Skip to main content
Erschienen in: Intensive Care Medicine 12/2007

01.12.2007 | Original

Accuracy of bedside capillary blood glucose measurements in critically ill patients

verfasst von: C. Dana Critchell, Vincent Savarese, Amy Callahan, Christine Aboud, Serge Jabbour, Paul Marik

Erschienen in: Intensive Care Medicine | Ausgabe 12/2007

Einloggen, um Zugang zu erhalten

Abstract

Objective

To compare the accuracy of fingerstick with laboratory venous plasma glucose measurements (laboratory glucose) in medical ICU patients and to determine the factors which interfere with the accuracy of fingerstick measurements.

Participants

The study included 80 consecutive patients aged 58 ± 7 years, BMI 29.5 ± 9.0, and APACHE II score 15 ± 6 (277 simultaneous paired measurements).

Measurements

This prospective observational study compared fingerstick measurements to simultaneously sampled laboratory glucose once a day in patients in our medical ICU (twice daily if on an insulin infusion). Data recorded included patient demographics, admission diagnoses, APACHE II score, BMI, daily hematocrit, arterial blood gasses, chemistry results, concomitant medications (including vasopressors and corticosteroids), and upper extremity edema. Accuracy was defined as the percentage of paired values not in accord (> 15 mg dl–1 / 0.83 mmol–1 l–1 difference for laboratory values < 75 mg dl–1 / 4.12 mmol–1 l–1 and > 20% difference for laboratory values ≥ 75 mg/dl). Outliers (blood glucose difference > 100 mg dl–1 / 5.56 mmol–1 l–1) were excluded from the correlation and distribution analyses.

Results

Mean fingerstick glucose was 129 ± 45 mg/dl (7.2 ± 2.5 mmol/l) and mean laboratory glucose 123 ± 44 mg/dl (6.8 ± 2.4 mmol/l). The correlation coefficient between the two values was 0.9110 (Clinical and Laboratory Standards Institute threshold 0.9751). The mean difference (bias) between the two methods was 8.6 ± 18.6 mg/dl (0.48 ± 1.0 mmol/l) and limits of agreement +45.8 and –28.6 mg/dl (+2.5 and –1.6 mmol/l). Fifty-three (19%) paired measurements in 22 patients were not in accord (CLSI threshold ≤ 5%). In 44 (83%) of these paired measurements fingerstick glucose was greater than laboratory glucose.

Conclusions

The findings suggest that capillary blood glucose as measured by fingerstick is inaccurate in critically ill ICU patients and does not meet the CLSI standard. It is unclear whether the sampling method, device used, or both contributed to this inaccuracy. The wide limits of agreement suggest that fingerstick measurements should be used with great caution in protocols of tight glycemic control.
Literatur
1.
Zurück zum Zitat van den Berghe G, Wouters P, Weekers F, Verwaest C, Bruyninckx F, Schetz M, Vlasselaers D, Ferdinande P, Lauwers P, Bouillon R (2001) Intensive insulin therapy in critically ill patients. N Engl J Med 345:1359–1367CrossRef van den Berghe G, Wouters P, Weekers F, Verwaest C, Bruyninckx F, Schetz M, Vlasselaers D, Ferdinande P, Lauwers P, Bouillon R (2001) Intensive insulin therapy in critically ill patients. N Engl J Med 345:1359–1367CrossRef
2.
Zurück zum Zitat Kanji S, Buffie J, Hutton B, Bunting PS, Singh A, McDonald K, Fergusson D, McIntyre LA, Hebert PC (2005) Reliability of point-of-care testing for glucose measurement in critically ill adults. Crit Care Med 33:2778–2785PubMedCrossRef Kanji S, Buffie J, Hutton B, Bunting PS, Singh A, McDonald K, Fergusson D, McIntyre LA, Hebert PC (2005) Reliability of point-of-care testing for glucose measurement in critically ill adults. Crit Care Med 33:2778–2785PubMedCrossRef
3.
Zurück zum Zitat Chakravarthy SB, Markewitz BA, Lehman C, Orme JF (2005) Accuracy of glucose measurements in critically ill patients in shock (abstract). Chest 128:220S Chakravarthy SB, Markewitz BA, Lehman C, Orme JF (2005) Accuracy of glucose measurements in critically ill patients in shock (abstract). Chest 128:220S
4.
Zurück zum Zitat Finkielman JD, Oyen LJ, Afessa B (2005) Agreement between bedside blood and plasma glucose measurement in the ICU setting. Chest 127:1749–1751PubMedCrossRef Finkielman JD, Oyen LJ, Afessa B (2005) Agreement between bedside blood and plasma glucose measurement in the ICU setting. Chest 127:1749–1751PubMedCrossRef
5.
Zurück zum Zitat Kulkarni A, Saxena M, Price G, O'Leary MJ, Jacques T, Myburgh JA (2005) Analysis of blood glucose measurements using capillary and arterial blood samples in intensive care patients. Intensive Care Med 31:142–145PubMedCrossRef Kulkarni A, Saxena M, Price G, O'Leary MJ, Jacques T, Myburgh JA (2005) Analysis of blood glucose measurements using capillary and arterial blood samples in intensive care patients. Intensive Care Med 31:142–145PubMedCrossRef
6.
Zurück zum Zitat Vriesendorp TM, van Santen S, DeVries JH, de Jonge E, Rosendaal FR, Schultz MJ, Hoekstra JB (2006) Predisposing factors for hypoglycemia in the intensive care unit. Crit Care Med 34:96–101PubMedCrossRef Vriesendorp TM, van Santen S, DeVries JH, de Jonge E, Rosendaal FR, Schultz MJ, Hoekstra JB (2006) Predisposing factors for hypoglycemia in the intensive care unit. Crit Care Med 34:96–101PubMedCrossRef
7.
Zurück zum Zitat van den Berghe G, Wilmer A, Hermans G, Meersseman W, Wouters PJ, Milants I, van Wijngaerden E, Bobbaers H, Bouillon R (2006) Intensive insulin therapy in the medical ICU. N Engl J Med 354:449–461CrossRef van den Berghe G, Wilmer A, Hermans G, Meersseman W, Wouters PJ, Milants I, van Wijngaerden E, Bobbaers H, Bouillon R (2006) Intensive insulin therapy in the medical ICU. N Engl J Med 354:449–461CrossRef
8.
Zurück zum Zitat Critchell C, Callahan A, Aboud C, Jabbour S, Marik PE (2006) Accuracy of bedside capillary blood glucose measurements in critically ill patients (abstract). Crit Care Med 34(Suppl A):68CrossRef Critchell C, Callahan A, Aboud C, Jabbour S, Marik PE (2006) Accuracy of bedside capillary blood glucose measurements in critically ill patients (abstract). Crit Care Med 34(Suppl A):68CrossRef
9.
Zurück zum Zitat Marik PE, Varon J (2007) Intensive insulin therapy in the ICU: is it now time to jump off the bandwagon? Resuscitation 2007:191–193CrossRef Marik PE, Varon J (2007) Intensive insulin therapy in the ICU: is it now time to jump off the bandwagon? Resuscitation 2007:191–193CrossRef
10.
Zurück zum Zitat Knaus WA, Draper EA, Wagner DP, Zimmerman JE (1985) APACHE II: a severity of disease classification system. Crit Care Med 13:818–828PubMedCrossRef Knaus WA, Draper EA, Wagner DP, Zimmerman JE (1985) APACHE II: a severity of disease classification system. Crit Care Med 13:818–828PubMedCrossRef
11.
Zurück zum Zitat Szczesny G, Olszewski WL (2003) The pathomechanism of posttraumatic edema of the lower limbs. II. Changes in the lymphatic system. J Trauma 55:350–354PubMedCrossRef Szczesny G, Olszewski WL (2003) The pathomechanism of posttraumatic edema of the lower limbs. II. Changes in the lymphatic system. J Trauma 55:350–354PubMedCrossRef
12.
Zurück zum Zitat Bland JM, Altman DG (1986) Statistical methods for assessing agreement between two methods of clinical measurement. Lancet I:307–310 Bland JM, Altman DG (1986) Statistical methods for assessing agreement between two methods of clinical measurement. Lancet I:307–310
13.
Zurück zum Zitat Atkin SH, Dasmahapatra A, Jaker MA, Chorost MI, Reddy S (1991) Fingerstick glucose determination in shock. Ann Intern Med 114:1020–1024PubMed Atkin SH, Dasmahapatra A, Jaker MA, Chorost MI, Reddy S (1991) Fingerstick glucose determination in shock. Ann Intern Med 114:1020–1024PubMed
14.
Zurück zum Zitat Sylvain HF, Pokorny ME, English SM, Benson NH, Whitley TW, Ferenczy CJ, Harrison JG (1995) Accuracy of fingerstick glucose values in shock patients. Am J Crit Care 4:44–48PubMed Sylvain HF, Pokorny ME, English SM, Benson NH, Whitley TW, Ferenczy CJ, Harrison JG (1995) Accuracy of fingerstick glucose values in shock patients. Am J Crit Care 4:44–48PubMed
15.
Zurück zum Zitat Brunkhorst FM, Wahl HG (2006) Blood glucose measurements in the critically ill: more than just a blood draw. Crit Care 10:178PubMedCrossRef Brunkhorst FM, Wahl HG (2006) Blood glucose measurements in the critically ill: more than just a blood draw. Crit Care 10:178PubMedCrossRef
16.
Zurück zum Zitat Wiener K (1995) Whole blood glucose: what are we actually measuring? Ann Clin Biochem 32:1–8PubMed Wiener K (1995) Whole blood glucose: what are we actually measuring? Ann Clin Biochem 32:1–8PubMed
17.
Zurück zum Zitat Sidebottom RA, Williams PR, Kanarek KS (1982) Glucose determinations in plasma and serum: potential error related to increased hematocrit. Clin Chem 28:190–192PubMed Sidebottom RA, Williams PR, Kanarek KS (1982) Glucose determinations in plasma and serum: potential error related to increased hematocrit. Clin Chem 28:190–192PubMed
18.
Zurück zum Zitat Fogh-Anderson N, D'Orazio P (1998) Proposal for standardizing direct-reading biosensors for blood glucose. Clin Chem 44:655–659 Fogh-Anderson N, D'Orazio P (1998) Proposal for standardizing direct-reading biosensors for blood glucose. Clin Chem 44:655–659
19.
Zurück zum Zitat Sazama K, Robertson EA, Chesler RA (1979) Is antiglycolysis required for routine glucose analysis? Clin Chem 25:2038–2039PubMed Sazama K, Robertson EA, Chesler RA (1979) Is antiglycolysis required for routine glucose analysis? Clin Chem 25:2038–2039PubMed
20.
Zurück zum Zitat Tang Z, Louie RF, Lee JH, Lee DM, Miller EE, Kost GJ (2001) Oxygen effects on glucose meter measurements with glucose dehydrogenase- and oxidase-based test strips for point-of-care testing. Crit Care Med 29:1062–1070PubMedCrossRef Tang Z, Louie RF, Lee JH, Lee DM, Miller EE, Kost GJ (2001) Oxygen effects on glucose meter measurements with glucose dehydrogenase- and oxidase-based test strips for point-of-care testing. Crit Care Med 29:1062–1070PubMedCrossRef
21.
Zurück zum Zitat Tang Z, Du X, Louie RF, Kost GJ (2000) Effects of drugs on glucose measurements with handheld glucose meters and a portable glucose analyzer. Am J Clin Pathol 113:75–86PubMedCrossRef Tang Z, Du X, Louie RF, Kost GJ (2000) Effects of drugs on glucose measurements with handheld glucose meters and a portable glucose analyzer. Am J Clin Pathol 113:75–86PubMedCrossRef
Metadaten
Titel
Accuracy of bedside capillary blood glucose measurements in critically ill patients
verfasst von
C. Dana Critchell
Vincent Savarese
Amy Callahan
Christine Aboud
Serge Jabbour
Paul Marik
Publikationsdatum
01.12.2007
Verlag
Springer-Verlag
Erschienen in
Intensive Care Medicine / Ausgabe 12/2007
Print ISSN: 0342-4642
Elektronische ISSN: 1432-1238
DOI
https://doi.org/10.1007/s00134-007-0835-4

Weitere Artikel der Ausgabe 12/2007

Intensive Care Medicine 12/2007 Zur Ausgabe

Announcements

Announcement

Update AINS

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.