Inter- and intra-observer reproducibility of quantitative analysis for FP-CIT SPECT in patients with DLB

verfasst von: Atsutaka Okizaki, Michihiro Nakayama, Kaori Nakajima, Takayuki Katayama, Takahiro Uno, Fumiyoshi Morikawa, Juichiro Naoe, Koji Takahashi

Erschienen in: Annals of Nuclear Medicine | Ausgabe 10/2017

Einloggen, um Zugang zu erhalten

Abstract

Background

Dopamine transporter single photon emission CT (DAT-SPECT) is useful in the evaluation of dementia with Lewy bodies (DLB). The specific binding ratio (SBR) is an index to measure DAT density. However, poorly reproducible cases are occasionally experienced in clinical practice. We hypothesized that distance-weighted histogram (DWH) may be useful to improve the reproducibility of SBR. The purpose of this study was to investigate inter- and intra-observer reproducibility of SBR with conventional and DWH methods, and to visually evaluate the precision in reference voxel of interest (VOI) definition using these methods.

Methods

This study included 50 adult patients with probable DLB. They underwent brain MRI, DAT-SPECT, and cerebral blood flow SPECT with N-isopropyl-p-[123I]iodoamphetamine (I-123 IMP). SBR of the striatum was calculated using conventional and DWH method. For inter-observer reproducibility validation, conventional and DWH SBR were independently evaluated by experienced nuclear medicine physicians; these measurements were repeated by the nuclear medicine physician to investigate intra-observer reproducibility.

Results

Proper reference VOI definition was achieved in 60.0% using conventional SBR and in 98.0% with DWH SBR. Both conventional and DWH SBR demonstrated good inter- and intra-observer reproducibility, however, there were statistically significant inter- and intra-observer variations with conventional SBR measurements. Average inter- and intra-observer errors of conventional SBR were 7.9 and 6.1%, respectively. Conversely, DWH SBR errors were not observed in all patients. Moreover, average inter- and intra-observer errors were significantly higher in conventional SBR with improper reference VOI definition than in that with proper reference VOI definition.

Conclusions

Although the reproducibility of conventional and DWH SBR was good, inter- and intra-observer bias could not be ignored in conventional SBR, particularly with improper reference VOI definition. Thus, DWH may be useful to improve inter- and intra-observer reproducibility of SBR.
Literatur
1.
Zurück zum Zitat McKeith I, O’Brien J, Walker Z, Tatsch K, Booij J, Darcourt J, et al. Sensitivity and specificity of dopamine transporter imaging with 123I-FP-CIT SPECT in dementia with Lewy bodies: a phase III, multicentre study. Lancet Neurol. 2007;6:305–13. doi:10.1016/s1474-4422(07)70057-1.CrossRefPubMed McKeith I, O’Brien J, Walker Z, Tatsch K, Booij J, Darcourt J, et al. Sensitivity and specificity of dopamine transporter imaging with 123I-FP-CIT SPECT in dementia with Lewy bodies: a phase III, multicentre study. Lancet Neurol. 2007;6:305–13. doi:10.​1016/​s1474-4422(07)70057-1.CrossRefPubMed
2.
Zurück zum Zitat O’Brien JT, Colloby S, Fenwick J, Williams ED, Firbank M, Burn D, et al. Dopamine transporter loss visualized with FP-CIT SPECT in the differential diagnosis of dementia with Lewy bodies. Arch Neurol. 2004;61:919–25. doi:10.1001/archneur.61.6.919.CrossRefPubMed O’Brien JT, Colloby S, Fenwick J, Williams ED, Firbank M, Burn D, et al. Dopamine transporter loss visualized with FP-CIT SPECT in the differential diagnosis of dementia with Lewy bodies. Arch Neurol. 2004;61:919–25. doi:10.​1001/​archneur.​61.​6.​919.CrossRefPubMed
4.
Zurück zum Zitat Suwijn SR, van Boheemen CJ, de Haan RJ, Tissingh G, Booij J, de Bie RM. The diagnostic accuracy of dopamine transporter SPECT imaging to detect nigrostriatal cell loss in patients with Parkinson’s disease or clinically uncertain parkinsonism: a systematic review. EJNMMI Res. 2015;5:12. doi:10.1186/s13550-015-0087-1.CrossRefPubMedPubMedCentral Suwijn SR, van Boheemen CJ, de Haan RJ, Tissingh G, Booij J, de Bie RM. The diagnostic accuracy of dopamine transporter SPECT imaging to detect nigrostriatal cell loss in patients with Parkinson’s disease or clinically uncertain parkinsonism: a systematic review. EJNMMI Res. 2015;5:12. doi:10.​1186/​s13550-015-0087-1.CrossRefPubMedPubMedCentral
5.
Zurück zum Zitat Cummings JL, Henchcliffe C, Schaier S, Simuni T, Waxman A, Kemp P. The role of dopaminergic imaging in patients with symptoms of dopaminergic system neurodegeneration. Brain J Neurol. 2011;134:3146–66. doi:10.1093/brain/awr177.CrossRef Cummings JL, Henchcliffe C, Schaier S, Simuni T, Waxman A, Kemp P. The role of dopaminergic imaging in patients with symptoms of dopaminergic system neurodegeneration. Brain J Neurol. 2011;134:3146–66. doi:10.​1093/​brain/​awr177.CrossRef
6.
Zurück zum Zitat Tossici-Bolt L, Hoffmann SM, Kemp PM, Mehta RL, Fleming JS. Quantification of [123I]FP-CIT SPECT brain images: an accurate technique for measurement of the specific binding ratio. Eur J Nucl Med Mol Imaging. 2006;33:1491–9. doi:10.1007/s00259-006-0155-x.CrossRefPubMed Tossici-Bolt L, Hoffmann SM, Kemp PM, Mehta RL, Fleming JS. Quantification of [123I]FP-CIT SPECT brain images: an accurate technique for measurement of the specific binding ratio. Eur J Nucl Med Mol Imaging. 2006;33:1491–9. doi:10.​1007/​s00259-006-0155-x.CrossRefPubMed
7.
8.
Zurück zum Zitat Booij J, Speelman JD, Horstink MW, Wolters EC. The clinical benefit of imaging striatal dopamine transporters with [123I]FP-CIT SPET in differentiating patients with presynaptic parkinsonism from those with other forms of parkinsonism. Eur J Nucl Med. 2001;28:266–72.CrossRefPubMed Booij J, Speelman JD, Horstink MW, Wolters EC. The clinical benefit of imaging striatal dopamine transporters with [123I]FP-CIT SPET in differentiating patients with presynaptic parkinsonism from those with other forms of parkinsonism. Eur J Nucl Med. 2001;28:266–72.CrossRefPubMed
9.
Zurück zum Zitat Papathanasiou N, Rondogianni P, Chroni P, Themistocleous M, Boviatsis E, Pedeli X, et al. Interobserver variability, and visual and quantitative parameters of (123)I-FP-CIT SPECT (DaTSCAN) studies. Ann Nucl Med. 2012;26:234–40. doi:10.1007/s12149-011-0564-1.CrossRefPubMed Papathanasiou N, Rondogianni P, Chroni P, Themistocleous M, Boviatsis E, Pedeli X, et al. Interobserver variability, and visual and quantitative parameters of (123)I-FP-CIT SPECT (DaTSCAN) studies. Ann Nucl Med. 2012;26:234–40. doi:10.​1007/​s12149-011-0564-1.CrossRefPubMed
10.
Zurück zum Zitat Murayama K, Nishiyama Y, Hirose Y, Abe M, Ohyu S, Ninomiya A, et al. Differentiating between central nervous system lymphoma and high-grade glioma using dynamic susceptibility contrast and dynamic contrast-enhanced mr imaging with histogram analysis. Magn Reson Med Sci. 2017. doi:10.2463/mrms.mp.2016-0113.PubMed Murayama K, Nishiyama Y, Hirose Y, Abe M, Ohyu S, Ninomiya A, et al. Differentiating between central nervous system lymphoma and high-grade glioma using dynamic susceptibility contrast and dynamic contrast-enhanced mr imaging with histogram analysis. Magn Reson Med Sci. 2017. doi:10.​2463/​mrms.​mp.​2016-0113.PubMed
12.
Zurück zum Zitat Aylward EH, Li Q, Habbak QR, Warren A, Pulsifer MB, Barta PE, et al. Basal ganglia volume in adults with down syndrome. Psychiatry Res. 1997;74:73–82.CrossRefPubMed Aylward EH, Li Q, Habbak QR, Warren A, Pulsifer MB, Barta PE, et al. Basal ganglia volume in adults with down syndrome. Psychiatry Res. 1997;74:73–82.CrossRefPubMed
13.
Zurück zum Zitat Shrout PE, Fleiss JL. Intraclass correlations: uses in assessing rater reliability. Psychol Bull. 1979;86:420–8.CrossRefPubMed Shrout PE, Fleiss JL. Intraclass correlations: uses in assessing rater reliability. Psychol Bull. 1979;86:420–8.CrossRefPubMed
14.
Zurück zum Zitat Weir JP. Quantifying test-retest reliability using the intraclass correlation coefficient and the SEM. J Strength Cond Res. 2005;19:231–40. doi:10.1519/15184.1.PubMed Weir JP. Quantifying test-retest reliability using the intraclass correlation coefficient and the SEM. J Strength Cond Res. 2005;19:231–40. doi:10.​1519/​15184.​1.PubMed
16.
Zurück zum Zitat Tsuchida T, Ballinger JR, Vines D, Kim YJ, Utsunomiya K, Lang AE, et al. Reproducibility of dopamine transporter density measured with 123I-FPCIT SPECT in normal control and Parkinson’s disease patients. Ann Nucl Med. 2004;18:609–16.CrossRefPubMed Tsuchida T, Ballinger JR, Vines D, Kim YJ, Utsunomiya K, Lang AE, et al. Reproducibility of dopamine transporter density measured with 123I-FPCIT SPECT in normal control and Parkinson’s disease patients. Ann Nucl Med. 2004;18:609–16.CrossRefPubMed
17.
Zurück zum Zitat Booij J, Habraken JB, Bergmans P, Tissingh G, Winogrodzka A, Wolters EC, et al. Imaging of dopamine transporters with iodine-123-FP-CIT SPECT in healthy controls and patients with Parkinson’s disease. J Nucl Med. 1998;39:1879–84.PubMed Booij J, Habraken JB, Bergmans P, Tissingh G, Winogrodzka A, Wolters EC, et al. Imaging of dopamine transporters with iodine-123-FP-CIT SPECT in healthy controls and patients with Parkinson’s disease. J Nucl Med. 1998;39:1879–84.PubMed
19.
Zurück zum Zitat Scherfler C, Schwarz J, Antonini A, Grosset D, Valldeoriola F, Marek K, et al. Role of DAT-SPECT in the diagnostic work up of parkinsonism. Movement Disord. 2007;22:1229–38. doi:10.1002/mds.21505.CrossRefPubMed Scherfler C, Schwarz J, Antonini A, Grosset D, Valldeoriola F, Marek K, et al. Role of DAT-SPECT in the diagnostic work up of parkinsonism. Movement Disord. 2007;22:1229–38. doi:10.​1002/​mds.​21505.CrossRefPubMed
20.
Zurück zum Zitat Vlaar AM, de Nijs T, Kessels AG, Vreeling FW, Winogrodzka A, Mess WH, et al. Diagnostic value of 123I-ioflupane and 123I-iodobenzamide SPECT scans in 248 patients with parkinsonian syndromes. Eur Neurol. 2008;59:258 – 66. doi:10.1159/000115640.CrossRefPubMed Vlaar AM, de Nijs T, Kessels AG, Vreeling FW, Winogrodzka A, Mess WH, et al. Diagnostic value of 123I-ioflupane and 123I-iodobenzamide SPECT scans in 248 patients with parkinsonian syndromes. Eur Neurol. 2008;59:258 – 66. doi:10.​1159/​000115640.CrossRefPubMed
Metadaten
Titel
Inter- and intra-observer reproducibility of quantitative analysis for FP-CIT SPECT in patients with DLB
verfasst von
Atsutaka Okizaki
Michihiro Nakayama
Kaori Nakajima
Takayuki Katayama
Takahiro Uno
Fumiyoshi Morikawa
Juichiro Naoe
Koji Takahashi
Publikationsdatum
14.09.2017
Verlag
Springer Japan
Erschienen in
Annals of Nuclear Medicine / Ausgabe 10/2017
Print ISSN: 0914-7187
Elektronische ISSN: 1864-6433
DOI
https://doi.org/10.1007/s12149-017-1209-9

Weitere Artikel der Ausgabe 10/2017

Annals of Nuclear Medicine 10/2017 Zur Ausgabe