Nerve localization for peripheral regional anesthesia

Recommendations of the German Society of Anaesthesiology and Intensive Care Medicine

verfasst von: T. Steinfeldt, MD, U. Schwemmer, T. Volk, M. Neuburger, T. Wiesmann, A.R. Heller, O. Vicent, A. Stanek, M. Franz, H. Wulf, P. Kessler

Erschienen in: Die Anaesthesiologie | Ausgabe 7/2014

Einloggen, um Zugang zu erhalten

Abstract

The German Society of Anaesthesiology and Intensive Care Medicine (Deutsche Gesellschaft für Anästhesiologie und Intensivmedizin, DGAI) established an expert panel to develop preliminary recommendations for nerve localization in peripheral regional anesthesia. Based on expert knowledge and the relatively limited data, the recommendations state how ultrasound and/or electrical nerve stimulation should be used in daily practice, and where and when local anesthetics should be injected. Moreover, it was defined under which conditions a peripheral nerve block under general anesthesia or deep sedation is applicable.
Regarding the use of ultrasound the expert opinion was that out-of-plane and in-plane-techniques can be considered equal with respect to patient safety. Nevertheless, the direct or indirect visualization of the needle tip has to be assured. The injection of local anesthetics has to be visualized. Injections into nerves or those requiring an injection pressure should be avoided. The sole use of electrical nerve stimulation or ultrasound for nerve localization is still a suitable option as well as their combined use. To avoid accidental intraneural needle placement, an electrical current threshold ≥ 0.5 mA should be used. Moreover, it was stated that peripheral nerve blocks or continuous nerve block techniques under sedation or general anesthesia are applicable in adult patients who are unable to tolerate the block being performed in an awake state or have difficulty cooperating.
This article is published in English.
Literatur
1.
Zurück zum Zitat Perlas A, Niazi A, McCartney C et al (2006) The sensitivity of motor response to nerve stimulation and paresthesia for nerve localization as evaluated by ultrasound. Reg Anesth Pain Med 31:445–450CrossRefPubMed Perlas A, Niazi A, McCartney C et al (2006) The sensitivity of motor response to nerve stimulation and paresthesia for nerve localization as evaluated by ultrasound. Reg Anesth Pain Med 31:445–450CrossRefPubMed
2.
Zurück zum Zitat Bollini CA, Urmey WF, Vascello L et al (2003) Relationship between evoked motor response and sensory paresthesia in interscalene brachial plexus block. Reg Anesth Pain Med 28:384–388CrossRefPubMed Bollini CA, Urmey WF, Vascello L et al (2003) Relationship between evoked motor response and sensory paresthesia in interscalene brachial plexus block. Reg Anesth Pain Med 28:384–388CrossRefPubMed
3.
Zurück zum Zitat Urmey WF, Stanton J (2002) Inability to consistently elicit a motor response following sensory paresthesia during interscalene block administration. Anesthesiology 96:552–554CrossRefPubMed Urmey WF, Stanton J (2002) Inability to consistently elicit a motor response following sensory paresthesia during interscalene block administration. Anesthesiology 96:552–554CrossRefPubMed
4.
Zurück zum Zitat Tsai TP, Vuckovic I, Dilberovic F et al (2008) Intensity of the stimulating current may not be a reliable indicator of intraneural needle placement. Reg Anesth Pain Med 33:207–210CrossRefPubMed Tsai TP, Vuckovic I, Dilberovic F et al (2008) Intensity of the stimulating current may not be a reliable indicator of intraneural needle placement. Reg Anesth Pain Med 33:207–210CrossRefPubMed
5.
Zurück zum Zitat Wiesmann T, Bornträger A, Vassiliou T et al (2014) Minimal current intensity to elicit an evoked motor response cannot discern between needle-nerve contact and intraneural needle insertion. Anesth Analg 118:681–686CrossRefPubMed Wiesmann T, Bornträger A, Vassiliou T et al (2014) Minimal current intensity to elicit an evoked motor response cannot discern between needle-nerve contact and intraneural needle insertion. Anesth Analg 118:681–686CrossRefPubMed
6.
Zurück zum Zitat Bigeleisen PE, Moayeri N, Groen GJ (2009) Extraneural versus intraneural stimulation thresholds during ultrasound-guided supraclavicular block. Anesthesiology 110:1235–1243CrossRefPubMed Bigeleisen PE, Moayeri N, Groen GJ (2009) Extraneural versus intraneural stimulation thresholds during ultrasound-guided supraclavicular block. Anesthesiology 110:1235–1243CrossRefPubMed
7.
Zurück zum Zitat Bigeleisen PE (2006) Nerve puncture and apparent intraneural injection during ultrasound-guided axillary block does not invariably result in neurologic injury. Anesthesiology 105:779–783CrossRefPubMed Bigeleisen PE (2006) Nerve puncture and apparent intraneural injection during ultrasound-guided axillary block does not invariably result in neurologic injury. Anesthesiology 105:779–783CrossRefPubMed
8.
Zurück zum Zitat Steinfeldt T, Schwemmer U, Volk T et al (2013) Handlungsempfehlung—Nervenlokalisation in der Regionalanästhesie. Anaesth Intensivmed 54:662–666 Steinfeldt T, Schwemmer U, Volk T et al (2013) Handlungsempfehlung—Nervenlokalisation in der Regionalanästhesie. Anaesth Intensivmed 54:662–666
9.
Zurück zum Zitat Magora F, Rozin R, Ben-Menachem Y et al (1969) Obturator nerve block: an evaluation of technique. Br J Anaesth 41:695–698CrossRefPubMed Magora F, Rozin R, Ben-Menachem Y et al (1969) Obturator nerve block: an evaluation of technique. Br J Anaesth 41:695–698CrossRefPubMed
10.
Zurück zum Zitat Sites BD, Brull R, Chan VW et al (2010) Artifacts and pitfall errors associated with ultrasound-guided regional anesthesia: Part II: a pictorial approach to understanding and avoidance. Reg Anesth Pain Med 35:S81–S92CrossRefPubMed Sites BD, Brull R, Chan VW et al (2010) Artifacts and pitfall errors associated with ultrasound-guided regional anesthesia: Part II: a pictorial approach to understanding and avoidance. Reg Anesth Pain Med 35:S81–S92CrossRefPubMed
11.
Zurück zum Zitat Schafhalter-Zoppoth I, McCulloch CE, Gray AT (2004) Ultrasound visibility of needles used for regional nerve block: an in vitro study. Reg Anesth Pain Med 29:480–488CrossRefPubMed Schafhalter-Zoppoth I, McCulloch CE, Gray AT (2004) Ultrasound visibility of needles used for regional nerve block: an in vitro study. Reg Anesth Pain Med 29:480–488CrossRefPubMed
12.
Zurück zum Zitat Steinfeldt T, Nimphius W, Werner T et al (2010) Nerve injury by needle nerve perforation in regional anaesthesia: does size matter? Br J Anaesth 104:245–253CrossRefPubMed Steinfeldt T, Nimphius W, Werner T et al (2010) Nerve injury by needle nerve perforation in regional anaesthesia: does size matter? Br J Anaesth 104:245–253CrossRefPubMed
13.
Zurück zum Zitat Chin KJ, Perlas A, Chan VW et al (2008) Needle visualization in ultrasound-guided regional anesthesia: challenges and solutions. Reg Anesth Pain Med 33:532–544PubMed Chin KJ, Perlas A, Chan VW et al (2008) Needle visualization in ultrasound-guided regional anesthesia: challenges and solutions. Reg Anesth Pain Med 33:532–544PubMed
14.
Zurück zum Zitat Maecken T, Zenz M, Grau T (2007) Ultrasound characteristics of needles for regional anesthesia. Reg Anesth Pain Med 32:440–447CrossRefPubMed Maecken T, Zenz M, Grau T (2007) Ultrasound characteristics of needles for regional anesthesia. Reg Anesth Pain Med 32:440–447CrossRefPubMed
15.
Zurück zum Zitat Reusz G, Sarkany P, Gal J, Csomos A (2014) Needle-related ultrasound artifacts and their importance in anaesthetic practice. Br J Anaesth 112:794–802CrossRefPubMed Reusz G, Sarkany P, Gal J, Csomos A (2014) Needle-related ultrasound artifacts and their importance in anaesthetic practice. Br J Anaesth 112:794–802CrossRefPubMed
16.
Zurück zum Zitat Hebard S, Hocking G (2011) Echogenic technology can improve needle visibility during ultrasound-guided regional anesthesia. Reg Anesth Pain Med 36:185–189CrossRefPubMed Hebard S, Hocking G (2011) Echogenic technology can improve needle visibility during ultrasound-guided regional anesthesia. Reg Anesth Pain Med 36:185–189CrossRefPubMed
17.
Zurück zum Zitat Guo S, Schwab A, McLeod G et al (2012) Echogenic regional anaesthesia needles: a comparison study in Thiel cadavers. Ultrasound Med Biol 38:702–707CrossRefPubMed Guo S, Schwab A, McLeod G et al (2012) Echogenic regional anaesthesia needles: a comparison study in Thiel cadavers. Ultrasound Med Biol 38:702–707CrossRefPubMed
18.
Zurück zum Zitat Ruiz A, Sala-Blanch X, Martinez-Ocón J et al (2014) Incidence of intraneural needle insertion in ultrasound-guided femoral nerve block: a comparison between the out-of-plane versus the in-plane approaches. Rev Esp Anestesiol Reanim 61:73–77CrossRefPubMed Ruiz A, Sala-Blanch X, Martinez-Ocón J et al (2014) Incidence of intraneural needle insertion in ultrasound-guided femoral nerve block: a comparison between the out-of-plane versus the in-plane approaches. Rev Esp Anestesiol Reanim 61:73–77CrossRefPubMed
19.
Zurück zum Zitat Moayeri N, Krediet AC, Welleweerd JC et al (2012) Early ultrasonographic detection of low-volume intraneural injection. Br J Anaesth 109:432–438CrossRefPubMed Moayeri N, Krediet AC, Welleweerd JC et al (2012) Early ultrasonographic detection of low-volume intraneural injection. Br J Anaesth 109:432–438CrossRefPubMed
20.
Zurück zum Zitat Tsui BC, Wagner A, Finucane B (2004) Electrophysiologic effect of injectates on peripheral nerve stimulation. Reg Anesth Pain Med 29:189–193CrossRefPubMed Tsui BC, Wagner A, Finucane B (2004) Electrophysiologic effect of injectates on peripheral nerve stimulation. Reg Anesth Pain Med 29:189–193CrossRefPubMed
21.
Zurück zum Zitat Sala-Blanch X, Vandepitte C, Laur JJ et al (2011) A practical review of perineural versus intraneural injections: a call for standard nomenclature. Int Anesthesiol Clin 49:1–12CrossRefPubMed Sala-Blanch X, Vandepitte C, Laur JJ et al (2011) A practical review of perineural versus intraneural injections: a call for standard nomenclature. Int Anesthesiol Clin 49:1–12CrossRefPubMed
22.
Zurück zum Zitat Vassiliou T, Müller HH, Ellert A et al (2013) High- versus low-stimulation current threshold for axillary plexus blocks: a prospective randomized triple-blinded noninferiority trial in 205 patients. Anesth Analg 116:247–254CrossRefPubMed Vassiliou T, Müller HH, Ellert A et al (2013) High- versus low-stimulation current threshold for axillary plexus blocks: a prospective randomized triple-blinded noninferiority trial in 205 patients. Anesth Analg 116:247–254CrossRefPubMed
23.
Zurück zum Zitat Sites BD, Gallagher J, Sparks M (2003) Ultrasound-guided popliteal block demonstrates an atypical motor response to nerve stimulation in 2 patients with diabetes mellitus. Reg Anesth Pain Med 28:479–482CrossRefPubMed Sites BD, Gallagher J, Sparks M (2003) Ultrasound-guided popliteal block demonstrates an atypical motor response to nerve stimulation in 2 patients with diabetes mellitus. Reg Anesth Pain Med 28:479–482CrossRefPubMed
24.
Zurück zum Zitat Adler AI, Boyko EJ, Ahroni JH et al (1997) Risk factors for diabetic peripheral sensory neuropathy. Results of the Seattle Prospective Diabetic Foot Study. Diabetes Care 20:1162–1167CrossRefPubMed Adler AI, Boyko EJ, Ahroni JH et al (1997) Risk factors for diabetic peripheral sensory neuropathy. Results of the Seattle Prospective Diabetic Foot Study. Diabetes Care 20:1162–1167CrossRefPubMed
25.
Zurück zum Zitat Neuburger M, Rotzinger M, Kaiser H (2001) Electric nerve stimulation in relation to impulse strength. A quantitative study of the distance of the electrode point to the nerve. Anaesthesist 50:181–186CrossRefPubMed Neuburger M, Rotzinger M, Kaiser H (2001) Electric nerve stimulation in relation to impulse strength. A quantitative study of the distance of the electrode point to the nerve. Anaesthesist 50:181–186CrossRefPubMed
26.
Zurück zum Zitat Tsui BC, Pillay JJ, Chu KT et al (2008) Electrical impedance to distinguish intraneural from extraneural needle placement in porcine nerves during direct exposure and ultrasound guidance. Anesthesiology 109:479–483CrossRefPubMed Tsui BC, Pillay JJ, Chu KT et al (2008) Electrical impedance to distinguish intraneural from extraneural needle placement in porcine nerves during direct exposure and ultrasound guidance. Anesthesiology 109:479–483CrossRefPubMed
27.
Zurück zum Zitat Ercole A (2008) The effect of injectate conductivity on the electric field with the nerve stimulator needle: a computer simulation. Anesth Analg 107:1427–1432CrossRefPubMed Ercole A (2008) The effect of injectate conductivity on the electric field with the nerve stimulator needle: a computer simulation. Anesth Analg 107:1427–1432CrossRefPubMed
28.
Zurück zum Zitat Tsui BC, Kropelin B (2005) The electrophysiological effect of dextrose 5 % in water on single-shot peripheral nerve stimulation. Anesth Analg 100:1837–1839CrossRefPubMed Tsui BC, Kropelin B (2005) The electrophysiological effect of dextrose 5 % in water on single-shot peripheral nerve stimulation. Anesth Analg 100:1837–1839CrossRefPubMed
29.
Zurück zum Zitat Neal JM, Bernards CM, Hadzic A et al (2008) ASRA practice advisory on neurologic complications in regional anesthesia and pain medicine. Reg Anesth Pain Med 33:404–415PubMedCentralCrossRefPubMed Neal JM, Bernards CM, Hadzic A et al (2008) ASRA practice advisory on neurologic complications in regional anesthesia and pain medicine. Reg Anesth Pain Med 33:404–415PubMedCentralCrossRefPubMed
30.
Zurück zum Zitat Klein SM, Melton MS, Grill WM et al (2012) Peripheral nerve stimulation in regional anesthesia. Reg Anesth Pain Med 37:383–392CrossRefPubMed Klein SM, Melton MS, Grill WM et al (2012) Peripheral nerve stimulation in regional anesthesia. Reg Anesth Pain Med 37:383–392CrossRefPubMed
31.
Zurück zum Zitat Koscielniak-Nielsen ZJ, Rassmussen H, Jepsen K (2001) Effect of impulse duration on patients’ perception of electrical stimulation and block effectiveness during axillary block in unsedated ambulatory patients. Reg Anesth Pain Med 26:428–433CrossRefPubMed Koscielniak-Nielsen ZJ, Rassmussen H, Jepsen K (2001) Effect of impulse duration on patients’ perception of electrical stimulation and block effectiveness during axillary block in unsedated ambulatory patients. Reg Anesth Pain Med 26:428–433CrossRefPubMed
32.
Zurück zum Zitat Hadzic A, Vloka JD, Claudio RE et al (2004) Electrical nerve localization: effects of cutaneous electrode placement and duration of the stimulus on motor response. Anesthesiology 100:1526–1530CrossRefPubMed Hadzic A, Vloka JD, Claudio RE et al (2004) Electrical nerve localization: effects of cutaneous electrode placement and duration of the stimulus on motor response. Anesthesiology 100:1526–1530CrossRefPubMed
33.
Zurück zum Zitat Engelhardt L, Grosse J, Birnbaum J et al (2007) Inhibition of a pacemaker during nerve stimulation for regional anaesthesia. Anaesthesia 62:1071–1074CrossRefPubMed Engelhardt L, Grosse J, Birnbaum J et al (2007) Inhibition of a pacemaker during nerve stimulation for regional anaesthesia. Anaesthesia 62:1071–1074CrossRefPubMed
34.
Zurück zum Zitat LaBan MM, Petty D, Hauser AM et al (1988) Peripheral nerve conduction stimulation: its effect on cardiac pacemakers. Arch Phys Med Rehabil 69:358–362PubMed LaBan MM, Petty D, Hauser AM et al (1988) Peripheral nerve conduction stimulation: its effect on cardiac pacemakers. Arch Phys Med Rehabil 69:358–362PubMed
35.
Zurück zum Zitat Neuburger M, Gültlinger O, Ass B et al (2005) Influence of blockades with local anesthetics on the stimulation ability of a nerve by peripheral nerve stimulation. Results of a randomized study. Anaesthesist 54:575–577CrossRefPubMed Neuburger M, Gültlinger O, Ass B et al (2005) Influence of blockades with local anesthetics on the stimulation ability of a nerve by peripheral nerve stimulation. Results of a randomized study. Anaesthesist 54:575–577CrossRefPubMed
36.
Zurück zum Zitat Orebaugh SL, Kentor ML, Williams BA (2012) Adverse outcomes associated with nerve stimulator-guided and ultrasound-guided peripheral nerve blocks by supervised trainees: update of a single-site database. Reg Anesth Pain Med 37:577–582PubMedCentralCrossRefPubMed Orebaugh SL, Kentor ML, Williams BA (2012) Adverse outcomes associated with nerve stimulator-guided and ultrasound-guided peripheral nerve blocks by supervised trainees: update of a single-site database. Reg Anesth Pain Med 37:577–582PubMedCentralCrossRefPubMed
37.
Zurück zum Zitat Barrington MJ, Watts SA, Gledhill SR et al (2009) Preliminary results of the Australasian Regional Anaesthesia Collaboration: a prospective audit of more than 7000 peripheral nerve and plexus blocks for neurologic and other complications. Reg Anesth Pain Med 34:534–541CrossRefPubMed Barrington MJ, Watts SA, Gledhill SR et al (2009) Preliminary results of the Australasian Regional Anaesthesia Collaboration: a prospective audit of more than 7000 peripheral nerve and plexus blocks for neurologic and other complications. Reg Anesth Pain Med 34:534–541CrossRefPubMed
38.
Zurück zum Zitat Beach ML, Sites BD, Gallagher JD (2006) Use of a nerve stimulator does not improve the efficacy of ultrasound-guided supraclavicular nerve blocks. J Clin Anesth 18:580–584CrossRefPubMed Beach ML, Sites BD, Gallagher JD (2006) Use of a nerve stimulator does not improve the efficacy of ultrasound-guided supraclavicular nerve blocks. J Clin Anesth 18:580–584CrossRefPubMed
39.
Zurück zum Zitat Habicher M, Ocken M, Birnbaum J et al (2009) Electrical nerve stimulation for peripheral nerve blocks. Ultrasound-guided needle positioning and effect of 5 % glucose injection. Anaesthesist 58:986–991CrossRefPubMed Habicher M, Ocken M, Birnbaum J et al (2009) Electrical nerve stimulation for peripheral nerve blocks. Ultrasound-guided needle positioning and effect of 5 % glucose injection. Anaesthesist 58:986–991CrossRefPubMed
40.
Zurück zum Zitat Brull R, McCartney CJ, Chan VW et al (2007) Neurological complications after regional anesthesia: contemporary estimates of risk. Anesth Analg 104:965–974CrossRefPubMed Brull R, McCartney CJ, Chan VW et al (2007) Neurological complications after regional anesthesia: contemporary estimates of risk. Anesth Analg 104:965–974CrossRefPubMed
41.
Zurück zum Zitat Bashein G, Robertson HT, Kennedy WF Jr (1985) Persistent phrenic nerve paresis following interscalene brachial plexus block. Anesthesiology 63:102–104CrossRefPubMed Bashein G, Robertson HT, Kennedy WF Jr (1985) Persistent phrenic nerve paresis following interscalene brachial plexus block. Anesthesiology 63:102–104CrossRefPubMed
42.
Zurück zum Zitat Kaufman BR, Nystrom E, Nath S et al (2000) Debilitating chronic pain syndromes after presumed intraneural injections. Pain 85:283–286CrossRefPubMed Kaufman BR, Nystrom E, Nath S et al (2000) Debilitating chronic pain syndromes after presumed intraneural injections. Pain 85:283–286CrossRefPubMed
43.
Zurück zum Zitat Auroy Y, Narchi P, Messiah A et al (1997) Serious complications related to regional anesthesia: results of a prospective survey in France. Anesthesiology 87:479–486CrossRefPubMed Auroy Y, Narchi P, Messiah A et al (1997) Serious complications related to regional anesthesia: results of a prospective survey in France. Anesthesiology 87:479–486CrossRefPubMed
44.
Zurück zum Zitat Bernards CM, Hadzic A, Suresh S et al (2008) Regional anesthesia in anesthetized or heavily sedated patients. Reg Anesth Pain Med 33:449–460CrossRefPubMed Bernards CM, Hadzic A, Suresh S et al (2008) Regional anesthesia in anesthetized or heavily sedated patients. Reg Anesth Pain Med 33:449–460CrossRefPubMed
45.
Zurück zum Zitat Kessler P, Steinfeldt T, Gogarten W et al (2013) Peripheral regional anesthesia in patients under general anesthesia: risk assessment with respect to parasthesia, injection pain and nerve damage. Anaesthesist 62:483–488CrossRefPubMed Kessler P, Steinfeldt T, Gogarten W et al (2013) Peripheral regional anesthesia in patients under general anesthesia: risk assessment with respect to parasthesia, injection pain and nerve damage. Anaesthesist 62:483–488CrossRefPubMed
46.
Zurück zum Zitat Mossetti V, Ivani G (2012) Controversial issues in pediatric regional anesthesia. Paediatr Anaesth 22:109–114CrossRefPubMed Mossetti V, Ivani G (2012) Controversial issues in pediatric regional anesthesia. Paediatr Anaesth 22:109–114CrossRefPubMed
Metadaten
Titel
Nerve localization for peripheral regional anesthesia
Recommendations of the German Society of Anaesthesiology and Intensive Care Medicine
verfasst von
T. Steinfeldt, MD
U. Schwemmer
T. Volk
M. Neuburger
T. Wiesmann
A.R. Heller
O. Vicent
A. Stanek
M. Franz
H. Wulf
P. Kessler
Publikationsdatum
01.07.2014
Verlag
Springer Berlin Heidelberg
Erschienen in
Die Anaesthesiologie / Ausgabe 7/2014
Print ISSN: 2731-6858
Elektronische ISSN: 2731-6866
DOI
https://doi.org/10.1007/s00101-014-2343-6

Weitere Artikel der Ausgabe 7/2014

Der Anaesthesist 7/2014 Zur Ausgabe

Kasuistiken

Uvula-Ödem

Update AINS

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.