To assess the intraday repeatability of macular architecture measurements in glaucomatous and non-glaucomatous patients using spectral-domain optical coherence tomography (SD-OCT) and to evaluate the independence from intraindividual intraocular pressure (IOP) fluctuations.
Methods
In this single-center, time-point comparison study, 88 eyes with glaucoma, 53 eyes with ocular hypertension (OHT), and 253 healthy eyes underwent two standardized SD-OCT and intraocular pressure (IOP) measurements on the same day with a 5-h time gap. Bland–Altman plots, intraclass correlation coefficients (ICC), and random-effects model were used to analyze repeatability of entire retinal thickness, retinal nerve fiber layer, ganglion cell layer, inner plexiform layer, and inner nuclear layer measurements.
Results
Intraday measurements were highly reproducible in all 3 groups. ICC were greater than 0.90, respectively. The pairwise comparisons of morphometric parameters showed a statistically significant difference (P < 0.001, respectively) between groups (glaucoma vs. control, glaucoma vs. OHT) and a significant influence of time points. No correlation was found between IOP fluctuations and morphometric parameters (P > 0.05, respectively), except for a weak positive correlation with GCL (rho = 0.109, P = 0.031).
Conclusions
The evaluation of macular morphometric parameters of SD-OCT showed a high intraday repeatability and an excellent degree of agreement in glaucoma, ocular hypertension, and healthy groups. The fixed effects of time points were statistically significant. Except for a weak positive correlation of ganglion cell layer, variability did not appear to be affected by intraday IOP changes. Additional research is required to fully understand the impact of IOP fluctuations on macular morphometric parameters, considering the small observed IOP changes.
Alexander C. Rokohl and Ludwig M. Heindl contributed equally to this work and share the senior authorship.
Adam Kopecky was affiliated as a visiting and associated researcher at the University of Cologne (regular visits between 2020 and 2021) during this research and he is currently affiliated with the Czech Republic.
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Glaucoma is a heterogeneous group of optic nerve diseases and leads to characteristic morphological changes of the optic nerve (increasing optic disc excavation), loss of retinal ganglion cells, and thinning of the retinal nerve fiber layer [1‐3]. The specific visual field changes are also the main features of this chronic progressive optic neuropathy. If left untreated, glaucoma leads to severe visual acuity loss even to permanent blindness [1, 4, 5].
Anzeige
Thanks to modern technological advances in the last decades, spectral-domain optical coherence tomography (SD-OCT) has become a significant and established examination technique in ophthalmology and the diagnosis of glaucoma in particular [6‐8]. SD-OCT measurements in glaucoma are based on Bruch's membrane opening and include key quantitative measurements such as an analysis of BMO-minimal rim width, circumpapillary retinal nerve fiber layer thickness, and macular architecture in the posterior pole [7, 9]. Compared with confocal scanning laser ophthalmoscopy (cSLO), which was the gold standard until recently, SD-OCT is a more reliable method for morphometric examination of the optic disc [6, 10‐16].
In diagnosing glaucoma with SD-OCT, the most significant and frequently used parameter is the assessment of the circumpapillary retinal nerve fiber layer thickness of the optic nerve head [15, 17‐20]. However, assessment of macular architecture (especially thickness of retinal nerve fiber layer, ganglion cell layer, and inner plexiform layer) can serve as an important additional diagnostic tool in the early diagnosis of glaucoma [17, 21‐24]. With more advanced and highly reproducible segmentation algorithms of all retinal layers in the macular region, this type of analysis is of interest for clinical use in the detection of glaucomatous lesions [9, 17, 25‐27].
In addition, with an increasing number of follow-up SD-OCT examinations in glaucoma thanks to its ubiquitous use, the logical question arises to what extent intraindividual measurement variability may influence the results of comparative progression analyses [2, 6, 17]. Repeated precise measurements are essential for a correct glaucoma progression analysis of intraindividual disease dynamics [2, 7].
There is a hypothesis that variations in the anatomical structures of the optic nerve are possible due to intraday variations in intraocular pressure (IOP) and blood flow of the optic nerve [19, 28]. Such fluctuations in anatomical structures may also be characteristic of the retina in the macula region and hinder the accurate detection of glaucoma progression depending on the time of day when SD-OCT coherence tomography is performed [17].
Anzeige
Knowing the presence and quantitative values of intraday variations of macular architecture, the true glaucomatous progression and measurement variability of morphometric parameters can be distinguished. However, there is a lack of confirmed data on this to date. Therefore, the purposes of this study are to assess the intraday repeatability of macular architecture measurements in glaucomatous and non-glaucomatous patients using spectral-domain optical coherence tomography and to evaluate the independence from intraindividual intraocular pressure fluctuations.
Materials and methods
The present single-center, time-point comparison study was conducted between February 2020 and June 2021 in a German primary ophthalmology center. All procedures performed in this study were following the ethical standards of the institutional and/or national research committee and with the 1964 Helsinki declaration and its later amendments or comparable ethical standards. Approval for the study was obtained from the Ethics Commission of Cologne University’s Faculty of Medicine (19–1663). Written informed consent was obtained from all participants.
In this study, SD-OCT and Goldmann applanation tonometry were performed at two different time points with a time interval of 5 h on the same day (morning and afternoon). The following inclusion criteria were considered: the ability to give consent or consent by the legal representative, sufficient fluency in the German language to understand the contents of this study, and good cooperation of the patient. The following patients were excluded from this study: underage patients, pregnant patients, patients with a current or past history of systemic diseases with ocular involvement, retinal vascular diseases, visual field defects of other origins than glaucoma, high myopia greater than 6.00 dpt, pathological changes of the macular region or the optic nerve, severe opacities of the refractive media. Patients were classified into the three diagnostic groups according to the 5th edition of the European Glaucoma Society (EGS) guidelines: glaucoma group, ocular hypertension (OHT) group, and healthy patients without glaucoma (control group) [29].
Spectral Domain optical coherence tomography was performed according to the general guidelines using the Glaucoma program (scan size in the macular region 25 × 30°, central position of the pattern, number of scans of the region 61, distance between the regions 120 µm). The SD-OCT Spectralis® Glaucoma Module Premium Edition (Heidelberg Engineering GmbH, Heidelberg, Germany) was used with the device operating software version 6.13. The focus of the study was posterior pole analysis (8 × 8 grid with 64 squares pro layer) and measurement of the following morphometric parameters: entire retinal thickness (RT), the thickness of retinal nerve fiber layer (RNFL), ganglion cell layer (GCL), inner plexiform layer (IPL), and inner nuclear layer (INL). The minimal accepted quality of SD-OCT scans was 15 dB.
Statistical analyses
Patient data were pseudonymized, analyzed, and graphed using IBM SPSS Statistics version 27.0 (IBM Corp. Armonk, NY, USA). Dependencies between groups, time points and macular morphometric parameters were examined using random-effects models. Type C intraclass correlation coefficients were calculated for quantifying the agreement of morphometric parameters. The Bland–Altman graph was used for graphical assessment. Spearman's rank correlation analyses were used to test correlations between morphometric parameters and IOP. The threshold for statistical significance was P < 0.05.
Results
Demographics
In this study, 221 patients (442 eyes) were enrolled. Both eyes of the participants were included in the study. The insufficient image quality of SD-OCT scans (i.e., the quality of scans less than 15 dB) led to exclusion of 18 eyes from the study. In addition, 15 patients (30 eyes) dropped out since they did not participate in the second examination. Subsequently, the 394 eyes (204 patients) with a mean age of 58.14 ± 12.57 years were included in the statistical analyses (Table 1).
Table 1
Demographic and baseline data of the subjects
Groups
Glaucoma group
(N = 88)
OHT group
(N = 53)
Control group
(N = 253)
Gender, n (%)
men
39 (44.3%)
16 (30.2%)
91 (36.0%)
women
49 (55.7%)
37 (69.8%)
162 (64.0%)
Eye, n (%)
right
43 (48.9%)
28 (52.8%)
128 (50.6%)
left
45 (51.1%)
25 (47.2%)
125 (49.4%)
Age, y
Mean ± SD
64.39 ± 8.88
58.51 ± 11.73
55.89 ± 13.12
Median
65.50
59.00
59.00
Lens status, n (%)
clear lens
43 (48.9%)
37 (69.8%)
189 (74.7%)
cataract
35 (39.7%)
15 (28.3%)
51 (20.2%)
pseudophakia
10 (11.4%)
1 (1.9%)
13 (5.1%)
BCVA (LogMAR)
Mean ± SD
0.05 ± 0.09
0.02 ± 0.05
0.01 ± 0.05
Spherical equivalent, D
Mean ± SD
0.24 ± 1.39
0.17 ± 1.78
0.24 ± 1.69
Range
-4.37 – 2.50
-5.50 – 4.75
-4.50 – 4.38
IOP, mmHg, Mean ± SD
first examination
16.06 ± 3.90
21.92 ± 2.82
15.64 ± 2.85
second examination
15.31 ± 3.74
20.32 ± 3.14
14.73 ± 2.94
maximal IOP in medical history
26.36 ± 5.85
24.64 ± 2.26
17.00 ± 2.62
Mean Deviation, dB
Mean ± SD
-3.30 ± 4.05
-1.47 ± 2.11
-1.33 ± 1.95
OHT ocular hypertension, dB decibel, IOP intraocular pressure, D diopters, SD standard deviation
The first SD-OCT examinations were performed in the morning between 7:10 a.m. and 12:43 p.m. (9:40 ± 1:12), whereas the second SD-OCT examinations were performed between 1:04 p.m. and 5:15 p.m. (3:17 ± 0:51). The difference between the first and second examination averaged 5 h 36 min (336 ± 60 min). The mean value of the focus scan was 0.46 ± 1.64 diopters in the first SD-OCT examination and 0.45 ± 1.63 diopters in the second SD-OCT examination. Table 2 provides mean of macular morphometric parameters and their absolute difference in two time points.
Table 2
Mean of morphometric parameters and their absolute difference in two time points
There was a good agreement of the SD-OCT measurements in all 3 groups and the macular morphometric measurements were repeatable between examinations at two different time points during the same day. The mean of the difference of morphometric parameters in two time points was in the glaucoma group for RT = 0.57 ± 2.50 µm, RNFL = 0.05 ± 1.26 µm, GCL = 0.05 ± 1.62 µm, IPL = 0.08 ± 0.54 µm, and INL = -0.05 ± 0.81 µm, in the OHT group for RT = 0.83 ± 1.80 µm, RNFL = -0.20 ± 1.36 µm, GCL = 0.17 ± 0.51 µm, IPL = 0.24 ± 0.56 µm, and INL = 0.01 ± 0.57 µm, in the control group for RT = 0.64 ± 1.68 µm, RNFL = -0.01 ± 0.93 µm, GCL = 0.11 ± 0.51 µm, IPL = 0.21 ± 0.58 µm, and INL = -0.10 ± 0.53 µm and mostly distributed within standard deviations (± 1.96 SD). The mean values were in a very narrow range. There was no relationship of differences and decreasing or increasing means of morphometric parameters (Fig. 1).
Fig. 1
Bland-Altmann plot showing the intraday repeatability of two measurements (n = 88) of the entire retinal thickness in the glaucoma group
The 95% confidence intervals of ICC in the glaucoma group were from 0.988 to 0.995 for RT (P < 0.001), from 0.990 to 0.996 for RNFL (P < 0.001), from 0.990 to 0.996 for GCL (P < 0.001), from 0.979 to 0.991 for IPL (P < 0.001), and from 0.962 to 0.984 for INL (P < 0.001), in the OHT group from 0.990 to 0.998 for RT (P < 0.001), from 0.967 to 0.989 for RNFL (P < 0.001), from 0.974 to 0.992 for GCL (P < 0.001), from 0.933 to 0.983 for IPL (P < 0.001) and from 0.952 to 0.984 for INL (P < 0.001), in the control group from 0.990 to 0.996 for RT (P < 0.001), from 0.987 to 0.992 for RNFL (P < 0.001), from 0.985 to 0.992 for GCL (P < 0.001), from 0.962 to 0.983 for IPL (P < 0.001), from 0.970 to 0.982 for INL (P < 0.001) (Table 3).
Table 3
Type C intraclass correlation coefficients of the macular morphometric parameters between two intraday measurements
Retinal layers
Glaucoma group
OHT group
Control group
Entire retinal thickness
0.992 (0.988–0.995)
0.996 (0.990–0.998)
0.994 (0.990–0.996)
Retinal nerve fiber layer
0.993 (0.990–0.996)
0.981 (0.967–0.989)
0.990 (0.987–0.992)
Ganglion cell layer
0.994 (0.990–0.996)
0.986 (0.974–0.992)
0.989 (0.985–0.992)
Inner plexiform layer
0.986 (0.979–0.991)
0.967 (0.933–0.983)
0.975 (0.962–0.983)
Inner nuclear layer
0.975 (0.962–0.984)
0.972 (0.952–0.984)
0.976 (0.970–0.982)
The values of these parameters reflect mean and 95% confidence intervals. OHT—ocular hypertension
Random-effects model
The fixed effect of the group variable was statistically significant (P < 0.001) in three groups. The pairwise comparisons of morphometric parameters (RT, RNFL, GCL, IPL) showed a statistically significant difference between the glaucoma group and the OHT group (P < 0.001) and between the glaucoma group and the control group (P < 0.001) and no statistically significant difference between the OHT group and the control group. The fixed effects for time points (RT: η2 = 0.1055, P < 0.001; GCL: η2 = 0.0374, P < 0.001; IPL: η2 = 0.0944, P < 0.001; INL: η2 = 0.0141, P = 0.018) were statistically significant, except in the case of RNFL (η2 = 0.0006, P = 0.635).
The Spearman's rank coefficient of correlation
The mean IOP was 16.58 ± 3.76 mmHg (min = 8 mmHg, max = 29 mmHg) at the first examination and 15.61 ± 3.67 mmHg (min = 6 mmHg, max = 28 mmHg) at the second examination. The intraindividual absolute mean difference of IOP between the two measurements at two different time points was 1.82 ± 1.66 mmHg. The intraindividual absolute mean variability of morphometric parameters among patients in all three groups was 1.49 ± 1.35 µm for RT, 0.70 ± 0.81 µm for RNFL, 0.39 ± 0.38 µm for GCL, 0.37 ± 0.47 µm for IPL and 0.44 ± 0.43 µm for INL. No statistically significant correlation was found between absolute intraindividual variability of IOP and absolute difference of RT (rho = 0.088, P = 0.082), RNFL (rho = 0.087, P = 0.085), IPL (rho = -0.035, P = 0.488) and INL (rho = 0.040, P = 0.435). A statistically significant weak positive correlation was found between the mean absolute difference of IOP and the mean absolute difference of GCL (rho = 0.109, P = 0.031).
Discussion
Timely detection of the progression of optical neuropathy is an important aspect of glaucoma management since early recognition of changes in optic nerve morphology and macular architecture can help clinicians to adjust patient treatment and prevent permanent vision loss [30]. Accurate and reliable SD-OCT measurements are crucial for a correct analysis of the glaucoma progression and dynamics of the disease [2, 7]. Difficulties in right diagnosing of the glaucoma progression include not only the complexity of distinguishing true changes associated with the disease from normal aging processes of the retina but also the variability of SD-OCT measurements themselves [17, 31].
Anzeige
In this study, we could show that the macular morphometric parameters (RT, RNFL, GCL, IPL, INL) have a high degree of agreement between the measurements in the morning and the afternoon. According to Bland–Altman plots and intraclass correlation coefficients, it can be concluded that the variations of the morphometric parameters during the day are not clinically significant and that the measurement results in all three groups (glaucoma, ocular hypertension, and healthy groups) speak for the overall excellent repeatability.
Ctori et al. also reported good repeatability and reproducibility of the individual measurements of the retinal layer thickness in young healthy volunteers. In this study, two investigators independently performed two SD-OCT examinations for each participant in a single visit. In the literature, there are also works devoted to the assessment of the variability of morphometric parameters of the optic nerve head. Enders et al. reported high intraday repeatability after assessing the variability of global BMO-MRW, global RNFL thickness, and global BMO-MRA at two different time points with a time gap of more than 5 h [19].
On the other hand, the analysis using random-effects model indicated a statistically significant influence of time points on a measurement of macular architecture. This variability between intraday measurements could be explained by the inability of performing an absolutely identical SD-OCT examination in practice, due to the patient cooperation during the examination and the skills of the staff performing tomography. The fluctuating quality of SD-OCT images can play a crucial role in the variation in the thickness of the retina and its layers since the correct determination of the boundaries of these layers is done automatically by the software of the OCT device. The higher the quality of the images, the more correctly the segmentation algorithms of the software work and, consequently, the more accurately thickness of retinal layers is detected. Each performed examination should also be verified by a physician and manual corrections should be made to the automatic segmentation of the retinal layers, if necessary [7, 32]. As has already been confirmed, manual segmentation correction increases the diagnostic power of another SD-OCT morphometric parameter such as circumpapillary RNFL to detect glaucomatous changes of the retina [9, 33, 34].
The absolute mean variability of the RT, RNFL, IPL, and INL did not correlate in our study with the absolute IOP variations on the two SD-OCT examinations performed. The Spearman correlation of the mean absolute difference of the GCL and the mean absolute difference of the IOP has a value of rho = 0.109 (P = 0.031) and can be described as weak positive correlation, because it is in the range of 0.1 to 0.3 [35]. However, it cannot be claimed why there is a weak positive correlation only with ganglion cell layer. Possibly, this layer of the retina is more susceptible to IOP variations than the other retinal layers or the entire retina itself. Further studies are necessary to gain a comprehensive understanding of the potential impact of IOP changes on macular morphometric parameters, due to the small detected intraday IOP fluctuations. Enders et al. found no significant correlation between intraindividual variability in morphometric parameters of the optic nerve head (global BMO-MRW, global RNFL thickness, global BMO-MRA) and intraindividual IOP changes in two examinations performed on the same day [19].
Anzeige
Limitations of this work include the relatively small number of patients in the glaucoma group and the OHT group. In this study, only the inner retinal layers were examined because of the already proven changes that occur in them by the progression of glaucoma [23, 24]. To obtain a complete picture of the possible intraindividual variations of macular morphometric parameters during the day, it would be useful to study all retinal layers. The good documentation of the morphometric parameters of the retina thanks to high-resolution SD-OCT is an advantage of this work. The variation of retinal vascular characteristics of the macular region and its possible influence on intraindividual variation in macular morphometric measurements should be investigated with the help of optical coherence tomography angiography (OCT-A) in future studies.
In summary, the data of this study showed that the macular morphometric parameters of SD-OCT demonstrated excellent repeatability in two measurements performed in the morning and afternoon (time interval of 5 h 36 min). According to intraclass correlation coefficients, the degree of agreement of each morphometric parameter in three studied groups in two time points was very good. The time points had a significant influence on studied variables. The fluctuation of thickness of the RT RNFL, IPL, and INL does not seem to be affected by the moderate variations of IOP. There was a weak positive correlation between the mean absolute difference of IOP and the mean absolute difference of GCL.
Declarations
Ethical approval
All procedures performed in studies involving human participants were following the ethical standards of the institutional and/or national research committee and with the 1964 Helsinki declaration and its later amendments or comparable ethical standards. This article does not contain any studies with animals performed by any of the authors.
Approval for the study was obtained from the Ethics Commission of Cologne University’s Faculty of Medicine (19–1663). Written informed consent was obtained from all participants.
Conflict of interest
All authors declare that they have no financial or proprietary interest in any materials or methods mentioned in the manuscript.
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Unsere Produktempfehlungen
e.Med Interdisziplinär
Kombi-Abonnement
Für Ihren Erfolg in Klinik und Praxis - Die beste Hilfe in Ihrem Arbeitsalltag
Mit e.MedInterdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de.
Gardiner SK, Ren R, Yang H, Fortune B, Burgoyne CF, Demirel S (2014) A method to estimate the amount of neuroretinal rim tissue in glaucoma: comparison with current methods for measuring rim area. Am J Ophthalmol 157:540-549 e541-542. https://doi.org/10.1016/j.ajo.2013.11.007CrossRefPubMed
Pazos M, Dyrda AA, Biarnés M, Gómez A, Martín C, Mora C, Fatti G, Antón A (2017) Diagnostic accuracy of spectralis SD OCT automated macular layers segmentation to discriminate normal from early glaucomatous eyes. Ophthalmology 124:1218–1228. https://doi.org/10.1016/j.ophtha.2017.03.044CrossRefPubMed
10.
Reis AS, O’Leary N, Yang H, Sharpe GP, Nicolela MT, Burgoyne CF, Chauhan BC (2012) Influence of clinically invisible, but optical coherence tomography detected, optic disc margin anatomy on neuroretinal rim evaluation. Invest Ophthalmol Vis Sci 53:1852–1860. https://doi.org/10.1167/iovs.11-9309CrossRefPubMedPubMedCentral
11.
Enders P, Schaub F, Hermann MM, Cursiefen C, Heindl LM (2017) Neuroretinal rim in non-glaucomatous large optic nerve heads: a comparison of confocal scanning laser tomography and spectral domain optical coherence tomography. Br J Ophthalmol 101:138–142. https://doi.org/10.1136/bjophthalmol-2015-307730CrossRefPubMed
12.
Enders P, Schaub F, Adler W, Hermann MM, Dietlein TS, Cursiefen C, Heindl LM, Medscape (2018) Bruch’s membrane opening-based optical coherence tomography of the optic nerve head: a useful diagnostic tool to detect glaucoma in macrodiscs. Eye (Lond) 32:314–323. https://doi.org/10.1038/eye.2017.306CrossRefPubMed
13.
Enders P, Schaub F, Adler W, Nikoluk R, Hermann MM, Heindl LM (2017) The use of Bruch’s membrane opening-based optical coherence tomography of the optic nerve head for glaucoma detection in microdiscs. Br J Ophthalmol 101:530–535. https://doi.org/10.1136/bjophthalmol-2016-308957CrossRefPubMed
14.
Enders P, Adler W, Kiessling D, Weber V, Schaub F, Hermann MM, Dietlein T, Cursiefen C, Heindl LM (2019) Evaluation of two-dimensional Bruch’s membrane opening minimum rim area for glaucoma diagnostics in a large patient cohort. Acta Ophthalmol 97:60–67. https://doi.org/10.1111/aos.13698CrossRefPubMed
15.
Chauhan BC, O’Leary N, AlMobarak FA, Reis ASC, Yang H, Sharpe GP, Hutchison DM, Nicolela MT, Burgoyne CF (2013) Enhanced detection of open-angle glaucoma with an anatomically accurate optical coherence tomography-derived neuroretinal rim parameter. Ophthalmology 120:535–543. https://doi.org/10.1016/j.ophtha.2012.09.055CrossRefPubMed
16.
Toshev AP, Lamparter J, Pfeiffer N, Hoffmann EM (2017) Bruch’s membrane opening-minimum rim width assessment with spectral-domain optical coherence tomography performs better than confocal scanning laser ophthalmoscopy in discriminating early glaucoma patients from control subjects. J Glaucoma 26:27–33. https://doi.org/10.1097/IJG.0000000000000532CrossRefPubMed
Ajtony C, Balla Z, Somoskeoy S, Kovacs B (2007) Relationship between visual field sensitivity and retinal nerve fiber layer thickness as measured by optical coherence tomography. Invest Ophthalmol Vis Sci 48:258–263. https://doi.org/10.1167/iovs.06-0410CrossRefPubMed
19.
Enders P, Bremen A, Schaub F, Hermann MM, Diestelhorst M, Dietlein T, Cursiefen C, Heindl LM (2017) Intraday repeatability of Bruch’s membrane opening-based neuroretinal rim measurements. Invest Ophthalmol Vis Sci 58:5195–5200. https://doi.org/10.1167/iovs.17-22812CrossRefPubMed
20.
He L, Yang H, Gardiner SK, Williams G, Hardin C, Strouthidis NG, Fortune B, Burgoyne CF (2014) Longitudinal detection of optic nerve head changes by spectral domain optical coherence tomography in early experimental glaucoma. Invest Ophthalmol Vis Sci 55:574–586. https://doi.org/10.1167/iovs.13-13245CrossRefPubMedPubMedCentral
21.
Unterlauft JD, Theilig T, Hasan S, Bohm MR, Rauscher F (2020) Analysis of glaucomatous changes of the macula using optical coherence tomography. Klin Monbl Augenheilkd 237:185–191. https://doi.org/10.1055/a-0808-4807CrossRefPubMed
Takayama K, Hangai M, Durbin M, Nakano N, Morooka S, Akagi T, Ikeda HO, Yoshimura N (2012) A novel method to detect local ganglion cell loss in early glaucoma using spectral-domain optical coherence tomography. Invest Ophthalmol Vis Sci 53:6904–6913. https://doi.org/10.1167/iovs.12-10210CrossRefPubMed
Jeoung JW, Choi YJ, Park KH, Kim DM (2013) Macular ganglion cell imaging study: glaucoma diagnostic accuracy of spectral-domain optical coherence tomography. Invest Ophthalmol Vis Sci 54:4422–4429. https://doi.org/10.1167/iovs.12-11273CrossRefPubMed
28.
Gardiner SK, Demirel S, Reynaud J, Fortune B (2016) Changes in retinal nerve fiber layer reflectance intensity as a predictor of functional progression in glaucoma. Invest Ophthalmol Vis Sci 57:1221–1227. https://doi.org/10.1167/iovs.15-18788CrossRefPubMedPubMedCentral
Fingeret M, Suh MH, Hood DC, Ritch R (2018) Heidelberg Engineering GmbH. Glaucoma imaging atlas. A diagnostic imaging guide for assessment and management. buch.one - Offsetdruckerei Karl Grammlich GmbH, Pliezhausen
31.
Knight OJ, Girkin CA, Budenz DL, Durbin MK, Feuer WJ (2012) Effect of race, age, and axial length on optic nerve head parameters and retinal nerve fiber layer thickness measured by Cirrus HD-OCT. Arch Ophthalmol 130:312–318. https://doi.org/10.1001/archopthalmol.2011.1576CrossRefPubMedPubMedCentral
32.
Heindl LM, Adler W, El-Malahi O, Schaub F, Hermann MM, Dietlein TS, Cursiefen C, Enders P (2018) The optimal diameter for circumpapillary retinal nerve fiber layer thickness measurement by SD-OCT in glaucoma. J Glaucoma 27:1086–1093. https://doi.org/10.1097/IJG.0000000000001027CrossRefPubMed
33.
Rüfer F, Bartsch JJ, Erb C, Riehl A, Zeitz PF (2016) Epiretinal membrane as a source of errors during the measurement of peripapillary nerve fibre thickness using spectral-domain optical coherence tomography (SD-OCT). Graefes Arch Clin Exp Ophthalmol 254:2017–2023. https://doi.org/10.1007/s00417-016-3453-4CrossRefPubMed
34.
Mansberger SL, Menda SA, Fortune BA, Gardiner SK, Demirel S (2017) Automated segmentation errors when using optical coherence tomography to measure retinal nerve fiber layer thickness in glaucoma. Am J Ophthalmol 174:1–8. https://doi.org/10.1016/j.ajo.2016.10.020CrossRefPubMed
35.
Cohen J (1988) The signifiance of product moment r. In: Statistical power analysis for the behavioral sciences, 2nd edn. Lawrence Erlbaum Associates, Publishers, Hillsdale, pp 79–81
Die Auswertung von Gesichtsfeldbefunden ermöglicht in vielen Fällen eine genaue Topodiagnostik und kann auch häufig direkt zu einer konkreten Verdachtsdiagnose führen. Im klinischen Alltag ist dabei die 30° automatische Rasterperimetrie die …
Das Glaukom ist eine der Hauptursachen für Sehbehinderung und Blindheit in Deutschland und weltweit. Der mögliche Einfluss von Ernährung sowie Nahrungsergänzungsmitteln auf den Verlauf und die Therapie dieser Erkrankung werden im klinischen Alltag …
Das trockene Auge ist eine der häufigsten Oberflächenerkrankungen des menschlichen Auges. Mit einer geschätzten Prävalenz von 5–30 % bei Erwachsenen ab dem 50. Lebensjahr gilt das trockene Auge mittlerweile als „Volkskrankheit“ [ 4 ].
Die mit dem humanen Leukozytenantigen(HLA)-B27 assoziierte, akute anteriore Uveitis (AAU) ist die häufigste Form nichtinfektiöser intraokulärer Entzündungen und wird als eigene klinische Entität gesehen. Es sind überwiegend junge Erwachsene im …