Comparison of two methods for enhanced continuous circulatory monitoring in patients with septic shock

verfasst von: F. Spöhr, P. Hettrich, H. Bauer, U. Haas, E. Martin, B. W. Böttiger

Erschienen in: Intensive Care Medicine | Ausgabe 10/2007

Einloggen, um Zugang zu erhalten

Abstract

Objective

To compare a modified pulmonary artery catheter (PAC) and pulse-contour analysis by the PiCCO (Pulsion Medical Systems, Munich, Germany) system for continuous assessment of cardiac output in patients with septic shock. In addition, to assess the relationships between an index of global end-diastolic volume (GEDV) derived by the PiCCO system with traditional PAC-derived indicators of filling: central venous pressure; pulmonary artery occlusion pressure; and right ventricular end-diastolic volume (RVEDV).

Design

Prospective cohort study.

Setting

Surgical intensive care unit of a university hospital.

Patients and participants

14 patients with septic shock.

Interventions

None.

Measurements and results

A significant correlation was found between continuous cardiac output by PAC (CCOPAC) and by pulse-contour analysis (r 2 = 0.714, p < 0.0001), accompanied by a bias of 0.1 l min−1 and a precision of 2.7 l min−1. The correlation between CCOPAC and cardiac output measured by transcardiopulmonary thermodilution was also significant (r 2 = 0.781, p < 0.0001). There was a bias for the two methods of 0.2 l min−1, and a precision of 2.2 l min−1. The GEDV showed no correlation with central venous pressure, pulmonary artery occlusion pressure, or RVEDV.

Conclusion

In patients with septic shock, the averaged bias in continuous measurement of cardiac output by both a modified pulmonary artery catheter and pulse-contour analysis was small, but variability was large. No correlation was found between GEDV and RVEDV. The clinical importance of different cardiac filling parameters needs further investigation.
Anhänge
Nur mit Berechtigung zugänglich
Literatur
1.
Zurück zum Zitat Nishikawa T, Dohi S (1993) Errors in the measurement of cardiac output by thermodilution. Can J Anaesth 40:142–153PubMed Nishikawa T, Dohi S (1993) Errors in the measurement of cardiac output by thermodilution. Can J Anaesth 40:142–153PubMed
2.
Zurück zum Zitat Mermel LA, Maki DG (1994) Infectious complications of Swan-Ganz pulmonary artery catheters. Pathogenesis, epidemiology, prevention, and management. Am J Respir Crit Care Med 149:1020–1036PubMed Mermel LA, Maki DG (1994) Infectious complications of Swan-Ganz pulmonary artery catheters. Pathogenesis, epidemiology, prevention, and management. Am J Respir Crit Care Med 149:1020–1036PubMed
3.
Zurück zum Zitat Boldt J, Menges T, Wollbruck M, Hammermann H, Hempelmann G (1994) Is continuous cardiac output measurement using thermodilution reliable in the critically ill patient? Crit Care Med 22:1913–1918PubMedCrossRef Boldt J, Menges T, Wollbruck M, Hammermann H, Hempelmann G (1994) Is continuous cardiac output measurement using thermodilution reliable in the critically ill patient? Crit Care Med 22:1913–1918PubMedCrossRef
4.
Zurück zum Zitat Böttiger BW, Soder M, Rauch H, Böhrer H, Motsch J, Bauer H, Martin E (1996) Semi-continuous versus injectate cardiac output measurement in intensive care patients after cardiac surgery. Intensive Care Med 22:312–318PubMed Böttiger BW, Soder M, Rauch H, Böhrer H, Motsch J, Bauer H, Martin E (1996) Semi-continuous versus injectate cardiac output measurement in intensive care patients after cardiac surgery. Intensive Care Med 22:312–318PubMed
5.
Zurück zum Zitat Yelderman M (1990) Continuous measurement of cardiac output with the use of stochastic system identification techniques. J Clin Monit 6:322–332PubMedCrossRef Yelderman M (1990) Continuous measurement of cardiac output with the use of stochastic system identification techniques. J Clin Monit 6:322–332PubMedCrossRef
6.
Zurück zum Zitat Vincent JL, Reuse C, Frank N, Contempre B, Kahn RJ (1989) Right ventricular dysfunction in septic shock: assessment by measurements of right ventricular ejection fraction using the thermodilution technique. Acta Anaesthesiol Scand 33:34–38PubMed Vincent JL, Reuse C, Frank N, Contempre B, Kahn RJ (1989) Right ventricular dysfunction in septic shock: assessment by measurements of right ventricular ejection fraction using the thermodilution technique. Acta Anaesthesiol Scand 33:34–38PubMed
7.
Zurück zum Zitat Bossert T, Gummert JF, Bittner HB, Barten M, Walther T, Falk V, Mohr FW (2006) Swan-Ganz catheter-induced severe complications in cardiac surgery: right ventricular perforation, knotting, and rupture of a pulmonary artery. J Cardiac Surg 21:292–295CrossRef Bossert T, Gummert JF, Bittner HB, Barten M, Walther T, Falk V, Mohr FW (2006) Swan-Ganz catheter-induced severe complications in cardiac surgery: right ventricular perforation, knotting, and rupture of a pulmonary artery. J Cardiac Surg 21:292–295CrossRef
8.
Zurück zum Zitat Gödje O, Höke K, Goetz AE, Felbinger TW, Reuter DA, Reichart B, Friedl R, Hannekum A, Pfeiffer UJ (2002) Reliability of a new algorithm for continuous cardiac output determination by pulse-contour analysis during hemodynamic instability. Crit Care Med 30:52–58PubMedCrossRef Gödje O, Höke K, Goetz AE, Felbinger TW, Reuter DA, Reichart B, Friedl R, Hannekum A, Pfeiffer UJ (2002) Reliability of a new algorithm for continuous cardiac output determination by pulse-contour analysis during hemodynamic instability. Crit Care Med 30:52–58PubMedCrossRef
9.
Zurück zum Zitat Lichtwarck-Aschoff M, Beale R, Pfeiffer UJ (1996) Central venous pressure, pulmonary artery occlusion pressure, intrathoracic blood volume, and right ventricular end-diastolic volume as indicators of cardiac preload. J Crit Care 11:180–188PubMedCrossRef Lichtwarck-Aschoff M, Beale R, Pfeiffer UJ (1996) Central venous pressure, pulmonary artery occlusion pressure, intrathoracic blood volume, and right ventricular end-diastolic volume as indicators of cardiac preload. J Crit Care 11:180–188PubMedCrossRef
10.
Zurück zum Zitat Hansen RM, Viquerat CE, Matthay MA, Wiener-Kronish JP, DeMarco T, Bahtia S, Marks JD, Botvinick EH, Chatterjee K (1986) Poor correlation between pulmonary arterial wedge pressure and left ventricular end-diastolic volume after coronary artery bypass graft surgery. Anesthesiology 64:764–770PubMed Hansen RM, Viquerat CE, Matthay MA, Wiener-Kronish JP, DeMarco T, Bahtia S, Marks JD, Botvinick EH, Chatterjee K (1986) Poor correlation between pulmonary arterial wedge pressure and left ventricular end-diastolic volume after coronary artery bypass graft surgery. Anesthesiology 64:764–770PubMed
11.
Zurück zum Zitat Kumar A, Anel R, Bunnell E, Habet K, Zanotti S, Marshall S, Neumann A, Ali A, Cheang M, Kavinsky C, Parrillo JE (2004) Pulmonary artery occlusion pressure and central venous pressure fail to predict ventricular filling volume, cardiac performance, or the response to volume infusion in normal subjects. Crit Care Med 32:691–699PubMedCrossRef Kumar A, Anel R, Bunnell E, Habet K, Zanotti S, Marshall S, Neumann A, Ali A, Cheang M, Kavinsky C, Parrillo JE (2004) Pulmonary artery occlusion pressure and central venous pressure fail to predict ventricular filling volume, cardiac performance, or the response to volume infusion in normal subjects. Crit Care Med 32:691–699PubMedCrossRef
12.
Zurück zum Zitat Reuse C, Vincent JL, Pinsky MR (1990) Measurements of right ventricular volumes during fluid challenge. Chest 98:1450–1454PubMed Reuse C, Vincent JL, Pinsky MR (1990) Measurements of right ventricular volumes during fluid challenge. Chest 98:1450–1454PubMed
13.
Zurück zum Zitat Michard F, Alaya S, Zarka V, Bahloul M, Richard C, Teboul JL (2003) Global end-diastolic volume as an indicator of cardiac preload in patients with septic shock. Chest 124:1900–1908PubMedCrossRef Michard F, Alaya S, Zarka V, Bahloul M, Richard C, Teboul JL (2003) Global end-diastolic volume as an indicator of cardiac preload in patients with septic shock. Chest 124:1900–1908PubMedCrossRef
14.
Zurück zum Zitat Levy MM, Fink MP, Marshall JC, Abraham E, Angus D, Cook D, Cohen J, Opal SM, Vincent JL, Ramsay G (2003) 2001 SCCM/ESICM/ACCP/ATS/SIS International Sepsis Definitions Conference. Crit Care Med 31:1250–1256PubMedCrossRef Levy MM, Fink MP, Marshall JC, Abraham E, Angus D, Cook D, Cohen J, Opal SM, Vincent JL, Ramsay G (2003) 2001 SCCM/ESICM/ACCP/ATS/SIS International Sepsis Definitions Conference. Crit Care Med 31:1250–1256PubMedCrossRef
15.
Zurück zum Zitat Wesseling K, de Wit B, Ty Smith N (1983) A simple device for the continuous measurement of cardiac output. Adv Cardiovasc Phys 5:16–52 Wesseling K, de Wit B, Ty Smith N (1983) A simple device for the continuous measurement of cardiac output. Adv Cardiovasc Phys 5:16–52
16.
Zurück zum Zitat Rödig G, Prasser C, Keyl C, Liebold A, Hobbhahn J (1999) Continuous cardiac output measurement: pulse contour analysis vs thermodilution technique in cardiac surgical patients. Br J Anaesth 82:525–530PubMed Rödig G, Prasser C, Keyl C, Liebold A, Hobbhahn J (1999) Continuous cardiac output measurement: pulse contour analysis vs thermodilution technique in cardiac surgical patients. Br J Anaesth 82:525–530PubMed
17.
Zurück zum Zitat Della Rocca G, Costa MG, Pompei L, Coccia C, Pietropaoli P (2002) Continuous and intermittent cardiac output measurement: pulmonary artery catheter versus aortic transpulmonary technique. Br J Anaesth 88:350–356PubMedCrossRef Della Rocca G, Costa MG, Pompei L, Coccia C, Pietropaoli P (2002) Continuous and intermittent cardiac output measurement: pulmonary artery catheter versus aortic transpulmonary technique. Br J Anaesth 88:350–356PubMedCrossRef
18.
Zurück zum Zitat Della Rocca G, Costa MG, Coccia C, Pompei L, Marco P di, Vilardi V, Pietropaoli P (2003) Cardiac output monitoring: aortic transpulmonary thermodilution and pulse contour analysis agree with standard thermodilution methods in patients undergoing lung transplantation. Can J Anaesth 50:707–711PubMed Della Rocca G, Costa MG, Coccia C, Pompei L, Marco P di, Vilardi V, Pietropaoli P (2003) Cardiac output monitoring: aortic transpulmonary thermodilution and pulse contour analysis agree with standard thermodilution methods in patients undergoing lung transplantation. Can J Anaesth 50:707–711PubMed
19.
Zurück zum Zitat Jakobsen CJ, Melsen NC, Andresen EB (1995) Continuous cardiac output measurements in the perioperative period. Acta Anaesthesiol Scand 39:485–488PubMed Jakobsen CJ, Melsen NC, Andresen EB (1995) Continuous cardiac output measurements in the perioperative period. Acta Anaesthesiol Scand 39:485–488PubMed
20.
Zurück zum Zitat Zink W, Nöll J, Rauch H, Bauer H, Desimone R, Martin E, Böttiger BW (2004) Continuous assessment of right ventricular ejection fraction: new pulmonary artery catheter versus transoesophageal echocardiography. Anaesthesia 59:1126–1132PubMedCrossRef Zink W, Nöll J, Rauch H, Bauer H, Desimone R, Martin E, Böttiger BW (2004) Continuous assessment of right ventricular ejection fraction: new pulmonary artery catheter versus transoesophageal echocardiography. Anaesthesia 59:1126–1132PubMedCrossRef
21.
Zurück zum Zitat Lichtenthal PR, Gordan D (1996) Testing the safety of Baxter continuous cardiac output monitoring system. J Clin Monit 12:243–249PubMedCrossRef Lichtenthal PR, Gordan D (1996) Testing the safety of Baxter continuous cardiac output monitoring system. J Clin Monit 12:243–249PubMedCrossRef
22.
Zurück zum Zitat Mielck F, Buhre W, Hanekop G, Tirilomis T, Hilgers R, Sonntag H (2003) Comparison of continuous cardiac output measurements in patients after cardiac surgery. J Cardiothorac Vasc Anesth 17:211–216PubMedCrossRef Mielck F, Buhre W, Hanekop G, Tirilomis T, Hilgers R, Sonntag H (2003) Comparison of continuous cardiac output measurements in patients after cardiac surgery. J Cardiothorac Vasc Anesth 17:211–216PubMedCrossRef
23.
Zurück zum Zitat Schmid ER, Schmidlin D, Tornic M, Seifert B (1999) Continuous thermodilution cardiac output: clinical validation against a reference technique of known accuracy. Intensive Care Med 25:166–172PubMedCrossRef Schmid ER, Schmidlin D, Tornic M, Seifert B (1999) Continuous thermodilution cardiac output: clinical validation against a reference technique of known accuracy. Intensive Care Med 25:166–172PubMedCrossRef
24.
Zurück zum Zitat Parrillo JE (1989) The cardiovascular pathophysiology of sepsis. Annu Rev Med 40:469–485PubMed Parrillo JE (1989) The cardiovascular pathophysiology of sepsis. Annu Rev Med 40:469–485PubMed
25.
Zurück zum Zitat Diebel LN, Wilson RF, Tagett MG, Kline RA (1992) End-diastolic volume. A better indicator of preload in the critically ill. Arch Surg 127:817–822PubMed Diebel LN, Wilson RF, Tagett MG, Kline RA (1992) End-diastolic volume. A better indicator of preload in the critically ill. Arch Surg 127:817–822PubMed
26.
Zurück zum Zitat Wagner JG, Leatherman JW (1998) Right ventricular end-diastolic volume as a predictor of the hemodynamic response to a fluid challenge. Chest 113:1048–1054PubMed Wagner JG, Leatherman JW (1998) Right ventricular end-diastolic volume as a predictor of the hemodynamic response to a fluid challenge. Chest 113:1048–1054PubMed
27.
Zurück zum Zitat Wiesenack C, Fiegl C, Keyser A, Laule S, Prasser C, Keyl C (2005) Continuously assessed right ventricular end-diastolic volume as a marker of cardiac preload and fluid responsiveness in mechanically ventilated cardiac surgical patients. Crit Care 9:R226–R233PubMedCrossRef Wiesenack C, Fiegl C, Keyser A, Laule S, Prasser C, Keyl C (2005) Continuously assessed right ventricular end-diastolic volume as a marker of cardiac preload and fluid responsiveness in mechanically ventilated cardiac surgical patients. Crit Care 9:R226–R233PubMedCrossRef
28.
Zurück zum Zitat Gödje O, Hoke K, Lamm P, Schmitz C, Thiel C, Weinert M, Reichart B (1998) Continuous, less invasive, hemodynamic monitoring in intensive care after cardiac surgery. Thorac Cardiovasc Surg 46:242–249PubMedCrossRef Gödje O, Hoke K, Lamm P, Schmitz C, Thiel C, Weinert M, Reichart B (1998) Continuous, less invasive, hemodynamic monitoring in intensive care after cardiac surgery. Thorac Cardiovasc Surg 46:242–249PubMedCrossRef
29.
Zurück zum Zitat Wiesenack C, Prasser C, Keyl C, Rodig G (2001) Assessment of intrathoracic blood volume as an indicator of cardiac preload: single transpulmonary thermodilution technique versus assessment of pressure preload parameters derived from a pulmonary artery catheter. J Cardiothorac Vasc Anesth 15:584–588PubMedCrossRef Wiesenack C, Prasser C, Keyl C, Rodig G (2001) Assessment of intrathoracic blood volume as an indicator of cardiac preload: single transpulmonary thermodilution technique versus assessment of pressure preload parameters derived from a pulmonary artery catheter. J Cardiothorac Vasc Anesth 15:584–588PubMedCrossRef
30.
Zurück zum Zitat Bindels AJ, van der Hoeven JG, Graafland AD, de Koning J, Meinders AE (2000) Relationships between volume and pressure measurements and stroke volume in critically ill patients. Crit Care 4:193–199PubMedCrossRef Bindels AJ, van der Hoeven JG, Graafland AD, de Koning J, Meinders AE (2000) Relationships between volume and pressure measurements and stroke volume in critically ill patients. Crit Care 4:193–199PubMedCrossRef
31.
Zurück zum Zitat Hofer CK, Furrer L, Matter-Ensner S, Maloigne M, Klaghofer R, Genoni M, Zollinger A (2005) Volumetric preload measurement by thermodilution: a comparison with transoesophageal echocardiography. Br J Anaesth 94:748–755PubMedCrossRef Hofer CK, Furrer L, Matter-Ensner S, Maloigne M, Klaghofer R, Genoni M, Zollinger A (2005) Volumetric preload measurement by thermodilution: a comparison with transoesophageal echocardiography. Br J Anaesth 94:748–755PubMedCrossRef
32.
Zurück zum Zitat Sakka SG, Bredle DL, Reinhart K, Meier-Hellmann A (1999) Comparison between intrathoracic blood volume and cardiac filling pressures in the early phase of hemodynamic instability of patients with sepsis or septic shock. J Crit Care 14:78–83PubMedCrossRef Sakka SG, Bredle DL, Reinhart K, Meier-Hellmann A (1999) Comparison between intrathoracic blood volume and cardiac filling pressures in the early phase of hemodynamic instability of patients with sepsis or septic shock. J Crit Care 14:78–83PubMedCrossRef
Metadaten
Titel
Comparison of two methods for enhanced continuous circulatory monitoring in patients with septic shock
verfasst von
F. Spöhr
P. Hettrich
H. Bauer
U. Haas
E. Martin
B. W. Böttiger
Publikationsdatum
01.10.2007
Verlag
Springer-Verlag
Erschienen in
Intensive Care Medicine / Ausgabe 10/2007
Print ISSN: 0342-4642
Elektronische ISSN: 1432-1238
DOI
https://doi.org/10.1007/s00134-007-0703-2

Weitere Artikel der Ausgabe 10/2007

Intensive Care Medicine 10/2007 Zur Ausgabe

Update AINS

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.