Performance of noninvasive ventilation modes on ICU ventilators during pressure support: a bench model study

verfasst von: Laurence Vignaux, Didier Tassaux, Philippe Jolliet

Erschienen in: Intensive Care Medicine | Ausgabe 8/2007

Einloggen, um Zugang zu erhalten

Abstract

Objective

Noninvasive ventilation (NIV) is often applied with ICU ventilators. However, leaks at the patient-ventilator interface interfere with several key ventilator functions. Many ICU ventilators feature an NIV-specific mode dedicated to preventing these problems. The present bench model study aimed to evaluate the performance of these modes.

Design and setting

Bench model study in an intensive care research laboratory of a university hospital.

Methods

Eight ICU ventilators, widely available in Europe and featuring an NIV mode, were connected by an NIV mask to a lung model featuring a plastic head to mimic NIV conditions, driven by an ICU ventilator imitating patient effort. Tests were conducted in the absence and presence of leaks, the latter condition with and without activation of the NIV mode. Trigger delay, trigger-associated inspiratory workload, and pressurization were tested in conditions of normal respiratory mechanics, and cycling was also assessed in obstructive and restrictive conditions.

Results

On most ventilators leaks led to an increase in trigger delay and workload, a decrease in pressurization, and delayed cycling. On most ventilators the NIV mode partly or totally corrected these problems, but with large variations between machines. Furthermore, on some ventilators the NIV mode worsened the leak-induced dysfunction.

Conclusions

The results of this bench-model NIV study confirm that leaks interfere with several key functions of ICU ventilators. Overall, NIV modes can correct part or all of this interference, but with wide variations between machines in terms of efficiency. Clinicians should be aware of these differences when applying NIV with an ICU ventilator.
Anhänge
Nur mit Berechtigung zugänglich
Literatur
1.
Zurück zum Zitat Peter JV, Moran JL, Hughes JP (2002) Noninvasive mechanical ventilation in acute respiratory failure—a meta-analysis update. Crit Care Med 30:555–562PubMedCrossRef Peter JV, Moran JL, Hughes JP (2002) Noninvasive mechanical ventilation in acute respiratory failure—a meta-analysis update. Crit Care Med 30:555–562PubMedCrossRef
2.
Zurück zum Zitat Liesching T, Kwok H, Hill NS (2003) Acute applications of noninvasive positive pressure ventilation. Chest 124:699–713PubMedCrossRef Liesching T, Kwok H, Hill NS (2003) Acute applications of noninvasive positive pressure ventilation. Chest 124:699–713PubMedCrossRef
3.
Zurück zum Zitat Ferrer M, Esquinas A, Leon M, Gonzalez G, Alarcon A, Torres A (2003) Noninvasive ventilation in severe hypoxemic respiratory failure. Am J Respir Crit Care Med 168:1438–1444PubMedCrossRef Ferrer M, Esquinas A, Leon M, Gonzalez G, Alarcon A, Torres A (2003) Noninvasive ventilation in severe hypoxemic respiratory failure. Am J Respir Crit Care Med 168:1438–1444PubMedCrossRef
4.
Zurück zum Zitat Antonelli M, Conti G, Moro ML, Esquinas A, Gonzalez-Diaz G, Confalonieri M, Pelaia P, Principi T, Gregoretti C, Beltrame F, Pennisi MA, Arcangeli A, Proietti R, Passariello M, Meduri GU (2001) Predictors of failure of noninvasive positive pressure ventilation in patients with acute hypoxemic respiratory failure: a multi-center study. Intensive Care Med 27:1718–1728PubMedCrossRef Antonelli M, Conti G, Moro ML, Esquinas A, Gonzalez-Diaz G, Confalonieri M, Pelaia P, Principi T, Gregoretti C, Beltrame F, Pennisi MA, Arcangeli A, Proietti R, Passariello M, Meduri GU (2001) Predictors of failure of noninvasive positive pressure ventilation in patients with acute hypoxemic respiratory failure: a multi-center study. Intensive Care Med 27:1718–1728PubMedCrossRef
5.
Zurück zum Zitat Carlucci A, Richard JC, Wysocki M, Lepage E, Brochard L (2001) Noninvasive versus conventional mechanical ventilation. An epidemiologic survey. Am J Respir Crit Care Med 163:874–880PubMed Carlucci A, Richard JC, Wysocki M, Lepage E, Brochard L (2001) Noninvasive versus conventional mechanical ventilation. An epidemiologic survey. Am J Respir Crit Care Med 163:874–880PubMed
6.
Zurück zum Zitat Bernstein G, Knodel E, Heldt GP (1995) Airway leak size in neonates and autocycling of three flow-triggered ventilators. Crit Care Med 23:1739–1744PubMedCrossRef Bernstein G, Knodel E, Heldt GP (1995) Airway leak size in neonates and autocycling of three flow-triggered ventilators. Crit Care Med 23:1739–1744PubMedCrossRef
7.
Zurück zum Zitat Calderini E, Confalonieri M, Puccio PG, Francavilla N, Stella L, Gregoretti C (1999) Patient-ventilator asynchrony during noninvasive ventilation: the role of expiratory trigger. Intensive Care Med 25:662–667PubMedCrossRef Calderini E, Confalonieri M, Puccio PG, Francavilla N, Stella L, Gregoretti C (1999) Patient-ventilator asynchrony during noninvasive ventilation: the role of expiratory trigger. Intensive Care Med 25:662–667PubMedCrossRef
8.
Zurück zum Zitat Prinianakis G, Delmastro M, Carlucci A, Ceriana P, Nava S (2004) Effect of varying the pressurisation rate during noninvasive pressure support ventilation. Eur Respir J 23:314–320PubMedCrossRef Prinianakis G, Delmastro M, Carlucci A, Ceriana P, Nava S (2004) Effect of varying the pressurisation rate during noninvasive pressure support ventilation. Eur Respir J 23:314–320PubMedCrossRef
9.
Zurück zum Zitat Schettino GP, Tucci MR, Sousa R, Valente Barbas CS, Passos Amato MB, Carvalho CR (2001) Mask mechanics and leak dynamics during noninvasive pressure support ventilation: a bench study. Intensive Care Med 27:1887–1891PubMedCrossRef Schettino GP, Tucci MR, Sousa R, Valente Barbas CS, Passos Amato MB, Carvalho CR (2001) Mask mechanics and leak dynamics during noninvasive pressure support ventilation: a bench study. Intensive Care Med 27:1887–1891PubMedCrossRef
10.
Zurück zum Zitat Tobin MJ, Jubran A, Laghi F (2001) Patient-ventilator interaction. Am J Respir Crit Care Med 163:1059–1063PubMed Tobin MJ, Jubran A, Laghi F (2001) Patient-ventilator interaction. Am J Respir Crit Care Med 163:1059–1063PubMed
11.
Zurück zum Zitat Kondili E, Prinianakis G, Georgopoulos D (2003) Patient-ventilator interaction. Br J Anaesth 91:106–119PubMedCrossRef Kondili E, Prinianakis G, Georgopoulos D (2003) Patient-ventilator interaction. Br J Anaesth 91:106–119PubMedCrossRef
12.
Zurück zum Zitat Moore MJ, Schmidt GA (2000) Non-invasive ventilation: why does it fail? In Vincent JL (ed) Yearbook of intensive care and emergency medicine. Springer, Berlin Heidelberg New York, pp 318–327 Moore MJ, Schmidt GA (2000) Non-invasive ventilation: why does it fail? In Vincent JL (ed) Yearbook of intensive care and emergency medicine. Springer, Berlin Heidelberg New York, pp 318–327
13.
Zurück zum Zitat Peerless JR, Davies A, Klein D, Yu D (1999) Skin complications in the intensive care unit. Clin Chest Med 20:453–467PubMedCrossRef Peerless JR, Davies A, Klein D, Yu D (1999) Skin complications in the intensive care unit. Clin Chest Med 20:453–467PubMedCrossRef
14.
Zurück zum Zitat Miyoshi E, Fujino Y, Uchiyama A, Mashimo T, Nishimura M (2005) Effects of gas leak on triggering function, humidification, and inspiratory oxygen fraction during noninvasive positive airway pressure ventilation. Chest 128:3691–3698PubMedCrossRef Miyoshi E, Fujino Y, Uchiyama A, Mashimo T, Nishimura M (2005) Effects of gas leak on triggering function, humidification, and inspiratory oxygen fraction during noninvasive positive airway pressure ventilation. Chest 128:3691–3698PubMedCrossRef
15.
Zurück zum Zitat Meyer TJ, Pressman MR, Benditt J, McCool FD, Millman RP, Natarajan R, Hill NS (1997) Air leaking through the mouth during nocturnal nasal ventilation: effect on sleep quality. Sleep 20:561–569PubMed Meyer TJ, Pressman MR, Benditt J, McCool FD, Millman RP, Natarajan R, Hill NS (1997) Air leaking through the mouth during nocturnal nasal ventilation: effect on sleep quality. Sleep 20:561–569PubMed
16.
Zurück zum Zitat Mehta S, McCool FD, Hill NS (2001) Leak compensation in positive pressure ventilators: a lung model study. Eur Respir J 17:259–267PubMedCrossRef Mehta S, McCool FD, Hill NS (2001) Leak compensation in positive pressure ventilators: a lung model study. Eur Respir J 17:259–267PubMedCrossRef
17.
Zurück zum Zitat Battisti A, Tassaux D, Janssens JP, Michotte JB, Jaber S, Jolliet P (2005) Performance characteristics of ten recent bilevel ventilators: a comparative bench study. Chest 127:1784–1792PubMedCrossRef Battisti A, Tassaux D, Janssens JP, Michotte JB, Jaber S, Jolliet P (2005) Performance characteristics of ten recent bilevel ventilators: a comparative bench study. Chest 127:1784–1792PubMedCrossRef
18.
Zurück zum Zitat Tassaux D, Strasser S, Fonseca S, Dalmas E, Jolliet P (2002) Comparative bench study of triggering, pressurization and cycling between the home ventilator VPAPII® and three ICU ventilators. Intensive Care Med 28:1254–1261PubMedCrossRef Tassaux D, Strasser S, Fonseca S, Dalmas E, Jolliet P (2002) Comparative bench study of triggering, pressurization and cycling between the home ventilator VPAPII® and three ICU ventilators. Intensive Care Med 28:1254–1261PubMedCrossRef
19.
Zurück zum Zitat Richard JC, Carlucci A, Breton L, Langlais N, Jaber S, Maggiore S, Fougere S, Harf A, Brochard L (2002) Bench testing of pressure support ventilation with three different generations of ventilators. Intensive Care Med 28:1049–1057PubMedCrossRef Richard JC, Carlucci A, Breton L, Langlais N, Jaber S, Maggiore S, Fougere S, Harf A, Brochard L (2002) Bench testing of pressure support ventilation with three different generations of ventilators. Intensive Care Med 28:1049–1057PubMedCrossRef
20.
Zurück zum Zitat Aslanian P, El Atrous S, Isabey D, Valente E, Corsi D, Harf A, Lemaire F, Brochard L (1998) Effects of flow triggering on breathing effort during partial ventilatory support. Am J Respir Crit Care Med 157:135–143PubMed Aslanian P, El Atrous S, Isabey D, Valente E, Corsi D, Harf A, Lemaire F, Brochard L (1998) Effects of flow triggering on breathing effort during partial ventilatory support. Am J Respir Crit Care Med 157:135–143PubMed
21.
Zurück zum Zitat Nava S, Bruschi C, Rubini F, Palo A, Iotti G, Braschi A (1995) Respiratory response and inspiratory effort during pressure support ventilation in COPD patients. Intensive Care Med 21:871–879PubMedCrossRef Nava S, Bruschi C, Rubini F, Palo A, Iotti G, Braschi A (1995) Respiratory response and inspiratory effort during pressure support ventilation in COPD patients. Intensive Care Med 21:871–879PubMedCrossRef
22.
Zurück zum Zitat Tassaux D, Michotte JB, Gainnier M, Gratadour P, Fonseca S, Jolliet P (2004) Expiratory trigger setting in pressure Support Ventilation: from mathematical model to bedside. Crit Care Med 32:1844–1850PubMedCrossRef Tassaux D, Michotte JB, Gainnier M, Gratadour P, Fonseca S, Jolliet P (2004) Expiratory trigger setting in pressure Support Ventilation: from mathematical model to bedside. Crit Care Med 32:1844–1850PubMedCrossRef
23.
Zurück zum Zitat Tassaux D, Gainnier M, Battisti A, Jolliet P (2005) Impact of expiratory trigger setting on delayed cycling and inspiratory muscle workload. Am J Respir Crit Care Med 172:1283–1289PubMedCrossRef Tassaux D, Gainnier M, Battisti A, Jolliet P (2005) Impact of expiratory trigger setting on delayed cycling and inspiratory muscle workload. Am J Respir Crit Care Med 172:1283–1289PubMedCrossRef
Metadaten
Titel
Performance of noninvasive ventilation modes on ICU ventilators during pressure support: a bench model study
verfasst von
Laurence Vignaux
Didier Tassaux
Philippe Jolliet
Publikationsdatum
01.08.2007
Verlag
Springer-Verlag
Erschienen in
Intensive Care Medicine / Ausgabe 8/2007
Print ISSN: 0342-4642
Elektronische ISSN: 1432-1238
DOI
https://doi.org/10.1007/s00134-007-0713-0

Weitere Artikel der Ausgabe 8/2007

Intensive Care Medicine 8/2007 Zur Ausgabe

Update AINS

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.