Abstract
Image quality in single photon emission computed tomography (SPECT) using a rotating gamma camera is dependent on the time course of the tracer in the field of view. If acquisition times are slow comapred to the tracer turnover, artifacts may occur in the reconstructed images. The properties of such artifacts were studied by computer simulation. Experimental projection data of point sources, cylindrical phantoms, and an anatomically realistic brain phantom were altered by sequentially weighting the projections with a function that varied exponentially or linearly with time. The observed distortion in the reconstructed images could be related to the ratio between the object activity variation and the camera rotation time. If the tracer concentration changed less than a factor of two during one camera rotation then little image distortion was visible although quantitatively the resolution was degraded.
If the object's activity variation with time is fast enough to produce noticeable distortion, the artifacts can be reduced by performing multiple, rapid, camera rotations, instead of one rotation, for the same total acquisition time. The proposed procedure is computer storage space intensive and takes longer to produce the transaxial images, but improves image quality.
Similar content being viewed by others
References
Brooks RA, Di Chiro G (1976) Principles of computer assisted tomography (CAT) in radiographic and radioisotope imaging. Phys Med Biol 21:689–732
Budinger TF, Gullberg GT, Huesman RH (1979) In: Herrmann GT (ed) Emission computed tomography—Reconstruction from projections, implementation and applications. Springer, Heidelberg New York, pp 147–246
Holden JE, Ip WR (1978) continuous time-dependence in computed tomography. Med Phys 5:4850–490
Ip WR, Holden JE, Winkler SS (1983) A study of the image discrepancies due to object time dependance in transmission and emission tomography. Phys Med Biol 28:952–953
Shepp LA, Vardi Y (1982) Maximum likelihood reconstruction for emission tomography. IEEE Trans Med Imaging MI 1:113–122
Wong DF, Links JM, Molliver ME, Hengst TC, Clifford CM, Buhle L, Bryan M, Stump M, Wagner HN Jr (1984) An anatomically realistic brain phantom for quantification with positron tomography. J Nucl Med 25:108
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Bok, B.D., Bice, A.N., Clausen, M. et al. Artifacts in camera based single photon emission tomography due to time activity variation. Eur J Nucl Med 13, 439–442 (1987). https://doi.org/10.1007/BF00281856
Received:
Accepted:
Issue Date:
DOI: https://doi.org/10.1007/BF00281856