03.07.2019 | Original Article

Influence of virtual monochromatic spectral image at different energy levels on motion artifact correction in dual-energy spectral coronary CT angiography

verfasst von: Yongjun Jia, Bingying Zhai, Taiping He, Yong Yu, Nan Yu, Haifeng Duan, Chuangbo Yang, Xirong Zhang

Erschienen in: Japanese Journal of Radiology | Ausgabe 9/2019

Einloggen, um Zugang zu erhalten

Abstract

Purpose

To investigate the influence of virtual monochromatic spectral (VMS) CT images at different energy levels on the effectiveness of a motion correction technique (SSF) in dual-energy Spectral coronary CT angiography (CCTA).

Materials and methods

29 cases suspected of or diagnosed with coronary artery disease underwent Spectral CCTA using a prospective ECG triggering with 250 ms padding time. SSF was applied to the determined least-motion phase to generate 6 additional sets of VMS images with energy levels from 40 to 100 keV. CT value and standard deviation (SD) in the aortic root and epicardial adipose tissue were measured. Image quality of the RCA, LAD and LCX was evaluated on a per-vessel basis in each patient. Two reviewers evaluated the artery using the score of the segment.

Results

The low energy VMS images increased CT value and image noise compared with higher-energy VMS images, except 90 keV and 100 keV. The CNR of 40–70 keV were higher than those of 80–100 keV (P < 0.05). The image quality scores for images at 50–80 keV were higher than those of 40, 90, and 100 keV (P < 0.05), and the VMS image quality at 50 keV and 60 keV with SSF was the highest.

Conclusion

SSF can effectively reduce the motion artifacts when coronary vessels have suitable contrast enhancement which can be achieved by adjusting energy levels of VMS images.
Literatur
1.
Zurück zum Zitat Feuchtner G, Loureiro R, Bezerra H, et al. Quantification of coronary stenosis by dual source computed tomography in patients: a comparative study with intravascular ultrasound and invasive angiography. Eur J Radiol. 2012;81:83–8.CrossRefPubMed Feuchtner G, Loureiro R, Bezerra H, et al. Quantification of coronary stenosis by dual source computed tomography in patients: a comparative study with intravascular ultrasound and invasive angiography. Eur J Radiol. 2012;81:83–8.CrossRefPubMed
2.
Zurück zum Zitat Ding Z, Friedman MH. Quantification of 3-D coronary arterial motion using clinical biplane cineangiograms. Int J Cardiovasc Imaging. 2000;16:331–46.CrossRef Ding Z, Friedman MH. Quantification of 3-D coronary arterial motion using clinical biplane cineangiograms. Int J Cardiovasc Imaging. 2000;16:331–46.CrossRef
3.
Zurück zum Zitat Araoz PA, Kirsch J, Primak AN, Braun NN, et al. Optimal image reconstruction phase at low and high heart rates in dual-source CT coronary angiography. Int J Cardiovasc Imaging. 2009;25:837–45.CrossRefPubMedPubMedCentral Araoz PA, Kirsch J, Primak AN, Braun NN, et al. Optimal image reconstruction phase at low and high heart rates in dual-source CT coronary angiography. Int J Cardiovasc Imaging. 2009;25:837–45.CrossRefPubMedPubMedCentral
4.
Zurück zum Zitat Leipsic J, Labounty TM, Hague CJ, et al. Effect of a novel vendor-specific motion-correction algorithm on image quality and diagnostic accuracy in persons undergoing coronary CT angiography without rate-control medications. J Cardiovasc Comput Tomogr. 2012;6:164–71.CrossRefPubMed Leipsic J, Labounty TM, Hague CJ, et al. Effect of a novel vendor-specific motion-correction algorithm on image quality and diagnostic accuracy in persons undergoing coronary CT angiography without rate-control medications. J Cardiovasc Comput Tomogr. 2012;6:164–71.CrossRefPubMed
5.
Zurück zum Zitat Qianwen L, Pengyu L, Zhuangzhi S, et al. Effect of a novel motion correction algorithm (SSF) on the image quality of coronary CTA with intermediate heart rates: segment-based and vessel-based analyses. Eur J Radiol. 2014;83:2024–32.CrossRef Qianwen L, Pengyu L, Zhuangzhi S, et al. Effect of a novel motion correction algorithm (SSF) on the image quality of coronary CTA with intermediate heart rates: segment-based and vessel-based analyses. Eur J Radiol. 2014;83:2024–32.CrossRef
6.
Zurück zum Zitat Cheng J, Yin Y, Wu H, et al. Optimal monochromatic energy levels in spectral CT pulmonary angiography for the evaluation of pulmonary embolism. PLoS One. 2013;8:e63140.CrossRefPubMedPubMedCentral Cheng J, Yin Y, Wu H, et al. Optimal monochromatic energy levels in spectral CT pulmonary angiography for the evaluation of pulmonary embolism. PLoS One. 2013;8:e63140.CrossRefPubMedPubMedCentral
7.
Zurück zum Zitat Liang J, Wang H, Xu L, et al. Impact of SSF on diagnostic performance of coronary computed tomography angiography within 1 heart beat in patients with high heart rate using a 256-row detector computed tomography. J Comput Assist Tomogr. 2017;42:54–61.CrossRef Liang J, Wang H, Xu L, et al. Impact of SSF on diagnostic performance of coronary computed tomography angiography within 1 heart beat in patients with high heart rate using a 256-row detector computed tomography. J Comput Assist Tomogr. 2017;42:54–61.CrossRef
8.
Zurück zum Zitat Lee H, Kim JA, Lee JS, et al. Impact of a vendor-specific motion-correction algorithm on image quality, interpretability, and diagnostic performance of daily routine coronary CT angiography: influence of heart rate on the effect of motion-correction. Int J Cardiovasc Imaging. 2014;30:1603–12.CrossRefPubMed Lee H, Kim JA, Lee JS, et al. Impact of a vendor-specific motion-correction algorithm on image quality, interpretability, and diagnostic performance of daily routine coronary CT angiography: influence of heart rate on the effect of motion-correction. Int J Cardiovasc Imaging. 2014;30:1603–12.CrossRefPubMed
9.
Zurück zum Zitat Achenbach S, Paul JF, Laurent F, et al. Comparative assessment of image quality for coronary CT angiography with iobitridol and two contrast agents with higher iodine concentrations: iopromide and iomeprol. A multicentre randomized double-blind trial. Eur Radiol. 2017;27:821–30.CrossRef Achenbach S, Paul JF, Laurent F, et al. Comparative assessment of image quality for coronary CT angiography with iobitridol and two contrast agents with higher iodine concentrations: iopromide and iomeprol. A multicentre randomized double-blind trial. Eur Radiol. 2017;27:821–30.CrossRef
10.
Zurück zum Zitat Leipsic J, Abbara S, Achenbach S, et al. SCCT guidelines for the interpretation and reporting of coronary CT angiography: a report of the Society of Cardiovascular Computed Tomography Guidelines Committee. J Cardiovasc Comput Tomogr. 2014;8:342–58.CrossRef Leipsic J, Abbara S, Achenbach S, et al. SCCT guidelines for the interpretation and reporting of coronary CT angiography: a report of the Society of Cardiovascular Computed Tomography Guidelines Committee. J Cardiovasc Comput Tomogr. 2014;8:342–58.CrossRef
11.
Zurück zum Zitat Hong C, Becker CR, Huber A, et al. ECG-gated reconstructed multi-detector row CT coronary angiography: effect of varying trigger delay on image quality. Radiology. 2001;220:712–7.CrossRefPubMed Hong C, Becker CR, Huber A, et al. ECG-gated reconstructed multi-detector row CT coronary angiography: effect of varying trigger delay on image quality. Radiology. 2001;220:712–7.CrossRefPubMed
12.
Zurück zum Zitat Li M, Zhang GM, Zhao JS, Jiang ZW, et al. Diagnostic performance of dual-source CT coronary angiography with and without heart rate control: systematic review and meta-analysis. Clin Radiol. 2014;69:163–71.CrossRefPubMed Li M, Zhang GM, Zhao JS, Jiang ZW, et al. Diagnostic performance of dual-source CT coronary angiography with and without heart rate control: systematic review and meta-analysis. Clin Radiol. 2014;69:163–71.CrossRefPubMed
13.
Zurück zum Zitat Hassanab A. Technical challenges of coronary CT angiography: today and tomorrow. Eur J Radiol. 2011;79(79):161–71.CrossRef Hassanab A. Technical challenges of coronary CT angiography: today and tomorrow. Eur J Radiol. 2011;79(79):161–71.CrossRef
14.
Zurück zum Zitat Andreini D, Pontone G, Mushtaq S, et al. Low-dose CT coronary angiography with a novel IntraCycle motion-correction algorithm in patients with high heart rate or heart rate variability. Eur Heart J Cardiovasc Imaging. 2015;16:1093–100.CrossRefPubMed Andreini D, Pontone G, Mushtaq S, et al. Low-dose CT coronary angiography with a novel IntraCycle motion-correction algorithm in patients with high heart rate or heart rate variability. Eur Heart J Cardiovasc Imaging. 2015;16:1093–100.CrossRefPubMed
15.
Zurück zum Zitat Machida H, Lin XZ, Fukui R, et al. Influence of the motion correction algorithm on the quality and interpretability of images of single-source 64-detector coronary CT angiography among patients grouped by heart rate. Jpn J Radiol. 2015;33:84–93.CrossRefPubMed Machida H, Lin XZ, Fukui R, et al. Influence of the motion correction algorithm on the quality and interpretability of images of single-source 64-detector coronary CT angiography among patients grouped by heart rate. Jpn J Radiol. 2015;33:84–93.CrossRefPubMed
16.
Zurück zum Zitat Shechter G, Resar JR, Mcveigh ER. Displacement and velocity of the coronary arteries: cardiac and respiratory motion. IEEE Trans Med Imaging MI. 2006;25:369–75.CrossRef Shechter G, Resar JR, Mcveigh ER. Displacement and velocity of the coronary arteries: cardiac and respiratory motion. IEEE Trans Med Imaging MI. 2006;25:369–75.CrossRef
17.
Zurück zum Zitat Ohana M, Labani A, Severac F, et al. Single source dual energy CT: what is the optimal monochromatic energy level for the analysis of the lung parenchyma? Eur J Radiol. 2017;88:163–70.CrossRefPubMed Ohana M, Labani A, Severac F, et al. Single source dual energy CT: what is the optimal monochromatic energy level for the analysis of the lung parenchyma? Eur J Radiol. 2017;88:163–70.CrossRefPubMed
18.
Zurück zum Zitat Wang XP, Wang B, Hou P, et al. Screening and comparison of polychromatic and monochromatic image reconstruction of abdominal arterial energy spectrum CT. J Biol Regul Homeost Agents. 2017;31:189–94.PubMed Wang XP, Wang B, Hou P, et al. Screening and comparison of polychromatic and monochromatic image reconstruction of abdominal arterial energy spectrum CT. J Biol Regul Homeost Agents. 2017;31:189–94.PubMed
Metadaten
Titel
Influence of virtual monochromatic spectral image at different energy levels on motion artifact correction in dual-energy spectral coronary CT angiography
verfasst von
Yongjun Jia
Bingying Zhai
Taiping He
Yong Yu
Nan Yu
Haifeng Duan
Chuangbo Yang
Xirong Zhang
Publikationsdatum
03.07.2019
Verlag
Springer Japan
Erschienen in
Japanese Journal of Radiology / Ausgabe 9/2019
Print ISSN: 1867-1071
Elektronische ISSN: 1867-108X
DOI
https://doi.org/10.1007/s11604-019-00852-0

Weitere Artikel der Ausgabe 9/2019

Japanese Journal of Radiology 9/2019 Zur Ausgabe

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.