Skip to main content
Erschienen in: Strahlentherapie und Onkologie 5/2017

Open Access 17.03.2017 | Original Article

Fully automated VMAT treatment planning for advanced-stage NSCLC patients

verfasst von: Giuseppe Della Gala, M.Sc., Maarten L. P. Dirkx, Ph.D., Nienke Hoekstra, M.D., Dennie Fransen, RTT, Nico Lanconelli, Ph.D., Marjan van de Pol, M.D., Ben J. M. Heijmen, Ph.D., Steven F. Petit, Ph.D.

Erschienen in: Strahlentherapie und Onkologie | Ausgabe 5/2017

Abstract

Purpose

To develop a fully automated procedure for multicriterial volumetric modulated arc therapy (VMAT) treatment planning (autoVMAT) for stage III/IV non-small cell lung cancer (NSCLC) patients treated with curative intent.

Materials and methods

After configuring the developed autoVMAT system for NSCLC, autoVMAT plans were compared with manually generated clinically delivered intensity-modulated radiotherapy (IMRT) plans for 41 patients. AutoVMAT plans were also compared to manually generated VMAT plans in the absence of time pressure. For 16 patients with reduced planning target volume (PTV) dose prescription in the clinical IMRT plan (to avoid violation of organs at risk tolerances), the potential for dose escalation with autoVMAT was explored.

Results

Two physicians evaluated 35/41 autoVMAT plans (85%) as clinically acceptable. Compared to the manually generated IMRT plans, autoVMAT plans showed statistically significant improved PTV coverage (V95% increased by 1.1% ± 1.1%), higher dose conformity (R50 reduced by 12.2% ± 12.7%), and reduced mean lung, heart, and esophagus doses (reductions of 0.9 Gy ± 1.0 Gy, 1.5 Gy ± 1.8 Gy, 3.6 Gy ± 2.8 Gy, respectively, all p < 0.001). To render the six remaining autoVMAT plans clinically acceptable, a dosimetrist needed less than 10 min hands-on time for fine-tuning. AutoVMAT plans were also considered equivalent or better than manually optimized VMAT plans. For 6/16 patients, autoVMAT allowed tumor dose escalation of 5–10 Gy.

Conclusion

Clinically deliverable, high-quality autoVMAT plans can be generated fully automatically for the vast majority of advanced-stage NSCLC patients. For a subset of patients, autoVMAT allowed for tumor dose escalation.
Hinweise
During part of the work Steven Petit was also affiliated to Massachusetts General Hospital—Harvard Medical School
In recent years, the use of volumetric modulated arc therapy (VMAT) for treatment of locally advanced non-small cell lung cancer (NSCLC) patients has grown. The main advantage compared to fixed-field intensity-modulated radiotherapy (IMRT) is its improved delivery efficiency. The shorter treatment time of VMAT allows for increased patient throughput, reduced intrafractional motion, and improved patient comfort [1].
IMRT and VMAT treatment planning with commercial treatment planning systems (TPS) is an iterative trial-and-error process. The procedure is time consuming and the quality of the final treatment plan may depend highly on the experience and skills of the dosimetrist, the complexity of the case, and the available time. In addition, there is no guarantee that the final treatment plan reflects the optimal dose distribution for a specific patient in terms of PTV coverage, doses in organs at risk (OAR), and the desired tradeoffs between them.
To increase the consistency and quality of treatment plans, and to reduce treatment planning time, there is a growing interest in automated treatment planning [216]. Different vendors offer so-called knowledge-based automated treatment planning solutions. These methods rely on a database of previously treated plans that are used to predict achievable dose objectives for a new patient. For NSCLC patients, several methods for automated treatment planning have been evaluated. Fogliata et al. used Varian’s RapidPlan (Varian Medical Systems, Palo Alto, CA, USA), which applies a database of previously treated plans [12]. Tol et al. used an automated interactive optimization procedure for VMAT plan optimization based on dose–volume constraints and dose–volume histogram analyses [4]. This method requires that the user knows the input priorities that result in acceptable treatment plans for the majority of patients. Zhang et al. and Quan et al. proposed a completely automated process based on in-house developed optimization algorithms [1, 5]. Beam angles were optimized using a library of plans from previously treated patients. For VMAT and IMRT planning techniques, they showed that automatically generated plans were generally preferred by oncologists when compared to manually generated plans. Recently, Kamran et al. investigated the benefit of IMRT planning in combination with multicriterial optimization [13]. Compared to manually generated plans without multicriterial optimization, OAR sparing could be significantly improved without comprising target coverage. Consequently, the multicriterial optimized plans were preferred by the physicians.
In the authors’ department, Erasmus-iCycle was developed, a system for fully automated multicriterial treatment plan generation. Currently, Erasmus-iCycle is clinically applied for IMRT and VMAT plan generation for prostate, head and neck, and cervical cancer patients [3, 6, 14]. The authors have previously demonstrated that automatically generated plans were of noninferior or higher quality compared to plans generated by an expert dosimetrist in the absence of time pressure [3].
The goal of this study was to configure, test, and implement Erasmus-iCycle for automated, multicriterial VMAT treatment planning (autoVMAT) for stage III/IV NSCLC patients treated with curative intent. For 41 consecutively treated patients, autoVMAT plans were compared to manually generated, clinically delivered IMRT plans using dosimetric indices and physicians’ scoring. For a subgroup of patients, autoVMAT plans were also compared to manually generated VMAT plans. Sixteen out of 41 patients had a reduced PTV dose prescription in the clinical plan to avoid violation of OAR tolerance doses. For these patients, the potential for dose escalation with autoVMAT was explored.

Materials and methods

Patients and treatment

Two sets of locally advanced stage III–IV NSCLC patients treated with curative intent with IMRT at the Erasmus MC Cancer Institute were included in this study. The first set, consisting of 7 patients treated in 2014, was used to configure the system for automated planning. The second set, consisting of 41 consecutive patients treated between January and August 2015, was used for validation. All clinically delivered IMRT plans consisted of 5–9 beams and were generated with the common manual trial-and-error planning approach, using the Monaco TPS version 5.0 (Elekta AB, Stockholm, Sweden).
For segmentation of the targets and OARs, the 50% exhale phase of a four-dimensional computed tomography (4DCT) scan was used. The clinical target volume (CTV) was defined by expanding the gross tumor volume (GTV) by a margin of 5 mm. If needed, the CTV was manually edited based on anatomic borders. Lymph node stations with affected nodes—as determined on diagnostic CT or positron-emission tomography (PET) scans, endoscopic ultrasound, or fine needle aspiration pathology examination—were also defined as CTV. To define the planning target volume (PTV), the CTVs of the primary tumor and lymph nodes were expanded by a margin of 1 cm in the lateral and ventral–dorsal directions and by a margin of 1–1.2 cm in the cranial–caudal direction, depending on respiratory tumor motion amplitude. According to protocol, the intended PTV dose prescription was 66 Gy. In practice, prescribed doses varied between 30 and 66 Gy, in order to avoid violation of clinical constraints. For 25 patients treated with concurrent chemotherapy, the dose was delivered in fractions of 2 Gy; 15 patients treated with sequential chemotherapy received 2.75 Gy per fraction. The other 8 patients, receiving a prescribed dose of 45 Gy or lower, were treated with 3 Gy per fraction.
Planning goals were to cover at least 95% of the PTV and 99% of the CTV with 95% of the prescribed dose, and 99% of the PTV with 90% of the prescribed dose, while sparing OARs as much as possible. The mean dose in the total lung volume minus GTV had to be kept below 20 Gy and the V20Gy (volume receiving 20 Gy or more) below 35%. Maximum allowed doses to the spinal cord and plexus brachialis were 46.3 Gy and 56.6 Gy (3 Gy/fraction), 49.3 Gy and 60 Gy (2.75 Gy/fraction), and 50 Gy and 66 Gy (2 Gy/fraction), respectively. The protocol also prescribed a V5Gy < 60% for lungs minus GTV and V45Gy < 25% for the esophagus, but these requirements did not determine whether a plan was deemed acceptable or not. A high level of dose conformity was desired.

Automated multicriterial treatment planning

The in-house developed Erasmus-iCycle TPS cannot be directly used for clinical plan generation. Therefore, automated multicriterial treatment planning was implemented as a two-step process, combing Erasmus-iCycle and the clinically used Monaco TPS. The process is described in detail by Voet et al. [3] and is briefly summarized here. In the first step, Erasmus-iCycle is used to automatically generate an equi-angular 23-beam IMRT plan, simulating VMAT delivery. Further increasing the number of beams increased calculation times but did not lead to improved plan quality [3]. Next, using the achieved constraint and objective values in the Erasmus-iCycle plan, a patient-specific Monaco template is automatically generated, followed by automated plan generation in Monaco based on this template, to result in a clinically deliverable VMAT plan that closely mimics the initial Erasmus-iCycle plan. The same system is used clinically for automated clinical plan generation for prostate, head and neck, and cervical cancer patients [3, 6, 14].
Plan generation with Erasmus-iCycle is based on a site-specific “wishlist”, with hard planning constraints (never to be violated) and prioritized objectives. The assigned objective priorities are used to automatically steer the multiobjective plan generation, ensuring clinically desirable tradeoffs between all treatment objectives. Generally, adequate PTV coverage has the highest priority, followed by the most important OAR objective.
In this study, all Erasmus-iCycle plans were generated by employing one fixed wishlist, with some minor variations to handle differences in dose prescription and fractionation (see above). An initial wishlist was composed based on previous clinical experience, the planning protocol, and wishes of treating physicians on how to improve clinically applied plans. It was used to automatically generate a plan for the first 7 patients included in this study. These plans were then evaluated together with physicians, and the wishlist was modified according to their input. This iterative procedure continued until no further improvements in plan quality were achieved for the 7 training patients. The final wishlist is shown in Table 1.
Table 1
Erasmus-iCycle wish list for advanced stage NSCLC. The wishlist consists of six hard constraints that are met per definition and 13 objectives that are optimized in order of priority. Once the goal of an objective is achieved or when further optimization is no longer possible, the optimizer fixes the achieved value of the objective (with a bit of slack) as a constraint and continues with the next objective. The first priority was proper CTV and PTV coverage, ensured by the use of the LTCP cost function with a sensitivity value of 0.8 [17]. When the spinal cord was close to the target or overlapped it, the objective was applied to the target volume from which the spinal cord expanded by 5 mm was subtracted. The purpose was to locate the acceptable underdosage to the PTV near the spinal cord. The third objective aims to reduce the maximum dose at 4 cm from the PTV to 60% of the prescribed dose. The next objective tries to reduce the mean lung dose to 15 Gy. Two more objectives (shell PTV + 1 cm and shell PTV + 3 cm) with priorities 5 and 6 realize a steep dose falloff outside the PTV. To minimize the esophagus volume treated to a high dose, an EUD objective were used. Objectives 8 to 13 aimed at further reducing the doses to the OARs. In the final phase of the optimization, all objectives are consecutively minimized to their fullest extent
Priority
Volume
Dose metrics
Limit/goal
Parameter
Constraint
Patient
Max
107% of prescribed dose
Constraint
CTV
Max
107% of prescribed dose
Constraint
PTV—(spinal cord + 5 mm)
Max
107% of prescribed dose
Constraint
Spinal cord + 3 mm
Max
47 Gy
Constraint
Plexus
Max
63 Gy
Constraint
Lungs—GTV
Mean
19 Gy
1
CTV
↓ LTCP
1
α = 0.8
2
PTV
↓ LTCP
0.5
α = 0.8
3
Patient—(PTV + 4 cm)
↓ Max
60% of prescribed dose
4
Lungs—GTV
↓ Mean
15 Gy
5
Shell PTV + 1 cm
↓ Max
90% of prescribed dose
6
Shell PTV + 3 cm
↓ Max
80% of prescribed dose
7
Esophagus
↓ EUD
44 Gy
k = 8
8
Esophagus
↓ Mean
15 Gy
9
Lungs—GTV
↓ Mean
5 Gy
10
Heart
↓ Mean
15 Gy
11
Spinal cord + 3 mm
↓ Max
42 Gy
12
Unspecified tissue
↓ Max
40 Gy
13
Plexus
↓ Mean
20 Gy
LTCP logarithmic tumor control probability, EUD equivalent uniform dose, CTV clinical target volume, GTV gross tumor volume, PTV planning target volume, NSCLC non-small cell lung cancer, OAR organs at risk. The arrows indicate that objectives are minimized

Comparison of autoVMAT and clinically applied IMRT plans

For the second group of 41 patients, the manually generated IMRT plans were compared with autoVMAT plans using dosimetric indices and by independent scoring by two physicians. For fair comparison, in this part of the investigation, the prescribed tumor dose for the autoVMAT plans was identical to that used in the corresponding clinical plans.

Comparison of automatically and manually generated VMAT plans

At the time of this study, the authors did not yet apply VMAT in clinical routine for advanced-stage NSCLC patients. Prior to clinical introduction, for 10/41 validation patients the autoVMAT plans were also compared with VMAT plans that were manually generated by an expert dosimetrist without planning time limitations. The dosimetrist had no prior knowledge of the Erasmus-iCycle plans of these patients.

Dose escalation

In clinical routine, when it was not possible to achieve the intended dose prescription of 66 Gy without violating at least one of the clinical hard constraints, the number of planned fractions, and consequently the prescribed dose, was reduced following a standard schedule: 1. reduction of the prescribed dose to 60 or 60.5 Gy (depending on fractionation schedule), 2. if plan generation was still unfeasible, further reduction to 55 Gy, or, if needed, to 45, 39, or 30 Gy delivered in 3 Gy per fraction. Intermediate dose levels were not clinically applied due to planning time limitations. For the 16 patients with a prescription dose lower than 66 Gy, it was attempted to escalate the dose using autoVMAT, following the standard schedule steps backwards: from 45 Gy to 55 Gy, from 55 Gy to 60.5 Gy, from 60 Gy or 60.5 Gy to 66 Gy. For fair comparison with the clinically applied procedure, intermediate dose escalations were not investigated.

Statistics

Plan conformity was quantified using R50, the ratio between the total volume receiving at least 50% of the prescribed dose and the volume of the PTV. Dose homogeneity in the PTV was measured using the homogeneity index (HI), defined as HI = (D2% − D98%)/D50%, where D2%, D98%, and D50% are the doses covering 2, 98, and 50% of the PTV, respectively. Differences in dosimetric indices were expressed as mean ±1 standard deviation. Statistical significance of differences in dosimetric indices was evaluated with two-tailed Wilcoxon signed-rank tests, considering p < 0.05 statistically significant.

Results

AutoVMAT and clinically applied IMRT plans

Two physicians judged 35/41 of the autoVMAT plans (85%) as clinically acceptable. These plans were considered equivalent or (as in most cases) better than the corresponding IMRT plans, due to a combination of better PTV coverage and dose conformity, as well as sparing of the lungs, heart, and esophagus (Fig. 1 and Table 2). PTV volume receiving at least 95% of the prescribed dose (V95%) was on average improved by 1.1% ± 1.1% (p < 0.001) and the near-minimum dose in the PTV (D99%) by 0.6 Gy ± 1.5 Gy (p = 0.004). The mean lung dose (lungs excluding GTV) was reduced by 0.9 Gy ± 1.0 Gy (p < 0.001), V20Gy by 1.3% ± 2.1% (p = 0.001), and V5Gy by 3.2% ± 7.6% (p = 0.023). The esophagus mean dose was on average 3.6 Gy ± 2.8 Gy (p < 0.001) lower and V45Gy was reduced by 4.6% ± 4.8% (p < 0.001). The heart mean dose was reduced by 1.5 Gy ± 1.8 Gy (p < 0.001).
Table 2
Mean values and uncertainties, reported as one standard deviation (1 SD), for the evaluated dose metrics of the clinical IMRT plans and the autoVMAT plans for the 35 study patients with acceptable autoVMAT plans. For all metrics, statically significant differences were observed in a two-tailed Wilcoxon signed-rank test
 
Clinical IMRT
autoVMAT
p-value Wilcoxon test
Dose metrics
Mean
SD
Mean
SD
PTV V95% (%)
97.0
1.4
98.2
1.1
<0.001
PTV D99% (Gy)
55.2
8.9
55.8
9.1
0.004
Conformity index R50 (–)
4.67
0.85
4.13
0.56
<0.001
Homogeneity index (–)
0.127
0.035
0.101
0.035
<0.001
Mean lung dose (Gy)
17.3
2.6
16.5
2.3
<0.001
Lungs V5Gy (%)
56.0
10.6
52.8
9.7
0.023
Lungs V20Gy (%)
29.4
4.8
28.1
4.4
0.001
Mean heart dose (Gy)
12.3
6.6
10.8
5.9
<0.001
Mean esophagus dose (Gy)
24.4
6.9
20.8
6.5
<0.001
Esophagus V45Gy (%)
27.0
13.2
22.4
11.9
<0.001
IMRT intensity-modulated radiotherapy, autoVMAT automated volumetric modulated arc therapy, PTV planning target volume. V95% is the PTV volume receiving at least 95% of the prescribed dose, D99% is the minimum dose delivered to 99% of the PTV. V5Gy, V20Gy and V45Gy are the volumes receiving respecively 5 Gy, 20 Gy and 45 Gy, or more
The autoVMAT plans were generally more conformal than the clinical IMRT plans as indicated by a reduction in R50 by, on average, 12.2% ± 12.7% (p < 0.001). Moreover, the PTV dose homogeneity improved by 0.025 ± 0.017 (p < 0.001), i. e., an average reduction of 20%.

Unacceptable plans

For 6 patients, the physicians deemed the autoVMAT plan clinically unacceptable due to violation of at least one clinical constraint. In 2 patients PTV coverage was not adequate (V95 < 95%, V90 < 99%); for 4 patients the MLD and/or lungs V20Gy exceeded clinical constraints. It took a dosimetrist less than 10 min hands-on time to manually fine-tune these plans in Monaco, rendering them acceptable. In contrast, 3–4 h were required to generate a VMAT plan from scratch.

AutoVMAT and manually generated VMAT plans

A radiation oncologist evaluated all autoVMAT plans as equivalent or better in quality compared to the manually generated VMAT plans. Average differences were not statistically significant, except for lungs V5Gy (reduction by 6.3% ± 3.3%; p = 0.008), esophagus mean dose (reduction by 2.2 ± 1.8 Gy; p = 0.012), and esophagus V45Gy (reduction by 3.2% ± 2.3%; p = 0.012), all in favor of autoVMAT plans. The manually generated plans were slightly more conformal, as expressed by a difference in R50 of 8% ± 10% (p = 0.038). The differences in dose metrics are shown in Fig. 2.

Dose escalation

For 6/16 patients with a prescription dose less than 66 Gy, a clinically acceptable autoVMAT plan with a higher prescribed tumor dose could be generated. In 2 patients dose escalation from 45 to 55 Gy was possible, in 2 further patients from 60 to 66 Gy. For 1 patient the dose could be escalated from 55 to 60.5 Gy and for another from 60.5 to 66 Gy. In all cases, the physicians preferred the plan with the higher tumor dose.

Discussion

This study developed a method for fully automated VMAT treatment planning for NSCLC patients based on Erasmus-iCycle. A single wishlist was used for all patients treated with the same protocol. The Erasmus-iCycle dose distributions were converted automatically into clinically deliverable Monaco VMAT plans. It was shown that autoVMAT plans were of equivalent or, as in most cases, higher quality than the clinically applied IMRT plans. This may be attributed partly to the delivery technique (VMAT vs. IMRT), as was also observed previously [1], and partly to the difficulty of manually selecting the optimal IMRT beam configuration for individual patients. AutoVMAT plans were also of a higher or equivalent quality when compared to VMAT plans generated by an expert dosimetrist in the absence of time constraints. Only in 6/41 cases were autoVMAT plans not clinically acceptable, and a dosimetrist needed only 10 min to make these plans fulfil the constraints. The automated procedure led to a decrease in hands-on treatment planning time of 3–4 h per patient compared to conventional VMAT planning. Based on these results, autoVMAT has been clinically applied for all advanced-stage NSCLC patients treated with curative intent at the authors’ hospital since March 2016.
In previous studies for prostate, head and neck, and cervical cancer [3, 6, 14], it was not observed that some of the autoVMAT plans were unacceptable and of a lesser quality than manually generated IMRT plans with the same prescribed dose. The authors believe that this may be attributed to the simple pencil beam dose calculation algorithm in Erasmus-iCycle [17]. For NSCLC patients, this may result in high-quality Erasmus-iCycle plans that cannot be accurately mimicked with the advanced Monte Carlo dose engine in Monaco. Therefore, in an ongoing project, the Monte Carlo engine used in Monaco is coupled to Erasmus-iCycle. It is expected that, in the near future, this will avoid generation of unacceptable plans for NSCLC patients as well.
For a considerable proportion of advanced-stage NSCLC patients, a prescribed dose of 66 Gy is not achievable. In these cases, the prescribed dose has to be lowered to yield acceptable OAR doses. This can be a tedious process and the clinical protocol was therefore to lower the prescribed dose in fixed steps of multiple fractions. With autoVMAT it is now feasible to apply individualized dose reductions, avoiding PTV doses lower than strictly needed.
Erasmus-iCycle for autoVMAT planning has a couple of advantages compared to other approaches for automated (VMAT) planning. First, it does not fully rely on a database of prior patients, as published knowledge-based approaches do [1, 4, 5, 12]. A small set of training patients (≤10) is only used as starting point for establishing a site-specific wishlist. When generating this wishlist, the explicit aim is to achieve a higher plan quality than in the clinical plans of the training patients. With this procedure, automatically generated plans for previously treated patients have, on average, a higher or at least non-inferior plan quality than the clinical plans [3], as also shown in this work. The need for a small set of training patients has an additional advantage in case of changes in clinical protocol, which would otherwise require a new large database of high-quality plans.

Conclusion

Fully automated VMAT treatment planning was implemented for advanced-stage NSCLC. For the vast majority of patients, autoVMAT plans were equivalent or better than manually generated plans. Moreover, for a subset of patients, autoVMAT allowed for tumor dose escalation within clinical constraints. Consequently, autoVMAT has been applied for NSCLC in routine clinical use at the authors’ hospital since March 2016.

Acknowledgements

The authors would like to thank Kathrin Frey for her help with the translation into German.

Compliance with ethical guidelines

Conflict of interest

The department of Radiation Oncology of Erasmus MC Cancer Institute has research collaborations with Elekta AB and Accuray. G. Della Gala, M.L.P. Dirkx, N. Hoekstra, D. Fransen, N. Lanconelli, M. van de Pol, and B.J.M. Heijmen declare that they have no competing interests. S.F. Petit receives financial support from the Dutch Cancer Society (KWF EMCR 2014-6667).
Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://​creativecommons.​org/​licenses/​by/​4.​0/​), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.

Unsere Produktempfehlungen

e.Med Interdisziplinär

Kombi-Abonnement

Für Ihren Erfolg in Klinik und Praxis - Die beste Hilfe in Ihrem Arbeitsalltag

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de.

e.Med Radiologie

Kombi-Abonnement

Mit e.Med Radiologie erhalten Sie Zugang zu CME-Fortbildungen des Fachgebietes Radiologie, den Premium-Inhalten der radiologischen Fachzeitschriften, inklusive einer gedruckten Radiologie-Zeitschrift Ihrer Wahl.

Strahlentherapie und Onkologie

Print-Titel

•Übersichten, Originalien, Kasuistiken

•Kommentierte Literatur aus der Radioonkologie, Strahlenbiologie und -physik

Literatur
1.
Zurück zum Zitat Quan EM, Chang JY, Liao Z et al (2012) Automated volumetric modulated arc therapy treatment planning for stage III lung cancer: how does it compare with intensity-modulated radio therapy? Int J Radiat Oncol Biol Phys 84:e69–76CrossRefPubMedPubMedCentral Quan EM, Chang JY, Liao Z et al (2012) Automated volumetric modulated arc therapy treatment planning for stage III lung cancer: how does it compare with intensity-modulated radio therapy? Int J Radiat Oncol Biol Phys 84:e69–76CrossRefPubMedPubMedCentral
2.
Zurück zum Zitat Bezjak A, Rumble RB, Rodrigues G, Hope A, Warde P, Members of the IIEP (2012) Intensity-modulated radiotherapy in the treatment of lung cancer. Clin Oncol (R Coll Radiol) 24:508–520CrossRef Bezjak A, Rumble RB, Rodrigues G, Hope A, Warde P, Members of the IIEP (2012) Intensity-modulated radiotherapy in the treatment of lung cancer. Clin Oncol (R Coll Radiol) 24:508–520CrossRef
3.
Zurück zum Zitat Voet PW, Dirkx ML, Breedveld S, Al-Mamgani A, Incrocci L, Heijmen BJ (2014) Fully automated volumetric modulated arc therapy plan generation for prostate cancer patients. Int J Radiat Oncol Biol Phys 88:1175–1179CrossRefPubMed Voet PW, Dirkx ML, Breedveld S, Al-Mamgani A, Incrocci L, Heijmen BJ (2014) Fully automated volumetric modulated arc therapy plan generation for prostate cancer patients. Int J Radiat Oncol Biol Phys 88:1175–1179CrossRefPubMed
4.
Zurück zum Zitat Tol JP, Dahele M, Delaney AR, Doornaert P, Slotman BJ, Verbakel WF (2016) Detailed evaluation of an automated approach to interactive optimization for volumetric modulated arc therapy plans. Med Phys 43:1818CrossRefPubMed Tol JP, Dahele M, Delaney AR, Doornaert P, Slotman BJ, Verbakel WF (2016) Detailed evaluation of an automated approach to interactive optimization for volumetric modulated arc therapy plans. Med Phys 43:1818CrossRefPubMed
5.
Zurück zum Zitat Zhang X, Li X, Quan EM, Pan X, Li Y (2011) A methodology for automatic intensity-modulated radiation treatment planning for lung cancer. Phys Med Biol 56:3873–3893CrossRefPubMed Zhang X, Li X, Quan EM, Pan X, Li Y (2011) A methodology for automatic intensity-modulated radiation treatment planning for lung cancer. Phys Med Biol 56:3873–3893CrossRefPubMed
6.
Zurück zum Zitat Voet PW, Dirkx ML, Breedveld S, Fransen D, Levendag PC, Heijmen BJ (2013) Toward fully automated multicriterial plan generation: a prospective clinical study. Int J Radiat Oncol Biol Phys 85:866–872CrossRefPubMed Voet PW, Dirkx ML, Breedveld S, Fransen D, Levendag PC, Heijmen BJ (2013) Toward fully automated multicriterial plan generation: a prospective clinical study. Int J Radiat Oncol Biol Phys 85:866–872CrossRefPubMed
7.
Zurück zum Zitat Amit G, Purdie TG (2015) Automated planning of breast radiotherapy using cone beam CT imaging. Med Phys 42:770–779CrossRefPubMed Amit G, Purdie TG (2015) Automated planning of breast radiotherapy using cone beam CT imaging. Med Phys 42:770–779CrossRefPubMed
8.
Zurück zum Zitat Sharpe MB, Moore KL, Orton CG (2014) Point/Counterpoint: Within the next ten years treatment planning will become fully automated without the need for human intervention. Med Phys 41:120601CrossRefPubMed Sharpe MB, Moore KL, Orton CG (2014) Point/Counterpoint: Within the next ten years treatment planning will become fully automated without the need for human intervention. Med Phys 41:120601CrossRefPubMed
9.
Zurück zum Zitat Wu B, McNutt T, Zahurak M et al (2012) Fully automated simultaneous integrated boosted-intensity modulated radiation therapy treatment planning is feasible for head-and-neck cancer: a prospective clinical study. Int J Radiat Oncol Biol Phys 84:e647–e653CrossRefPubMed Wu B, McNutt T, Zahurak M et al (2012) Fully automated simultaneous integrated boosted-intensity modulated radiation therapy treatment planning is feasible for head-and-neck cancer: a prospective clinical study. Int J Radiat Oncol Biol Phys 84:e647–e653CrossRefPubMed
10.
Zurück zum Zitat Fiege J, McCurdy B, Potrebko P, Champion H, Cull A (2011) PARETO: A novel evolutionary optimization approach to multiobjective IMRT planning. Med Phys 38:5217–5229CrossRefPubMed Fiege J, McCurdy B, Potrebko P, Champion H, Cull A (2011) PARETO: A novel evolutionary optimization approach to multiobjective IMRT planning. Med Phys 38:5217–5229CrossRefPubMed
11.
Zurück zum Zitat Boylan C, Rowbottom C (2014) A bias-free, automated planning tool for technique comparison in radiotherapy – application to nasopharyngeal carcinoma treatments. J Appl Clin Med Phys 15:4530CrossRefPubMed Boylan C, Rowbottom C (2014) A bias-free, automated planning tool for technique comparison in radiotherapy – application to nasopharyngeal carcinoma treatments. J Appl Clin Med Phys 15:4530CrossRefPubMed
12.
Zurück zum Zitat Fogliata A, Belosi F, Clivio A et al (2014) On the pre-clinical validation of a commercial model-based optimisation engine: application to volumetric modulated arc therapy for patients with lung or prostate cancer. Radiother Oncol 113:385–391CrossRefPubMed Fogliata A, Belosi F, Clivio A et al (2014) On the pre-clinical validation of a commercial model-based optimisation engine: application to volumetric modulated arc therapy for patients with lung or prostate cancer. Radiother Oncol 113:385–391CrossRefPubMed
13.
Zurück zum Zitat Kamran S, Mueller BS, Paetzold P et al (2016) Multi-criteria optimization achieves superior normal tissue sparing in a planning study of intensity-modulated radiation therapy for RTOG 1308-eligible non-small cell lung cancer patients. Radiother Oncol 116:515–520CrossRef Kamran S, Mueller BS, Paetzold P et al (2016) Multi-criteria optimization achieves superior normal tissue sparing in a planning study of intensity-modulated radiation therapy for RTOG 1308-eligible non-small cell lung cancer patients. Radiother Oncol 116:515–520CrossRef
14.
Zurück zum Zitat Sharfo AW, Voet PW, Breedveld S, Mens JW, Hoogeman MS, Heijmen BJ (2015) Comparison of VMAT and IMRT strategies for cervical cancer patients using automated planning. Radiother Oncol 114:395–401CrossRefPubMed Sharfo AW, Voet PW, Breedveld S, Mens JW, Hoogeman MS, Heijmen BJ (2015) Comparison of VMAT and IMRT strategies for cervical cancer patients using automated planning. Radiother Oncol 114:395–401CrossRefPubMed
15.
Zurück zum Zitat Krayenbuehl J, Norton I, Studer G, Guckenberger M (2015) Evaluation of an automated knowledge based treatment planning system for head and neck. Radiat Oncol 10:226CrossRefPubMedPubMedCentral Krayenbuehl J, Norton I, Studer G, Guckenberger M (2015) Evaluation of an automated knowledge based treatment planning system for head and neck. Radiat Oncol 10:226CrossRefPubMedPubMedCentral
16.
Zurück zum Zitat Hazell I, Bzdusek K, Kumar P et al (2016) Automatic planning of head and neck treatment plans. J Appl Clin Med Phys 17(1):272–282CrossRefPubMed Hazell I, Bzdusek K, Kumar P et al (2016) Automatic planning of head and neck treatment plans. J Appl Clin Med Phys 17(1):272–282CrossRefPubMed
17.
Zurück zum Zitat Breedveld S, Storchi PR, Voet PW, Heijmen BJ (2012) iCycle: Integrated, multicriterial beam angle, and profile optimization for generation of coplanar and noncoplanar IMRT plans. Med Phys 39:951–963CrossRefPubMed Breedveld S, Storchi PR, Voet PW, Heijmen BJ (2012) iCycle: Integrated, multicriterial beam angle, and profile optimization for generation of coplanar and noncoplanar IMRT plans. Med Phys 39:951–963CrossRefPubMed
Metadaten
Titel
Fully automated VMAT treatment planning for advanced-stage NSCLC patients
verfasst von
Giuseppe Della Gala, M.Sc.
Maarten L. P. Dirkx, Ph.D.
Nienke Hoekstra, M.D.
Dennie Fransen, RTT
Nico Lanconelli, Ph.D.
Marjan van de Pol, M.D.
Ben J. M. Heijmen, Ph.D.
Steven F. Petit, Ph.D.
Publikationsdatum
17.03.2017
Verlag
Springer Berlin Heidelberg
Erschienen in
Strahlentherapie und Onkologie / Ausgabe 5/2017
Print ISSN: 0179-7158
Elektronische ISSN: 1439-099X
DOI
https://doi.org/10.1007/s00066-017-1121-1

Weitere Artikel der Ausgabe 5/2017

Strahlentherapie und Onkologie 5/2017 Zur Ausgabe

Update Onkologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.