Intra-cardiac ultrasound guided approach for catheter ablation of typical right free wall accessory pathways

verfasst von: Matevž Jan, Tine Prolič Kalinšek, Jernej Štublar, Matija Jelenc, Andrej Pernat, David Žižek, Nikola Lakič

Erschienen in: BMC Cardiovascular Disorders | Ausgabe 1/2020

Abstract

Background

Right free wall accessory pathways (AP) are difficult to treat with catheter ablation as ablation catheter (AC) instability at the tricuspid annulus often precludes successful procedure.
The aim of our study was to test a novel intra-cardiac echocardiography (ICE) guided technique for AC placement. Feasibility and success rates were observed.

Methods

Eight consecutive patients (aged 29 ± 21 years, 4 female) with Wolff-Parkinson-White syndrome and a right free wall AP were included in the study. ICE, three-dimensional (3D) electro-anatomic mapping (EAM) system, and a steerable long sheath were used together with either an irrigated or a non-irrigated tip radio-frequency AC to achieve a “loop” manoeuvre which provided AC tip stability at the ventricular aspect of the tricuspid annulus. X-ray fluoroscopy was not used.

Results

Three patients had an anterior and five had a lateral location of the right free wall AP. Procedures were successful in all patients, without recurrences during the mean follow-up of 397 ± 363 days. Average procedural duration was 90 ± 31 min. On average, 6.6 ± 5.7 ablations were needed. Average time to terminate AP conduction after the start of ablation was 4.8 ± 4.2 s. In five patients (62%) AP conduction was successfully terminated with the first ablation. There were no procedural complications.

Conclusions

The novel ICE-guided approach with concomitant use of the steerable sheath and the 3D EAM system for zero-fluoroscopy mapping and ablation of the right free wall APs proved feasible and resulted in excellent acute and long-term outcomes.
Hinweise

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Abkürzungen
AP
Accessory pathway
AC
Ablation catheter
3D
Three dimensional
EAM
Electro-anatomic mapping
ICE
Intra-cardiac echocardiography
ECG
Electro-cardiogram
AVRT
Atrio-ventricular reentry tachycardia
RF
Radio-frequency
EP
Electrophysiology

Background

Typical right free wall accessory pathway (AP) mediated tachycardias are relatively rare and associated with higher procedural failure rates compared to left-sided and septal AP locations [15]. Different catheter ablation (CA) methods have been proposed to increase acute and long-term success of right free wall AP ablation. Those methods were mostly modifications of mapping approach, including use of a three-dimensional (3D) electro-anatomic mapping (EAM) system, use of multipolar diagnostic catheter for detailed tricuspid annulus mapping, use of detailed atrial insertion mapping and others [68]. Recently, fluoroscopy-guided ablation with catheter tip positioning on the ventricular insertion of the AP, beneath the tricuspid valve, was shown to result in excellent procedural and long-term success [9, 10].
The purpose of our work was to assess feasibility of a novel intra-cardiac echocardiography (ICE) guided technique for CA of the right free wall APs without the use of fluoroscopy.

Methods

Study population

In this retrospective analysis we included 8 consecutive patients with typical right free wall accessory pathway which underwent radio-frequency (RF) CA between April 2017 and August 2019. All patients had preexcitation clearly visible in the 12 lead electro-cardiogram (ECG) recording and had history of palpitations and supra-ventricular tachycardia recorded on the ECG recording on at least one occasion. Patients with previous CA were also included. Clinical, laboratory examinations and echocardiography were performed before the procedure.

General procedural characteristics

Most procedures were performed in conscious sedation, except in paediatric patients, where general anaesthesia was used. Right femoral vein punctures were performed to access the heart. All procedures were performed with the three dimensional (3D) electro-anatomic mapping (EAM) system (NavX™, Abbott, Abbott Park, Illinois, USA) and intra-cardiac echocardiography (AcuNav™, Biosense Webster, Irvine, California, USA) without the use of fluoroscopy.

Initial mapping with the 3D EAM system

As previously described, a 10-polar diagnostic catheter was inserted through the femoral vein and advanced into the right atrium without any use of fluoroscopy and only with the aid of the 3D EAM system [1114]. Afterwards, a limited 3D anatomy of the right atrium was obtained, position of the His bundle was tagged on the map and finally the catheter was advanced into the coronary sinus. An RF ablation catheter (D curve FlexAbility™, Abbott, Abbott Park, Illinois, USA or D curve Celsius® Biosense Webster, Irvine, California, USA) was advanced into the right atrium through a steerable sheath (large curve Agilis™, Abbott, Abbott Park, Illinois, USA) in a similar manner as the diagnostic catheter. Ventriculo-atrial and atrio-ventricular AP conduction was tested before the start of mapping. Mapping was performed in sinus rhythm or during atrial pacing, depending on which resulted in more preexcitation of the ventricles. A detailed local anatomical and activation map at the earliest ventricular activation at the tricuspid annulus was created. When atrio-ventricular reentry tachycardia (AVRT) was induced and sustained the tricuspid annulus was mapped for the earliest atrial activation.

Use of the steerable long sheath and ICE for catheter positioning - the “loop” manoeuvre

Once the initial activation mapping revealed the free wall location of the accessory pathway a stable position of the ablation catheter tip was obtained with the following manoeuvre:
  • ICE probe was advanced into the right atrium in a fluoroless manner, using the ICE image for guidance.
  • Under ICE guidance the steerable long sheath was bent in the septal direction in the case of lateral location of the accessory pathway or in the direction of the posterior tricuspid annulus in the case of anterior location of the accessory pathway.
  • Under ICE guidance the ablation catheter was advanced and bent in the direction of the location of the accessory pathway, with its curvature being the opposite of the curvature of the long sheath, with both together forming an open S shape loop.
  • Under ICE guidance, slight counterclockwise long sheath rotation and with precise increase or decrease of the curvature of both the steerable sheath and the ablation catheter the tip of the ablation catheter was positioned just beneath the tricuspid annulus.
A schematic example of the “loop” manoeuvre in the case of an anterior right free wall accessory pathway is shown in Fig. 1.

Detailed mapping and ablation

After a stable position of the ablation catheter tip was obtained some additional mapping points for accurate accessory pathway location were obtained and were typically located slightly toward the ventricular side of the map compared to the mapping points obtained during initial mapping. Finally, RF ablation was performed with 40 W when a non-irrigated tip ablation catheter was used and with 30 W when an irrigated tip ablation catheter was used, both in temperature controlled mode, with cutoff set to 60 and 43 °C, respectively. Ventriculo-atrial and atrio-ventricular AP conduction was tested after successful ablation.

Procedural parameters

Procedural time was defined as time from the insertion of the venous sheaths to their removal at the end of the ablation procedure. Ablation time was defined as the time spent delivering the RF energy to the myocardium and was automatically recorded by the electrophysiology (EP) recording system. Time needed to achieve accessory pathway conduction block after the start of RF energy delivery was recorded. Number of all ablations per procedure and type of ablation catheter used (irrigated tip versus non-irrigated tip) were recorded. Voltage amplitude of atrial and ventricular endocardial signals at the successful ablation site were also recorded.

Procedural success and adverse events

Procedural endpoint was defined as anterograde and retrograde block of accessory pathway conduction. All patients had 12 lead ECG recordings immediately after the procedure, the next day, 1 month after hospital discharge and at the scheduled outpatient clinic follow up approximately 6 months after the procedure. Additional 12 lead ECG recordings and 24 h Holter ECG recordings were obtained according to the discretion of the physician performing the follow up.
Peri-procedural adverse events resulting in prolonged or repeat hospitalisation, those requiring surgical intervention or resulting in patient’s long-term disability, were monitored within 30-days from the ablation procedure.

Results

The study population included 4 males (50%) and 4 females, aged 29 ± 21 years. Detailed baseline patient characteristics are shown in Table 1. 3D EAM system, ICE and long steerable sheath were used in all procedures. X-ray fluoroscopy was not used. In all cases bidirectional AP conduction was present.
Table 1
Baseline patient characteristics
Patient number
age range (years)
BMI
Previous procedures (number)
1
10–19
22,0
0
2
10–19
17,3
0
3
40–49
27,1
0
4
10–19
19,0
0
5
30–39
23,1
0
6
20–29
23,9
0
7
70–79
32,7
1
8
10–19
19,5
0
BMI body mass index
Right free wall was arbitrarily divided into four regions depending on the orientation of the loop and the tip of the ablation catheter as seen with the ICE (Fig. 1): 3 patients (37%) had anterior location of the accessory pathway with the loop of the ablation catheter oriented vertically and the tip being lateral from the location of the His bundle (Fig. 2); 1 patient (12%) had an anterolateral location of the accessory pathway with the loop and the tip of the ablation catheter oriented oblique toward the upper right part of the tricuspid annulus (Fig. 3); 3 patients (37%) had a lateral location of the accessory pathway with the loop and the tip being oriented horizontally (Fig. 4); 1 patient (12%) had a postero-lateral location of the accessory pathway with the loop and the tip of the ablation catheter oriented oblique toward the lower right part of the tricuspid annulus (Fig. 5).
One patient with the anterior AP had a failed cryoablation attempt during the same procedure. One patient with postero-lateral AP had a previous RF ablation procedure.
Ventricular insertion of the accessory pathway was mapped during sinus rhythm in 7 (87%) patients (Fig. 6, Panels C and D). Only in one patient mapping was performed during ongoing orthodromic AVRT, with annotation of the earliest far-field atrial electrogram recorded at the ventricular side of the tricuspid annulus (Fig. 6, Panels A and B).
Average procedural time was 90 ± 31 min. On average 6.6 ± 5.7 ablations were needed with an average ablation time of 236 ± 109 s. In 5 patients (62%) the first RF ablation was already successful in blocking the AP conduction. Average time to AP conduction block after start of the successful RF application was 4.8 ± 4.2 s. Average endocardial atrial signal amplitude at the successful ablation site was 0.16 ± 0.13 mV with an average atrial to ventricular voltage amplitude ratio of 0.15. Detailed procedural parameters are shown in Table 2.
Table 2
Procedural parameters
Patient number
Procedural duration (min)
AP location
I/NI RF or Cryo
TT (s)
Mapping in SR or AVRT
Number of RF lesions
Number of RF lesions before AP block
RF ablation time (s)
Atrial signal amplitude at successful ablation site (mV)
Atrial to ventricular signal ratio at successful ablation site
1
120
A
NI
10
SR
20
17
447
0.30
0.3
2
80
A
NI
2
SR
3
0
140
0.37
0.3
3
115
L
NI
1
SR
9
0
283
0.05
0.01
4
65
AL
I
4
AVRT
4
0
203
0.08
0.08
5
145
A
Cryo failed, I
2
SR
4
0
145
0.15
0.06
6
60
L
I
4
SR
4
2
193
0.09
0.17
7
65
PL
I
3
SR
3
0
146
0.06
0.17
8
75
L
NI
13
SR
6
1
331
0.10
0.1
AP accessory pathway, I/NI irrigated or non-irrigated tip ablation catheter, RF radio-frequency, Cryo cryoablation, TT time to termination of AP conduction after start of ablation, SR sinus rhythm, AVRT atrio-ventricular reentry tachycardia
Acute endpoint was reached in all procedures and there were no complications. During the average follow-up of 397 ± 363 days all patients (100%) were free of preexcitation and had no clinical tachycardias.

Discussion

Our novel ICE-guided technique for CA of the right free wall APs showed promising acute and long-term results. In addition, there were no peri-procedural adverse events.
In general, catheter instability is the major reason for unsuccessful CA of the right free wall accessory pathways. Catheter instability at the tricuspid annulus produces difficulties in mapping the atrial or ventricular insertion of the accessory pathway and failure to apply adequate RF current [15]. To some extent, when using transfemoral approach, catheter instability can be overcome with mapping the atrial insertion during ongoing orthodromic AVRT or ventricular pacing using the 3D EAM or a linear multipolar catheter positioned around the tricuspid annulus [68]. According to recent reports, the fluoroscopy-guided transjugular approach can help with stabilization of the ablation catheter tip in case of right anteroseptal, anterior and anterolateral accessory pathways, resulting in high procedural success rates and no recurrences [16, 17]. In our study we introduce a novel “loop” manoeuvre, which enabled transfemoral mapping of the ventricular aspect of the tricuspid annulus, which was mainly performed during sinus rhythm (ante-grade activation of the accessory pathway). In one patient we were able to map the atrial activation at the ventricular side of the annulus during ongoing orthodromic AVRT. However, atrial activation mapping from the ventricular side of the annulus can be less accurate due to the low amplitude of far-field atrial signals. Clearly, the near-field and large amplitude ventricular signals are more appealing to map and are the preferred target when mapping the ventricular aspect of the tricuspid valve annulus.
Steerable long sheath and ablation catheter placement was performed mainly under ICE control, which enabled accurate positioning of the ablation catheter tip to the target area beneath the tricuspid valve. This approach showed to be very useful in a zero-fluoroscopy setting since relying mostly on 3D EAM system and endocardial electrograms could prove misleading due to possible low voltage atrial signals at the ventricular aspect of the annulus. However, 3D EAM system seems to be an indispensable mapping tool, since ICE enables the operator to visualise detailed anatomy, but cannot show overall anatomy and activation pattern during catheter ablation of the right free wall APs.
Although the sub-annular region of the tricuspid valve is regarded as a possible low flow area, the use of non-irrigated RF ablation catheter proved satisfactory. There were no issues with temperature surge and power delivery in the four procedures that were performed with the non-irrigated ablation catheter.
Similar fluoroscopy-guided technique of sub-valvular mapping and ablation was recently described [9, 10]. Both studies included a small number of patients with right free wall AP. In the study by Wieczorek et al. there were 7 patients with right free wall AP. Steerable long sheath was only used when deemed necessary and AP block was achieved in approximately 6 s of RF ablation with average procedural time of 63 min and 18 min of fluoroscopy. In the second study by Yang et al., the 3D EAM system was used in only one of 7 patients with right free wall AP. They were able to reach AP block on average in approximately 6 s of RF ablation. Procedural time was 86 min and fluoroscopy time was 9 min. Compared to both studies we were able to reach slightly shorter RF ablation times, which could be the result of ICE-guided visual control of the catheter tip orientation, location and contact with the sub-valvular region at the ventricular aspect of the tricuspid valve. On the other hand, slightly longer duration of procedures in our study could be the result of more time-consuming catheter positioning using the 3D EAM and ICE.
There are some obvious limitations of our study. First, a small number of patients limits the value of our conclusions. Second, mapping and ablation was mostly targeting the ventricular insertion of the right free wall APs and our findings cannot be translated to atrial insertion mapping and ablation. Third, mapping was mostly performed during sinus rhythm and in patients with bidirectional AP conduction. Finally, our mapping approach might not be adequate in patients with only retrograde AP conduction.

Conclusions

The novel ICE-guided approach with concomitant use of the steerable sheath and the 3D EAM system for zero-fluoroscopy mapping and ablation of the right free wall APs proved feasible and resulted in excellent acute and long-term outcomes.

Acknowledgments

Not applicable.
The study complies with the Declaration of Helsinki. The study protocol was approved by The National Medical Ethics Committee (Komisija za medicinsko etiko Republike Slovenije).
All patients (or their parents in case of paediatric patients) gave written informed consent before ablation procedures were performed.
Not applicable.

Competing interests

Matevž Jan received honoraria from Biosense Webster. Other authors report no competing interests.
Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://​creativecommons.​org/​licenses/​by/​4.​0/​. The Creative Commons Public Domain Dedication waiver (http://​creativecommons.​org/​publicdomain/​zero/​1.​0/​) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Literatur
1.
Zurück zum Zitat Jackman WM, Wang XZ, Friday KJ, Roman CA, Moulton KP, Beckman KJ, McClelland JH, Twidale N, Hazlitt HA, Prior MI, et al. Catheter ablation of accessory atrioventricular pathways (Wolff-Parkinson-white syndrome) by radiofrequency current. N Engl J Med. 1991;324(23):1605–11.CrossRef Jackman WM, Wang XZ, Friday KJ, Roman CA, Moulton KP, Beckman KJ, McClelland JH, Twidale N, Hazlitt HA, Prior MI, et al. Catheter ablation of accessory atrioventricular pathways (Wolff-Parkinson-white syndrome) by radiofrequency current. N Engl J Med. 1991;324(23):1605–11.CrossRef
2.
Zurück zum Zitat Xie B, Heald SC, Camm AJ, Rowland E, Ward DE. Radiofrequency catheter ablation of accessory atrioventricular pathways: primary failure and recurrence of conduction. Heart. 1997;77(4):363–8.CrossRef Xie B, Heald SC, Camm AJ, Rowland E, Ward DE. Radiofrequency catheter ablation of accessory atrioventricular pathways: primary failure and recurrence of conduction. Heart. 1997;77(4):363–8.CrossRef
3.
Zurück zum Zitat Langberg JJ, Calkins H, Kim YN, Sousa J, el -Atassi R, Leon A, Borganelli M, Kalbfleisch SJ, Morady F. Recurrence of conduction in accessory atrioventricular connections after initially successful radiofrequency catheter ablation. J Am Coll Cardiol. 1992;19(7):1588–92.CrossRef Langberg JJ, Calkins H, Kim YN, Sousa J, el -Atassi R, Leon A, Borganelli M, Kalbfleisch SJ, Morady F. Recurrence of conduction in accessory atrioventricular connections after initially successful radiofrequency catheter ablation. J Am Coll Cardiol. 1992;19(7):1588–92.CrossRef
4.
Zurück zum Zitat Twidale N, Wang XZ, Beckman KJ, McClelland JH, Moulton KP, Prior MI, Hazlitt HA, Lazzara R, Jackman WM. Factors associated with recurrence of accessory pathway conduction after radiofrequency catheter ablation. Pacing Clin Electrophysiol. 1991;14(11 Pt 2):2042–8.CrossRef Twidale N, Wang XZ, Beckman KJ, McClelland JH, Moulton KP, Prior MI, Hazlitt HA, Lazzara R, Jackman WM. Factors associated with recurrence of accessory pathway conduction after radiofrequency catheter ablation. Pacing Clin Electrophysiol. 1991;14(11 Pt 2):2042–8.CrossRef
5.
Zurück zum Zitat Chen TH, Tsai ML, Chang PC, Wo HT, Chou CC, Wen MS, Wang CC, Yeh SJ, Wu D. Risk factors of recurrence and complication in radiofrequency catheter ablation of atrioventricular reentrant tachycardia in children and adolescents. Cardiol Young. 2013;23(5):682–91.CrossRef Chen TH, Tsai ML, Chang PC, Wo HT, Chou CC, Wen MS, Wang CC, Yeh SJ, Wu D. Risk factors of recurrence and complication in radiofrequency catheter ablation of atrioventricular reentrant tachycardia in children and adolescents. Cardiol Young. 2013;23(5):682–91.CrossRef
6.
Zurück zum Zitat Wong T, Hussain W, Markides V, Gorog DA, Wright I, Peters NS, Davies DW. Ablation of difficult right-sided accessory pathways aided by mapping of tricuspid annular activation using a halo catheter : halo-mapping of right sided accessory pathways. J Interv Card Electrophysiol. 2006;16(3):175–82.CrossRef Wong T, Hussain W, Markides V, Gorog DA, Wright I, Peters NS, Davies DW. Ablation of difficult right-sided accessory pathways aided by mapping of tricuspid annular activation using a halo catheter : halo-mapping of right sided accessory pathways. J Interv Card Electrophysiol. 2006;16(3):175–82.CrossRef
7.
Zurück zum Zitat Chen ML, Yang B, Ju WZ, Chen HW, Chen C, Hou XF, Zhang F, Cui JY, Zhang CH, Shan QJ, Zou JG, Sun JP, Xi YT, Cheng J, Cao KJ. Right-sided free wall accessory pathway refractory to conventional catheter ablation: lessons from 3-dimensional electroanatomic mapping. J Cardiovasc Electrophysiol. 2010;21(12):1317–24.CrossRef Chen ML, Yang B, Ju WZ, Chen HW, Chen C, Hou XF, Zhang F, Cui JY, Zhang CH, Shan QJ, Zou JG, Sun JP, Xi YT, Cheng J, Cao KJ. Right-sided free wall accessory pathway refractory to conventional catheter ablation: lessons from 3-dimensional electroanatomic mapping. J Cardiovasc Electrophysiol. 2010;21(12):1317–24.CrossRef
8.
Zurück zum Zitat Long DY, Dong JZ, Liu XP, Tang RB, Ning M, Gao LY, Yu RH, Fang DP, Jiang CX, Yuan YQ, Sang CH, Yin XD, Chen G, Zhang XY, Liang C, Ma CS. Ablation of right-sided accessory pathways with atrial insertion far from the tricuspid annulus using an electroanatomical mapping system. J Cardiovasc Electrophysiol. 2011;22(5):499–505.CrossRef Long DY, Dong JZ, Liu XP, Tang RB, Ning M, Gao LY, Yu RH, Fang DP, Jiang CX, Yuan YQ, Sang CH, Yin XD, Chen G, Zhang XY, Liang C, Ma CS. Ablation of right-sided accessory pathways with atrial insertion far from the tricuspid annulus using an electroanatomical mapping system. J Cardiovasc Electrophysiol. 2011;22(5):499–505.CrossRef
9.
Zurück zum Zitat Wieczorek M, Tajtaraghi S, Hoeltgen R. A subvalvular catheter approach for radiofrequency ablation of right-sided accessory pathways. J Interv Card Electrophysiol. 2019;56(1):29–36.CrossRef Wieczorek M, Tajtaraghi S, Hoeltgen R. A subvalvular catheter approach for radiofrequency ablation of right-sided accessory pathways. J Interv Card Electrophysiol. 2019;56(1):29–36.CrossRef
10.
Zurück zum Zitat Yang J, Yang G, Chen H, Ju W, Yang B, Zhang F, Jin Y, Cao J, Qian L, Chen M. An alternative under-valve approach to ablate right-sided accessory pathways. Heart Rhythm. 2019;16(1):51–6.CrossRef Yang J, Yang G, Chen H, Ju W, Yang B, Zhang F, Jin Y, Cao J, Qian L, Chen M. An alternative under-valve approach to ablate right-sided accessory pathways. Heart Rhythm. 2019;16(1):51–6.CrossRef
11.
Zurück zum Zitat Jan M, Žižek D, Rupar K, Mazić U, Kuhelj D, Lakič N, Geršak B. Fluoroless catheter ablation of various right and left sided supra-ventricular tachycardias in children and adolescents. Int J Cardiovasc Imaging. 2016;32(11):1609–16.CrossRef Jan M, Žižek D, Rupar K, Mazić U, Kuhelj D, Lakič N, Geršak B. Fluoroless catheter ablation of various right and left sided supra-ventricular tachycardias in children and adolescents. Int J Cardiovasc Imaging. 2016;32(11):1609–16.CrossRef
12.
Zurück zum Zitat Razminia M, Willoughby MC, Demo H, Keshmiri H, Wang T, D'Silva OJ, ZheutlinTA JH, Okhumale P, Kehoe RF. Fluoroless catheter ablation of cardiac a rhythmias: a 5-year experience. Pacing Clin Electrophysiol. 2017;40(4):425–33.CrossRef Razminia M, Willoughby MC, Demo H, Keshmiri H, Wang T, D'Silva OJ, ZheutlinTA JH, Okhumale P, Kehoe RF. Fluoroless catheter ablation of cardiac a rhythmias: a 5-year experience. Pacing Clin Electrophysiol. 2017;40(4):425–33.CrossRef
13.
Zurück zum Zitat Sánchez JM, Yanics MA, Wilson P, Doshi A, Kurian T, Pieper S. Fluoroless catheter ablation in adults: a single center experience. J Interv Card Electrophysiol. 2016;45(2):199–207.CrossRef Sánchez JM, Yanics MA, Wilson P, Doshi A, Kurian T, Pieper S. Fluoroless catheter ablation in adults: a single center experience. J Interv Card Electrophysiol. 2016;45(2):199–207.CrossRef
14.
Zurück zum Zitat Reddy VY, Morales G, Ahmed H, Neuzil P, Dukkipati S, Kim S, Clemens J, D’Avila A. Catheter ablation of atrial fibrillation without the use of fluoroscopy. Heart Rhythm. 2010;7(11):1644–53.CrossRef Reddy VY, Morales G, Ahmed H, Neuzil P, Dukkipati S, Kim S, Clemens J, D’Avila A. Catheter ablation of atrial fibrillation without the use of fluoroscopy. Heart Rhythm. 2010;7(11):1644–53.CrossRef
15.
Zurück zum Zitat Morady F, Strickberger A, Man KC, Daoud E, Niebauer M, Goyal R, Harvey M, Bogun F. Reasons for prolonged or failed attempts at radiofrequency catheter ablation of accessory pathways. J Am Coll Cardiol. 1996;27(3):683–9.CrossRef Morady F, Strickberger A, Man KC, Daoud E, Niebauer M, Goyal R, Harvey M, Bogun F. Reasons for prolonged or failed attempts at radiofrequency catheter ablation of accessory pathways. J Am Coll Cardiol. 1996;27(3):683–9.CrossRef
16.
Zurück zum Zitat DiLorenzo MP, Pass RH, Nappo L, Ceresnak SR. Ablating the anteroseptal accessory pathway-ablation via the right internal jugular vein may improve safety and efficacy. J Interv Card Electrophysiol. 2012;35(3):293–9.CrossRef DiLorenzo MP, Pass RH, Nappo L, Ceresnak SR. Ablating the anteroseptal accessory pathway-ablation via the right internal jugular vein may improve safety and efficacy. J Interv Card Electrophysiol. 2012;35(3):293–9.CrossRef
17.
Zurück zum Zitat Ergül Y, Özgür S, Şahin GT, Kafalı HC, Çelebi SB, Bay B, Güzeltaş A. The Transjugular approach: an alternative route to improve ablation success in right anteriorly and Anterolaterally-located supraventricular tachycardia substrates in children. Pediatr Cardiol. 2019+;40(3):477–82.CrossRef Ergül Y, Özgür S, Şahin GT, Kafalı HC, Çelebi SB, Bay B, Güzeltaş A. The Transjugular approach: an alternative route to improve ablation success in right anteriorly and Anterolaterally-located supraventricular tachycardia substrates in children. Pediatr Cardiol. 2019+;40(3):477–82.CrossRef
Metadaten
Titel
Intra-cardiac ultrasound guided approach for catheter ablation of typical right free wall accessory pathways
verfasst von
Matevž Jan
Tine Prolič Kalinšek
Jernej Štublar
Matija Jelenc
Andrej Pernat
David Žižek
Nikola Lakič
Publikationsdatum
01.12.2020
Verlag
BioMed Central
Erschienen in
BMC Cardiovascular Disorders / Ausgabe 1/2020
Elektronische ISSN: 1471-2261
DOI
https://doi.org/10.1186/s12872-020-01494-1

Weitere Artikel der Ausgabe 1/2020

BMC Cardiovascular Disorders 1/2020 Zur Ausgabe

Update Kardiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.