Accuracy of invisalign® treatments in the anterior tooth region

First results

verfasst von: Dr. E. Krieger, J. Seiferth, I. Saric, B.A. Jung, H. Wehrbein

Erschienen in: Journal of Orofacial Orthopedics / Fortschritte der Kieferorthopädie | Ausgabe 2/2011

Einloggen, um Zugang zu erhalten

Abstract

Aim

To analyse (a) to what extent the pretreatment model at the beginning of the treatment corresponds to the initial position in the ClinCheck® and (b) to what extent the predicted treatment result corresponds to the actual result of the therapy at the end of the treatment.

Material and methods

Pre- and posttreatment models as well as the initial and final position of the ClinCheck with a total of 35 patients aged between 15 and 59 were measured; all of whom were treated by using the Invisalign® technology (Invisalign®, Align Technology, S.C., Calif., USA). The measurement of the initial and final models was conducted by using an electronic digital calliper rule, i.e. that of ClinCheck® using the measurement tool ToothMeasure® of Invisalign® Software. The following parameters in the anterior region were measured: Overjet, Overbite, dental midline shift.

Results

Pretreament models and the initial ClinChecks® revealed slight deviations in the parameters overjet 0.08 mm (standard deviation (SD) 0.3), overbite 0.3 mm (SD 0.4) and dental midline deviation 0.1mm (SD 0.4). The final model and the final ClinCheck® revealed larger deviations: the differences for the Overjet were on average 0.4 mm (SD 0.7), Overbite 0.9 mm (SD 0.9) and dental midline shift 0.4 mm (SD 0.5).

Conclusions

The IT-based transmission of mal-aligned teeth into the ClinCheck® presentation provides sufficiently good accuracy. Tooth corrections in the vertical plane were more difficult to realize. A vertical overcorrection in the final ClinCheck®, a case refinement at the end of the treatment or additional measures (e.g. horizontal beveled attachments or vertical elastics) seems useful to achieve the individually intended therapeutic goal.
Literatur
1.
Zurück zum Zitat Align Technology. The Invisalign® reference guide. http://passthrough.fw-notify.net/download/620383/http://www.aligntechinstitute.com/files/pdf/foreign/RefGuide_G.pdf Align Technology. The Invisalign® reference guide. http://​passthrough.​fw-notify.​net/​download/​620383/​http://www.aligntechinstitute.com/files/pdf/foreign/RefGuide_G.pdf
2.
Zurück zum Zitat Align Technology. Attachment protocol summary. http://www. aligntechinstitute.com/GetHelp/Documents/pdf Align Technology. Attachment protocol summary. http://​www.​ aligntechinstitute.com/GetHelp/Documents/pdf
3.
Zurück zum Zitat Bollen AM, Huang G, King G et al (2003) Activation time and material stiffness of sequential removable orthodontic appliances. Part 1: Ability to complete treatment. Am J Orthod Dentofacial Orthop 124:496–501PubMedCrossRef Bollen AM, Huang G, King G et al (2003) Activation time and material stiffness of sequential removable orthodontic appliances. Part 1: Ability to complete treatment. Am J Orthod Dentofacial Orthop 124:496–501PubMedCrossRef
4.
Zurück zum Zitat Boyd RL (2005) Surgical-orthodontic treatment of two skeletal Class III patients with Invisalign® and fixed appliances. J Clin Orthod 39:245–258PubMed Boyd RL (2005) Surgical-orthodontic treatment of two skeletal Class III patients with Invisalign® and fixed appliances. J Clin Orthod 39:245–258PubMed
5.
Zurück zum Zitat Boyd RL (2007) Complex orthodontic treatment using a new protocol for the Invisalign® appliance. J Clin Orthod 41:525–547PubMed Boyd RL (2007) Complex orthodontic treatment using a new protocol for the Invisalign® appliance. J Clin Orthod 41:525–547PubMed
6.
Zurück zum Zitat Boyd RL (2008) Esthetic orthodontic treatment using the invisalign® appliance for moderate to complex malocclusions. J Dent Educ 72:948–967PubMed Boyd RL (2008) Esthetic orthodontic treatment using the invisalign® appliance for moderate to complex malocclusions. J Dent Educ 72:948–967PubMed
7.
Zurück zum Zitat Clements KM, Bollen AM, Huang G et al (2003) Activation time and material stiffness of sequential removable orthodontic appliances. Part 2: dental improvements. Am J Orthod Dentofacial Orthop 124:502–508PubMedCrossRef Clements KM, Bollen AM, Huang G et al (2003) Activation time and material stiffness of sequential removable orthodontic appliances. Part 2: dental improvements. Am J Orthod Dentofacial Orthop 124:502–508PubMedCrossRef
8.
Zurück zum Zitat Djeu G, Shelton C, Maganzini A (2005) Outcome assessment of Invisalign® and traditional orthodontic treatment compared with the American Board of Orthodontics objective grading system. Am J Orthod Dentofacial Orthop 128:292–298PubMedCrossRef Djeu G, Shelton C, Maganzini A (2005) Outcome assessment of Invisalign® and traditional orthodontic treatment compared with the American Board of Orthodontics objective grading system. Am J Orthod Dentofacial Orthop 128:292–298PubMedCrossRef
9.
Zurück zum Zitat Eliades T, Pratsinis H, Athanasiou AE et al (2009) Cytotoxicity and estrogenicity of Invisalign® appliances. Am J Orthod Dentofacial Orthop 136:100–103PubMedCrossRef Eliades T, Pratsinis H, Athanasiou AE et al (2009) Cytotoxicity and estrogenicity of Invisalign® appliances. Am J Orthod Dentofacial Orthop 136:100–103PubMedCrossRef
10.
Zurück zum Zitat Gracco A, Mazzoli A, Favoni O et al (2009) Short-term chemical and physical changes in invisalign® appliances. Aust Orthod J 25:34–40PubMed Gracco A, Mazzoli A, Favoni O et al (2009) Short-term chemical and physical changes in invisalign® appliances. Aust Orthod J 25:34–40PubMed
11.
Zurück zum Zitat Hönn M, Göz G (2006) A premolar extraction case using the Invisalign® system. J Orofac Orthop 67:385–394PubMedCrossRef Hönn M, Göz G (2006) A premolar extraction case using the Invisalign® system. J Orofac Orthop 67:385–394PubMedCrossRef
13.
Zurück zum Zitat Kravitz ND, Kusnoto B, Agran B, Viana G (2008) Influence of attachments and interproximal reduction on the accuracy of canine rotation with Invisalign®. A prospective clinical study. Angle Orthod 78:682–687PubMedCrossRef Kravitz ND, Kusnoto B, Agran B, Viana G (2008) Influence of attachments and interproximal reduction on the accuracy of canine rotation with Invisalign®. A prospective clinical study. Angle Orthod 78:682–687PubMedCrossRef
14.
Zurück zum Zitat Kravitz ND, Kusnoto B, BeGole E et al (2009) How well does Invisalign® work? A prospective clinical study evaluating the efficacy of tooth movement with Invisalign®. Am J Orthod Dentofacial Orthop 135:27–35PubMedCrossRef Kravitz ND, Kusnoto B, BeGole E et al (2009) How well does Invisalign® work? A prospective clinical study evaluating the efficacy of tooth movement with Invisalign®. Am J Orthod Dentofacial Orthop 135:27–35PubMedCrossRef
15.
Zurück zum Zitat Kuo E, Miller RJ (2003) Automated custom-manufacturing technology in orthodontics. Am J Orthod Dentofacial Orthop 123:578–581PubMedCrossRef Kuo E, Miller RJ (2003) Automated custom-manufacturing technology in orthodontics. Am J Orthod Dentofacial Orthop 123:578–581PubMedCrossRef
16.
Zurück zum Zitat Meier B, Wiemer KB, Miethke RR (2003) Invisalign®-patient profiling. Analysis of a prospective survey. J Orofac Orthop 64:352–358PubMedCrossRef Meier B, Wiemer KB, Miethke RR (2003) Invisalign®-patient profiling. Analysis of a prospective survey. J Orofac Orthop 64:352–358PubMedCrossRef
17.
Zurück zum Zitat Melkos AB (2005) Advances in digital technology and orthodontics: a reference to the Invisalign® method. Med Sci Monit 11:PI 39– PI 42 Melkos AB (2005) Advances in digital technology and orthodontics: a reference to the Invisalign® method. Med Sci Monit 11:PI 39– PI 42
18.
Zurück zum Zitat Miller RJ, Duong TT, Derakhshan M (2002) Lower incisor extraction treatment with the Invisalign® system. J Clin Orthod 36:95–102PubMed Miller RJ, Duong TT, Derakhshan M (2002) Lower incisor extraction treatment with the Invisalign® system. J Clin Orthod 36:95–102PubMed
19.
Zurück zum Zitat Nguyen CV, Chen J (2006) Three-dimensional superimposition tool. In: Tuncay OC (ed) The Invisalign® system. Quintessence, New Malden, p 12–32 Nguyen CV, Chen J (2006) Three-dimensional superimposition tool. In: Tuncay OC (ed) The Invisalign® system. Quintessence, New Malden, p 12–32
20.
Zurück zum Zitat Schupp W, Haubrich J, Neumann I (2010) Invisalign (®) treatment of patients with craniomandibular disorders. Int Orthod 8:253–267PubMed Schupp W, Haubrich J, Neumann I (2010) Invisalign (®) treatment of patients with craniomandibular disorders. Int Orthod 8:253–267PubMed
21.
Zurück zum Zitat Schuster S, Eliades G, Zinelis S et al (2004) Structural conformation and leaching from in vitro aged and retrieved Invisalign® appliances. Am J Orthod Dentofacial Orthop 126:725–728PubMedCrossRef Schuster S, Eliades G, Zinelis S et al (2004) Structural conformation and leaching from in vitro aged and retrieved Invisalign® appliances. Am J Orthod Dentofacial Orthop 126:725–728PubMedCrossRef
22.
Zurück zum Zitat Vlaskalic V, Boyd RL (2002) Clinical evolution of the Invisalign® appliance. J Calif Dent Assoc 30:769–776PubMed Vlaskalic V, Boyd RL (2002) Clinical evolution of the Invisalign® appliance. J Calif Dent Assoc 30:769–776PubMed
23.
Metadaten
Titel
Accuracy of invisalign® treatments in the anterior tooth region
First results
verfasst von
Dr. E. Krieger
J. Seiferth
I. Saric
B.A. Jung
H. Wehrbein
Publikationsdatum
01.04.2011
Verlag
Springer-Verlag
Erschienen in
Journal of Orofacial Orthopedics / Fortschritte der Kieferorthopädie / Ausgabe 2/2011
Print ISSN: 1434-5293
Elektronische ISSN: 1615-6714
DOI
https://doi.org/10.1007/s00056-011-0017-4

Weitere Artikel der Ausgabe 2/2011

Journal of Orofacial Orthopedics / Fortschritte der Kieferorthopädie 2/2011 Zur Ausgabe

Informationen

DGKFO-Seiten

Newsletter

Bestellen Sie unseren kostenlosen Newsletter Update Zahnmedizin und bleiben Sie gut informiert – ganz bequem per eMail.