Comparison of a new device for the continuous intra-gastric measurement of intra-abdominal pressure (CiMon) with direct intra-peritoneal measurements in cirrhotic patients during paracentesis

verfasst von: Valentin Becker, Roland M. Schmid, Andreas Umgelter

Erschienen in: Intensive Care Medicine | Ausgabe 5/2009

Einloggen, um Zugang zu erhalten

Abstract

Objective

To evaluate a new commercially available system for the intra-gastric measurement of intra-abdominal pressure (IAP).

Design and setting

Prospective interventional trial in an intensive care unit of a university hospital.

Patients

Ten patients with cirrhosis and tense ascites scheduled for paracentesis and instrumented with a ballon-tipped nasogastric tube.

Intervention

Intermittent paracentesis with repeat paired measurements of IAP.

Measurements and results

Intra-abdominal pressure was measured directly through the paracentesis catheter using a scaled measurement gauge. Simultaneously, intra-gastric measurements were performed with the CiMon device (Pulsion Medical Systems, Munich, Germany), by means of a ballon-tipped nasogastric tube connected to a pressure transducer. The range of IAP was 6.5–28 mmHg for direct measurements and 3.7–16 mmHg for measurements obtained with the CiMon device. The regression equation according to Passing and Bablok for the two methods had an intercept of 0.054 (95% CI–1.475–1.669) and a slope of 0.674 (95% CI 0.538–0.813). The cusum test showed a significant deviation from linearity (P < 0.05). In the Bland–Altman test bias and limits of agreement were – 4.9 (±6.8) mmHg or, expressed as percentage of the average, −44% (±49%).

Conclusion

Intra-gastric measurments of IAP using the CiMon device were not reliable enough to be clinically useful in these ascitic patients.
Anhänge
Nur mit Berechtigung zugänglich
Literatur
1.
Zurück zum Zitat Malbrain ML, Chiumello D, Pelosi P, Wilmer A, Brienza N, Malcangi V, Bihari D, Innes R, Cohen J, Singer P, Japiassu A, Kurtop E, De Keulenaer BL, Daelemans R, Del Turco M, Cosimini P, Ranieri M, Jacquet L, Laterre PF, Gattinoni L (2004) Prevalence of intra-abdominal hypertension in critically ill patients: a multicentre epidemiological study. Intensive Care Med 30:822–829PubMedCrossRef Malbrain ML, Chiumello D, Pelosi P, Wilmer A, Brienza N, Malcangi V, Bihari D, Innes R, Cohen J, Singer P, Japiassu A, Kurtop E, De Keulenaer BL, Daelemans R, Del Turco M, Cosimini P, Ranieri M, Jacquet L, Laterre PF, Gattinoni L (2004) Prevalence of intra-abdominal hypertension in critically ill patients: a multicentre epidemiological study. Intensive Care Med 30:822–829PubMedCrossRef
2.
Zurück zum Zitat Malbrain ML, Chiumello D, Pelosi P, Bihari D, Innes R, Ranieri VM, Del Turco M, Wilmer A, Brienza N, Malcangi V, Cohen J, Japiassu A, De Keulenaer BL, Daelemans R, Jacquet L, Laterre PF, Frank G, de Souza P, Cesana B, Gattinoni L (2005) Incidence and prognosis of intraabdominal hypertension in a mixed population of critically ill patients: a multiple-center epidemiological study. Crit Care Med 33:315–322PubMedCrossRef Malbrain ML, Chiumello D, Pelosi P, Bihari D, Innes R, Ranieri VM, Del Turco M, Wilmer A, Brienza N, Malcangi V, Cohen J, Japiassu A, De Keulenaer BL, Daelemans R, Jacquet L, Laterre PF, Frank G, de Souza P, Cesana B, Gattinoni L (2005) Incidence and prognosis of intraabdominal hypertension in a mixed population of critically ill patients: a multiple-center epidemiological study. Crit Care Med 33:315–322PubMedCrossRef
3.
Zurück zum Zitat Vidal MG, Ruiz Weisser J, Gonzalez F, Toro MA, Loudet C, Balasini C, Canales H, Reina R, Estenssoro E (2008) Incidence and clinical effects of intra-abdominal hypertension in critically ill patients. Crit Care Med 36:1823–1831PubMedCrossRef Vidal MG, Ruiz Weisser J, Gonzalez F, Toro MA, Loudet C, Balasini C, Canales H, Reina R, Estenssoro E (2008) Incidence and clinical effects of intra-abdominal hypertension in critically ill patients. Crit Care Med 36:1823–1831PubMedCrossRef
4.
Zurück zum Zitat Umgelter A, Reindl W, Wagner KS, Franzen M, Stock K, Schmid RM, Huber W (2007) Effects of plasma expansion with albumin and paracentesis on haemodynamics and kidney function in critically ill cirrhotic patients with tense ascites and hepatorenal syndrome: a prospective uncontrolled trial. Crit Care 12:R4CrossRef Umgelter A, Reindl W, Wagner KS, Franzen M, Stock K, Schmid RM, Huber W (2007) Effects of plasma expansion with albumin and paracentesis on haemodynamics and kidney function in critically ill cirrhotic patients with tense ascites and hepatorenal syndrome: a prospective uncontrolled trial. Crit Care 12:R4CrossRef
5.
Zurück zum Zitat Malbrain ML, Cheatham ML, Kirkpatrick A, Sugrue M, Parr M, De Waele J, Balogh Z, Leppäniemi A, Olvera C, Ivatury R, D’Amours S, Wendon J, Hillman K, Johansson K, Kolkman KAW (2006) Results from the international conference of experts on intra-abdominal hypertension and abdominal compartment syndrome. I. Definitions. Intensive Care Med 32:1722–1732PubMedCrossRef Malbrain ML, Cheatham ML, Kirkpatrick A, Sugrue M, Parr M, De Waele J, Balogh Z, Leppäniemi A, Olvera C, Ivatury R, D’Amours S, Wendon J, Hillman K, Johansson K, Kolkman KAW (2006) Results from the international conference of experts on intra-abdominal hypertension and abdominal compartment syndrome. I. Definitions. Intensive Care Med 32:1722–1732PubMedCrossRef
6.
Zurück zum Zitat Schachtrupp A, Wauters J, Wilmer A (2007) What is the best animal model for ACS? Acta Clin Belg Suppl 1:225–232 Schachtrupp A, Wauters J, Wilmer A (2007) What is the best animal model for ACS? Acta Clin Belg Suppl 1:225–232
8.
Zurück zum Zitat Malbrain ML (2004) Different techniques to measure intra-abdominal pressure (IAP): time for a critical re-appraisal. Intensive Care Med 30:357–371PubMedCrossRef Malbrain ML (2004) Different techniques to measure intra-abdominal pressure (IAP): time for a critical re-appraisal. Intensive Care Med 30:357–371PubMedCrossRef
9.
Zurück zum Zitat Cheatham ML, Malbrain ML, Kirkpatrick A, Sugrue M, Parr M, De Waele J, Balogh Z, Leppäniemi A, Olvera C, Ivatury R, D’Amours S, Wendon J, Hillman K, Wilmer A (2007) Results from the international conference of experts on intra-abdominal hypertension and abdominal compartment syndrome. II. Recommendations. Intensive Care Med 33:951–962PubMedCrossRef Cheatham ML, Malbrain ML, Kirkpatrick A, Sugrue M, Parr M, De Waele J, Balogh Z, Leppäniemi A, Olvera C, Ivatury R, D’Amours S, Wendon J, Hillman K, Wilmer A (2007) Results from the international conference of experts on intra-abdominal hypertension and abdominal compartment syndrome. II. Recommendations. Intensive Care Med 33:951–962PubMedCrossRef
10.
Zurück zum Zitat Cheatham ML, Safcsak K (1998) Intraabdominal pressure: a revised method for measurement. J Am Coll Surg 186:594–595PubMedCrossRef Cheatham ML, Safcsak K (1998) Intraabdominal pressure: a revised method for measurement. J Am Coll Surg 186:594–595PubMedCrossRef
11.
Zurück zum Zitat Zengerink I, McBeth PB, Zygun DA, Ranson K, Ball CG, Laupland KB, Widder S, Kirkpatrick AW (2008) Validation and experience with a simple continuous intra-abdominal pressure measurement technique in a multidisciplinary medical/surgical critical care unit. J Trauma 64:1159–1164PubMedCrossRef Zengerink I, McBeth PB, Zygun DA, Ranson K, Ball CG, Laupland KB, Widder S, Kirkpatrick AW (2008) Validation and experience with a simple continuous intra-abdominal pressure measurement technique in a multidisciplinary medical/surgical critical care unit. J Trauma 64:1159–1164PubMedCrossRef
12.
Zurück zum Zitat Passing H, Bablok W (1983) A new biometrical procedure for testing the equality of measurements from two different analytical methods. Application of linear regression procedures for method comparison studies in clinical chemistry, Part I. J Clin Chem Clin Biochem 21:709–720PubMed Passing H, Bablok W (1983) A new biometrical procedure for testing the equality of measurements from two different analytical methods. Application of linear regression procedures for method comparison studies in clinical chemistry, Part I. J Clin Chem Clin Biochem 21:709–720PubMed
13.
Zurück zum Zitat Bland JM, Altman DG (1999) Measuring agreement in method comparison studies. Stat Methods Med Res 8:135–160PubMedCrossRef Bland JM, Altman DG (1999) Measuring agreement in method comparison studies. Stat Methods Med Res 8:135–160PubMedCrossRef
14.
Zurück zum Zitat Bradley SE, Bradley GP (1947) The effect of increased intra-abdominal pressure on renal function in man. J Clin Invest 26:1010–1022PubMedCrossRef Bradley SE, Bradley GP (1947) The effect of increased intra-abdominal pressure on renal function in man. J Clin Invest 26:1010–1022PubMedCrossRef
15.
Zurück zum Zitat Sugrue M, Jones F, Deane SA, Bishop G, Bauman A, Hillman K (1999) Intra-abdominal hypertension is an independent cause of postoperative renal impairment. Arch Surg 134:1082–1085PubMedCrossRef Sugrue M, Jones F, Deane SA, Bishop G, Bauman A, Hillman K (1999) Intra-abdominal hypertension is an independent cause of postoperative renal impairment. Arch Surg 134:1082–1085PubMedCrossRef
16.
Zurück zum Zitat Dalfino L, Tullo L, Donadio I, Malcangi V, Brienza N (2007) Intra-abdominal hypertension and acute renal failure in critically ill patients. Intensive Care Med 34:707–713PubMedCrossRef Dalfino L, Tullo L, Donadio I, Malcangi V, Brienza N (2007) Intra-abdominal hypertension and acute renal failure in critically ill patients. Intensive Care Med 34:707–713PubMedCrossRef
17.
Zurück zum Zitat Malbrain ML, De laet I, Viaene D, Schoonheydt K, Dits H (2008) In vitro validation of a novel method for continuous intra-abdominal pressure monitoring. Intensive Care Med 34:740–745PubMedCrossRef Malbrain ML, De laet I, Viaene D, Schoonheydt K, Dits H (2008) In vitro validation of a novel method for continuous intra-abdominal pressure monitoring. Intensive Care Med 34:740–745PubMedCrossRef
18.
Zurück zum Zitat Collee GG, Lomax DM, Ferguson C, Hanson GC (1993) Bedside measurement of intra-abdominal pressure (IAP) via an indwelling naso-gastric tube: clinical validation of the technique. Intensive Care Med 19:478–480PubMedCrossRef Collee GG, Lomax DM, Ferguson C, Hanson GC (1993) Bedside measurement of intra-abdominal pressure (IAP) via an indwelling naso-gastric tube: clinical validation of the technique. Intensive Care Med 19:478–480PubMedCrossRef
19.
Zurück zum Zitat Davis PJ, Koottayi S, Taylor A, Butt WW (2005) Comparison of indirect methods of measuring intra-abdominal pressure in children. Intensive Care Med 32:471–475CrossRef Davis PJ, Koottayi S, Taylor A, Butt WW (2005) Comparison of indirect methods of measuring intra-abdominal pressure in children. Intensive Care Med 32:471–475CrossRef
Metadaten
Titel
Comparison of a new device for the continuous intra-gastric measurement of intra-abdominal pressure (CiMon) with direct intra-peritoneal measurements in cirrhotic patients during paracentesis
verfasst von
Valentin Becker
Roland M. Schmid
Andreas Umgelter
Publikationsdatum
01.05.2009
Verlag
Springer-Verlag
Erschienen in
Intensive Care Medicine / Ausgabe 5/2009
Print ISSN: 0342-4642
Elektronische ISSN: 1432-1238
DOI
https://doi.org/10.1007/s00134-009-1451-2

Weitere Artikel der Ausgabe 5/2009

Intensive Care Medicine 5/2009 Zur Ausgabe

Update AINS

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.