Intermediate Results of iStent or iStent inject Implantation Combined with Cataract Surgery in a Real-World Setting: A Longitudinal Retrospective Study

verfasst von: Ricardo Augusto Paletta Guedes, Daniela Marcelo Gravina, Jonathan Clive Lake, Vanessa Maria Paletta Guedes, Alfredo Chaoubah

Erschienen in: Ophthalmology and Therapy | Ausgabe 1/2019

Abstract

Introduction

In this real-world, retrospective, comparative study we evaluated 6-month performance and safety in consecutive eyes following implantation of the iStent® or iStent inject® trabecular micro-bypass device with concomitant cataract surgery.

Methods

Performance outcomes included intraocular pressure (IOP) reduction; glaucoma medication reduction; proportions of eyes achieving an IOP of < 18, < 16, < 14, or < 12 mmHg; and proportions of eyes on 0, 1, 2, or ≥ 3 medications. Safety outcomes included adverse events, secondary surgeries, and best-corrected visual acuity (BCVA).

Results

A total of 73 eyes with open-angle glaucoma and cataract were included in the study; of these, 38 eyes were implanted with the iStent device and 35 were implanted with the iStent inject device. The two groups of patients had similar baseline characteristics, with the exception of mean age and medication burden (both higher in patients receiving the iStent inject device); over 90% of eyes in both groups had early glaucoma. At 6 months after surgery, mean IOP had fallen from 16.5 ± 3.9 to 13.9 ± 2.3 mmHg in eyes with the iStent implant (p < 0.001), and from 17.3 ± 3.0 to 12.7 ± 1.8 mmHg in those with the iStent inject implant (p < 0.001). This reduction was significantly greater in the iStent inject eyes than in the iStent eyes (26.6 vs. 15.8%) (p = 0.005). Significantly more eyes receiving the iStent inject device compared to the iStent device achieved an IOP of < 18 mmHg at 6 months post surgery (100 vs. 86.8%) (p = 0.033). Average medication usage was reduced from 1.8 to 0.4 medications in iStent eyes (p < 0.001) and from 2.3 to 0.4 medications in iStent inject eyes (p < 0.001). Over 70% of eyes in both groups became medication-free by 6 months post implantation. Adverse events in iStent eyes were mild and resulted in no sequelae; two iStent eyes underwent non-penetrating deep sclerectomy during follow-up. No complications or secondary surgeries were noted in iStent inject eyes. All eyes in both groups maintained or showed improved BCVA versus baseline.

Conclusion

Significant and safe IOP and medication reductions were observed after iStent or iStent inject implantation with concomitant cataract surgery. Trends toward greater effectiveness and fewer adverse events were observed with the iStent inject device compared with the iStent device.

Funding

Article processing charges were provided by Glaukos Corporation.
Hinweise

Enhanced digital features

To view enhanced digital features for this article go to https://​doi.​org/​10.​6084/​m9.​figshare.​7618775.

Introduction

The permanent optic nerve damage associated with glaucoma makes it the leading cause of irreversible blindness worldwide. All existing therapies, both medical and surgical, aim to lower intraocular pressure (IOP), which remains the primary risk factor linked to glaucoma progression and visual field decline. Indeed, considerable data support the strong relationship between reduced IOP and reduced glaucoma progression and vision loss [14]; this was quantified in the landmark Early Manifest Glaucoma Trial (EMGT) to be approximately 10% decreased risk of glaucoma progression per every 1 mmHg IOP reduction [1].
Historically, the glaucoma treatment landscape has consisted of medications and/or laser trabeculoplasty as initial treatment, while incisional filtering surgeries, such as trabeculectomy, non-penetrating deep sclerectomy, and tube implants, have rounded out the more invasive end of the treatment spectrum [5]. The effectiveness of medications may be limited by local and systemic side effects, poor adherence rates, difficulty with instillation, complex dosing regimens, and ocular surface hypersensitivities [610], while the utility of laser trabeculoplasty is curbed by its waning treatment effect over time [11]. Filtering surgeries yield considerable IOP reduction, but carry risks that include hypotony, infection, bleb-related complications, and/or choroidal detachment [1214]. Over the past two decades, micro-invasive glaucoma surgery (MIGS) has gained an increasing role in glaucoma treatment, and it may be particularly useful in patients whose disease lies between the aforementioned two extremes on the treatment spectrum.
A substantial body of peer-reviewed evidence has been amassed on the iStent implant (Glaukos Corp., San Clemente, CA, USA), which is the first MIGS device to be approved by the U.S. Food and Drug Administration, and its more recent iteration, the iStent inject implant (Glaukos Corp.) [1543]. These studies have demonstrated sizable, durable reductions in IOP and medication burden, combined with favorable safety. Results have been achieved in a variety of clinical settings, including with and without cataract surgery, in mild to severe glaucoma, in primary open-angle glaucoma as well as pseudoexfoliative glaucoma and ocular hypertension, in controlled clinical trials and single-surgeon case series, in comparative and non-comparative studies, and in evaluations of single or multiple stents.
Either the iStent (containing one stent) or iStent inject (containing two stents, each with the newer design) devices (Fig. 1) were implanted in the eyes included in the present study. Both devices are designed to decrease IOP by creating a patent pathway for aqueous humor to exit the anterior chamber through the trabecular meshwork into Schlemm’s canal. The inclusion of both devices in the cohort allowed us to compare not only one versus two stents, but more importantly to assess the additional benefit of the iStent inject device itself. Such additional benefit may emerge due to the presence of four lateral outlet lumens per stent to allow aqueous fluid to exit from the anterior chamber, potentially maximizing the number of clock-hours of outflow, and also due to the device’s greater procedural efficiency, which eases the learning curve for surgeons and conceivably results in more straightforward, complication-free implantation.
The rationale behind the use of multiple versus single stents evolved from preclinical and laboratory studies to clinical trials and single-surgeon case series. Laboratory investigations by Hunter et al. [44] and Bahler et al. [45, 46], for example, and supportive biomechanical models by Johnstone et al. [47], have demonstrated decreased outflow resistance and reduced IOP after single and double stent placement. In these studies, the bulk of IOP and resistance reduction was achieved after the first stent, with incremental benefit achieved with additional stent placement.
Clinical studies have mirrored these preclinical findings. A prospective randomized trial by Katz et al. [33] compared one, two, and three iStents in a standalone procedure (i.e., without concomitant cataract surgery), and Belovay et al. [39] and El Wardani et al. [40] evaluated two or three iStents during cataract surgery. The results of these three studies confirmed that most IOP reduction is owed to the first stent and that each additional stent yields further IOP and medication reduction.
The present study represents a natural evolution from these prior investigations, with side-by-side outcomes of both the iStent and iStent inject devices in a clinical setting. In this study we assessed the safety and effectiveness outcomes during 6 months following the implantation of either the iStent or iStent inject device in combination with cataract surgery in eyes with mild to moderate open-angle glaucoma. To our knowledge, this is the first study with head-to-head data on the two devices, as observed in a real-world setting in the hands of a single surgeon at a single site.

Methods

In this longitudinal retrospective study, we evaluated all consecutive eyes implanted with either the iStent or iStent inject device with concomitant cataract surgery from June 2017 to April 2018. All surgeries were performed in a single center by the same surgeon (R.G.). The choice between implants was based on their availability in the Brazilian market. From June 2017 to November 2017, only the iStent trabecular bypass was commercially available for use, while use of the iStent inject device steadily increased after its approval by the Brazilian regulatory agency in November 2017. Since the level of experience can confound surgical outcomes, we excluded the first ten iStent cases and the first five iStent inject cases from this eye cohort. Thus, all eyes included in this study were unaffected by the surgeon’s learning curve.
Eligible patients were over 18 years of age and had open-angle glaucoma, cataract requiring surgery, and the need for IOP and/or medication reduction; follow-up data were available for the entire 6 months following surgery. Eyes with missing data, shorter follow-up than 6 months, and/or significant coexisting ocular diagnoses that could confound the effect of the surgery were excluded.
Main effectiveness outcomes included percentage IOP reduction, percentage medication reduction, and proportion of patients achieving different levels of IOP control at the end of the follow-up (IOP < 18 mmHg, < 16 mmHg, < 14 mmHg, or < 12 mmHg). Safety outcomes included adverse events, secondary surgical interventions, and best-corrected visual acuity (BCVA). IOP was measured by Goldmann applanation, and BCVA was measured using a standard Snellen chart. Proper positioning of the iStent and iStent inject was verified by gonioscopy examination, which was performed both during surgery and at all clinical examinations throughout the follow-up.
Chi-square test and Student’s t test were used for the analysis of categorical and numerical variables, respectively. Statistical analysis was performed using SPSS statistical software (IBM Corp., Armonk, NY, USA), with a p value of 0.05 as the threshold for statistical significance.
All procedures were in accordance with the Institutional Review Board of the Paletta Guedes Eye Institute and with the 1964 Helsinki declaration and its later amendments or comparable ethical standards. Given that this was a retrospective outcomes analysis that included only patients from the surgeon’s real-world clinical population who already had received treatment, formal clinical trial registration was not required. As it was a retrospective study based on information on the patients’ medical records, informed consent from study participants was not necessary.

Results

Demographics and Baseline Ocular Characteristics

A total of 73 eyes met inclusion criteria and comprised the study population. These included 38 eyes receiving an iStent implant and 35 eyes receiving an iStent inject for which 6 months of postoperative follow-up data were available. The mean ± standard deviation age was 70.9 ± 9.0 years. The majority of patients were female (65.8%). Glaucoma stage was early, moderate and advanced in 97.3, 1.4, and 1.4% of study eyes, respectively.
Table 1 presents the preoperative comparison between groups. Groups were statistically comparable in baseline IOP, laterality (right or left eye), glaucoma stage, and baseline visual acuity. The two implant groups differed in their mean age (patients receiving the iStent inject device were an average of 8 years older) and the number of medications at baseline (patients receiving the iStent inject device had a higher mean number of medications).
Table 1
Demographic and baseline ocular characteristics according to the iStent and iStent inject groups
Patient characteristics
iStent device (n = 38 eyes)
iStent inject device (n = 35 eyes)
p value (between-group comparison)
Age (years)
67.1 ± 8.8
75.1 ± 7.1
< 0.001a
Baseline IOP (mmHg)
16.5 ± 3.9
17.3 ± 3.0
0.275a
Baseline number of medications
1.8 ± 0.9
2.3 ± 1.0
0.019a
Race
 Caucasian
60.5%
82.6%
0.122b
 African descent
39.5%
17.4%
Gender
 Male
28.9%
40.0%
0.227b
 Female
71.1%
60.0%
Glaucoma stage
 Early
100%
94.2%
0.189b
 Moderate
0.0%
2.9%
 Advanced
0.0%
2.9%
Baseline visual acuity
 20/30 or better
47.4%
34.3%
0.101b
 20/40 up to (but not including) 20/200
52.6%
57.1%
 20/200 or worse
0.0%
8.6%
Values in table are presented as the mean ± standard deviation (SD) or as the percentage, as appropriate
IOP Intraocular pressure
aStudent’s t test
bChi-square test

Intraocular Pressure

Mean IOP in the iStent group was reduced from 16.5 ± 3.9 mmHg at baseline to 13.9 ± 2.3 mmHg at 6 months post implantation(p < 0.001). In the iStent inject group, mean IOP dropped from 17.3 ± 3.0 at baseline to 12.7 ± 1.8 mmHg at 6 months post implantation (p < 0.001). Mean percentage IOP reduction was significantly greater in eyes receiving the iStent inject device (26.6%) than in those receiving the iStent implant (15.8%) (p = 0.005).
Figure 2 shows the mean IOP through the 6 months of follow-up for the overall eye cohort. Figure 3 and Table 2 show the mean IOP through the 6 months of follow-up for each study group separately. Statistically significant reductions in IOP versus baseline were achieved at all time points in the iStent inject group, and at all but two time points (Days 15 and 30) in the iStent group. IOP distribution into different IOP thresholds at the end of follow-up is shown in Table 3. Although baseline IOP was statistically similar between groups, at 6 months post surgery, significantly more iStent inject eyes than iStent eyes achieved an IOP of < 18 mmHg (p = 0.033) and IOP of <  12 mmHg (p = 0.04). More eyes with the iStent inject implant also achieved the other IOP thresholds at 6 months, but the between-group difference was not statistically significant at these latter thresholds (84.2 and 55.3% of iStent eyes vs. 88.6 and 71.4% of iStent inject eyes achieved a final IOP < 16 and < 14 mmHg, respectively; p > 0.05).
Table 2
Intraocular pressure from baseline to 6 months post surgery according to type of implant
Time Point
IOP (mmHg), iStent device (n = 38 eyes)a
IOP (mmHg), iStent inject device (n = 35 eyes)a
p valueb (between-group comparison of IOP at each time point)
Baseline
16.5 ± 3.9
17.3 ± 3.0
0.275
Day 1
13.7 ± 3.3 (p = 0.001)
11.9 ± 3.0 (p < 0.001)
0.022
Day 15
15.3 ± 4.2 (p = 0.135)
14.0 ± 3.2 (p < 0.001)
0.133
Day 30
15.2 ± 4.1 (p = 0.074)
13.9 ± 3.0 (p < 0.001)
0.152
Month 3
13.8 ± 2.8 (p < 0.001)
12.9 ± 1.8 (p < 0.001)
0.116
Month 6
13.9 ± 2.3 (p < 0.001)
12.7 ± 1.8 (p < 0.001)
0.014
Values in table are presented as the mean ± SD
aPaired Student’s t test (comparison vs. baseline); i.e., p value in each cell for IOP at that time point vs. baseline
bStudent’s t test (comparison between groups)
Table 3
Proportional analysis of intraocular pressure at 6 months post surgery according to type of implant
IOP threshold at 6 months
iStent device (%)
iStent inject device (%)
p valuea (between-group comparison of percentage of eyes at each IOP threshold)
IOP < 18 mmHg
86.8
100.0
0.033
IOP < 16 mmHg
84.2
88.6
0.422
IOP < 14 mmHg
55.3
71.4
0.118
IOP < 12 mmHg
7.9
25.7
0.040
aChi-square test (comparison between groups)

Medications

In eyes implanted with the iStent device, the mean number of medications reduced from 1.8 at baseline to 0.4 medications at 6 months post surgery (77.8% reduction; p < 0.001); in eyes implanted with the iStent inject device, medication usage reduced from 2.3 to 0.4 medications over the same time period (82.6% reduction; p < 0.001) (Fig. 4). Prior to surgery (i.e. at baseline) 63.1 and 74.3% of eyes in the iStent and iStent inject groups, respectively, were on ≥ 2 glaucoma medications, and   5.3 and 0% of eyes, respectively, were medication-free (Fig. 5a, b). By 6 months post surgery, a similarly high number of eyes in both groups had become medication-free (71.1% in iStent eyes and 74.3% in iStent inject eyes; p = 0.482) (Fig. 5a, b). The medication regimen postoperative was either decreased or maintained versus the preoperative (baseline) regimen for all eyes in both implant groups.

Safety

Both implant groups demonstrated favorable safety. Intraoperatively, there were two cases of technical difficulty with stent implantation (iStent eyes) that resulted in no sequelae and two cases of mild hemorrhagic reflux at the implantation site that resolved without intervention by postoperative Week 1. No eyes with the iStent inject device experienced an intraoperative complication (difference in intraoperative complication rate between groups: p = 0.142, not significant). Gonioscopy was performed intraoperatively and at all postoperative examinations; no significant problems in stent position were identified in either group at any point during the study. During the entire 6-month follow-up, all eyes in both groups maintained or improved their visual acuity versus baseline (preoperative), which is consistent with expectations for phacoemulsification cataract surgery. A minority of eyes (42.9%) presented a visual acuity of 20/30 or better at baseline, whereas 85.7% had that visual acuity level at the 6-month follow-up time point.
Only one eye (2.6%) with the iStent implant had an adverse event during the entire 6 months of follow-up. This one eye had peripheral anterior synechiae resulting in stent occlusion at 3 months post surgery, which was corrected in-office by Nd:YAG laser iridotomy and resulted in no further sequelae. No adverse events were reported for eyes with the iStent inject implant.
During the follow-up, two eyes with the iStent implant underwent secondary surgery (non-penetrating deep sclerectomy). In one of these eyes, the IOP became elevated after systemic steroid medication was given for asthma; no associated stent malfunction was noted. The IOP remained elevated in this eye despite topical medications, and systemic steroid withdrawal was not medically safe, so the decision was made to undertake surgery. In the second eye, IOP elevation occurred at approximately 2 months post surgery and did not have any known cause or stent malfunction. The patient had dry eye and medication hypersensitivities, so surgery was considered instead of additional medication to manage the IOP. There were no secondary surgeries in the iStent inject group.

Discussion

This real-world, single-surgeon study provides the first side-by-side comparative data on iStent and iStent inject devices in eyes with glaucoma. Outcomes for 6 months following surgery showed favorable safety and meaningful reductions in IOP and medication, with a directional trend toward better outcomes with the iStent inject device versus the iStent device. These findings were achieved in a real-world clinical setting and included all-comers, rather than a selection of only certain patients to yield positive outcomes. This approach allows for greater generalizability to physicians and patients seeking a suitable glaucoma treatment.
Clinically and statistically significant IOP reductions from baseline were observed in both implant groups at Month 6 (p < 0.001). These reductions are particularly notable given the relatively well-controlled IOP in both groups preoperatively (16.5 and 17.3 mmHg, respectively), which predisposes to modest postoperative IOP reductions [30].
The postoperative 6-month IOP range of approximately 12.5–14.0 mmHg is also significant given that these ab-interno (inside the eye) stent procedures act on the trabecular outflow pathway, where IOP reaches the physiologic lower limit of episcleral venous pressure. The potential for these trabecular stents to produce IOP reduction is not expected to be as dramatic as suprachoroidal or subconjunctival procedures; however, their trabecular location also underlies their lower safety risks compared to suprachoroidal or subconjunctival surgeries.
Although our limited sample size prevents us from drawing definitive conclusions, directional trends can be discerned in our comparison of the postoperative IOP and medications for the two groups. The IOP reduction in eyes with the iStent inject implant was approximately double that of those eyes receiving the iStent implant (26.6 vs. 15.8%, respectively; p = 0.005). Several factors may contribute to this significant difference. As shown in preclinical studies, and subsequently borne out in the clinical arena [33, 39, 40, 4447], the greatest reduction in IOP and outflow resistance comes from the first stent, with additional stents resulting in incrementally greater reductions in outflow resistance and IOP beyond those achieved by the first.
Both groups in our study were able to reduce their medication burden by over 80%, with a decrease of 1.4 medications in the iStent group and a decrease of 1.9 medications in the iStent inject group. The medication regimen from preoperative (baseline) to 6 months post surgery was either decreased or maintained in all eyes in both groups, and over 70% of eyes in both groups became medication-free by 6 months. The most salient benefits of reducing medication reliance are threefold: (1) medication effectiveness is dependent upon treatment adherence, which is known to be low in the majority of glaucoma patients; (2) topical medications result in local and systemic side effects and ocular surface compromise; (3) and medications carry costs associated with the drops themselves and with the additional caregiving and office visits needed to monitor medication regimens [610]. This latter drawback has been included in several recent population-based cost-effectiveness analyses of iStent surgery, which have shown the durable cost–benefit of stent surgery versus alternative treatment options [4851].
The safety profile of both groups was excellent, corroborating the existing body of evidence showing favorable safety of these devices [1543]. All eyes maintained or improved visual acuity from their preoperative (baseline) levels, suggesting no negative impact of stent implantation upon the visual improvement expected with cataract surgery. There were no reports of the complications seen with filtering surgeries, such as hypotony, choroidal detachment, endophthalmitis, bleb-related complications, or infection [1214]. During the 6 months of postoperative follow-up, adverse events occurred in the iStent group only, and these were limited and resulted in no sequelae. Two eyes with iStent implants had secondary surgery (non-penetrating deep sclerectomy) due to IOP elevation with no associated stent malfunction. There were no adverse events or secondary surgeries recorded in the iStent inject group.
Although both the iStent and iStent inject devices performed well in this cohort of eyes, the safety and effectiveness outcomes in the two groups suggest a treatment advantage favoring iStent inject over iStent. This advantage may be attributed to several characteristics of the device design. First, the presence of two stents, rather than one, in the trabecular meshwork allows for greater IOP-reducing potential, as shown in prior clinical and preclinical studies. Second, each individual iStent inject stent has four lateral outflow lumens, with the aim to maximize the number of clock-hours accessed for outflow. Finally, the iStent inject implantation procedure is considered to be more straightforward and efficient than that of the iStent, leading to a flatter and shorter learning curve for surgeons. This procedural elegance also conceivably allows for fewer complications with the iStent inject device than with the iStent device, as was observed in this study.
Because all patients underwent combined trabecular microbypass stent–cataract surgery, it was not possible to separate the IOP effect of cataract extraction versus stent implantation. Thus, patients’ postoperative outcomes should be understood in the context of expected IOP reductions after phacoemulsification alone. As established in the literature, post-phacoemulsification IOP reduction is estimated to be < 2 mmHg [52, 53] and/or to be 16.5% reduced [54] versus baseline IOP. Eyes with higher preoperative IOP are known to have greater postoperative IOP reduction [54]. In the present study, the majority of eyes had early glaucoma and relatively well-controlled preoperative IOP (16.5 and 17.3 mmHg in the iStent and iStent inject eyes, respectively), thereby predisposing to smaller reductions during the postoperative period. Despite this modest expectation, the study observed IOP reductions of 2.6–4.6 mmHg in the two groups (p < 0.001 for both); and equally importantly, these IOP reductions were achieved while also reducing the medication burden by 1.4–1.9 medications in the two groups (p < 0.001 for both).
Several limitations to our study should be mentioned. There was a modest number of eyes in each group. The study was intended to be a retrospective, non-comparative, preliminary outcomes analysis for iStent and iStent inject technologies in Brazil; however, future research could include prospective comparative trials, trials with larger sample sizes, and/or multicenter trials. All surgeries were completed by a single surgeon, and the groups were drawn from the surgeon’s real-life clinical population and thus were not identical; however, this real-world format may enhance applicability to current clinical settings by including a heterogeneous patient base unconstrained by the inclusion/exclusion criteria of traditional product-registration trials. There was no control group of cataract surgery only for comparison to the stent–cataract surgery groups. However, given that IOP and medications are numeric measurements, and medication decisions were made by the same doctor before and after surgery, patients’ preoperative IOP and medication number may be considered to be valid comparators for postoperative values. No medication washouts were undertaken, as this could compromise the ocular health of these real-world glaucoma patients. However, given that pre- and post-operative IOP and medications were reported in the same eyes, patients’ preoperative values were considered to be reasonable comparators for postoperative efficacy. This study covers outcomes during a 6-month postoperative period, as the iStent and iStent inject devices are relatively new technologies in Brazil. Thus, the results represent some of the first available data available in Brazil and for this surgeon. Future reports may include data with longer periods of follow-up.

Conclusions

These preliminary real-world data demonstrate significant reductions in IOP and number of anti-glaucomatous eye drops after implantation of either the iStent or iStent inject trabecular micro-bypass stent model. Implantation of the iStent inject device led to higher relative reductions in IOP and medication number and fewer adverse events versus implantation of the iStent device. A prospective study with a larger population and longer follow-up is necessary to validate these findings.

Acknowledgements

We thank the participants of the study.

Funding

No financial sponsorship was received for the work in this study. Article processing charges were funded by Glaukos Corporation. All authors had full access to all of the data in this study and take complete responsibility for the integrity of the data and accuracy of the data analysis.

Editorial Assistance

Editorial Assistance in the form of Medical Writing was conducted by Dr Dana Hornbeak from Glaukos Inc. Since all authors of the manuscript are native in the Portuguese language, Dr Dana Hornbeak had the courtesy to revise the English language in the text. Neither Dr. Hornbeak or Glaukos Inc. charged us for doing this service.

Authorship

All named authors meet the International Committee of Medical Journal Editors (ICMJE) criteria for authorship for this article, take responsibility for the integrity of the work as a whole, and have given their approval for this version to be published.

Disclosures

Ricardo Augusto Paletta Guedes has the following disclosures: consultant to Glaukos; Alcon; Allergan; OphtaVision Brasil. Daniela Marcelo Gravina, Jonathan Clive Lake, Vanessa Maria Paletta Guedes and Alfredo Chaoubah have nothing to disclose.

Compliance with Ethics Guidelines

All procedures were in accordance with the Institutional Review Board of the Paletta Guedes Eye Institute and with the 1964 Helsinki declaration and its later amendments or comparable ethical standards. Given that this was a retrospective outcomes analysis, including only patients from the surgeon’s real-world clinical population who already had received treatment, formal clinical trial registration was not required. As it was a retrospective study, based on information on the patients’ medical records, informed consent from study participants was not necessary.

Data Availability

The datasets generated during and/or analyzed during the current study are available from the corresponding author on reasonable request.

Open Access

This article is distributed under the terms of the Creative Commons Attribution-NonCommercial 4.0 International License (http://​creativecommons.​org/​licenses/​by-nc/​4.​0/​), which permits any noncommercial use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.
Literatur
1.
Zurück zum Zitat Leske MC, Heijl A, Hyman L, Bengtsson B, Komaroff E. Factors for progression and glaucoma treatment: the Early Manifest Glaucoma Trial. Curr Opin Ophthalmol. 2004;15:102–6.PubMedCrossRef Leske MC, Heijl A, Hyman L, Bengtsson B, Komaroff E. Factors for progression and glaucoma treatment: the Early Manifest Glaucoma Trial. Curr Opin Ophthalmol. 2004;15:102–6.PubMedCrossRef
2.
Zurück zum Zitat Heijl A, Leske MC, Bengtsson B, et al. Reduction of intraocular pressure and glaucoma progression: results from the Early Manifest Glaucoma Trial. Arch Ophthalmol. 2002;120:1268–1279. Heijl A, Leske MC, Bengtsson B, et al. Reduction of intraocular pressure and glaucoma progression: results from the Early Manifest Glaucoma Trial. Arch Ophthalmol. 2002;120:1268–1279.
3.
Zurück zum Zitat Kass MA, Heuer DK, Higginbotham EJ, et al. The Ocular Hypertension Treatment Study: a randomized trial determines that topical ocular hypotensive medication delays or prevents the onset of primary open-angle glaucoma. Arch Ophthalmol 2002;120:701–713 (discussion 829–830). Kass MA, Heuer DK, Higginbotham EJ, et al. The Ocular Hypertension Treatment Study: a randomized trial determines that topical ocular hypotensive medication delays or prevents the onset of primary open-angle glaucoma. Arch Ophthalmol 2002;120:701–713 (discussion 829–830).
4.
Zurück zum Zitat The Advanced Glaucoma Intervention Study (AGIS): 7. The relationship between control of intraocular pressure and visual field deterioration. Am J Ophthalmol 2000;130:429–440. The Advanced Glaucoma Intervention Study (AGIS): 7. The relationship between control of intraocular pressure and visual field deterioration. Am J Ophthalmol 2000;130:429–440.
6.
Zurück zum Zitat Nordstrom BL, Friedman DS, Mozaffari E, Quigley HA, Walker AM. Persistence and adherence with topical glaucoma therapy. Am J Ophthalmol. 2005;140(4):598–606.PubMedCrossRef Nordstrom BL, Friedman DS, Mozaffari E, Quigley HA, Walker AM. Persistence and adherence with topical glaucoma therapy. Am J Ophthalmol. 2005;140(4):598–606.PubMedCrossRef
7.
Zurück zum Zitat Tsai JC. A comprehensive perspective on patient adherence to topical glaucoma therapy. Ophthalmology. 2009;116:S30–6.PubMedCrossRef Tsai JC. A comprehensive perspective on patient adherence to topical glaucoma therapy. Ophthalmology. 2009;116:S30–6.PubMedCrossRef
8.
Zurück zum Zitat Newman-Casey PA, Robin AL, Blachley T, et al. The most common barriers to glaucoma medication adherence: a cross-sectional survey. Ophthalmology. 2015;122(7):1308–16.PubMedPubMedCentralCrossRef Newman-Casey PA, Robin AL, Blachley T, et al. The most common barriers to glaucoma medication adherence: a cross-sectional survey. Ophthalmology. 2015;122(7):1308–16.PubMedPubMedCentralCrossRef
9.
Zurück zum Zitat Baudouin C, Liang H, Hamard P, et al. The ocular surface of glaucoma patients treated over the long term expresses inflammatory markers related to both T-helper 1 and T-helper 2 pathways. Ophthalmology. 2008;115(1):109–15.PubMedCrossRef Baudouin C, Liang H, Hamard P, et al. The ocular surface of glaucoma patients treated over the long term expresses inflammatory markers related to both T-helper 1 and T-helper 2 pathways. Ophthalmology. 2008;115(1):109–15.PubMedCrossRef
10.
Zurück zum Zitat Leung EW, Medeiros FA, Weinreb RN. Prevalence of ocular surface disease in glaucoma patients. J Glaucoma. 2008;17(5):350–5.PubMedCrossRef Leung EW, Medeiros FA, Weinreb RN. Prevalence of ocular surface disease in glaucoma patients. J Glaucoma. 2008;17(5):350–5.PubMedCrossRef
11.
Zurück zum Zitat Glaucoma Laser Trial Research Group. The Glaucoma Laser Trial (GLT) and glaucoma laser trial follow-up study: 7 Results. Am J Ophthalmol. 1995;120:718–31.CrossRef Glaucoma Laser Trial Research Group. The Glaucoma Laser Trial (GLT) and glaucoma laser trial follow-up study: 7 Results. Am J Ophthalmol. 1995;120:718–31.CrossRef
12.
Zurück zum Zitat Gedde SJ, Herndon LW, Brandt JD, Budenz DL, Feuer WJ, Schiffman JC; Tube Versus Trabeculectomy Study Group. Postoperative complications in the Tube Versus Trabeculectomy (TVT) study during five years of follow-up. Am J Ophthalmol. 2012;153:804–814. Gedde SJ, Herndon LW, Brandt JD, Budenz DL, Feuer WJ, Schiffman JC; Tube Versus Trabeculectomy Study Group. Postoperative complications in the Tube Versus Trabeculectomy (TVT) study during five years of follow-up. Am J Ophthalmol. 2012;153:804–814.
13.
Zurück zum Zitat Rulli E, Biagioli E, Riva I, et al. Efficacy and safety of trabeculectomy vs nonpenetrating surgical procedures: a systematic review and meta-analysis. JAMA Ophthalmol. 2013;131(12):1573–82.PubMedCrossRef Rulli E, Biagioli E, Riva I, et al. Efficacy and safety of trabeculectomy vs nonpenetrating surgical procedures: a systematic review and meta-analysis. JAMA Ophthalmol. 2013;131(12):1573–82.PubMedCrossRef
14.
Zurück zum Zitat Jampel HD, Musch DC, Gillespie BW, Lichter PR, Wright MM, Guire KE; Collaborative Initial Glaucoma Treatment Study Group. Collaborative Initial Glaucoma Treatment Study Group. Perioperative complications of trabeculectomy in the collaborative initial glaucoma treatment study (CIGTS). Am J Ophthalmol. 2005;140(1):16-22. Jampel HD, Musch DC, Gillespie BW, Lichter PR, Wright MM, Guire KE; Collaborative Initial Glaucoma Treatment Study Group. Collaborative Initial Glaucoma Treatment Study Group. Perioperative complications of trabeculectomy in the collaborative initial glaucoma treatment study (CIGTS). Am J Ophthalmol. 2005;140(1):16-22.
15.
Zurück zum Zitat Samuelson TW, Katz LJ, Wells JM, et al. Randomized evaluation of the trabecular micro-bypass stent with phacoemulsification in patients with glaucoma and cataract. Ophthalmol. 2011;118:459–67.CrossRef Samuelson TW, Katz LJ, Wells JM, et al. Randomized evaluation of the trabecular micro-bypass stent with phacoemulsification in patients with glaucoma and cataract. Ophthalmol. 2011;118:459–67.CrossRef
16.
Zurück zum Zitat Craven ER, Katz LJ, Wells JM, Giamporcaro JE, iStent Study G. Cataract surgery with trabecular micro-bypass stent implantation in patients with mild-to-moderate open-angle glaucoma and cataract: two-year follow-up. J Cataract Refract Surg. 2012;38(8):1339–1345. Craven ER, Katz LJ, Wells JM, Giamporcaro JE, iStent Study G. Cataract surgery with trabecular micro-bypass stent implantation in patients with mild-to-moderate open-angle glaucoma and cataract: two-year follow-up. J Cataract Refract Surg. 2012;38(8):1339–1345.
17.
Zurück zum Zitat Arriola-Villalobos P, Martinez-de-la-Casa JM, Diaz-Valle D, et al. Combined iStent trabecular micro-bypass stent implantation and phacoemulsification for coexisting open-angle glaucoma and cataract: a long-term study. Br J Ophthalmol. 2012;96:645–9.PubMedCrossRef Arriola-Villalobos P, Martinez-de-la-Casa JM, Diaz-Valle D, et al. Combined iStent trabecular micro-bypass stent implantation and phacoemulsification for coexisting open-angle glaucoma and cataract: a long-term study. Br J Ophthalmol. 2012;96:645–9.PubMedCrossRef
18.
Zurück zum Zitat Arriola-Villalobos P, Martinez-de-la-Casa JM, Diaz-Valle D, Morales-Fernandez L, Fernandez-Perez C, Garcia-Feijoo J. Glaukos iStent inject® trabecular micro-bypass implantation associated with cataract surgery in patients with coexisting cataract and open-angle glaucoma or ocular hypertension: a long-term study. J Ophthalmol. 2016;2016:1056573.PubMedPubMedCentral Arriola-Villalobos P, Martinez-de-la-Casa JM, Diaz-Valle D, Morales-Fernandez L, Fernandez-Perez C, Garcia-Feijoo J. Glaukos iStent inject® trabecular micro-bypass implantation associated with cataract surgery in patients with coexisting cataract and open-angle glaucoma or ocular hypertension: a long-term study. J Ophthalmol. 2016;2016:1056573.PubMedPubMedCentral
19.
Zurück zum Zitat Fea AM. Phacoemulsification versus phacoemulsification with micro-bypass stent implantation in primary open-angle glaucoma. J Cataract Refract Surg. 2010;36:407–12.PubMedCrossRef Fea AM. Phacoemulsification versus phacoemulsification with micro-bypass stent implantation in primary open-angle glaucoma. J Cataract Refract Surg. 2010;36:407–12.PubMedCrossRef
20.
Zurück zum Zitat Fea AM, Belda JI, Rękas M, et al. Prospective unmasked randomized evaluation of the iStent inject® versus two ocular hypotensive agents in patients with primary open-angle glaucoma. Clin Ophthalmol. 2014;8:875–82.PubMedPubMedCentral Fea AM, Belda JI, Rękas M, et al. Prospective unmasked randomized evaluation of the iStent inject® versus two ocular hypotensive agents in patients with primary open-angle glaucoma. Clin Ophthalmol. 2014;8:875–82.PubMedPubMedCentral
21.
Zurück zum Zitat Fea AM, Consolandi G, Zola M, et al. Micro-Bypass implantation for primary open-angle glaucoma combined with phacoemulsification: 4-year follow-up. J Ophthalmol. 2015;26(2015):795357. Fea AM, Consolandi G, Zola M, et al. Micro-Bypass implantation for primary open-angle glaucoma combined with phacoemulsification: 4-year follow-up. J Ophthalmol. 2015;26(2015):795357.
22.
Zurück zum Zitat Donnenfeld ED, Solomon KD, Voskanyan L, et al.. A prospective 3-year follow-up trial of implantation of two trabecular microbypass stents in open-angle glaucoma. Clin Ophthalmol. 2015;9:2057–65.PubMedPubMedCentral Donnenfeld ED, Solomon KD, Voskanyan L, et al.. A prospective 3-year follow-up trial of implantation of two trabecular microbypass stents in open-angle glaucoma. Clin Ophthalmol. 2015;9:2057–65.PubMedPubMedCentral
23.
Zurück zum Zitat Chang DF, Donnenfeld ED, Katz LJ, et al. Efficacy of two trabecular micro-bypass stents combined with topical travoprost in open-angle glaucoma not controlled on two preoperative medications: 3-year follow-up. Clin Ophthalmol. 2017;11:523–8.PubMedPubMedCentralCrossRef Chang DF, Donnenfeld ED, Katz LJ, et al. Efficacy of two trabecular micro-bypass stents combined with topical travoprost in open-angle glaucoma not controlled on two preoperative medications: 3-year follow-up. Clin Ophthalmol. 2017;11:523–8.PubMedPubMedCentralCrossRef
24.
Zurück zum Zitat Voskanyan L, Garćia-Feijoo J, Belda JI, Fea A, Junemann A, Baudouin C. Synergy Study Group. Prospective, unmasked evaluation of the iStent® inject system for open-angle glaucoma: synergy trial. Adv Ther. 2014; 31(2):189–201. Voskanyan L, Garćia-Feijoo J, Belda JI, Fea A, Junemann A, Baudouin C. Synergy Study Group. Prospective, unmasked evaluation of the iStent® inject system for open-angle glaucoma: synergy trial. Adv Ther. 2014; 31(2):189–201.
25.
Zurück zum Zitat Ferguson T, Swan R, Ibach M, Schweitzer J, Sudhagoni R, Berdahl JP. Evaluation of a trabecular microbypass stent with cataract extraction in severe primary open-angle glaucoma. J Glaucoma. 2018;27(1):71–6.PubMed Ferguson T, Swan R, Ibach M, Schweitzer J, Sudhagoni R, Berdahl JP. Evaluation of a trabecular microbypass stent with cataract extraction in severe primary open-angle glaucoma. J Glaucoma. 2018;27(1):71–6.PubMed
26.
Zurück zum Zitat Myers JS, Masood I, Hornbeak DM, et al. Prospective evaluation of two iStent trabecular stents, one iStent Supra suprachoroidal stent and postoperative prostaglandina in refractory glaucoma: 4-year outcomes. Adv Ther. 2018;35:395–407.PubMedPubMedCentralCrossRef Myers JS, Masood I, Hornbeak DM, et al. Prospective evaluation of two iStent trabecular stents, one iStent Supra suprachoroidal stent and postoperative prostaglandina in refractory glaucoma: 4-year outcomes. Adv Ther. 2018;35:395–407.PubMedPubMedCentralCrossRef
27.
Zurück zum Zitat Gallardo MJ, Supnet RA, Giamporcaro JE, Hornbeak DM. Outcomes of combined trabecular micro-bypass and phacoemulsification in a predominantly Hispanic patient population. Clin Ophthalmol. 2016;10:1931–7.PubMedPubMedCentralCrossRef Gallardo MJ, Supnet RA, Giamporcaro JE, Hornbeak DM. Outcomes of combined trabecular micro-bypass and phacoemulsification in a predominantly Hispanic patient population. Clin Ophthalmol. 2016;10:1931–7.PubMedPubMedCentralCrossRef
28.
Zurück zum Zitat Neuhann TH. Trabecular micro-bypass stent implantation during small incision cataract surgery for open angle glaucoma or ocular hypertension: long term results. J Cataract Refract Surg. 2015;41:2664–71.PubMedCrossRef Neuhann TH. Trabecular micro-bypass stent implantation during small incision cataract surgery for open angle glaucoma or ocular hypertension: long term results. J Cataract Refract Surg. 2015;41:2664–71.PubMedCrossRef
29.
Zurück zum Zitat Ferguson TJ, Swan R, Ibach M, Schweitzer J, Sudhagoni R, Berdahl JP. Trabecular microbypass stent implantation with cataract extraction in pseudoexfoliation glaucoma. J Cataract Refract Surg. 2017;43(5):622–6.PubMedCrossRef Ferguson TJ, Swan R, Ibach M, Schweitzer J, Sudhagoni R, Berdahl JP. Trabecular microbypass stent implantation with cataract extraction in pseudoexfoliation glaucoma. J Cataract Refract Surg. 2017;43(5):622–6.PubMedCrossRef
30.
Zurück zum Zitat Ferguson TJ, Berdakl JP, Schweitzer JA, Sudhagoni RG. Clinical evaluation of a trabecular micro-bypass stent with phacoemulsification in patients with open-angle glaucoma and cataract. Clin Ophthamol. 2016;10:1767–73.CrossRef Ferguson TJ, Berdakl JP, Schweitzer JA, Sudhagoni RG. Clinical evaluation of a trabecular micro-bypass stent with phacoemulsification in patients with open-angle glaucoma and cataract. Clin Ophthamol. 2016;10:1767–73.CrossRef
31.
Zurück zum Zitat Ferguson T, Berdahl J, Schweitzer J, Sudhagoni R. Evaluation of a trabecular micro- bypass stent in pseudophakic patients with open-angle glaucoma. J Glaucoma. 2016;25:896–900.PubMedCrossRef Ferguson T, Berdahl J, Schweitzer J, Sudhagoni R. Evaluation of a trabecular micro- bypass stent in pseudophakic patients with open-angle glaucoma. J Glaucoma. 2016;25:896–900.PubMedCrossRef
32.
Zurück zum Zitat Vold SD, Voskanyan L, Tetz M, et al. Newly diagnosed primary open-angle glaucoma randomized to 2 trabecular bypass stents or prostaglandin: outcomes through 36 months. Ophthalmol Ther. 2016;5(2):161–72.PubMedPubMedCentralCrossRef Vold SD, Voskanyan L, Tetz M, et al. Newly diagnosed primary open-angle glaucoma randomized to 2 trabecular bypass stents or prostaglandin: outcomes through 36 months. Ophthalmol Ther. 2016;5(2):161–72.PubMedPubMedCentralCrossRef
33.
Zurück zum Zitat Katz LJ, Erb C, Carceller Guillamet A, . Long-term tirated IOP control with one, two or three trabecular micro-bypass stents in open-angle glaucoma subjetcs on topical hypotensive medication: 42-month outcomes. Clin Ophthalmol. 2018;12:255–62.PubMedPubMedCentralCrossRef Katz LJ, Erb C, Carceller Guillamet A, . Long-term tirated IOP control with one, two or three trabecular micro-bypass stents in open-angle glaucoma subjetcs on topical hypotensive medication: 42-month outcomes. Clin Ophthalmol. 2018;12:255–62.PubMedPubMedCentralCrossRef
34.
Zurück zum Zitat Klamann MK, Gonnermann J, Pahlitzsch M, et al. iStent inject in phakic open angle glaucoma. Graefes Arch Clin Exp Ophthalmol. 2015;253:941–7.PubMedCrossRef Klamann MK, Gonnermann J, Pahlitzsch M, et al. iStent inject in phakic open angle glaucoma. Graefes Arch Clin Exp Ophthalmol. 2015;253:941–7.PubMedCrossRef
35.
Zurück zum Zitat Seibold LK, Gamett KM, Kennedy JB, et al.. Outcomes after combined phacoemulsification and trabecular microbypass stent implantation in controlled open-angle glaucoma. J Cataract Refract Surg. 2016;42:1332–8.PubMedCrossRef Seibold LK, Gamett KM, Kennedy JB, et al.. Outcomes after combined phacoemulsification and trabecular microbypass stent implantation in controlled open-angle glaucoma. J Cataract Refract Surg. 2016;42:1332–8.PubMedCrossRef
36.
Zurück zum Zitat Berdahl J, Voskanyan L, Myers JS, et al. Implantation of two second-generation trabecular micro-bypass stents and topical travoprost in open-angle glaucoma not controlled on two preoperative medications: 18-month follow-up. Clin Exp Ophthalmol. 2017;45(8):797–802. Berdahl J, Voskanyan L, Myers JS, et al. Implantation of two second-generation trabecular micro-bypass stents and topical travoprost in open-angle glaucoma not controlled on two preoperative medications: 18-month follow-up. Clin Exp Ophthalmol. 2017;45(8):797–802.
37.
Zurück zum Zitat Lindstrom R, Lewis R, Hornbeak DM, et al. Outcomes following implantation of two second-generation trabecular micro-bypass stents in patients with open-angle glaucoma on one medication: 18-month follow-up. Adv Ther. 2016;33(11):2082–90.PubMedPubMedCentralCrossRef Lindstrom R, Lewis R, Hornbeak DM, et al. Outcomes following implantation of two second-generation trabecular micro-bypass stents in patients with open-angle glaucoma on one medication: 18-month follow-up. Adv Ther. 2016;33(11):2082–90.PubMedPubMedCentralCrossRef
38.
Zurück zum Zitat Hengerer FH, Auffarth GU, Riffel C, Conrad-Hengerer I. Prospective, non-randomized, 36-month study of second-generation trabecular micro-bypass stents with phacoemulsification in various types of glaucoma. Ophthalmol Ther. 2018;7(2):405–15. .PubMedPubMedCentralCrossRef Hengerer FH, Auffarth GU, Riffel C, Conrad-Hengerer I. Prospective, non-randomized, 36-month study of second-generation trabecular micro-bypass stents with phacoemulsification in various types of glaucoma. Ophthalmol Ther. 2018;7(2):405–15. .PubMedPubMedCentralCrossRef
39.
Zurück zum Zitat Belovay GW, Naqi A, Chan BJ, Rateb M, Ahmed II. Using multiple trabecular micro-bypass stents in cataract patients to treat open-angle glaucoma. J Cataract Refract Surg. 2012;38(11):1911–7.PubMedCrossRef Belovay GW, Naqi A, Chan BJ, Rateb M, Ahmed II. Using multiple trabecular micro-bypass stents in cataract patients to treat open-angle glaucoma. J Cataract Refract Surg. 2012;38(11):1911–7.PubMedCrossRef
40.
Zurück zum Zitat El Wardani M, Bergin C, Achache F, Sharkawi E. Evaluating the trabecular micro-bypass stent combined with phacoemulsification compared to phacoemulsification alone. Klin Monatsbl Augenheilkd. 2015;232:442–5.PubMedCrossRef El Wardani M, Bergin C, Achache F, Sharkawi E. Evaluating the trabecular micro-bypass stent combined with phacoemulsification compared to phacoemulsification alone. Klin Monatsbl Augenheilkd. 2015;232:442–5.PubMedCrossRef
41.
Zurück zum Zitat Malvankar-Mehta MS, Chen YN, Iordanous Y, Wang WW, Costella J, Hutnik CM. iStent as a solo procedure for glaucoma patients: a systematic review and meta-analysis. PLoS One. 2015;10(5):e0128146.PubMedPubMedCentralCrossRef Malvankar-Mehta MS, Chen YN, Iordanous Y, Wang WW, Costella J, Hutnik CM. iStent as a solo procedure for glaucoma patients: a systematic review and meta-analysis. PLoS One. 2015;10(5):e0128146.PubMedPubMedCentralCrossRef
42.
Zurück zum Zitat Malvankar-Mehta MS, Iordanous Y, Chen YN, et al. iStent with phacoemulsification versus phacoemulsification alone for patients with glaucoma and cataract: a meta-analysis. PLoS One. 2015;10(7):e0131770.PubMedPubMedCentralCrossRef Malvankar-Mehta MS, Iordanous Y, Chen YN, et al. iStent with phacoemulsification versus phacoemulsification alone for patients with glaucoma and cataract: a meta-analysis. PLoS One. 2015;10(7):e0131770.PubMedPubMedCentralCrossRef
43.
Zurück zum Zitat Popovic M, Campos-Moller X, Saheb H, K Ahmed II. Efficacy and adverse event profile of the iStent and iStent inject trabecular micro-bypass for open-angle glaucoma: a meta-analysis. J Curr Glaucoma Pract. 2018;12(2):67–84. Popovic M, Campos-Moller X, Saheb H, K Ahmed II. Efficacy and adverse event profile of the iStent and iStent inject trabecular micro-bypass for open-angle glaucoma: a meta-analysis. J Curr Glaucoma Pract. 2018;12(2):67–84.
44.
Zurück zum Zitat Hunter K, Fjield T, Heitzmann H, Shandas R, Kahook M. Characterization of micro-invasive trabecular bypass stents by ex vivo perfusion and computational flow modeling. Clin Ophthalmol. 2014;8:499–506.PubMedPubMedCentral Hunter K, Fjield T, Heitzmann H, Shandas R, Kahook M. Characterization of micro-invasive trabecular bypass stents by ex vivo perfusion and computational flow modeling. Clin Ophthalmol. 2014;8:499–506.PubMedPubMedCentral
45.
Zurück zum Zitat Bahler CK, Smedley GT, Zhou J, Johnson DH. Trabecular bypass stents decrease intraocular pressure in cultured human anterior segments. Am J Ophthal. 2004;138(6):988–94.PubMedCrossRef Bahler CK, Smedley GT, Zhou J, Johnson DH. Trabecular bypass stents decrease intraocular pressure in cultured human anterior segments. Am J Ophthal. 2004;138(6):988–94.PubMedCrossRef
46.
Zurück zum Zitat Bahler C, Hann C, Fjield T, Haffner D, Heitzmann H, Fautsch MP. Second-generation trabecular meshwork bypass stent (iStent inject) increases outflow facility in cultured human anterior segments. Am J Ophthal. 2012;153(6):1206–13.PubMedCrossRef Bahler C, Hann C, Fjield T, Haffner D, Heitzmann H, Fautsch MP. Second-generation trabecular meshwork bypass stent (iStent inject) increases outflow facility in cultured human anterior segments. Am J Ophthal. 2012;153(6):1206–13.PubMedCrossRef
47.
Zurück zum Zitat Johnstone MA. The aqueous outflow system as a mechanical pump: evidence from examination of tissue and aqueous movement in human and non-human primates. J Glaucoma. 2004;13(5):421–38.PubMedCrossRef Johnstone MA. The aqueous outflow system as a mechanical pump: evidence from examination of tissue and aqueous movement in human and non-human primates. J Glaucoma. 2004;13(5):421–38.PubMedCrossRef
48.
Zurück zum Zitat Ordóñez JE, Ordóñez A, Osorio UM. Cost-effectiveness analysis of iStent trabecular micro-bypass stent for patients with open-angle glaucoma in Colombia. Curr Med Res Opin. 2018;10:1–12. Ordóñez JE, Ordóñez A, Osorio UM. Cost-effectiveness analysis of iStent trabecular micro-bypass stent for patients with open-angle glaucoma in Colombia. Curr Med Res Opin. 2018;10:1–12.
49.
Zurück zum Zitat Tan SZ, Au L. Manchester iStent study: 3-year results and cost analysis. Eye (Lond). 2016;30(10):1365–70.CrossRef Tan SZ, Au L. Manchester iStent study: 3-year results and cost analysis. Eye (Lond). 2016;30(10):1365–70.CrossRef
50.
Zurück zum Zitat Berdahl JP, Khatana AK, Katz LJ, Herndon L, Layton AJ, Yu TM, Bauer MJ, Cantor LB. Cost-comparison of two trabecular micro-bypass stents versus selective laser trabeculoplasty or medications only for intraocular pressure control for patients with open-angle glaucoma. J Med Econ. 2017;20(7):760–6.PubMedCrossRef Berdahl JP, Khatana AK, Katz LJ, Herndon L, Layton AJ, Yu TM, Bauer MJ, Cantor LB. Cost-comparison of two trabecular micro-bypass stents versus selective laser trabeculoplasty or medications only for intraocular pressure control for patients with open-angle glaucoma. J Med Econ. 2017;20(7):760–6.PubMedCrossRef
51.
Zurück zum Zitat Ngan K, Fraser E, Buller S, Buller A. A cost minimization analysis comparing iStent accompanying cataract surgery and selective laser trabeculoplasty versus topical glaucoma medications in a public healthcare setting in New Zealand. Graefes Arch Clin Exp Ophthalmol. 2018;256(11):2181-9. Ngan K, Fraser E, Buller S, Buller A. A cost minimization analysis comparing iStent accompanying cataract surgery and selective laser trabeculoplasty versus topical glaucoma medications in a public healthcare setting in New Zealand. Graefes Arch Clin Exp Ophthalmol. 2018;256(11):2181-9.
52.
Zurück zum Zitat Vizzeri G, Weinreb RN. Cataract surgery and glaucoma. Curr Opin Ophthalmol. 2010;21(1):20–4.PubMedCrossRef Vizzeri G, Weinreb RN. Cataract surgery and glaucoma. Curr Opin Ophthalmol. 2010;21(1):20–4.PubMedCrossRef
53.
Zurück zum Zitat Friedman DS, Jampel HD, Lubomski LH, et al. Surgical strategies for coexisting glaucoma and cataract: an evidence-based update. Ophthalmology. 2002;109:1902–13.PubMedCrossRef Friedman DS, Jampel HD, Lubomski LH, et al. Surgical strategies for coexisting glaucoma and cataract: an evidence-based update. Ophthalmology. 2002;109:1902–13.PubMedCrossRef
54.
Zurück zum Zitat Mansberger SL, Gordon MO, Jampel H, et al. Reduction in intraocular pressure after cataract extraction: the Ocular Hypertension Treatment Study. Ophthalmology. 2012;119:1826–31.PubMedPubMedCentralCrossRef Mansberger SL, Gordon MO, Jampel H, et al. Reduction in intraocular pressure after cataract extraction: the Ocular Hypertension Treatment Study. Ophthalmology. 2012;119:1826–31.PubMedPubMedCentralCrossRef
Metadaten
Titel
Intermediate Results of iStent or iStent inject Implantation Combined with Cataract Surgery in a Real-World Setting: A Longitudinal Retrospective Study
verfasst von
Ricardo Augusto Paletta Guedes
Daniela Marcelo Gravina
Jonathan Clive Lake
Vanessa Maria Paletta Guedes
Alfredo Chaoubah
Publikationsdatum
05.02.2019
Verlag
Springer Healthcare
Erschienen in
Ophthalmology and Therapy / Ausgabe 1/2019
Print ISSN: 2193-8245
Elektronische ISSN: 2193-6528
DOI
https://doi.org/10.1007/s40123-019-0166-x

Weitere Artikel der Ausgabe 1/2019

Ophthalmology and Therapy 1/2019 Zur Ausgabe

Leitlinien kompakt für die Innere Medizin

Mit medbee Pocketcards sicher entscheiden.

Seit 2022 gehört die medbee GmbH zum Springer Medizin Verlag

Update Innere Medizin

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.