite the very different implant designs and methods of fixation, all of the implants tested reached very similar peak forces in each of the test materials used.

Conclusion

The results demonstrated that implant design only influences the pattern of failure, and that the peak forces reached by each implant are determined by the quality of the bone or test material into which they are placed. However, altering the force–displacement curve or pattern of failure may be enough to improve the fixation achieved and to provide an increased resistance to cut out.
Literatur
1.
Zurück zum Zitat Baumgaertner MR, Curtin SL, Lindskog DM, Keggi JM. The value of the tip apex distance in predicting failure of fixation of peritrochanteric fractures of the hip. J Bone Joint Surg Am. 1995;77:1058–64. PubMed Baumgaertner MR, Curtin SL, Lindskog DM, Keggi JM. The value of the tip apex distance in predicting failure of fixation of peritrochanteric fractures of the hip. J Bone Joint Surg Am. 1995;77:1058–64. PubMed
2.
Zurück zum Zitat Koval KJ, Zuckerman JD. Hip fractures: a practical guide to management. Berlin: Springer; 1999. Koval KJ, Zuckerman JD. Hip fractures: a practical guide to management. Berlin: Springer; 1999.
3.
Zurück zum Zitat Davis TRC, Sher JL, Horsman A, Simpson M, Porter BB, Checketts RG. Intertrochanteric femoral fractures. J Bone Joint Surg Br. 1990;72-B(1):26–31. Davis TRC, Sher JL, Horsman A, Simpson M, Porter BB, Checketts RG. Intertrochanteric femoral fractures. J Bone Joint Surg Br. 1990;72-B(1):26–31.
4.
Zurück zum Zitat James ETR, Hunter GA. The treatment of intertrochanteric fractures—a review article. Injury. 1983;14(5):421–31. PubMedCrossRef James ETR, Hunter GA. The treatment of intertrochanteric fractures—a review article. Injury. 1983;14(5):421–31. PubMedCrossRef
5.
Zurück zum Zitat Schipper IB, Marti RK, Van der Werken C. Unstable trochanteric femoral fractures: extramedullary or intramedullary fixation—review of literature. Injury. 2004;35(2):142–51. PubMedCrossRef Schipper IB, Marti RK, Van der Werken C. Unstable trochanteric femoral fractures: extramedullary or intramedullary fixation—review of literature. Injury. 2004;35(2):142–51. PubMedCrossRef
6.
Zurück zum Zitat Chirodian N, Arch B, Parker MJ. Sliding hip screw fixation of trochanteric hip fractures: outcome of 1024 procedures. Injury. 2005;36(6):793–800. PubMedCrossRef Chirodian N, Arch B, Parker MJ. Sliding hip screw fixation of trochanteric hip fractures: outcome of 1024 procedures. Injury. 2005;36(6):793–800. PubMedCrossRef
7.
Zurück zum Zitat Lustenberger A, Bekic J, Ganz R. Rotational instability of trochanteric fractures fixed with the dynamic hip screw. A radiological analysis. Unfallchirurg. 1995;98(10):514–7. German (PMID:7502084). PubMed Lustenberger A, Bekic J, Ganz R. Rotational instability of trochanteric fractures fixed with the dynamic hip screw. A radiological analysis. Unfallchirurg. 1995;98(10):514–7. German (PMID:7502084). PubMed
8.
Zurück zum Zitat Gibson D, Keogh C, Morris S. A biomechanical study comparing the dynamic hip screw with an X-Bolt in an unstable intertrochanteric fracture model of the proximal femur. Presented at: BOA/IOA Combined Meeting; 2011 Sept 14; Dublin, Ireland. Gibson D, Keogh C, Morris S. A biomechanical study comparing the dynamic hip screw with an X-Bolt in an unstable intertrochanteric fracture model of the proximal femur. Presented at: BOA/IOA Combined Meeting; 2011 Sept 14; Dublin, Ireland.
9.
Zurück zum Zitat Richards RH, Evans G, Egan J, Shearer JR. The AO dynamic hip screw and the Pugh sliding nail in femoral head fixation. J Bone Joint Surg Br. 1990;72(5):794–6. PubMed Richards RH, Evans G, Egan J, Shearer JR. The AO dynamic hip screw and the Pugh sliding nail in femoral head fixation. J Bone Joint Surg Br. 1990;72(5):794–6. PubMed
10.
Zurück zum Zitat Rosenblum SF, Zuckerman JD, Kummer FJ, Tam BS. A biomechanical evaluation of the Gamma nail. J Bone Joint Surg Br. 1992;74(3):352–7. PubMed Rosenblum SF, Zuckerman JD, Kummer FJ, Tam BS. A biomechanical evaluation of the Gamma nail. J Bone Joint Surg Br. 1992;74(3):352–7. PubMed
11.
Zurück zum Zitat Aminian A, Gao F, Fedoriw WW, Zhang LQ, Kalainov DM, Merk BR. Vertically oriented femoral neck fractures: mechanical analysis of four fixation techniques. J Orthop Trauma. 2007;21(8):544–8. PubMedCrossRef Aminian A, Gao F, Fedoriw WW, Zhang LQ, Kalainov DM, Merk BR. Vertically oriented femoral neck fractures: mechanical analysis of four fixation techniques. J Orthop Trauma. 2007;21(8):544–8. PubMedCrossRef
12.
Zurück zum Zitat Patel PSD, Shepherd DET, Hukins DWL. Compressive properties of commercially available polyurethane foams as mechanical models for osteoporotic human cancellous bone. BMC Musculoskelet Disord. 2008;9:137. PubMedCrossRef Patel PSD, Shepherd DET, Hukins DWL. Compressive properties of commercially available polyurethane foams as mechanical models for osteoporotic human cancellous bone. BMC Musculoskelet Disord. 2008;9:137. PubMedCrossRef
13.
Zurück zum Zitat Koval KJ, Zuckerman JD. Chapter 2. In: Hip fractures: a practical guide to management. Berlin: Springer; 1999. Koval KJ, Zuckerman JD. Chapter 2. In: Hip fractures: a practical guide to management. Berlin: Springer; 1999.
Metadaten
Titel
Influence of implant design on the method of failure for three implants designed for use in the treatment of intertrochanteric fractures: the dynamic hip screw (DHS), DHS blade and X-BOLT
verfasst von
F. O’Neill
T. McGloughlin
B. Lenehan
F. Condon
J. C. Coffey
M. Walsh
Publikationsdatum
01.06.2013
Verlag
Springer-Verlag
Erschienen in
European Journal of Trauma and Emergency Surgery / Ausgabe 3/2013
Print ISSN: 1863-9933
Elektronische ISSN: 1863-9941
DOI
https://doi.org/10.1007/s00068-013-0257-7

Weitere Artikel der Ausgabe 3/2013

European Journal of Trauma and Emergency Surgery 3/2013 Zur Ausgabe

Arthropedia

Grundlagenwissen der Arthroskopie und Gelenkchirurgie. Erweitert durch Fallbeispiele, DICOM-Daten, Videos und Abbildungen. » Jetzt entdecken

Neu im Fachgebiet Orthopädie und Unfallchirurgie

Update Orthopädie und Unfallchirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.