Interactive Flying Frustums (IFFs): spatially aware surgical data visualization

verfasst von: Javad Fotouhi, Mathias Unberath, Tianyu Song, Wenhao Gu, Alex Johnson, Greg Osgood, Mehran Armand, Nassir Navab

Erschienen in: International Journal of Computer Assisted Radiology and Surgery | Ausgabe 6/2019

Einloggen, um Zugang zu erhalten

Abstract

Purpose:

As the trend toward minimally invasive and percutaneous interventions continues, the importance of appropriate surgical data visualization becomes more evident. Ineffective interventional data display techniques that yield poor ergonomics that hinder hand–eye coordination, and therefore promote frustration which can compromise on-task performance up to adverse outcome. A very common example of ineffective visualization is monitors attached to the base of mobile C-arm X-ray systems.

Methods:

We present a spatially and imaging geometry-aware paradigm for visualization of fluoroscopic images using Interactive Flying Frustums (IFFs) in a mixed reality environment. We exploit the fact that the C-arm imaging geometry can be modeled as a pinhole camera giving rise to an 11-degree-of-freedom view frustum on which the X-ray image can be translated while remaining valid. Visualizing IFFs to the surgeon in an augmented reality environment intuitively unites the virtual 2D X-ray image plane and the real 3D patient anatomy. To achieve this visualization, the surgeon and C-arm are tracked relative to the same coordinate frame using image-based localization and mapping, with the augmented reality environment being delivered to the surgeon via a state-of-the-art optical see-through head-mounted display.

Results:

The root-mean-squared error of C-arm source tracking after hand–eye calibration was determined as \(0.43^{\circ }\pm 0.34^{\circ }\) and \(4.6\pm 2.7\,\hbox {mm}\) in rotation and translation, respectively. Finally, we demonstrated the application of spatially aware data visualization for internal fixation of pelvic fractures and percutaneous vertebroplasty.

Conclusion:

Our spatially aware approach to transmission image visualization effectively unites patient anatomy with X-ray images by enabling spatial image manipulation that abides image formation. Our proof-of-principle findings indicate potential applications for surgical tasks that mostly rely on orientational information such as placing the acetabular component in total hip arthroplasty, making us confident that the proposed augmented reality concept can pave the way for improving surgical performance and visuo-motor coordination in fluoroscopy-guided surgery.
Literatur
1.
Zurück zum Zitat Alambeigi F, Wang Y, Sefati S, Gao C, Murphy RJ, Iordachita I, Taylor RH, Khanuja H, Armand M (2017) A curved-drilling approach in core decompression of the femoral head osteonecrosis using a continuum manipulator. IEEE Robot Autom Lett 2(3):1480–1487CrossRef Alambeigi F, Wang Y, Sefati S, Gao C, Murphy RJ, Iordachita I, Taylor RH, Khanuja H, Armand M (2017) A curved-drilling approach in core decompression of the femoral head osteonecrosis using a continuum manipulator. IEEE Robot Autom Lett 2(3):1480–1487CrossRef
2.
Zurück zum Zitat Andress S, Johnson A, Unberath M, Winkler AF, Yu K, Fotouhi J, Weidert S, Osgood G, Navab N (2018) On-the-fly augmented reality for orthopedic surgery using a multimodal fiducial. J Med Imaging 5(2):021209CrossRef Andress S, Johnson A, Unberath M, Winkler AF, Yu K, Fotouhi J, Weidert S, Osgood G, Navab N (2018) On-the-fly augmented reality for orthopedic surgery using a multimodal fiducial. J Med Imaging 5(2):021209CrossRef
3.
Zurück zum Zitat Assarsson U, Moller T (2000) Optimized view frustum culling algorithms for bounding boxes. J Graph Tools 5(1):9–22CrossRef Assarsson U, Moller T (2000) Optimized view frustum culling algorithms for bounding boxes. J Graph Tools 5(1):9–22CrossRef
4.
Zurück zum Zitat Atria C, Last L, Packard N, Noo F (2018) Cone beam tomosynthesis fluoroscopy: a new approach to 3d image guidance. In: Medical imaging 2018: image-guided procedures, robotic interventions, and modeling, vol 10576. International Society for Optics and Photonics, p 105762V Atria C, Last L, Packard N, Noo F (2018) Cone beam tomosynthesis fluoroscopy: a new approach to 3d image guidance. In: Medical imaging 2018: image-guided procedures, robotic interventions, and modeling, vol 10576. International Society for Optics and Photonics, p 105762V
6.
Zurück zum Zitat Barr JD, Barr MS, Lemley TJ, McCann RM (2000) Percutaneous vertebroplasty for pain relief and spinal stabilization. Spine 25(8):923–928CrossRefPubMed Barr JD, Barr MS, Lemley TJ, McCann RM (2000) Percutaneous vertebroplasty for pain relief and spinal stabilization. Spine 25(8):923–928CrossRefPubMed
7.
Zurück zum Zitat Boszczyk BM, Bierschneider M, Panzer S, Panzer W, Harstall R, Schmid K, Jaksche H (2006) Fluoroscopic radiation exposure of the kyphoplasty patient. Eur Spine J 15(3):347–355CrossRefPubMed Boszczyk BM, Bierschneider M, Panzer S, Panzer W, Harstall R, Schmid K, Jaksche H (2006) Fluoroscopic radiation exposure of the kyphoplasty patient. Eur Spine J 15(3):347–355CrossRefPubMed
8.
Zurück zum Zitat Cardin MA, Wang JX, Plewes DB (2005) A method to evaluate human spatial coordination interfaces for computer-assisted surgery. Med Image Comput Comput Assist Interv MICCAI 2005:9–16 Cardin MA, Wang JX, Plewes DB (2005) A method to evaluate human spatial coordination interfaces for computer-assisted surgery. Med Image Comput Comput Assist Interv MICCAI 2005:9–16
9.
Zurück zum Zitat Chen X, Xu L, Wang Y, Wang H, Wang F, Zeng X, Wang Q, Egger J (2015) Development of a surgical navigation system based on augmented reality using an optical see-through head-mounted display. J Biomed Inform 55:124–131CrossRefPubMed Chen X, Xu L, Wang Y, Wang H, Wang F, Zeng X, Wang Q, Egger J (2015) Development of a surgical navigation system based on augmented reality using an optical see-through head-mounted display. J Biomed Inform 55:124–131CrossRefPubMed
10.
Zurück zum Zitat Chimenti PC, Mitten DJ (2015) Google glass as an alternative to standard fluoroscopic visualization for percutaneous fixation of hand fractures: a pilot study. Plast Reconstr Surg 136(2):328–330CrossRefPubMed Chimenti PC, Mitten DJ (2015) Google glass as an alternative to standard fluoroscopic visualization for percutaneous fixation of hand fractures: a pilot study. Plast Reconstr Surg 136(2):328–330CrossRefPubMed
11.
Zurück zum Zitat Deib G, Johnson A, Unberath M, Yu K, Andress S, Qian L, Osgood G, Navab N, Hui F, Gailloud P (2018) Image guided percutaneous spine procedures using an optical see-through head mounted display: proof of concept and rationale. J Neurointerv Surg 10:1187–1191CrossRefPubMed Deib G, Johnson A, Unberath M, Yu K, Andress S, Qian L, Osgood G, Navab N, Hui F, Gailloud P (2018) Image guided percutaneous spine procedures using an optical see-through head mounted display: proof of concept and rationale. J Neurointerv Surg 10:1187–1191CrossRefPubMed
12.
Zurück zum Zitat Endres F, Hess J, Engelhard N, Sturm J, Cremers D, Burgard W (2012) An evaluation of the RGB-D SLAM system. In: 2012 IEEE international conference on robotics and automation (ICRA). IEEE, pp 1691–1696 Endres F, Hess J, Engelhard N, Sturm J, Cremers D, Burgard W (2012) An evaluation of the RGB-D SLAM system. In: 2012 IEEE international conference on robotics and automation (ICRA). IEEE, pp 1691–1696
13.
Zurück zum Zitat Fischer M, Fuerst B, Lee SC, Fotouhi J, Habert S, Weidert S, Euler E, Osgood G, Navab N (2016) Preclinical usability study of multiple augmented reality concepts for K-wire placement. Int J Comput Assist Radiol Surg 11(6):1007–1014CrossRefPubMed Fischer M, Fuerst B, Lee SC, Fotouhi J, Habert S, Weidert S, Euler E, Osgood G, Navab N (2016) Preclinical usability study of multiple augmented reality concepts for K-wire placement. Int J Comput Assist Radiol Surg 11(6):1007–1014CrossRefPubMed
14.
Zurück zum Zitat Fotouhi J, Alexander CP, Unberath M, Taylor G, Lee SC, Fuerst B, Johnson A, Osgood GM, Taylor RH, Khanuja H, Armand M, Navab N (2018) Plan in 2-d, execute in 3-d: an augmented reality solution for cup placement in total hip arthroplasty. J Med Imaging 5(2):021205CrossRef Fotouhi J, Alexander CP, Unberath M, Taylor G, Lee SC, Fuerst B, Johnson A, Osgood GM, Taylor RH, Khanuja H, Armand M, Navab N (2018) Plan in 2-d, execute in 3-d: an augmented reality solution for cup placement in total hip arthroplasty. J Med Imaging 5(2):021205CrossRef
15.
Zurück zum Zitat Fotouhi J, Fuerst B, Johnson A, Lee SC, Taylor R, Osgood G, Navab N, Armand M (2017) Pose-aware C-arm for automatic re-initialization of interventional 2d/3d image registration. Int J Comput Assist Radiol Surg 12(7):1221–1230CrossRefPubMedPubMedCentral Fotouhi J, Fuerst B, Johnson A, Lee SC, Taylor R, Osgood G, Navab N, Armand M (2017) Pose-aware C-arm for automatic re-initialization of interventional 2d/3d image registration. Int J Comput Assist Radiol Surg 12(7):1221–1230CrossRefPubMedPubMedCentral
16.
Zurück zum Zitat Gafoor S, Schulz P, Heuer L, Matic P, Franke J, Bertog S, Reinartz M, Vaskelyte L, Hofmann I, Sievert H (2015) Use of echonavigator, a novel echocardiography–fluoroscopy overlay system, for transseptal puncture and left atrial appendage occlusion. J Interv Cardiol 28(2):215–217CrossRefPubMed Gafoor S, Schulz P, Heuer L, Matic P, Franke J, Bertog S, Reinartz M, Vaskelyte L, Hofmann I, Sievert H (2015) Use of echonavigator, a novel echocardiography–fluoroscopy overlay system, for transseptal puncture and left atrial appendage occlusion. J Interv Cardiol 28(2):215–217CrossRefPubMed
17.
Zurück zum Zitat Georgel PF, Schroeder P, Navab N (2009) Navigation tools for viewing augmented cad models. IEEE Comput Graph Appl 29(6):65–73CrossRefPubMed Georgel PF, Schroeder P, Navab N (2009) Navigation tools for viewing augmented cad models. IEEE Comput Graph Appl 29(6):65–73CrossRefPubMed
18.
Zurück zum Zitat Hajek J, Unberath M, Fotouhi J, Bier B, Lee SC, Osgood G, Maier A, Armand M, Navab N (2018) Closing the calibration loop: an inside-out-tracking paradigm for augmented reality in orthopedic surgery. arXiv preprint arXiv:1803.08610 Hajek J, Unberath M, Fotouhi J, Bier B, Lee SC, Osgood G, Maier A, Armand M, Navab N (2018) Closing the calibration loop: an inside-out-tracking paradigm for augmented reality in orthopedic surgery. arXiv preprint arXiv:​1803.​08610
19.
Zurück zum Zitat Hanna GB, Shimi SM, Cuschieri A (1998) Task performance in endoscopic surgery is influenced by location of the image display. Ann Surg 227(4):481CrossRefPubMedPubMedCentral Hanna GB, Shimi SM, Cuschieri A (1998) Task performance in endoscopic surgery is influenced by location of the image display. Ann Surg 227(4):481CrossRefPubMedPubMedCentral
20.
Zurück zum Zitat Hartley R, Zisserman A (2003) Multiple view geometry in computer vision. Cambridge University Press, Cambridge Hartley R, Zisserman A (2003) Multiple view geometry in computer vision. Cambridge University Press, Cambridge
21.
Zurück zum Zitat Hott JS, Deshmukh VR, Klopfenstein JD, Sonntag VK, Dickman CA, Spetzler RF, Papadopoulos SM (2004) Intraoperative Iso-C C-arm navigation in craniospinal surgery: the first 60 cases. Neurosurgery 54(5):1131–1137CrossRefPubMed Hott JS, Deshmukh VR, Klopfenstein JD, Sonntag VK, Dickman CA, Spetzler RF, Papadopoulos SM (2004) Intraoperative Iso-C C-arm navigation in craniospinal surgery: the first 60 cases. Neurosurgery 54(5):1131–1137CrossRefPubMed
22.
Zurück zum Zitat Joskowicz L, Hazan EJ (2016) Computer aided orthopaedic surgery: incremental shift or paradigm change? Med Image Anal 33:84–90CrossRefPubMed Joskowicz L, Hazan EJ (2016) Computer aided orthopaedic surgery: incremental shift or paradigm change? Med Image Anal 33:84–90CrossRefPubMed
23.
Zurück zum Zitat Liebergall M, Mosheiff R, Joskowicz L (2006) Computer-aided orthopaedic surgery in skeletal trauma. In: Bucholz RW, Heckman JD, Court-Brown CM (eds) Rockwood and greens fractures in adults, 6th edn. Lippincott, Philadelphia, pp 739–767 Liebergall M, Mosheiff R, Joskowicz L (2006) Computer-aided orthopaedic surgery in skeletal trauma. In: Bucholz RW, Heckman JD, Court-Brown CM (eds) Rockwood and greens fractures in adults, 6th edn. Lippincott, Philadelphia, pp 739–767
24.
Zurück zum Zitat Mason A, Paulsen R, Babuska JM, Rajpal S, Burneikiene S, Nelson EL, Villavicencio AT (2014) The accuracy of pedicle screw placement using intraoperative image guidance systems: a systematic review. J Neurosurg Spine 20(2):196–203CrossRefPubMed Mason A, Paulsen R, Babuska JM, Rajpal S, Burneikiene S, Nelson EL, Villavicencio AT (2014) The accuracy of pedicle screw placement using intraoperative image guidance systems: a systematic review. J Neurosurg Spine 20(2):196–203CrossRefPubMed
25.
Zurück zum Zitat Miller DL, Vañó E, Bartal G, Balter S, Dixon R, Padovani R, Schueler B, Cardella JF, De Baère T (2010) Occupational radiation protection in interventional radiology: a joint guideline of the cardiovascular and interventional radiology society of Europe and the society of interventional radiology. Cardiovasc Intervent Radiol 33(2):230–239CrossRefPubMed Miller DL, Vañó E, Bartal G, Balter S, Dixon R, Padovani R, Schueler B, Cardella JF, De Baère T (2010) Occupational radiation protection in interventional radiology: a joint guideline of the cardiovascular and interventional radiology society of Europe and the society of interventional radiology. Cardiovasc Intervent Radiol 33(2):230–239CrossRefPubMed
26.
Zurück zum Zitat Navab N, Mitschke M (2001) Method and apparatus using a virtual detector for three-dimensional reconstruction from x-ray images. US Patent 6,236,704 Navab N, Mitschke M (2001) Method and apparatus using a virtual detector for three-dimensional reconstruction from x-ray images. US Patent 6,236,704
27.
Zurück zum Zitat Qian L, Deguet A, Kazanzides P (2018) Arssist: augmented reality on a head-mounted display for the first assistant in robotic surgery. Healthc Technol Lett 5:194–200CrossRefPubMedPubMedCentral Qian L, Deguet A, Kazanzides P (2018) Arssist: augmented reality on a head-mounted display for the first assistant in robotic surgery. Healthc Technol Lett 5:194–200CrossRefPubMedPubMedCentral
28.
Zurück zum Zitat Qian L, Unberath M, Yu K, Fuerst B, Johnson A, Navab N, Osgood G (2017) Towards virtual monitors for image guided interventions-real-time streaming to optical see-through head-mounted displays. arXiv preprint arXiv:1710.00808 Qian L, Unberath M, Yu K, Fuerst B, Johnson A, Navab N, Osgood G (2017) Towards virtual monitors for image guided interventions-real-time streaming to optical see-through head-mounted displays. arXiv preprint arXiv:​1710.​00808
29.
Zurück zum Zitat Routt MC Jr, Nork SE, Mills WJ (2000) Percutaneous fixation of pelvic ring disruptions. Clin Orthop Relat Res (1976–2007) 375:15–29CrossRef Routt MC Jr, Nork SE, Mills WJ (2000) Percutaneous fixation of pelvic ring disruptions. Clin Orthop Relat Res (1976–2007) 375:15–29CrossRef
30.
Zurück zum Zitat Sefati S, Alambeigi F, Iordachita I, Armand M, Murphy RJ (2016) FBG-based large deflection shape sensing of a continuum manipulator: manufacturing optimization. In: 2016 IEEE SENSORS. IEEE, pp 1–3 Sefati S, Alambeigi F, Iordachita I, Armand M, Murphy RJ (2016) FBG-based large deflection shape sensing of a continuum manipulator: manufacturing optimization. In: 2016 IEEE SENSORS. IEEE, pp 1–3
31.
Zurück zum Zitat Sefati S, Pozin M, Alambeigi F, Iordachita I, Taylor RH, Armand M (2017) A highly sensitive fiber Bragg Grating shape sensor for continuum manipulators with large deflections. In: 2017 IEEE SENSORS. IEEE, pp 1–3 Sefati S, Pozin M, Alambeigi F, Iordachita I, Taylor RH, Armand M (2017) A highly sensitive fiber Bragg Grating shape sensor for continuum manipulators with large deflections. In: 2017 IEEE SENSORS. IEEE, pp 1–3
32.
Zurück zum Zitat Starr A, Jones A, Reinert C, Borer D (2001) Preliminary results and complications following limited open reduction and percutaneous screw fixation of displaced fractures of the acetabulum. Injury 32:SA45–50CrossRefPubMed Starr A, Jones A, Reinert C, Borer D (2001) Preliminary results and complications following limited open reduction and percutaneous screw fixation of displaced fractures of the acetabulum. Injury 32:SA45–50CrossRefPubMed
33.
Zurück zum Zitat Suzuki T, Soma K, Shindo M, Minehara H, Itoman M (2008) Anatomic study for pubic medullary screw insertion. J Orthop Surg 16(3):321–325CrossRef Suzuki T, Soma K, Shindo M, Minehara H, Itoman M (2008) Anatomic study for pubic medullary screw insertion. J Orthop Surg 16(3):321–325CrossRef
34.
Zurück zum Zitat Synowitz M, Kiwit J (2006) Surgeons radiation exposure during percutaneous vertebroplasty. J Neurosurg Spine 4(2):106–109CrossRefPubMed Synowitz M, Kiwit J (2006) Surgeons radiation exposure during percutaneous vertebroplasty. J Neurosurg Spine 4(2):106–109CrossRefPubMed
35.
Zurück zum Zitat Theocharopoulos N, Perisinakis K, Damilakis J, Papadokostakis G, Hadjipavlou A, Gourtsoyiannis N (2003) Occupational exposure from common fluoroscopic projections used in orthopaedic surgery. JBJS 85(9):1698–1703CrossRef Theocharopoulos N, Perisinakis K, Damilakis J, Papadokostakis G, Hadjipavlou A, Gourtsoyiannis N (2003) Occupational exposure from common fluoroscopic projections used in orthopaedic surgery. JBJS 85(9):1698–1703CrossRef
36.
Zurück zum Zitat Theologis A, Burch S, Pekmezci M (2016) Placement of iliosacral screws using 3d image-guided (O-arm) technology and stealth navigation: comparison with traditional fluoroscopy. Bone Joint J 98(5):696–702CrossRefPubMed Theologis A, Burch S, Pekmezci M (2016) Placement of iliosacral screws using 3d image-guided (O-arm) technology and stealth navigation: comparison with traditional fluoroscopy. Bone Joint J 98(5):696–702CrossRefPubMed
37.
38.
Zurück zum Zitat Tsai RY, Lenz RK (1989) A new technique for fully autonomous and efficient 3d robotics hand/eye calibration. IEEE Trans Robot Autom 5(3):345–358CrossRef Tsai RY, Lenz RK (1989) A new technique for fully autonomous and efficient 3d robotics hand/eye calibration. IEEE Trans Robot Autom 5(3):345–358CrossRef
39.
Zurück zum Zitat Tucker E, Fotouhi J, Unberath M, Lee SC, Fuerst B, Johnson A, Armand M, Osgood GM, Navab N (2018) Towards clinical translation of augmented orthopedic surgery: from pre-op ct to intra-op x-ray via RGBD sensing. In: Medical imaging 2018: imaging informatics for healthcare, research, and applications, vol 10579. International Society for Optics and Photonics, p 105790J Tucker E, Fotouhi J, Unberath M, Lee SC, Fuerst B, Johnson A, Armand M, Osgood GM, Navab N (2018) Towards clinical translation of augmented orthopedic surgery: from pre-op ct to intra-op x-ray via RGBD sensing. In: Medical imaging 2018: imaging informatics for healthcare, research, and applications, vol 10579. International Society for Optics and Photonics, p 105790J
40.
Zurück zum Zitat Unberath M, Fotouhi J, Hajek J, Maier A, Osgood G, Taylor R, Armand M, Navab N (2018) Augmented reality-based feedback for technician-in-the-loop C-arm repositioning. Healthc Technol Lett 5:143–147CrossRefPubMedPubMedCentral Unberath M, Fotouhi J, Hajek J, Maier A, Osgood G, Taylor R, Armand M, Navab N (2018) Augmented reality-based feedback for technician-in-the-loop C-arm repositioning. Healthc Technol Lett 5:143–147CrossRefPubMedPubMedCentral
41.
Zurück zum Zitat Vorraber W, Voessner S, Stark G, Neubacher D, DeMello S, Bair A (2014) Medical applications of near-eye display devices: an exploratory study. Int J Surg 12(12):1266–1272CrossRefPubMed Vorraber W, Voessner S, Stark G, Neubacher D, DeMello S, Bair A (2014) Medical applications of near-eye display devices: an exploratory study. Int J Surg 12(12):1266–1272CrossRefPubMed
42.
Zurück zum Zitat Yoon JW, Chen RE, Han PK, Si P, Freeman WD, Pirris SM (2016) Technical feasibility and safety of an intraoperative head-up display device during spine instrumentation. Int J Med Robot Comput Assist Surg 13:e1770CrossRef Yoon JW, Chen RE, Han PK, Si P, Freeman WD, Pirris SM (2016) Technical feasibility and safety of an intraoperative head-up display device during spine instrumentation. Int J Med Robot Comput Assist Surg 13:e1770CrossRef
Metadaten
Titel
Interactive Flying Frustums (IFFs): spatially aware surgical data visualization
verfasst von
Javad Fotouhi
Mathias Unberath
Tianyu Song
Wenhao Gu
Alex Johnson
Greg Osgood
Mehran Armand
Nassir Navab
Publikationsdatum
12.03.2019
Verlag
Springer International Publishing
Erschienen in
International Journal of Computer Assisted Radiology and Surgery / Ausgabe 6/2019
Print ISSN: 1861-6410
Elektronische ISSN: 1861-6429
DOI
https://doi.org/10.1007/s11548-019-01943-z

Weitere Artikel der Ausgabe 6/2019

International Journal of Computer Assisted Radiology and Surgery 6/2019 Zur Ausgabe

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.