A method for monitoring and improving patient: ventilator interaction

verfasst von: Magdy Younes, Laurent Brochard, Salvatore Grasso, John Kun, Jordi Mancebo, Marco Ranieri, Jean-Christophe Richard, Henry Younes

Erschienen in: Intensive Care Medicine | Ausgabe 8/2007

Einloggen, um Zugang zu erhalten

Abstract

Objective

To evaluate a new approach for monitoring and improving patient-ventilator interaction that utilizes a signal generated by the equation of motion, using improvised values for resistance and elastance obtained noninvasively.

Design and setting

Observational study in intensive care units in five European centers.

Patients

We studied 21 stable patients instrumented with esophageal/gastric catheters for a previous study and ventilated alternately with pressure support (PSV) and proportional assist (PAV) ventilation with a Tyco 840 ventilator.

Measurements and results

Previously recorded digital files were analyzed in real-time by a prototype incorporating the new technology (PVI monitor, YRT, Winnipeg, Canada). Actual onsets (PDI-TONSET) and ends (PDI-TEND) of inspiratory efforts, ineffective efforts, and patient respiratory rate were identified visually from transdiaphragmatic or calculated respiratory muscle pressure. Monitor-identified TONSET occurred 0.107 ± 0.074 s after PDI-TONSET, substantially less than trigger delay observed with conventional triggering (0.326 ± 0.086 s). End of effort was identified 0.097 ± 0.096 s after PDI-TEND, significantly less than actual cycling-off delay during PSV (0.486 ± 0.307 s) or PAV (0.277 ± 0.084 s). The monitor detected 80% of ineffective efforts. There was excellent agreement between monitor-estimated respiratory rate and actual patient rate over a wide range (17–59/min) of patient rates (mean ( ± SD) of difference −0.2 ± 1.9/min for pressure support and 0.2 ± 0.9/min for proportional assist) even when large discrepancies existed (> 35/min) between patient and ventilator rates.

Conclusions

The proposed approach should make it possible to improve patient-ventilator interaction and to obtain accurate estimates of true patient respiratory rate when there is nonsynchrony.
Anhänge
Nur mit Berechtigung zugänglich
Literatur
1.
Zurück zum Zitat Leung P, Jubran A, Tobin MJ (1997) Comparison of assisted ventilator modes on triggering, patient effort, and dyspnea. Am J Respir Crit Care Med 155:1940–1948CrossRef Leung P, Jubran A, Tobin MJ (1997) Comparison of assisted ventilator modes on triggering, patient effort, and dyspnea. Am J Respir Crit Care Med 155:1940–1948CrossRef
2.
Zurück zum Zitat Tobin MJ, Jubran A, Laghi F (2001) Patient-ventilator interaction. Am J Respir Crit Care Med 163:1059–1063CrossRef Tobin MJ, Jubran A, Laghi F (2001) Patient-ventilator interaction. Am J Respir Crit Care Med 163:1059–1063CrossRef
3.
Zurück zum Zitat Kondili E, Prinianakis G, Georgopoulos D (2003) Patient-ventilator interaction. Br J Anaesthesia 91:106–119CrossRef Kondili E, Prinianakis G, Georgopoulos D (2003) Patient-ventilator interaction. Br J Anaesthesia 91:106–119CrossRef
4.
Zurück zum Zitat Prinianakis G, Kondili E, Georgopoulos D (2003) Effects of the flow waveform method of triggering and cycling on patient-ventilator interaction during pressure support. Intensive Care Med 29:1950–1959CrossRef Prinianakis G, Kondili E, Georgopoulos D (2003) Effects of the flow waveform method of triggering and cycling on patient-ventilator interaction during pressure support. Intensive Care Med 29:1950–1959CrossRef
5.
Zurück zum Zitat Giannouli E, Webster K, Roberts D, Younes M (1999) Response of ventilator-dependent patients to different levels of pressure support and proportional assist. Am J Respir Crit Care Med 159:1716–1725CrossRef Giannouli E, Webster K, Roberts D, Younes M (1999) Response of ventilator-dependent patients to different levels of pressure support and proportional assist. Am J Respir Crit Care Med 159:1716–1725CrossRef
6.
Zurück zum Zitat Thille AW, Rodriguez P, Cabello B, Lellouche F, Brochard L (2006) Patient-ventilator asynchrony during assisted mechanical ventilation. Intensive Care Med 32:1515–1522CrossRef Thille AW, Rodriguez P, Cabello B, Lellouche F, Brochard L (2006) Patient-ventilator asynchrony during assisted mechanical ventilation. Intensive Care Med 32:1515–1522CrossRef
7.
Zurück zum Zitat Younes M (2006) Proportional assist ventilation. In: Tobin MJ (ed) Principles and practice of mechanical ventilation, 2nd edn. McGraw-Hill, New York, pp 335–364 Younes M (2006) Proportional assist ventilation. In: Tobin MJ (ed) Principles and practice of mechanical ventilation, 2nd edn. McGraw-Hill, New York, pp 335–364
8.
Zurück zum Zitat Younes M (2003) Method and device for monitoring and improving patient-ventilator interaction. PCT application #PCT/CA 2003/000976 Younes M (2003) Method and device for monitoring and improving patient-ventilator interaction. PCT application #PCT/CA 2003/000976
9.
Zurück zum Zitat Mead J, Agostoni E (1964) Dynamics of breathing. In: Fenn WO, Rahn H (eds) Handbook of physiology: respiration. American Physiological Society. Washington, pp 411–427 Mead J, Agostoni E (1964) Dynamics of breathing. In: Fenn WO, Rahn H (eds) Handbook of physiology: respiration. American Physiological Society. Washington, pp 411–427
10.
Zurück zum Zitat Younes M (1992) Proportional assist ventilation, a new approach to ventilatory support. Theory. Am Rev Respir Dis 145:114–120CrossRef Younes M (1992) Proportional assist ventilation, a new approach to ventilatory support. Theory. Am Rev Respir Dis 145:114–120CrossRef
11.
Zurück zum Zitat Grasso S, Ranieri VM, Brochard L, Richard GC, Mancebo J, Slutsky AS, Younes M (2001) Closed loop proportional assist ventilation (PAV): results of a phase II multicenter trial. Am J Respir Crit Care Med 163:303A Grasso S, Ranieri VM, Brochard L, Richard GC, Mancebo J, Slutsky AS, Younes M (2001) Closed loop proportional assist ventilation (PAV): results of a phase II multicenter trial. Am J Respir Crit Care Med 163:303A
12.
Zurück zum Zitat Campbell EJM (1958) The respiratory muscles and mechanics of breathing. Loyd-Luke, London Campbell EJM (1958) The respiratory muscles and mechanics of breathing. Loyd-Luke, London
13.
Zurück zum Zitat Appendini L, Patessio A, Zanaboni S, Carone M, Gukov B, Donner CF, Rossi A (1994) Physiologic effects of positive end-expiratory pressure and mask pressure support during exacerbations of chronic obstructive pulmonary disease. Am J Respir Crit Care Med 149:1069–1076CrossRef Appendini L, Patessio A, Zanaboni S, Carone M, Gukov B, Donner CF, Rossi A (1994) Physiologic effects of positive end-expiratory pressure and mask pressure support during exacerbations of chronic obstructive pulmonary disease. Am J Respir Crit Care Med 149:1069–1076CrossRef
14.
Zurück zum Zitat Younes M (1993) Patient-ventilator interaction with pressure-assisted modalities of ventilatory support. Semin Respir Med 14:299–322CrossRef Younes M (1993) Patient-ventilator interaction with pressure-assisted modalities of ventilatory support. Semin Respir Med 14:299–322CrossRef
15.
Zurück zum Zitat Mancebo J (2003) Triggering and cycling off during pressure support ventilation: simplicity or sophistication? Intensive Care Med 29:1871–1872CrossRef Mancebo J (2003) Triggering and cycling off during pressure support ventilation: simplicity or sophistication? Intensive Care Med 29:1871–1872CrossRef
16.
Zurück zum Zitat Imanaka H, Nishimura M, Takeuchi M, Kimball WR, Yahagi N, Kumon K (2000) Autotriggering caused by cardiogenic oscillation during flow-triggered mechanical ventilation. Crit Care Med 28:402–407CrossRef Imanaka H, Nishimura M, Takeuchi M, Kimball WR, Yahagi N, Kumon K (2000) Autotriggering caused by cardiogenic oscillation during flow-triggered mechanical ventilation. Crit Care Med 28:402–407CrossRef
17.
Zurück zum Zitat Younes M, Remmers JE (1981) Control of tidal volume and respiratory frequency. In: Hornbein T (ed) Control of breathing. Lung biology in health and disease, vol. 17 Dekker, New York. pp 617–667 Younes M, Remmers JE (1981) Control of tidal volume and respiratory frequency. In: Hornbein T (ed) Control of breathing. Lung biology in health and disease, vol. 17 Dekker, New York. pp 617–667
18.
Zurück zum Zitat Agostoni E, Citterio G, D'Angelo E (1979) Decay rate of inspiratory muscle pressure during expiration in man. Respir Physiol 36:269–285CrossRef Agostoni E, Citterio G, D'Angelo E (1979) Decay rate of inspiratory muscle pressure during expiration in man. Respir Physiol 36:269–285CrossRef
19.
Zurück zum Zitat Younes M. Kun J, Masiowski B, Webster K, Roberts D (2001) A method for noninvasive determination of inspiratory resistance during proportional assist ventilation. Am J Respir Crit Care Med 163:829–839CrossRef Younes M. Kun J, Masiowski B, Webster K, Roberts D (2001) A method for noninvasive determination of inspiratory resistance during proportional assist ventilation. Am J Respir Crit Care Med 163:829–839CrossRef
20.
Zurück zum Zitat Younes M, Webster K, Kun J, Roberts D, Masiowski B (2001) A method for measuring passive elastance during proportional assist ventilation. Am J Respir Crit Care Med 164:50–60CrossRef Younes M, Webster K, Kun J, Roberts D, Masiowski B (2001) A method for measuring passive elastance during proportional assist ventilation. Am J Respir Crit Care Med 164:50–60CrossRef
21.
Zurück zum Zitat Fernandez F, Mendez M, Younes M (1999) Effect of ventilator flow-rate on respiratory timing in normal subjects. Am J Respir Crit Care Med 159:710–719CrossRef Fernandez F, Mendez M, Younes M (1999) Effect of ventilator flow-rate on respiratory timing in normal subjects. Am J Respir Crit Care Med 159:710–719CrossRef
22.
Zurück zum Zitat Nava S, Bruschi C, Rubini F, Palo A, Iotti G, Braschi A (1995) Respiratory response to inspiratory effort during pressure support ventilation in COPD patients. Intensive Care Med 21:871–879CrossRef Nava S, Bruschi C, Rubini F, Palo A, Iotti G, Braschi A (1995) Respiratory response to inspiratory effort during pressure support ventilation in COPD patients. Intensive Care Med 21:871–879CrossRef
23.
Zurück zum Zitat Yamada Y, Shigeta M, Suwa K, Hanaoka K (1994) Respiratory muscle pressure analysis in pressure-support ventilation. J Appl Physiol 77:2237–2243CrossRef Yamada Y, Shigeta M, Suwa K, Hanaoka K (1994) Respiratory muscle pressure analysis in pressure-support ventilation. J Appl Physiol 77:2237–2243CrossRef
24.
Zurück zum Zitat Younes M, Kun J, Webster K, Roberts D (2002) Response of ventilator dependent patients to delayed opening of exhalation valve. Am J Respir Crit Care Med 166:21–30CrossRef Younes M, Kun J, Webster K, Roberts D (2002) Response of ventilator dependent patients to delayed opening of exhalation valve. Am J Respir Crit Care Med 166:21–30CrossRef
Metadaten
Titel
A method for monitoring and improving patient: ventilator interaction
verfasst von
Magdy Younes
Laurent Brochard
Salvatore Grasso
John Kun
Jordi Mancebo
Marco Ranieri
Jean-Christophe Richard
Henry Younes
Publikationsdatum
01.08.2007
Verlag
Springer Berlin Heidelberg
Erschienen in
Intensive Care Medicine / Ausgabe 8/2007
Print ISSN: 0342-4642
Elektronische ISSN: 1432-1238
DOI
https://doi.org/10.1007/s00134-007-0681-4

Weitere Artikel der Ausgabe 8/2007

Intensive Care Medicine 8/2007 Zur Ausgabe

Announcements

Announcements

Update AINS

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.