01.08.2011 | Original

Continuous arterial pressure waveform monitoring in pediatric cardiac transplant, cardiomyopathy and pulmonary hypertension patients

verfasst von: Sarena Teng, Jon Kaufman, Zhaoxing Pan, Angela Czaja, Heather Shockley, Eduardo da Cruz

Erschienen in: Intensive Care Medicine | Ausgabe 8/2011

Einloggen, um Zugang zu erhalten

Abstract

Purpose

A continuous cardiac output monitor based on arterial pressure waveform (FloTrac™/Vigileo™; Edwards Lifesciences, Irvine, CA) is now approved for use in adults but not in children. This device is minimally invasive, calculates cardiac output continuously and in real time, and is easy to use. Our study sought to validate the FloTrac™ with the pulmonary artery catheter (PAC) intermittent thermodilution technique in pediatric cardiac patients.

Methods

This was a prospective pilot study comparing cardiac output measurements obtained via the FloTrac™ and arterial pressure waveform analysis with intermittent thermodilution. Subjects carried the diagnosis of pulmonary hypertension or cardiomyopathy, or were in the postoperative course after orthotopic heart transplantation.

Results

Enrolled in the study were 31 subjects, and 136 data points were obtained. The age range was 8 months to 16 years. The mean body surface area (BSA) was 1.1 m2. Bland-Altman plots for the mean cardiac outputs of all subjects with a BSA ≥1 m2 showed limits of agreement of −2.7 to 8.0 l/min (±5.4 l/min). Patients with a BSA ≤1 m2 demonstrated even wider limits of agreement (±8.5 l/min). The intraclass correlation for the PAC was 0.929 and 0.992 for the FloTrac™.

Conclusion

There was poor agreement between the PAC and FloTrac™ in measuring cardiac output in a population of children with pulmonary hypertension or cardiomyopathy, or after cardiac transplantation. This is in contrast to adult studies published thus far. This suggests that the utility of the FloTrac™ and measurements obtained from arterial pulse wave analysis in children is uncertain at this time.
Literatur
1.
Zurück zum Zitat Swan HJ, Ganz W, Forrester J, Marcus H, Diamond G, Chonette D (1970) Catheterization of the heart in man with use of a flow-directed balloon-tipped catheter. N Engl J Med 283:447–451PubMedCrossRef Swan HJ, Ganz W, Forrester J, Marcus H, Diamond G, Chonette D (1970) Catheterization of the heart in man with use of a flow-directed balloon-tipped catheter. N Engl J Med 283:447–451PubMedCrossRef
2.
Zurück zum Zitat Connors AF Jr, Speroff T, Dawson NV, Thomas C, Harrell FE Jr, Wagner D, Desbiens N, Goldman L, Wu AW, Califf RM, Fulkerson WJ Jr, Vidaillet H, Broste S, Bellamy P, Lynn J, Knaus WA (1996) The effectiveness of right heart catheterization in the initial care of critically ill patients. SUPPORT Investigators. JAMA 276:889–897PubMedCrossRef Connors AF Jr, Speroff T, Dawson NV, Thomas C, Harrell FE Jr, Wagner D, Desbiens N, Goldman L, Wu AW, Califf RM, Fulkerson WJ Jr, Vidaillet H, Broste S, Bellamy P, Lynn J, Knaus WA (1996) The effectiveness of right heart catheterization in the initial care of critically ill patients. SUPPORT Investigators. JAMA 276:889–897PubMedCrossRef
3.
Zurück zum Zitat The ESCAPE investigators; ESCAPE Study Coordinators (2005) Evaluation study of congestive heart failure and pulmonary artery catheter effectiveness: the ESCAPE trial. JAMA 294:1625–1633CrossRef The ESCAPE investigators; ESCAPE Study Coordinators (2005) Evaluation study of congestive heart failure and pulmonary artery catheter effectiveness: the ESCAPE trial. JAMA 294:1625–1633CrossRef
4.
Zurück zum Zitat Harvey S, Harrison DA, Singer M, Ashcroft J, Jones CM, Elbourne D, Brampton W, Williams D, Young D, Rowan K; PAC-Man study collaboration (2005) Assessment of the clinical effectiveness of pulmonary artery catheters in management of patients in intensive care (PAC-Man): a randomised controlled trial. Lancet 366:472–477PubMedCrossRef Harvey S, Harrison DA, Singer M, Ashcroft J, Jones CM, Elbourne D, Brampton W, Williams D, Young D, Rowan K; PAC-Man study collaboration (2005) Assessment of the clinical effectiveness of pulmonary artery catheters in management of patients in intensive care (PAC-Man): a randomised controlled trial. Lancet 366:472–477PubMedCrossRef
5.
Zurück zum Zitat Shah MR, Hasselblad V, Stevenson LW, Binanay C, O’Connor CM, Sopko G, Califf RM (2005) Impact of the pulmonary artery catheter in critically ill patients: meta-analysis of randomized clinical trials. JAMA 294:1664–1670PubMedCrossRef Shah MR, Hasselblad V, Stevenson LW, Binanay C, O’Connor CM, Sopko G, Califf RM (2005) Impact of the pulmonary artery catheter in critically ill patients: meta-analysis of randomized clinical trials. JAMA 294:1664–1670PubMedCrossRef
6.
Zurück zum Zitat National Heart, Lung, and Blood Institute Acute Respiratory Distress Syndrome (ARDS) Clinical Trials Network; Wheeler AP, Bernard GR, Thompson BT, Schoenfeld D, Wiedemann HP, de Boisblanc B, Connors AF Jr, Hite RD, Harabin AL (2006) Pulmonary-artery versus central venous catheter to guide treatment of acute lung injury. Fluid and Catheter Treatment Trial (FACTT). N Engl J Med 354:2213–2224PubMedCrossRef National Heart, Lung, and Blood Institute Acute Respiratory Distress Syndrome (ARDS) Clinical Trials Network; Wheeler AP, Bernard GR, Thompson BT, Schoenfeld D, Wiedemann HP, de Boisblanc B, Connors AF Jr, Hite RD, Harabin AL (2006) Pulmonary-artery versus central venous catheter to guide treatment of acute lung injury. Fluid and Catheter Treatment Trial (FACTT). N Engl J Med 354:2213–2224PubMedCrossRef
7.
Zurück zum Zitat Squara P, Cecconi M, Rhodes A, Singer M, Chiche JD (2009) Tracking changes in cardiac output: methodological considerations for the validation of monitoring devices. Intensive Care Med 35:1801–1808PubMedCrossRef Squara P, Cecconi M, Rhodes A, Singer M, Chiche JD (2009) Tracking changes in cardiac output: methodological considerations for the validation of monitoring devices. Intensive Care Med 35:1801–1808PubMedCrossRef
8.
Zurück zum Zitat Manecke GR Jr, Auger WR (2007) Cardiac output determination from the arterial pressure wave: clinical testing of a novel algorithm that does not require calibration. J Cardiothorac Vasc Anesth 21:3–7PubMedCrossRef Manecke GR Jr, Auger WR (2007) Cardiac output determination from the arterial pressure wave: clinical testing of a novel algorithm that does not require calibration. J Cardiothorac Vasc Anesth 21:3–7PubMedCrossRef
9.
Zurück zum Zitat Mayer J, Boldt J, Schollhorn T, Rohm JD, Menistu AM, Suttner S (2007) Semi-invasive monitoring of cardiac output by a new device using arterial pressure waveform analysis: a comparison with intermittent pulmonary artery thermodilution in patients undergoing cardiac surgery. Br J Anaesth 98:176–182PubMedCrossRef Mayer J, Boldt J, Schollhorn T, Rohm JD, Menistu AM, Suttner S (2007) Semi-invasive monitoring of cardiac output by a new device using arterial pressure waveform analysis: a comparison with intermittent pulmonary artery thermodilution in patients undergoing cardiac surgery. Br J Anaesth 98:176–182PubMedCrossRef
10.
Zurück zum Zitat Opdam HI, Wan L, Bellomo R (2006) A pilot assessment of the FloTrac cardiac output monitoring system. Intensive Care Med 33:344–349PubMedCrossRef Opdam HI, Wan L, Bellomo R (2006) A pilot assessment of the FloTrac cardiac output monitoring system. Intensive Care Med 33:344–349PubMedCrossRef
11.
Zurück zum Zitat Sander M, Spies CD, Grubitzscj H, Foer A, Muller M, von Heymann C (2006) Comparison of uncalibrated arterial waveform analysis in cardiac surgery patients with thermodilution cardiac output measurements. Crit Care 10:R164PubMedCrossRef Sander M, Spies CD, Grubitzscj H, Foer A, Muller M, von Heymann C (2006) Comparison of uncalibrated arterial waveform analysis in cardiac surgery patients with thermodilution cardiac output measurements. Crit Care 10:R164PubMedCrossRef
12.
Zurück zum Zitat McGee WT, Horswell JL, Calderon J, Janvier G, Van Severen T, Van den Berghe G, Kozikowski L (2007) Validation of a continuous arterial pressure-based cardiac output measurement: a multicenter, prospective clinical trial. Crit Care 11:R105PubMedCrossRef McGee WT, Horswell JL, Calderon J, Janvier G, Van Severen T, Van den Berghe G, Kozikowski L (2007) Validation of a continuous arterial pressure-based cardiac output measurement: a multicenter, prospective clinical trial. Crit Care 11:R105PubMedCrossRef
13.
Zurück zum Zitat Saaka SG, Kozieras J, Theumer O, van Hout N (2007) Measurement of cardiac output: a comparison between transpulmonary thermodilution and uncalibrated pulse contour analysis. Br J Anaesth 99:337–342CrossRef Saaka SG, Kozieras J, Theumer O, van Hout N (2007) Measurement of cardiac output: a comparison between transpulmonary thermodilution and uncalibrated pulse contour analysis. Br J Anaesth 99:337–342CrossRef
14.
Zurück zum Zitat Della Rocca G, Costa MG, Coccia C, Pompei L, Di Marco P, Vilardi V, Pietropaoli P (2003) Cardiac output monitoring: aortic transpulmonary thermodilution and pulse contour analysis agree with standard thermodilution methods in patients undergoing lung transplantation. Can J Anaesth 50:707–711PubMedCrossRef Della Rocca G, Costa MG, Coccia C, Pompei L, Di Marco P, Vilardi V, Pietropaoli P (2003) Cardiac output monitoring: aortic transpulmonary thermodilution and pulse contour analysis agree with standard thermodilution methods in patients undergoing lung transplantation. Can J Anaesth 50:707–711PubMedCrossRef
15.
Zurück zum Zitat Felbinger TW, Reuter DA, Eltzschig HK, Bayerlein J, Goetx AE (2005) Cardiac index measurements during rapid preload changes: a comparison of pulmonary artery thermodilution with arterial pulse contour analysis. J Clin Anesth 17:241–248PubMedCrossRef Felbinger TW, Reuter DA, Eltzschig HK, Bayerlein J, Goetx AE (2005) Cardiac index measurements during rapid preload changes: a comparison of pulmonary artery thermodilution with arterial pulse contour analysis. J Clin Anesth 17:241–248PubMedCrossRef
16.
Zurück zum Zitat Gödje O, Höke K, Goetz AE, Felbinger TW, Reuter DA, Reichart B, Friedl R, Hannekum A, Pfeiffer UJ (2002) Reliability of a new algorithm for continuous cardiac output determination by pulse-contour analysis during haemodynamic instability. Crit Care Med 30:52–58PubMedCrossRef Gödje O, Höke K, Goetz AE, Felbinger TW, Reuter DA, Reichart B, Friedl R, Hannekum A, Pfeiffer UJ (2002) Reliability of a new algorithm for continuous cardiac output determination by pulse-contour analysis during haemodynamic instability. Crit Care Med 30:52–58PubMedCrossRef
17.
Zurück zum Zitat Halvorsen PS, Espinoza A, Lundblad R, Cvancarova M, Hol PK, Fosse E, Tønnessen TI (2006) Agreement between PiCCO pulse-contour analysis, pulmonal artery thermodilution and transthoracic thermodilution during off-pump coronary artery by-pass surgery. Acta Anaesthesiol Scand 50:1050–1057PubMedCrossRef Halvorsen PS, Espinoza A, Lundblad R, Cvancarova M, Hol PK, Fosse E, Tønnessen TI (2006) Agreement between PiCCO pulse-contour analysis, pulmonal artery thermodilution and transthoracic thermodilution during off-pump coronary artery by-pass surgery. Acta Anaesthesiol Scand 50:1050–1057PubMedCrossRef
18.
Zurück zum Zitat Cecconi M, Dawson D, Grounds RM, Rhodes A (2009) Lithium dilution cardiac output measurement in the critically ill patient: determination of precision of the technique. Intensive Care Med 35:498–504PubMedCrossRef Cecconi M, Dawson D, Grounds RM, Rhodes A (2009) Lithium dilution cardiac output measurement in the critically ill patient: determination of precision of the technique. Intensive Care Med 35:498–504PubMedCrossRef
19.
Zurück zum Zitat Knirsch W, Kretschmar O, Tomaske M, Stutz K, Nagdyman N, Balmer C, Schmitz A, Bettex D, Berger F, Bauersfeld U, Weiss M (2008) Cardiac output measurement in children: comparison of the ultrasound cardiac output monitor with thermodilution cardiac output measurement. Intensive Care Med 34:1060–1064PubMedCrossRef Knirsch W, Kretschmar O, Tomaske M, Stutz K, Nagdyman N, Balmer C, Schmitz A, Bettex D, Berger F, Bauersfeld U, Weiss M (2008) Cardiac output measurement in children: comparison of the ultrasound cardiac output monitor with thermodilution cardiac output measurement. Intensive Care Med 34:1060–1064PubMedCrossRef
20.
Zurück zum Zitat Notterman DA, Castello FV, Steinberg C, Greenwald BM, O’Loughlin JE, Gold JP (1989) A comparison of thermodilution and pulsed Doppler cardiac output measurement in critically ill children. J Pediatr 115:554–560PubMedCrossRef Notterman DA, Castello FV, Steinberg C, Greenwald BM, O’Loughlin JE, Gold JP (1989) A comparison of thermodilution and pulsed Doppler cardiac output measurement in critically ill children. J Pediatr 115:554–560PubMedCrossRef
21.
Zurück zum Zitat Kim J, Dreyer J, Chang A (2006) Arterial pulse wave analysis: an accurate means of determining cardiac output in children. Pediatr Crit Care Med 7:532–535PubMedCrossRef Kim J, Dreyer J, Chang A (2006) Arterial pulse wave analysis: an accurate means of determining cardiac output in children. Pediatr Crit Care Med 7:532–535PubMedCrossRef
22.
Zurück zum Zitat Bland JM, Altman DG (1986) Statistical methods for assessing agreement between two methods of clinical measurement. Lancet i:307–310CrossRef Bland JM, Altman DG (1986) Statistical methods for assessing agreement between two methods of clinical measurement. Lancet i:307–310CrossRef
23.
Zurück zum Zitat Critchley LA, Critchley JA (1999) A meta-analysis of studies using bias and precision statistics to compare cardiac output techniques. J Clin Monit Comput 115:85–91CrossRef Critchley LA, Critchley JA (1999) A meta-analysis of studies using bias and precision statistics to compare cardiac output techniques. J Clin Monit Comput 115:85–91CrossRef
24.
Zurück zum Zitat Barnhart HX, Haber MJ, Lin LI (2007) An overview on agreement with continuous measurements. J Biopharm Stat 17:529–569PubMedCrossRef Barnhart HX, Haber MJ, Lin LI (2007) An overview on agreement with continuous measurements. J Biopharm Stat 17:529–569PubMedCrossRef
25.
Zurück zum Zitat Mayer J, Boldt J, Poland R (2009) Continuous arterial pressure waveform-based cardiac output using the FloTrac/Vigileo: a review and meta-analysis. J Cardiothorac Vasc Anesth 23:401–406PubMedCrossRef Mayer J, Boldt J, Poland R (2009) Continuous arterial pressure waveform-based cardiac output using the FloTrac/Vigileo: a review and meta-analysis. J Cardiothorac Vasc Anesth 23:401–406PubMedCrossRef
26.
Zurück zum Zitat Manecke GR (2005) Edward’s FloTrac sensor and Vigileo Monitor: easy, accurate reliable cardiac output assessment using the arterial pulse wave. Expert Rev Med Devices 2:523–527PubMedCrossRef Manecke GR (2005) Edward’s FloTrac sensor and Vigileo Monitor: easy, accurate reliable cardiac output assessment using the arterial pulse wave. Expert Rev Med Devices 2:523–527PubMedCrossRef
27.
Zurück zum Zitat Button D, Weibel L, Reuthebuch O, Genoni M, Zollinger A, Hofer CK (2007) Clinical evaluation of the FloTrac/Vigileo system and two established continuous cardiac output monitoring devices in patients undergoing cardiac surgery. Br J Anaesth 99:329–336PubMedCrossRef Button D, Weibel L, Reuthebuch O, Genoni M, Zollinger A, Hofer CK (2007) Clinical evaluation of the FloTrac/Vigileo system and two established continuous cardiac output monitoring devices in patients undergoing cardiac surgery. Br J Anaesth 99:329–336PubMedCrossRef
28.
Zurück zum Zitat Calamandrei M, Mirabile L, Muschetta S (2008) Assessment of cardiac output in children: a comparison between the pressure recording analytical method and Doppler echocardiography. Pediatr Crit Care Med 9:310–312PubMedCrossRef Calamandrei M, Mirabile L, Muschetta S (2008) Assessment of cardiac output in children: a comparison between the pressure recording analytical method and Doppler echocardiography. Pediatr Crit Care Med 9:310–312PubMedCrossRef
29.
Zurück zum Zitat Ronco R, Riquelme C (2008) Cardiac output measurement in children: what is lacking? Pediatr Crit Care Med 9:333–334PubMedCrossRef Ronco R, Riquelme C (2008) Cardiac output measurement in children: what is lacking? Pediatr Crit Care Med 9:333–334PubMedCrossRef
30.
Zurück zum Zitat Cecconi M, Rhodes A, Poloniecki J, Della Rocca G, Grounds RM (2009) Bench to bedside review: the importance of the precision of the reference technique in method comparison studies – with specific reference to the measurement of cardiac output. Crit Care 13:201–206PubMedCrossRef Cecconi M, Rhodes A, Poloniecki J, Della Rocca G, Grounds RM (2009) Bench to bedside review: the importance of the precision of the reference technique in method comparison studies – with specific reference to the measurement of cardiac output. Crit Care 13:201–206PubMedCrossRef
Metadaten
Titel
Continuous arterial pressure waveform monitoring in pediatric cardiac transplant, cardiomyopathy and pulmonary hypertension patients
verfasst von
Sarena Teng
Jon Kaufman
Zhaoxing Pan
Angela Czaja
Heather Shockley
Eduardo da Cruz
Publikationsdatum
01.08.2011
Verlag
Springer-Verlag
Erschienen in
Intensive Care Medicine / Ausgabe 8/2011
Print ISSN: 0342-4642
Elektronische ISSN: 1432-1238
DOI
https://doi.org/10.1007/s00134-011-2252-y

Weitere Artikel der Ausgabe 8/2011

Intensive Care Medicine 8/2011 Zur Ausgabe

Update AINS

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.