Abstract

Background

Myocardial perfusion imaging (MPI) with 82Rb PET/CT is increasingly utilized in the evaluation of coronary artery disease with high diagnostic accuracy. Various softwares for data processing have been developed over the years with conflicting data regarding their reproducibility. In this study, we compared the quantitative results of myocardial perfusion and exam classification from three different softwares.

Methods

Data from consecutive patients who have undergone rest/stress 82Rb PET/CT MPI at the Royal Brompton & Harefield Trust, London, were analyzed. All data were processed using the Corridor 4DM (Invia, Ann Arbor, Michigan, USA), QPET (Cedars-Sinai, Los Angeles, California, USA), and SyngoMBF (Siemens Healthineers, Erlangen, Germany). The software packages addressed Lortie tracer kinetic model and region of interest (ROI) extraction correction option.

Statistics

A repeated-measures ANOVA with a Greenhouse–Geisser correction was performed with post hoc tests using Bonferroni correction. For intersoftware variability, Pearson correlation and intraclass correlation coefficients (ICC) were calculated. Bland–Altman assessed limit of agreement. Cohen’s Kappa assessed agreement in the classification of exams as normal or abnormal using an MFR cut-off value of 2.0. A P value of less than 0.05 was considered statistically significant.

Results

Data from 55 patients were analyzed. The mean values of myocardial blood flow (MBF) and myocardial perfusion reserve (MFR) were statistically significantly different among the softwares (P < 0.05). Corridor4DM had considerably lower values of MFR and classified a more substantial number of exams as abnormal (MFR: 2.21 ± 0.7, 2.4 ± 0.8, and 1.98 ± 0.8; and 18, 15, and 31 exams were abnormal for Syngo, QPET, and Corridor4DM, respectively). Accordingly, kappa agreement was moderate for Syngo vs QPET (k > 0.5), but minimal for Corridor4DM in comparison to its pairs (k < 0.4).

Conclusion

Users should be cautious when using different software interchangeably as systematic differences amongst them may introduce more extensive quantitative variation which could be clinically significant.
Anhänge
Nur mit Berechtigung zugänglich
Literatur
1.
Zurück zum Zitat Sampson UK, Dorbala S, Limaye A, Kwong R, Di Carli MF. Diagnostic accuracy of rubidium-82 myocardial perfusion imaging with hybrid positron emission tomography/computed tomography in the detection of coronary artery disease. J Am Coll Cardiol. 2007;49:1052-8.CrossRef Sampson UK, Dorbala S, Limaye A, Kwong R, Di Carli MF. Diagnostic accuracy of rubidium-82 myocardial perfusion imaging with hybrid positron emission tomography/computed tomography in the detection of coronary artery disease. J Am Coll Cardiol. 2007;49:1052-8.CrossRef
2.
Zurück zum Zitat Farhad H, Dunet V, Bachelard K, Allenbach G, Kaufmann PA, Prior JO. Added prognostic value of myocardial blood flow quantitation in rubidium-82 positron emission tomography imaging. Eur Heart J Cardiovasc Imaging. 2013;14:1203-10.CrossRef Farhad H, Dunet V, Bachelard K, Allenbach G, Kaufmann PA, Prior JO. Added prognostic value of myocardial blood flow quantitation in rubidium-82 positron emission tomography imaging. Eur Heart J Cardiovasc Imaging. 2013;14:1203-10.CrossRef
3.
Zurück zum Zitat Ziadi MC, Dekemp RA, Williams KA, et al. Impaired myocardial flow reserve on rubidium-82 positron emission tomography imaging predicts adverse outcomes in patients assessed for myocardial ischemia. J Am Coll Cardiol. 2011;58:740-8.CrossRef Ziadi MC, Dekemp RA, Williams KA, et al. Impaired myocardial flow reserve on rubidium-82 positron emission tomography imaging predicts adverse outcomes in patients assessed for myocardial ischemia. J Am Coll Cardiol. 2011;58:740-8.CrossRef
4.
Zurück zum Zitat Bravo PE, Chien D, Javadi M, Merrill J, Bengel FM. Reference ranges for LVEF and LV volumes from electrocardiographically gated 82Rb cardiac PET/CT using commercially available software. J Nucl Med. 2010;51:898-905.CrossRef Bravo PE, Chien D, Javadi M, Merrill J, Bengel FM. Reference ranges for LVEF and LV volumes from electrocardiographically gated 82Rb cardiac PET/CT using commercially available software. J Nucl Med. 2010;51:898-905.CrossRef
5.
Zurück zum Zitat Tahari AK, Lee A, Rajaram M, et al. Absolute myocardial flow quantification with (82)Rb PET/CT: Comparison of different software packages and methods. Eur J Nucl Med Mol Imaging. 2014;41:126-35.CrossRef Tahari AK, Lee A, Rajaram M, et al. Absolute myocardial flow quantification with (82)Rb PET/CT: Comparison of different software packages and methods. Eur J Nucl Med Mol Imaging. 2014;41:126-35.CrossRef
6.
Zurück zum Zitat Dekemp RA, Declerck J, Klein R, et al. Multisoftware reproducibility study of stress and rest myocardial blood flow assessed with 3D dynamic PET/CT and a 1-tissue-compartment model of 82Rb kinetics. J Nucl Med. 2013;54:571-7.CrossRef Dekemp RA, Declerck J, Klein R, et al. Multisoftware reproducibility study of stress and rest myocardial blood flow assessed with 3D dynamic PET/CT and a 1-tissue-compartment model of 82Rb kinetics. J Nucl Med. 2013;54:571-7.CrossRef
7.
Zurück zum Zitat Nesterov SV, Deshayes E, Sciagra R, et al. Quantification of myocardial blood flow in absolute terms using (82)Rb PET imaging: The RUBY-10 Study. JACC Cardiovasc Imaging. 2014;7:1119-27.CrossRef Nesterov SV, Deshayes E, Sciagra R, et al. Quantification of myocardial blood flow in absolute terms using (82)Rb PET imaging: The RUBY-10 Study. JACC Cardiovasc Imaging. 2014;7:1119-27.CrossRef
8.
Zurück zum Zitat Herzog BA, Husmann L, Valenta I, et al. Long-term prognostic value of 13 N-ammonia myocardial perfusion positron emission tomography added value of coronary flow reserve. J Am Coll Cardiol. 2009;54:150-6.CrossRef Herzog BA, Husmann L, Valenta I, et al. Long-term prognostic value of 13 N-ammonia myocardial perfusion positron emission tomography added value of coronary flow reserve. J Am Coll Cardiol. 2009;54:150-6.CrossRef
Metadaten
Titel
Intersoftware variability impacts classification of cardiac PET exams
verfasst von
Joaquim Barreto Oliveira
Yew Min Sen, MD
Kshama Wechalekar, MD
Publikationsdatum
01.12.2019
Verlag
Springer International Publishing
Erschienen in
Journal of Nuclear Cardiology / Ausgabe 6/2019
Print ISSN: 1071-3581
Elektronische ISSN: 1532-6551
DOI
https://doi.org/10.1007/s12350-018-1444-z

Weitere Artikel der Ausgabe 6/2019

Journal of Nuclear Cardiology 6/2019 Zur Ausgabe

Update Kardiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.