Intraoperative panoramic image using alignment grid, is it accurate?
verfasst von:
T. Apivatthakakul, M. Duanghakrung, S. Luevitoonvechkit, S. Patumasutra
Erschienen in:
Archives of Orthopaedic and Trauma Surgery
|
Ausgabe 7/2013
Einloggen, um Zugang zu erhalten
Abstract
Background
Minimally invasive orthopedic trauma surgery relies heavily on intraoperative fluoroscopic images to evaluate the quality of fracture reduction and fixation. However, fluoroscopic images have a narrow field of view and often cannot visualize the entire long bone axis.
Objectives
To compare the coronal femoral alignment between conventional X-rays to that achieved with a new method of acquiring a panoramic intraoperative image.
Materials and methods
Twenty-four cadaveric femurs with simple diaphyseal fractures were fixed with an angulated broad DCP to create coronal plane malalignment. An intraoperative alignment grid was used to help stitch different fluoroscopic images together to produce a panoramic image. A conventional X-ray of the entire femur was then performed. The coronal plane angulation in the panoramic images was then compared to the conventional X-rays using a Wilcoxon signed rank test.
Results
The mean angle measured from the panoramic view was 173.9° (range 169.3°–178.0°) with median of 173.2°. The mean angle measured from the conventional X-ray was 173.4° (range 167.7°–178.7°) with a median angle of 173.5°. There was no significant difference between both methods of measurement (P = 0.48).
Conclusion
Panoramic images produced by stitching fluoroscopic images together with help of an alignment grid demonstrated the same accuracy at evaluating the coronal plane alignment of femur fractures as conventional X-rays.