Erschienen in:
01.06.2006 | Brief Report
Continuous cardiac output monitoring after cardiopulmonary bypass: a comparison with bolus thermodilution measurement
verfasst von:
Karim Bendjelid, Nicolas Schütz, Peter M. Suter, Jacques-Andre Romand
Erschienen in:
Intensive Care Medicine
|
Ausgabe 6/2006
Einloggen, um Zugang zu erhalten
Abstract
Objective
The interchangeability of continuous measurement of
cardiac output (CO) with the traditional bolus method in patients after
cardiopulmonary bypass (CPB) is uncertain.
Design
Prospective observational clinical study.
Setting
A 20-bed surgical ICU at a university
hospital.
Patients
Fourteen deeply sedated, ventilated, post-cardiac surgery
patients, all equipped with a pulmonary artery catheter.
Interventions
Six hours after the end of the CPB, 56 simultaneous
bolus and continuous measurements were compared by a linear regression
analysis and Bland–Altman analysis. Bolus CO was estimated by averaging
triplicate injections of 10 ml room-temperature NaCl 0.9%, delivered
randomly during the respiratory cycle. A stringent maximum difference of
0.55 l min—1 (about 10% of the mean bolus measured) was considered
as a clinically acceptable agreement between the two types of measurements.
To be interchangeable the limits of agreement (± 2 SD of the mean
difference between the two methods) should not exceed the chosen acceptable
difference.
Measurements and results
Continuous was correlated with bolus CO,
with a correlation coefficient of r2 = 0.68. (p< 0.01). The Bland–Altman
analysis demonstrated an objective mean bias of 0.33 ± 0.6 l min–1
(confidence interval of –0.87 – 1.58) with 34% of measured values
falling outside of the clinically acceptable limits.
Conclusion
Our results suggest that, in the first 6 h after CPB,
continuous and bolus CO determinations are not interchangeable; one third of
the values obtained by continuous CO fell outside the strict limits of
clinically useful precision.