Reducing glycaemic variability in type 1 diabetes self-management with a continuous glucose monitoring system based on wired enzyme technology

verfasst von: T. Danne, H. W. de Valk, T. Kracht, K. Walte, R. Geldmacher, L. Sölter, W. von dem Berge, Z. K. Welsh, J. R. Bugler, K. Lange, O. Kordonouri

Erschienen in: Diabetologia | Ausgabe 8/2009

Einloggen, um Zugang zu erhalten

Abstract

Aims/hypothesis

This study was designed to investigate the use and impact of a continuous glucose monitoring system (the FreeStyle Navigator) under home-use conditions in the self-management of type 1 diabetes.

Methods

A 20 day masked phase, when real-time data and alarms were not available, was compared with a subsequent 40 day unmasked phase for a number of specified measures of glycaemic variability. HbA1c (measured by DCA 2000) and a hypoglycaemia fear survey were recorded at the start and end of the study.

Results

The study included 48 patients with type 1 diabetes (mean age 35.7 ± 10.9, range 18–61 years; diabetes duration 17.0 ± 9.5 years). Two patients did not complete the study for personal reasons. Comparing masked (all 20 days) and unmasked (last 20 days) phases, the following reductions were seen: time outside euglycaemia from 11.0 to 9.5 h/day (p = 0.002); glucose SD from 3.5 to 3.2 mmol/l (p < 0.001); hyperglycaemic time (>10.0 mmol/l) from 10.3 to 8.9 h/day (p = 0.0035); mean amplitude of glycaemic excursions (peak to nadir) down by 10% (p < 0.001); high blood glucose index down by 18% (p = 0.0014); and glycaemic risk assessment diabetes equation score down by 12% (p = 0.0013). Hypoglycaemic time (<3.9 mmol/l) decreased from 0.70 to 0.64 h/day without statistical significance (p > 0.05). Mean HbA1c fell from 7.6 ± 1.1% at baseline to 7.1 ± 1.1% (p < 0.001). In the hypoglycaemia fear survey, the patients tended to take less snacks at night-time after wearing the sensor.

Conclusions/interpretation

Home use of a continuous glucose monitoring system has a positive effect on the self-management of diabetes. Thus, continuous glucose monitoring may be a useful tool to decrease glycaemic variability.
Literatur
1.
Zurück zum Zitat Golicki DT, Golicka D, Groele L, Pankowska E (2008) Continuous glucose monitoring system in children with type 1 diabetes mellitus: a systematic review and meta-analysis. Diabetologia 51:233–240PubMedCrossRef Golicki DT, Golicka D, Groele L, Pankowska E (2008) Continuous glucose monitoring system in children with type 1 diabetes mellitus: a systematic review and meta-analysis. Diabetologia 51:233–240PubMedCrossRef
2.
Zurück zum Zitat Danne T, Lange K, Kordonouri O (2008) Real-time glucose sensors in children and adolescents with type-1 diabetes. Horm Res 70:193–202PubMedCrossRef Danne T, Lange K, Kordonouri O (2008) Real-time glucose sensors in children and adolescents with type-1 diabetes. Horm Res 70:193–202PubMedCrossRef
3.
Zurück zum Zitat Feldman B, Brazg R, Schwartz S, Weinstein R (2003) A continuous glucose sensor based on wired enzyme technology: results from a 3-day trial in patients with type 1 diabetes. Diabetes Technol Ther 5:769–779PubMedCrossRef Feldman B, Brazg R, Schwartz S, Weinstein R (2003) A continuous glucose sensor based on wired enzyme technology: results from a 3-day trial in patients with type 1 diabetes. Diabetes Technol Ther 5:769–779PubMedCrossRef
4.
Zurück zum Zitat Juvenile Diabetes Research Foundation Continuous Glucose Monitoring Study Group, Tamborlane WV, Beck RW, Bode BW et al (2008) Continuous glucose monitoring and intensive treatment of type 1 diabetes. N Engl J Med 359:1464–1476CrossRef Juvenile Diabetes Research Foundation Continuous Glucose Monitoring Study Group, Tamborlane WV, Beck RW, Bode BW et al (2008) Continuous glucose monitoring and intensive treatment of type 1 diabetes. N Engl J Med 359:1464–1476CrossRef
5.
Zurück zum Zitat Monnier L, Colette C (2008) Glycemic variability: should we and can we prevent it? Diabetes Care 31:S150–S154PubMedCrossRef Monnier L, Colette C (2008) Glycemic variability: should we and can we prevent it? Diabetes Care 31:S150–S154PubMedCrossRef
6.
Zurück zum Zitat Krinsley JS (2008) Glycemic variability: a strong independent predictor of mortality in critically ill patients. Crit Care Med 36:3008–3013PubMedCrossRef Krinsley JS (2008) Glycemic variability: a strong independent predictor of mortality in critically ill patients. Crit Care Med 36:3008–3013PubMedCrossRef
7.
Zurück zum Zitat Service FJ, Molnar GD, Rosevear, Ackermann E, Gatweood LC, Taylor WF (1970) Mean amplitude of glycemic excursions, a measure of diabetic instability. Diabetes 19:644–655 Service FJ, Molnar GD, Rosevear, Ackermann E, Gatweood LC, Taylor WF (1970) Mean amplitude of glycemic excursions, a measure of diabetic instability. Diabetes 19:644–655
8.
Zurück zum Zitat Monnier L, Mas E, Ginet C et al (2006) Activation of oxidative stress by acute glucose fluctuations compared with sustained chronic hyperglycemia in patients with type 2 diabetes. JAMA 295:1681–1687PubMedCrossRef Monnier L, Mas E, Ginet C et al (2006) Activation of oxidative stress by acute glucose fluctuations compared with sustained chronic hyperglycemia in patients with type 2 diabetes. JAMA 295:1681–1687PubMedCrossRef
9.
Zurück zum Zitat McDonnell CM, Northam EA, Donath SM, Werther GA, Cameron FJ (2007) Hyperglycemia and externalizing behavior in children with type 1 diabetes. Diabetes Care 30:2211–2215PubMedCrossRef McDonnell CM, Northam EA, Donath SM, Werther GA, Cameron FJ (2007) Hyperglycemia and externalizing behavior in children with type 1 diabetes. Diabetes Care 30:2211–2215PubMedCrossRef
10.
Zurück zum Zitat Frier BM (2008) How hypoglycaemia can affect the life of a person with diabetes. Diabetes Metab Res Rev 24:87–92PubMedCrossRef Frier BM (2008) How hypoglycaemia can affect the life of a person with diabetes. Diabetes Metab Res Rev 24:87–92PubMedCrossRef
11.
Zurück zum Zitat Reach G (2008) Continuous glucose monitoring and diabetes health outcomes: a critical appraisal. Diabetes Technol Ther 10:69–80PubMedCrossRef Reach G (2008) Continuous glucose monitoring and diabetes health outcomes: a critical appraisal. Diabetes Technol Ther 10:69–80PubMedCrossRef
12.
Zurück zum Zitat Danne T, Lange K, Kordonouri O (2007) New developments in the treatment of type 1 diabetes in children. Arch Dis Child 92:15–19CrossRef Danne T, Lange K, Kordonouri O (2007) New developments in the treatment of type 1 diabetes in children. Arch Dis Child 92:15–19CrossRef
13.
Zurück zum Zitat Fava S (2008) Role of postprandial hyperglycemia in cardiovascular disease. Expert Rev Cardiovasc Ther 6:859–872PubMedCrossRef Fava S (2008) Role of postprandial hyperglycemia in cardiovascular disease. Expert Rev Cardiovasc Ther 6:859–872PubMedCrossRef
14.
Zurück zum Zitat Brownlee M, Hirsch IB (2006) Glycemic variability: a hemoglobin A1c-independent risk factor for diabetic complications. JAMA 295:1707–1708PubMedCrossRef Brownlee M, Hirsch IB (2006) Glycemic variability: a hemoglobin A1c-independent risk factor for diabetic complications. JAMA 295:1707–1708PubMedCrossRef
15.
Zurück zum Zitat Hirsch IB, Armstrong D, Bergenstal RM et al (2008) Clinical application of emerging sensor technologies in diabetes management: consensus guidelines for continuous glucose monitoring (CGM). Diabetes Technol Ther 10:232–244PubMedCrossRef Hirsch IB, Armstrong D, Bergenstal RM et al (2008) Clinical application of emerging sensor technologies in diabetes management: consensus guidelines for continuous glucose monitoring (CGM). Diabetes Technol Ther 10:232–244PubMedCrossRef
16.
Zurück zum Zitat Cox DJ, Irvine A, Gonder-Frederick L, Nowacek G, Butterfield J (1987) Fear of hypoglycemia: quantification, validation, and utilization. Diabetes Care 10:617–621PubMedCrossRef Cox DJ, Irvine A, Gonder-Frederick L, Nowacek G, Butterfield J (1987) Fear of hypoglycemia: quantification, validation, and utilization. Diabetes Care 10:617–621PubMedCrossRef
17.
Zurück zum Zitat Kubiak T, Hermanns N, Schreckling HJ, Kulzer B, Haak T (2006) Evaluation of a self-management-based patient education programme for the treatment and prevention of hypoglycaemia-related problems in type 1 diabetes. Patient Educ Couns 60:228–234PubMedCrossRef Kubiak T, Hermanns N, Schreckling HJ, Kulzer B, Haak T (2006) Evaluation of a self-management-based patient education programme for the treatment and prevention of hypoglycaemia-related problems in type 1 diabetes. Patient Educ Couns 60:228–234PubMedCrossRef
18.
Zurück zum Zitat Bode B, Silver M, Weiss R, Martin K (2008) Evaluation of a continuous glucose system for home-use conditions. Manag Care 17:40–45PubMed Bode B, Silver M, Weiss R, Martin K (2008) Evaluation of a continuous glucose system for home-use conditions. Manag Care 17:40–45PubMed
19.
Zurück zum Zitat Clarke WL, Breton M, Brayman K, McCall A (2005) Quantifying temporal glucose variability in diabetes via continuous glucose monitoring: mathematical methods and clinical application. Diabetes Technol Ther 7:849–862PubMedCrossRef Clarke WL, Breton M, Brayman K, McCall A (2005) Quantifying temporal glucose variability in diabetes via continuous glucose monitoring: mathematical methods and clinical application. Diabetes Technol Ther 7:849–862PubMedCrossRef
20.
Zurück zum Zitat Hill NR, Hindmarsh PC, Stevens RJ, Stratton IM, Levy JC, Matthews DR (2007) A method for assessing quality of control from glucose profiles. Diabet Med 24:753–758PubMedCrossRef Hill NR, Hindmarsh PC, Stevens RJ, Stratton IM, Levy JC, Matthews DR (2007) A method for assessing quality of control from glucose profiles. Diabet Med 24:753–758PubMedCrossRef
21.
Zurück zum Zitat Deiss D, Bolinder J, Riveline JP et al (2006) Improved glycemic control in poorly controlled patients with type 1 diabetes using real-time continuous glucose monitoring. Diabetes Care 29:2730–2732PubMedCrossRef Deiss D, Bolinder J, Riveline JP et al (2006) Improved glycemic control in poorly controlled patients with type 1 diabetes using real-time continuous glucose monitoring. Diabetes Care 29:2730–2732PubMedCrossRef
22.
Zurück zum Zitat Halvorson M, Carpenter S, Kaiserman K, Kaufman F (2007) A pilot trial in pediatrics with the sensor-augmented pump: combining real-time continuous glucose monitoring with the insulin pump. J Pediatr 150:103–105PubMedCrossRef Halvorson M, Carpenter S, Kaiserman K, Kaufman F (2007) A pilot trial in pediatrics with the sensor-augmented pump: combining real-time continuous glucose monitoring with the insulin pump. J Pediatr 150:103–105PubMedCrossRef
23.
Zurück zum Zitat Hirsch IB, Abelseth J, Bode BW et al (2008) Sensor-augmented insulin pump therapy: results of the first randomized treat-to-target study. Diabetes Technol Ther 10:377–383PubMedCrossRef Hirsch IB, Abelseth J, Bode BW et al (2008) Sensor-augmented insulin pump therapy: results of the first randomized treat-to-target study. Diabetes Technol Ther 10:377–383PubMedCrossRef
24.
Zurück zum Zitat The Juvenile Diabetes Research Foundation Continuous Glucose Monitoring Study Group (2009) Continuous glucose monitoring and type 1 diabetes. N Engl J Med 360:190–192CrossRef The Juvenile Diabetes Research Foundation Continuous Glucose Monitoring Study Group (2009) Continuous glucose monitoring and type 1 diabetes. N Engl J Med 360:190–192CrossRef
25.
Zurück zum Zitat Wilson DM, Beck RW, Tamborlane WV et al; DirecNet Study Group (2007) The accuracy of the FreeStyle Navigator continuous glucose monitoring system in children with type 1 diabetes. Diabetes Care 30:59–64CrossRef Wilson DM, Beck RW, Tamborlane WV et al; DirecNet Study Group (2007) The accuracy of the FreeStyle Navigator continuous glucose monitoring system in children with type 1 diabetes. Diabetes Care 30:59–64CrossRef
26.
Zurück zum Zitat Wilson D, Chase HP, Kollman C, Diabetes Research in Children Network (DirecNet) Study Group et al (2008) Low-fat vs. high-fat bedtime snacks in children and adolescents with type 1 diabetes. Pediatr Diabetes 9:320–325PubMedCrossRef Wilson D, Chase HP, Kollman C, Diabetes Research in Children Network (DirecNet) Study Group et al (2008) Low-fat vs. high-fat bedtime snacks in children and adolescents with type 1 diabetes. Pediatr Diabetes 9:320–325PubMedCrossRef
27.
Zurück zum Zitat The Diabetes Research in Children Network (DirecNet) Study Group (2008) The relationship of A1c to glucose concentrations in children with type 1 diabetes: assessments by high frequency glucose determinations by sensors. Diabetes Care 31:381–385CrossRef The Diabetes Research in Children Network (DirecNet) Study Group (2008) The relationship of A1c to glucose concentrations in children with type 1 diabetes: assessments by high frequency glucose determinations by sensors. Diabetes Care 31:381–385CrossRef
Metadaten
Titel
Reducing glycaemic variability in type 1 diabetes self-management with a continuous glucose monitoring system based on wired enzyme technology
verfasst von
T. Danne
H. W. de Valk
T. Kracht
K. Walte
R. Geldmacher
L. Sölter
W. von dem Berge
Z. K. Welsh
J. R. Bugler
K. Lange
O. Kordonouri
Publikationsdatum
01.08.2009
Verlag
Springer-Verlag
Erschienen in
Diabetologia / Ausgabe 8/2009
Print ISSN: 0012-186X
Elektronische ISSN: 1432-0428
DOI
https://doi.org/10.1007/s00125-009-1408-6

Weitere Artikel der Ausgabe 8/2009

Diabetologia 8/2009 Zur Ausgabe

Leitlinien kompakt für die Innere Medizin

Mit medbee Pocketcards sicher entscheiden.

Seit 2022 gehört die medbee GmbH zum Springer Medizin Verlag

Update Innere Medizin

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.