LI-RADS technical requirements for CT, MRI, and contrast-enhanced ultrasound
- 22.09.2017
- Verfasst von
- Avinash R. Kambadakone
- Alice Fung
- Rajan T. Gupta
- Thomas A. Hope
- Kathryn J. Fowler
- Andrej Lyshchik
- Karthik Ganesan
- Vahid Yaghmai
- Alexander R. Guimaraes
- Dushyant V. Sahani
- Frank H. Miller
- Erschienen in
- Abdominal Radiology | Ausgabe 1/2018
Abstract
Accurate detection and characterization of liver observations to enable HCC diagnosis and staging using LI-RADS requires a technically adequate imaging exam. To help achieve this objective, LI-RADS has proposed technical requirements for CT, MR, and contrast-enhanced ultrasound of liver. This article reviews the technical requirements for liver imaging, including the description of minimum acceptable technical standards, such as the scanner hardware requirements, recommended dynamic imaging phases, and common technical challenges of liver imaging.
Anzeige
- Titel
- LI-RADS technical requirements for CT, MRI, and contrast-enhanced ultrasound
- Verfasst von
-
Avinash R. Kambadakone
Alice Fung
Rajan T. Gupta
Thomas A. Hope
Kathryn J. Fowler
Andrej Lyshchik
Karthik Ganesan
Vahid Yaghmai
Alexander R. Guimaraes
Dushyant V. Sahani
Frank H. Miller
- Publikationsdatum
- 22.09.2017
- Verlag
- Springer US
- Erschienen in
-
Abdominal Radiology / Ausgabe 1/2018
Print ISSN: 2366-004X
Elektronische ISSN: 2366-0058 - DOI
- https://doi.org/10.1007/s00261-017-1325-y
Dieser Inhalt ist nur sichtbar, wenn du eingeloggt bist und die entsprechende Berechtigung hast.