Intraoperative panoramic image using alignment grid, is it accurate?

verfasst von: T. Apivatthakakul, M. Duanghakrung, S. Luevitoonvechkit, S. Patumasutra

Erschienen in: Archives of Orthopaedic and Trauma Surgery | Ausgabe 7/2013

Einloggen, um Zugang zu erhalten

Abstract

Background

Minimally invasive orthopedic trauma surgery relies heavily on intraoperative fluoroscopic images to evaluate the quality of fracture reduction and fixation. However, fluoroscopic images have a narrow field of view and often cannot visualize the entire long bone axis.

Objectives

To compare the coronal femoral alignment between conventional X-rays to that achieved with a new method of acquiring a panoramic intraoperative image.

Materials and methods

Twenty-four cadaveric femurs with simple diaphyseal fractures were fixed with an angulated broad DCP to create coronal plane malalignment. An intraoperative alignment grid was used to help stitch different fluoroscopic images together to produce a panoramic image. A conventional X-ray of the entire femur was then performed. The coronal plane angulation in the panoramic images was then compared to the conventional X-rays using a Wilcoxon signed rank test.

Results

The mean angle measured from the panoramic view was 173.9° (range 169.3°–178.0°) with median of 173.2°. The mean angle measured from the conventional X-ray was 173.4° (range 167.7°–178.7°) with a median angle of 173.5°. There was no significant difference between both methods of measurement (P = 0.48).

Conclusion

Panoramic images produced by stitching fluoroscopic images together with help of an alignment grid demonstrated the same accuracy at evaluating the coronal plane alignment of femur fractures as conventional X-rays.
Literatur
1.
Zurück zum Zitat Ricci WM, Bellabarba C, Lewis R, Evanoff B, Herscovici D, Dipasquale T, Sanders R (2001) Angular malalignment after intramedullary nailing of femoral shaft fractures. J Orthop Trauma 15(2):90–95PubMedCrossRef Ricci WM, Bellabarba C, Lewis R, Evanoff B, Herscovici D, Dipasquale T, Sanders R (2001) Angular malalignment after intramedullary nailing of femoral shaft fractures. J Orthop Trauma 15(2):90–95PubMedCrossRef
2.
Zurück zum Zitat Saleh M, Klein W, Harriman P (1991) An intraoperatively applicable roentgen grid for assessing alignment of the long bones. Zentralbl Chir 116(14):859–865PubMed Saleh M, Klein W, Harriman P (1991) An intraoperatively applicable roentgen grid for assessing alignment of the long bones. Zentralbl Chir 116(14):859–865PubMed
3.
Zurück zum Zitat Farouk O, Krettek C, Miclau T, Schandelmaier P, Tscherne H (1998) Effects of percutaneous and conventional plating techniques on the blood supply to the femur. Arch Orthop Trauma Surg 117(8):438–441PubMedCrossRef Farouk O, Krettek C, Miclau T, Schandelmaier P, Tscherne H (1998) Effects of percutaneous and conventional plating techniques on the blood supply to the femur. Arch Orthop Trauma Surg 117(8):438–441PubMedCrossRef
4.
Zurück zum Zitat Liodakis E, Kenawey M, Liodaki E, Mommsen P, Krettek C, Hankemeier S (2010) The axis-board: an alternative to the cable technique for intraoperative assessment of lower limb alignment. Technol Health Care 18(3):165–171PubMed Liodakis E, Kenawey M, Liodaki E, Mommsen P, Krettek C, Hankemeier S (2010) The axis-board: an alternative to the cable technique for intraoperative assessment of lower limb alignment. Technol Health Care 18(3):165–171PubMed
5.
Zurück zum Zitat Yaniv Z, Joskowicz L (2004) Long bone panoramas from fluoroscopic X-ray images. IEEE Trans Med Imaging 23(1):26–35PubMedCrossRef Yaniv Z, Joskowicz L (2004) Long bone panoramas from fluoroscopic X-ray images. IEEE Trans Med Imaging 23(1):26–35PubMedCrossRef
6.
Zurück zum Zitat Messmer P, Matthews F, Wullschleger C, Hngli R, Regazzoni P, Jacob A (2006) Image fusion for intraoperative control of axis in long bone fracture treatment. Eur J Trauma 32(6):555–561CrossRef Messmer P, Matthews F, Wullschleger C, Hngli R, Regazzoni P, Jacob A (2006) Image fusion for intraoperative control of axis in long bone fracture treatment. Eur J Trauma 32(6):555–561CrossRef
7.
Zurück zum Zitat Wang L, Traub J, Weidert S, Heining SM, Euler E, Navab N (2010) Parallax-free intra-operative X-ray image stitching. Med Image Anal 14(5):674–686PubMedCrossRef Wang L, Traub J, Weidert S, Heining SM, Euler E, Navab N (2010) Parallax-free intra-operative X-ray image stitching. Med Image Anal 14(5):674–686PubMedCrossRef
8.
Zurück zum Zitat Zlowodzki M, Vogt D, Cole PA, Kregor PJ (2007) Plating of femoral shaft fractures: open reduction and internal fixation versus submuscular fixation. J Trauma 63(5):1061–1065PubMedCrossRef Zlowodzki M, Vogt D, Cole PA, Kregor PJ (2007) Plating of femoral shaft fractures: open reduction and internal fixation versus submuscular fixation. J Trauma 63(5):1061–1065PubMedCrossRef
9.
Zurück zum Zitat Apivatthakakul T, Chiewcharntanakit S (2009) Minimally invasive plate osteosynthesis (MIPO) in the treatment of the femoral shaft fracture where intramedullary nailing is not indicated. Int Orthop 33(4):1119–1126PubMedCrossRef Apivatthakakul T, Chiewcharntanakit S (2009) Minimally invasive plate osteosynthesis (MIPO) in the treatment of the femoral shaft fracture where intramedullary nailing is not indicated. Int Orthop 33(4):1119–1126PubMedCrossRef
10.
Zurück zum Zitat Dewaele P, Vuylsteke P, Van de Velde S, Schoeters E (1999) Full-leg/full-spine image stitching, a new and accurate CR-based imaging technique. Proceedings of SPIE—the international society for optical engineering 3661(1):131–138 Dewaele P, Vuylsteke P, Van de Velde S, Schoeters E (1999) Full-leg/full-spine image stitching, a new and accurate CR-based imaging technique. Proceedings of SPIE—the international society for optical engineering 3661(1):131–138
11.
Zurück zum Zitat Geijer H, Beckman K, Jonsson B, Andersson T, Persliden J (2001) Digital radiography of scoliosis with a scanning method: initial evaluation. Radiology 218(2):402–410PubMed Geijer H, Beckman K, Jonsson B, Andersson T, Persliden J (2001) Digital radiography of scoliosis with a scanning method: initial evaluation. Radiology 218(2):402–410PubMed
Metadaten
Titel
Intraoperative panoramic image using alignment grid, is it accurate?
verfasst von
T. Apivatthakakul
M. Duanghakrung
S. Luevitoonvechkit
S. Patumasutra
Publikationsdatum
01.07.2013
Verlag
Springer-Verlag
Erschienen in
Archives of Orthopaedic and Trauma Surgery / Ausgabe 7/2013
Print ISSN: 0936-8051
Elektronische ISSN: 1434-3916
DOI
https://doi.org/10.1007/s00402-013-1765-y

Weitere Artikel der Ausgabe 7/2013

Archives of Orthopaedic and Trauma Surgery 7/2013 Zur Ausgabe

Arthropedia

Grundlagenwissen der Arthroskopie und Gelenkchirurgie. Erweitert durch Fallbeispiele, Videos und Abbildungen. 
» Jetzt entdecken

Update Orthopädie und Unfallchirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.