Bench test evaluation of volume delivered by modern ICU ventilators during volume-controlled ventilation

verfasst von: Aissam Lyazidi, Arnaud W. Thille, Guillaume Carteaux, Fabrice Galia, Laurent Brochard, Jean-Christophe M. Richard

Erschienen in: Intensive Care Medicine | Ausgabe 12/2010

Einloggen, um Zugang zu erhalten

Abstract

Purpose

During volume-controlled ventilation, part of the volume delivered is compressed into the circuit. To correct for this phenomenon, modern ventilators use compensation algorithms. Humidity and temperature also influence the delivered volume.

Methods

In a bench study at a research laboratory in a university hospital, we compared nine ICU ventilators equipped with compensation algorithms, one with a proximal pneumotachograph and one without compensation. Each ventilator was evaluated under normal, obstructive, and restrictive conditions of respiratory mechanics. For each condition, three tidal volumes (V T) were set (300, 500, and 800 ml), with and without an inspiratory pause. The insufflated volume and the volume delivered at the Y-piece were measured independently, without a humidification device, under ambient temperature and pressure and dry gas conditions. We computed the actually delivered V T to the lung under body temperature and pressure and saturated water vapour conditions (BTPS).

Results

For target V T values of 300, 500, and 800 ml, actually delivered V T under BTPS conditions ranged from 261 to 396 ml (−13 to +32%), from 437 to 622 ml (−13 to +24%), and from 681 to 953 ml (−15 to +19%), respectively (p < 0.01). Respiratory system mechanics and application of an inspiratory pause significantly affected actually delivered V T. Assuming a set V T of 6 ml/kg of predicted body weight, a difference of 1–2 ml/kg with actually delivered V T would be commonly observed.

Conclusion

The difference between preset V T and actually delivered V T is clinically meaningful and differs across modern ICU ventilators.
Anhänge
Nur mit Berechtigung zugänglich
Literatur
1.
Zurück zum Zitat (2000) Ventilation with lower tidal volumes as compared with traditional tidal volumes for acute lung injury and the acute respiratory distress syndrome. The Acute Respiratory Distress Syndrome Network. N Engl J Med 342:1301-1308 (2000) Ventilation with lower tidal volumes as compared with traditional tidal volumes for acute lung injury and the acute respiratory distress syndrome. The Acute Respiratory Distress Syndrome Network. N Engl J Med 342:1301-1308
2.
Zurück zum Zitat Cote CJ, Petkau AJ, Ryan JF, Welch JP (1983) Wasted ventilation measured in vitro with eight anesthetic circuits with and without inline humidification. Anesthesiology 59:442–446CrossRefPubMed Cote CJ, Petkau AJ, Ryan JF, Welch JP (1983) Wasted ventilation measured in vitro with eight anesthetic circuits with and without inline humidification. Anesthesiology 59:442–446CrossRefPubMed
3.
Zurück zum Zitat Forbat AF, Her C (1980) Correction for gas compression in mechanical ventilators. Anesth Analg 59:488–493CrossRefPubMed Forbat AF, Her C (1980) Correction for gas compression in mechanical ventilators. Anesth Analg 59:488–493CrossRefPubMed
4.
Zurück zum Zitat Lyazidi A, Richard JC-M, Thille AW, Besson O, Brochard L (2008) Are tidal volume delivered by the ventilator and received by the patient similar? Intensive Care Med 34:A0756 Lyazidi A, Richard JC-M, Thille AW, Besson O, Brochard L (2008) Are tidal volume delivered by the ventilator and received by the patient similar? Intensive Care Med 34:A0756
5.
Zurück zum Zitat Lellouche F, Taille S, Maggiore SM, Qader S, L’Her E, Deye N, Brochard L (2004) Influence of ambient and ventilator output temperatures on performance of heated-wire humidifiers. Am J Respir Crit Care Med 170:1073–1079CrossRefPubMed Lellouche F, Taille S, Maggiore SM, Qader S, L’Her E, Deye N, Brochard L (2004) Influence of ambient and ventilator output temperatures on performance of heated-wire humidifiers. Am J Respir Crit Care Med 170:1073–1079CrossRefPubMed
6.
Zurück zum Zitat Ferreira JC, Chipman DW, Kacmarek RM (2008) Trigger performance of mid-level ICU mechanical ventilators during assisted ventilation: a bench study. Intensive Care Med 34:1669–1675CrossRefPubMed Ferreira JC, Chipman DW, Kacmarek RM (2008) Trigger performance of mid-level ICU mechanical ventilators during assisted ventilation: a bench study. Intensive Care Med 34:1669–1675CrossRefPubMed
7.
Zurück zum Zitat Thille AW, Lyazidi A, Richard JC, Galia F, Brochard L (2009) A bench study of intensive care unit ventilators: new versus old and turbine-based versus compressed gas-based ventilators. Intensive Care Med 35:1368–1376CrossRefPubMed Thille AW, Lyazidi A, Richard JC, Galia F, Brochard L (2009) A bench study of intensive care unit ventilators: new versus old and turbine-based versus compressed gas-based ventilators. Intensive Care Med 35:1368–1376CrossRefPubMed
8.
Zurück zum Zitat Richard JC, Carlucci A, Breton L, Langlais N, Jaber S, Maggiore S, Fougere S, Harf A, Brochard L (2002) Bench testing of pressure support ventilation with three different generations of ventilators. Intensive Care Med 28:1049–1057CrossRefPubMed Richard JC, Carlucci A, Breton L, Langlais N, Jaber S, Maggiore S, Fougere S, Harf A, Brochard L (2002) Bench testing of pressure support ventilation with three different generations of ventilators. Intensive Care Med 28:1049–1057CrossRefPubMed
9.
Zurück zum Zitat Brochard L, Roudot-Thoraval F, Roupie E, Delclaux C, Chastre J, Fernandez-Mondejar E, Clementi E, Mancebo J, Factor P, Matamis D, Ranieri M, Blanch L, Rodi G, Mentec H, Dreyfuss D, Ferrer M, Brun-Buisson C, Tobin M, Lemaire F (1998) Tidal volume reduction for prevention of ventilator-induced lung injury in acute respiratory distress syndrome. The multicenter trail group on tidal volume reduction in ARDS. Am J Respir Crit Care Med 158:1831–1838PubMed Brochard L, Roudot-Thoraval F, Roupie E, Delclaux C, Chastre J, Fernandez-Mondejar E, Clementi E, Mancebo J, Factor P, Matamis D, Ranieri M, Blanch L, Rodi G, Mentec H, Dreyfuss D, Ferrer M, Brun-Buisson C, Tobin M, Lemaire F (1998) Tidal volume reduction for prevention of ventilator-induced lung injury in acute respiratory distress syndrome. The multicenter trail group on tidal volume reduction in ARDS. Am J Respir Crit Care Med 158:1831–1838PubMed
10.
Zurück zum Zitat Stewart TE, Meade MO, Cook DJ, Granton JT, Hodder RV, Lapinsky SE, Mazer CD, McLean RF, Rogovein TS, Schouten BD, Todd TR, Slutsky AS (1998) Evaluation of a ventilation strategy to prevent barotrauma in patients at high risk for acute respiratory distress syndrome. Pressure- and volume-limited Ventilation Strategy Group. N Engl J Med 338:355–361CrossRefPubMed Stewart TE, Meade MO, Cook DJ, Granton JT, Hodder RV, Lapinsky SE, Mazer CD, McLean RF, Rogovein TS, Schouten BD, Todd TR, Slutsky AS (1998) Evaluation of a ventilation strategy to prevent barotrauma in patients at high risk for acute respiratory distress syndrome. Pressure- and volume-limited Ventilation Strategy Group. N Engl J Med 338:355–361CrossRefPubMed
11.
Zurück zum Zitat Brower RG, Shanholtz CB, Fessler HE, Shade DM, White P Jr, Wiener CM, Teeter JG, Dodd-o JM, Almog Y, Piantadosi S (1999) Prospective, randomized, controlled clinical trial comparing traditional versus reduced tidal volume ventilation in acute respiratory distress syndrome patients. Crit Care Med 27:1492–1498CrossRefPubMed Brower RG, Shanholtz CB, Fessler HE, Shade DM, White P Jr, Wiener CM, Teeter JG, Dodd-o JM, Almog Y, Piantadosi S (1999) Prospective, randomized, controlled clinical trial comparing traditional versus reduced tidal volume ventilation in acute respiratory distress syndrome patients. Crit Care Med 27:1492–1498CrossRefPubMed
12.
Zurück zum Zitat Cannon ML, Cornell J, Tripp-Hamel DS, Gentile MA, Hubble CL, Meliones JN, Cheifetz IM (2000) Tidal volumes for ventilated infants should be determined with a pneumotachometer placed at the endotracheal tube. Am J Respir Crit Care Med 162:2109–2112PubMed Cannon ML, Cornell J, Tripp-Hamel DS, Gentile MA, Hubble CL, Meliones JN, Cheifetz IM (2000) Tidal volumes for ventilated infants should be determined with a pneumotachometer placed at the endotracheal tube. Am J Respir Crit Care Med 162:2109–2112PubMed
13.
Zurück zum Zitat Castle RA, Dunne CJ, Mok Q, Wade AM, Stocks J (2002) Accuracy of displayed values of tidal volume in the pediatric intensive care unit. Crit Care Med 30:2566–2574CrossRefPubMed Castle RA, Dunne CJ, Mok Q, Wade AM, Stocks J (2002) Accuracy of displayed values of tidal volume in the pediatric intensive care unit. Crit Care Med 30:2566–2574CrossRefPubMed
14.
Zurück zum Zitat Neve V, Leclerc F, Noizet O, Vernoux S, Leteurtre S, Forget P, Sadik A, Riou Y (2003) Influence of respiratory system impedance on volume and pressure delivered at the Y piece in ventilated infants. Pediatr Crit Care Med 4:418–425CrossRefPubMed Neve V, Leclerc F, Noizet O, Vernoux S, Leteurtre S, Forget P, Sadik A, Riou Y (2003) Influence of respiratory system impedance on volume and pressure delivered at the Y piece in ventilated infants. Pediatr Crit Care Med 4:418–425CrossRefPubMed
15.
Zurück zum Zitat Fujita Y, Imanaka H, Fujino Y, Takeuchi M, Tomita T, Mashimo T, Nishimura M (2006) Effect of humidifying devices on the measurement of tidal volume by mechanical ventilators. J Anesth 20:166–172CrossRefPubMed Fujita Y, Imanaka H, Fujino Y, Takeuchi M, Tomita T, Mashimo T, Nishimura M (2006) Effect of humidifying devices on the measurement of tidal volume by mechanical ventilators. J Anesth 20:166–172CrossRefPubMed
Metadaten
Titel
Bench test evaluation of volume delivered by modern ICU ventilators during volume-controlled ventilation
verfasst von
Aissam Lyazidi
Arnaud W. Thille
Guillaume Carteaux
Fabrice Galia
Laurent Brochard
Jean-Christophe M. Richard
Publikationsdatum
01.12.2010
Verlag
Springer-Verlag
Erschienen in
Intensive Care Medicine / Ausgabe 12/2010
Print ISSN: 0342-4642
Elektronische ISSN: 1432-1238
DOI
https://doi.org/10.1007/s00134-010-2044-9

Weitere Artikel der Ausgabe 12/2010

Intensive Care Medicine 12/2010 Zur Ausgabe

Update AINS

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.