Antegrade versus retrograde screw fixation of anterior column acetabular fractures: a biomechanical in vitro study

verfasst von: Georg Osterhoff, Dag Wulsten, Satish Babu, Mark Heyland, Carlotta Pari

Erschienen in: European Journal of Trauma and Emergency Surgery | Ausgabe 5/2021

Einloggen, um Zugang zu erhalten

Abstract

Background

To compare the mechanical strength of antegrade versus retrograde lag screw fixation of anterior column acetabular fractures.

Methods

Standardised anterior column fractures were created in synthetic pelvis models and stabilised by either antegrade (ANTE, n = 4) or retrograde (RETRO, n = 4) anterior column screw fixation. In a validated setup, a cyclic loading protocol was applied with increasing axial force (750 cycles, 250–750 N) followed by load to failure. Construct survival, energy absorbed, construct stiffness, and load to failure were assessed. Descriptive and opto-metric methods were used to describe the mode of failure.

Results

All constructs failed with loads below 1500 N. With regard to energy absorbed until failure, the ANTE group resisted to 3.763 × 105 N*cycles (range 3.760 × 105–3.763 × 105) and the RETRO group to 3.762 × 105 N*cycles (range 3.761 × 105–3.765 × 105; p = 1.0). The load to failure was 1254 N (range 977–1299) in the ANTE group and 1234 N (range 1087–1456) in the RETRO group (p = 1.0). Construct stiffness with 250 N was not different between the two groups (ANTE 192 N/mm vs. RETRO 215 N/mm, p = 0.486). In all samples, the mode of failure was a transiliac fracture with screw breakout due to rotation of the pubic fragment around the axis of the screw with a range of rotational motion [ROM] during cyclic testing of 0.96° in one ANTE sample and 0.82° in one RETRO sample for 750 N, and ROM at failure of 2.53° in one ANTE sample and 2.23° in one RETRO sample. There was some plastic deformation of the screws in all cases but no breakage.

Conclusions

In this in vitro mechanical study, antegrade screw fixation of an anterior column acetabular fracture was not different in construct survival, load to failure, stiffness, and mode of failure when compared to retrograde screw fixation.
Literatur
4.
Zurück zum Zitat Starr AJ, Reinert CM, Jones AL. Percutaneous fixation of the columns of the acetabulum: a new technique. J Orthop Trauma. 1998;12(1):51–8.CrossRefPubMed Starr AJ, Reinert CM, Jones AL. Percutaneous fixation of the columns of the acetabulum: a new technique. J Orthop Trauma. 1998;12(1):51–8.CrossRefPubMed
10.
Zurück zum Zitat Schäffler A, Döbele S, Stuby F, et al. Die neue anatomische Flügelplatte für osteoporotische Azetabulumfrakturen: biomechanische Testung und erste klinische Erfahrungen (A new anatomical wing plate for osteoporotic acetabular fractures: biomechanical testing and first clinical experience). Z Orthop Unfall. 2014;152(1):26–322. https://doi.org/10.1055/s-0033-1360351.CrossRefPubMed Schäffler A, Döbele S, Stuby F, et al. Die neue anatomische Flügelplatte für osteoporotische Azetabulumfrakturen: biomechanische Testung und erste klinische Erfahrungen (A new anatomical wing plate for osteoporotic acetabular fractures: biomechanical testing and first clinical experience). Z Orthop Unfall. 2014;152(1):26–322. https://​doi.​org/​10.​1055/​s-0033-1360351.CrossRefPubMed
16.
Zurück zum Zitat Chow S, Shao J, Wang H. Sample size calculations in clinical research, 2nd edn. CRC biostatistics series. London: Chapman & Hall; 2008. Chow S, Shao J, Wang H. Sample size calculations in clinical research, 2nd edn. CRC biostatistics series. London: Chapman & Hall; 2008.
18.
Zurück zum Zitat Bergmann G, Deuretzbacher G, Heller M, et al. Hip contact forces and gait patterns from routine activities. J Biomech. 2001;34(7):859–71.CrossRefPubMed Bergmann G, Deuretzbacher G, Heller M, et al. Hip contact forces and gait patterns from routine activities. J Biomech. 2001;34(7):859–71.CrossRefPubMed
22.
Zurück zum Zitat Starr AJ, Jones AL, Reinert CM, et al. Preliminary results and complications following limited open reduction and percutaneous screw fixation of displaced fractures of the acetabulum. Injury. 2001;32(Suppl 1):SA45–SA50.CrossRefPubMed Starr AJ, Jones AL, Reinert CM, et al. Preliminary results and complications following limited open reduction and percutaneous screw fixation of displaced fractures of the acetabulum. Injury. 2001;32(Suppl 1):SA45–SA50.CrossRefPubMed
Metadaten
Titel
Antegrade versus retrograde screw fixation of anterior column acetabular fractures: a biomechanical in vitro study
verfasst von
Georg Osterhoff
Dag Wulsten
Satish Babu
Mark Heyland
Carlotta Pari
Publikationsdatum
29.10.2019
Verlag
Springer Berlin Heidelberg
Erschienen in
European Journal of Trauma and Emergency Surgery / Ausgabe 5/2021
Print ISSN: 1863-9933
Elektronische ISSN: 1863-9941
DOI
https://doi.org/10.1007/s00068-019-01255-6

Weitere Artikel der Ausgabe 5/2021

European Journal of Trauma and Emergency Surgery 5/2021 Zur Ausgabe

Arthropedia

Grundlagenwissen der Arthroskopie und Gelenkchirurgie. Erweitert durch Fallbeispiele, Videos und Abbildungen. 
» Jetzt entdecken

Update Orthopädie und Unfallchirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.