Internal fixation of intertrochanteric fractures using the percutaneous compression plate: the event of cutting out and its prediction by the tip-apex-distance

verfasst von: B. Schmidt-Rohlfing, M. Hofman, N. Heussen

Erschienen in: European Orthopaedics and Traumatology | Ausgabe 4/2015

Einloggen, um Zugang zu erhalten

Abstract

Introduction

In a recent study, the analyzed percutaneous compression plate (PCCP) data identified the tip-apex-distance (TAD) as the most valuable risk factor for the reoperation rate and in particular for the event of cutting out of the cephalic screws after internal fixation of intertrochanteric fractures. However, it is unknown which potential distance between the tip of the screws and the apex of the femoral head is most valuable in predicting the event of cutting out. As the PCCP device has two cephalic screws, there are at least three different distances which can be measured, including the distance between the tip of the lower screw and the apex of the femoral head, the distance between the center of the line between the two cephalic screws and the apex, and finally between the tip of the upper screw and the apex. Our study was motivated by the question which measurement can provide the best results for the prediction of the event of cutting out of the cephalic screws.

Patients and methods

The digitalized postoperative X-ray films of 174 patients with 174 PCCP implants were analyzed, and we obtained three different measurements for every patient. The primary outcome was the proportion of those patients with a cutting out of the screws. A univariate logistic regression was performed to assess the impact of the TAD on the event of cutting out. To describe the degree of consistency among measurements, we calculated an intra-class correlation coefficient (ICC) and the corresponding 95 % confidence limits. For each of the different measurements, we performed a receiver operating characteristic (ROC) curve analysis and calculated the area under the curve (AUC) and the corresponding 95 % confidence limits. In order to identify the most valuable distance measure, we added pairwise comparisons of the AUCs.

Results

Out of 174 patients, six patients (3.4 %) were reoperated due to cutting out of the cephalic screws. In the univariate logistic regression model, the p value was 0.0012 (OR 0.874; 95 % CI 0.806 to 0.948). The ROC curve analysis showed similar results for the different measurements; however, the highest AUC value was obtained for the center position.

Conclusions

In conclusion, all three measurements provide similar results and are suitable to predict the risk of cutting out of the cephalic screws.
Literatur
1.
Zurück zum Zitat Gotfried Y (2000) Percutaneous compression plating of intertrochanteric hip fractures. J Orthop Trauma 14:490–495CrossRefPubMed Gotfried Y (2000) Percutaneous compression plating of intertrochanteric hip fractures. J Orthop Trauma 14:490–495CrossRefPubMed
2.
Zurück zum Zitat Shen J, Luo F, Sun D, Huang Q, Xu J, Dong S, Xie Z (2015) Mid-term results after treatment of intertrochanteric femoral fractures with percutaneous compression plate (PCCP). Injury 46:347–357CrossRefPubMed Shen J, Luo F, Sun D, Huang Q, Xu J, Dong S, Xie Z (2015) Mid-term results after treatment of intertrochanteric femoral fractures with percutaneous compression plate (PCCP). Injury 46:347–357CrossRefPubMed
3.
Zurück zum Zitat Langford J, Pillai G, Ugliailoro AD et al (2011) Perioperative lateral trochanteric wall fractures: sliding hip screw versus percutaneous compression plate for intertrochanteric hip fractures. J Orthop Trauma 25:191–195CrossRefPubMed Langford J, Pillai G, Ugliailoro AD et al (2011) Perioperative lateral trochanteric wall fractures: sliding hip screw versus percutaneous compression plate for intertrochanteric hip fractures. J Orthop Trauma 25:191–195CrossRefPubMed
4.
Zurück zum Zitat Palm H, Jacobsen S, Sonne-Holm S, Gebuhr P, Hip Fracture Study Group (2007) Integrity of the lateral femoral wall in intertrochanteric hip fractures: an important predictor of a reoperation. J Bone Joint Surg Am 89:470–475CrossRefPubMed Palm H, Jacobsen S, Sonne-Holm S, Gebuhr P, Hip Fracture Study Group (2007) Integrity of the lateral femoral wall in intertrochanteric hip fractures: an important predictor of a reoperation. J Bone Joint Surg Am 89:470–475CrossRefPubMed
5.
Zurück zum Zitat Brandt S, Lefever S, Janzing H et al (2002) Percutaneous compression plating (PCCP) versus the dynamic hipscrew for pertrochanteric hip fractures: preliminary results. Injury 33:413–418CrossRefPubMed Brandt S, Lefever S, Janzing H et al (2002) Percutaneous compression plating (PCCP) versus the dynamic hipscrew for pertrochanteric hip fractures: preliminary results. Injury 33:413–418CrossRefPubMed
6.
Zurück zum Zitat Janzing H, Houben B, Brandt S et al (2002) The Gotfried PerCutaneous Compression Plate versus the Dynamic Hip Screw in the treatment of pertrochanteric hip fractures: minimal invasive treatment reduces operative time and postoperative pain. J Trauma 52:293–298CrossRefPubMed Janzing H, Houben B, Brandt S et al (2002) The Gotfried PerCutaneous Compression Plate versus the Dynamic Hip Screw in the treatment of pertrochanteric hip fractures: minimal invasive treatment reduces operative time and postoperative pain. J Trauma 52:293–298CrossRefPubMed
7.
Zurück zum Zitat Peyser A, Weil Y, Brocke L et al (2007) A prospective, randomised study comparing the percutaneous compression plate and the compression hip screw for the treatment of intertrochanteric fractures of the hip. J Bone Joint Surg (Br) 88:1210–1217CrossRef Peyser A, Weil Y, Brocke L et al (2007) A prospective, randomised study comparing the percutaneous compression plate and the compression hip screw for the treatment of intertrochanteric fractures of the hip. J Bone Joint Surg (Br) 88:1210–1217CrossRef
8.
Zurück zum Zitat Yang E, Qureshi S, Trokhan S et al (2011) Gotfried percutaneous compression plating compared with sliding hip screw fixation of intertrochanteric hip fractures: a prospective randomized study. J Bone Joint Surg Am 92:942–947 Yang E, Qureshi S, Trokhan S et al (2011) Gotfried percutaneous compression plating compared with sliding hip screw fixation of intertrochanteric hip fractures: a prospective randomized study. J Bone Joint Surg Am 92:942–947
9.
Zurück zum Zitat Zhang L, Shen J, Yu S, Huang Q, Xie Z (2014) Percutaneous compression plate versus dynamic hip screw for treatment of intertrochanteric Hip fractures: a meta-analyse of five randomized controlled trials. ScientificWorldJournal 10:512512 Zhang L, Shen J, Yu S, Huang Q, Xie Z (2014) Percutaneous compression plate versus dynamic hip screw for treatment of intertrochanteric Hip fractures: a meta-analyse of five randomized controlled trials. ScientificWorldJournal 10:512512
10.
Zurück zum Zitat Cheng T, Zhang GY, Liu T et al (2012) A meta-analysis of percutaneous compression plate 343 versus sliding hip screw for the management of intertrochanteric fractures of the hip. J Trauma Acute Care Surg 72:1435–1443CrossRefPubMed Cheng T, Zhang GY, Liu T et al (2012) A meta-analysis of percutaneous compression plate 343 versus sliding hip screw for the management of intertrochanteric fractures of the hip. J Trauma Acute Care Surg 72:1435–1443CrossRefPubMed
11.
Zurück zum Zitat Panesar S, Mirza S, Bharadwaj G et al (2008) The percutaneous compression plate versus the dynamic hip screw: a meta-analysis. Acta Orthop Belg 74:38–48PubMed Panesar S, Mirza S, Bharadwaj G et al (2008) The percutaneous compression plate versus the dynamic hip screw: a meta-analysis. Acta Orthop Belg 74:38–48PubMed
12.
Zurück zum Zitat Baumgaertner M, Curtin S, Lindskog D et al (1995) The value of the tip-apex distance in predicting failure of fixation of pertrochanteric fractures of the hip. J Bone Joint Surg Am 77:1058–1064PubMed Baumgaertner M, Curtin S, Lindskog D et al (1995) The value of the tip-apex distance in predicting failure of fixation of pertrochanteric fractures of the hip. J Bone Joint Surg Am 77:1058–1064PubMed
13.
Zurück zum Zitat Schmidt-Rohlfing B, Heussen N, Knobe M, Pfeifer R, Kaneshige JR, Pape HC (2013) Reoperation rate after internal fixation of intertrochanteric femur fractures with the percutaneous compression plate: what are the risk factors? J Orthop Trauma 27:312–317CrossRefPubMed Schmidt-Rohlfing B, Heussen N, Knobe M, Pfeifer R, Kaneshige JR, Pape HC (2013) Reoperation rate after internal fixation of intertrochanteric femur fractures with the percutaneous compression plate: what are the risk factors? J Orthop Trauma 27:312–317CrossRefPubMed
14.
Zurück zum Zitat Hanley J, McNeil B (1982) The meaning and use of the area under a receiver operating characteristic (ROC) curve. Radiology 143:29–36CrossRefPubMed Hanley J, McNeil B (1982) The meaning and use of the area under a receiver operating characteristic (ROC) curve. Radiology 143:29–36CrossRefPubMed
15.
Zurück zum Zitat Zweig MH, Campbell G (1993) Receiver-operating characteristic (ROC) plots: a fundamental evaluation tool in clinical medicine. Clin Chem 39:561–577PubMed Zweig MH, Campbell G (1993) Receiver-operating characteristic (ROC) plots: a fundamental evaluation tool in clinical medicine. Clin Chem 39:561–577PubMed
Metadaten
Titel
Internal fixation of intertrochanteric fractures using the percutaneous compression plate: the event of cutting out and its prediction by the tip-apex-distance
verfasst von
B. Schmidt-Rohlfing
M. Hofman
N. Heussen
Publikationsdatum
01.12.2015
Verlag
Springer Berlin Heidelberg
Erschienen in
European Orthopaedics and Traumatology / Ausgabe 4/2015
Print ISSN: 1867-4569
Elektronische ISSN: 1867-4577
DOI
https://doi.org/10.1007/s12570-015-0327-x

Weitere Artikel der Ausgabe 4/2015

European Orthopaedics and Traumatology 4/2015 Zur Ausgabe

Arthropedia

Grundlagenwissen der Arthroskopie und Gelenkchirurgie. Erweitert durch Fallbeispiele, Videos und Abbildungen. 
» Jetzt entdecken

Update Orthopädie und Unfallchirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.