Skip to main content
Erschienen in: Intensive Care Medicine 4/2009

01.04.2009 | Original

Adult ICU ventilators to provide neonatal ventilation: a lung simulator study

verfasst von: Andrew D. Marchese, Daniel Chipman, Pedro de la Oliva, Robert M. Kacmarek

Erschienen in: Intensive Care Medicine | Ausgabe 4/2009

Einloggen, um Zugang zu erhalten

Abstract

Background

Traditionally, specific ventilators have been manufactured to only provide neonatal mechanical ventilation. However, many of the current generation of ICU ventilators also include a neonatal mode.

Methods

Using the IngMar ASL5000 lung simulator the Puritan Bennett 840, the Maquet Servo i, the Viasys AVEA, the GE Engström, the Drager Evita XL and Babylog 8000 Plus were evaluated during assisted ventilation in the pressure assist/control mode. Three lung mechanics were set: resistance 50 cmH2O/L/s, compliance 2 mL/cmH2O; resistance 100 cmH2O/L/s, compliance 1 mL/cmH2O; and resistance 150 cmH2O/L/s, compliance 0.5 mL/cmH2O. A maximum negative pressure drop of 4 and 7 cmH2O was achieved during simulated inspirations. Each ventilator was evaluated with PEEP 5 cmH2O, peak pressure 20 cmH2O and inspiratory time 0.3 s and with PEEP 10 cmH2O, peak pressure 30 cmH2O and inspiratory time 0.4 s. Each ventilator setting was then repeated with a leak of 0.3 L/min at a constant pressure of 5 cmH2O.

Results

Overall each of the 5 ICU ventilators responded faster or greater than the Babylog with respect to: pressure to trigger (except the Servo i), time to trigger (except the Evita XL), time between trigger and return of pressure to baseline, time from start of breath to 90% of peak pressure (except the Avea) and pressure time product of breath activation. Expiratory tidal volume was also greater with all ICU ventilators except the Avea. Variation in mechanics, leak, PEEP and muscular effort had little effect on these differences.

Conclusion

All ICU ventilators tested were able to at least equal the performance of the Babylog 8000 Plus on all variables evaluated.
Anhänge
Nur mit Berechtigung zugänglich
Literatur
1.
Zurück zum Zitat Mushin WW (1980) Automatic ventilation of the lungs, 3rd edn. England Blackwell Scientific Publications Ltd, Oxford Mushin WW (1980) Automatic ventilation of the lungs, 3rd edn. England Blackwell Scientific Publications Ltd, Oxford
2.
Zurück zum Zitat Egan DE (1981) Fundamentals of inhalation therapy, 1st edn. The C. V. Mosby Co., St. Louis, p 280 Egan DE (1981) Fundamentals of inhalation therapy, 1st edn. The C. V. Mosby Co., St. Louis, p 280
3.
Zurück zum Zitat McPherson SP, Spearman CB (1981) Respiratory therapy equipment, 2nd edn. The C. V. Mosby Co., St. Louis, p 361 McPherson SP, Spearman CB (1981) Respiratory therapy equipment, 2nd edn. The C. V. Mosby Co., St. Louis, p 361
4.
Zurück zum Zitat Kirby RR, Robison EJ, Schulz JC, deLemos R (1971) A new pediatric volume ventilator. Anesth Analg (Cleve.) 50:533–537 Kirby RR, Robison EJ, Schulz JC, deLemos R (1971) A new pediatric volume ventilator. Anesth Analg (Cleve.) 50:533–537
5.
Zurück zum Zitat Understanding the IMVbird. From 876 (1975) Bird Corp. Palm Springs, Calif Understanding the IMVbird. From 876 (1975) Bird Corp. Palm Springs, Calif
6.
Zurück zum Zitat Marchese AD, Chipman D, De La Oliva P, Kacmarek RM (2007) Triggering and initial gas delivery: the Babylog vs. ICU ventilators in the neo-mode (Abstract). Resp Care 52:1566 Marchese AD, Chipman D, De La Oliva P, Kacmarek RM (2007) Triggering and initial gas delivery: the Babylog vs. ICU ventilators in the neo-mode (Abstract). Resp Care 52:1566
7.
Zurück zum Zitat IngMar Medical, Ltd. (2006) ASL5000TM Active servo lung computerized breathing simulator and ventilator test instrument user’s manual. IngMar Medical, Ltd., Pittsburgh (PDF version of document downloaded March 05, 2007) IngMar Medical, Ltd. (2006) ASL5000TM Active servo lung computerized breathing simulator and ventilator test instrument user’s manual. IngMar Medical, Ltd., Pittsburgh (PDF version of document downloaded March 05, 2007)
8.
Zurück zum Zitat Kapasi M, Fujino Y, Kirmse M, Catlin EA, Kacmarek RM (2001) Effort and work of breathing in neonates during assisted, patient triggered ventilation. Ped Crit Care 2:9–16CrossRef Kapasi M, Fujino Y, Kirmse M, Catlin EA, Kacmarek RM (2001) Effort and work of breathing in neonates during assisted, patient triggered ventilation. Ped Crit Care 2:9–16CrossRef
9.
Zurück zum Zitat Nishimura M, Hess D, Kacmarek RM (1995) The response of flow-triggered infant ventilators. Am J Res Crit Care Med 152:1901–1909 Nishimura M, Hess D, Kacmarek RM (1995) The response of flow-triggered infant ventilators. Am J Res Crit Care Med 152:1901–1909
10.
Zurück zum Zitat Hummler HD, Gerhardt T, Gonzalez A, Bolivar J, Clause N, Evertt R, Bancalari E (1996) Patient-triggered ventilation in neonates: comparison of a flow-and impedence-triggered system. Am J Resp Crit Care Med 154:1049–1054PubMed Hummler HD, Gerhardt T, Gonzalez A, Bolivar J, Clause N, Evertt R, Bancalari E (1996) Patient-triggered ventilation in neonates: comparison of a flow-and impedence-triggered system. Am J Resp Crit Care Med 154:1049–1054PubMed
11.
Zurück zum Zitat Simbruner G, Gregory GA (1981) Performance of neonatal ventilators; the effect of changes in resistance and compliance. Crit Care Med 9:509–515PubMedCrossRef Simbruner G, Gregory GA (1981) Performance of neonatal ventilators; the effect of changes in resistance and compliance. Crit Care Med 9:509–515PubMedCrossRef
12.
Zurück zum Zitat Gappa M, Pillow J, Allen J, Mayer O, Stocks J (2006) Lung Function tests in neonates and infants with chronic lung disease: Lung and chest-wall mechanics. Pediatr Pulmonol 41:291–317PubMedCrossRef Gappa M, Pillow J, Allen J, Mayer O, Stocks J (2006) Lung Function tests in neonates and infants with chronic lung disease: Lung and chest-wall mechanics. Pediatr Pulmonol 41:291–317PubMedCrossRef
13.
Zurück zum Zitat Berstein G, Knodel E, Gergory H (1995) Airway leak size in neonates and auto-cycling of three flow-triggered ventilators. Crit Care Med 23:1739–1744CrossRef Berstein G, Knodel E, Gergory H (1995) Airway leak size in neonates and auto-cycling of three flow-triggered ventilators. Crit Care Med 23:1739–1744CrossRef
14.
Zurück zum Zitat Sharma A, Milner AD, Greenough A (2007) Performance of neonatal ventilators in volume targeted ventilation mode. Acta Paediatr 96:176–180PubMedCrossRef Sharma A, Milner AD, Greenough A (2007) Performance of neonatal ventilators in volume targeted ventilation mode. Acta Paediatr 96:176–180PubMedCrossRef
Metadaten
Titel
Adult ICU ventilators to provide neonatal ventilation: a lung simulator study
verfasst von
Andrew D. Marchese
Daniel Chipman
Pedro de la Oliva
Robert M. Kacmarek
Publikationsdatum
01.04.2009
Verlag
Springer-Verlag
Erschienen in
Intensive Care Medicine / Ausgabe 4/2009
Print ISSN: 0342-4642
Elektronische ISSN: 1432-1238
DOI
https://doi.org/10.1007/s00134-008-1332-0

Weitere Artikel der Ausgabe 4/2009

Intensive Care Medicine 4/2009 Zur Ausgabe

Update AINS

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.